RANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3
K130236 · Radiometer Medical Aps · JJS · Mar 15, 2013 · Clinical Chemistry
Device Facts
Record ID
K130236
Device Name
RANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3
Applicant
Radiometer Medical Aps
Product Code
JJS · Clinical Chemistry
Decision Date
Mar 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil
Device Story
Range+ QUALICHECK is an assayed quality control system consisting of three levels of aqueous solutions containing biological buffers, salts, glucose, lactate, dyes, and preservatives. It is used to verify the accuracy and precision of Radiometer ABL700, ABL800/8x7, ABL77, ABL80, and ABL90 series analyzers. The solution is equilibrated with carbon dioxide and oxygen. The healthcare provider uses the control solution by conditioning ampoules at 18-32°C, then running the sample on the analyzer. The analyzer measures the parameters and compares results against the provided control ranges. This process ensures the analyzer is functioning correctly, allowing the clinician to rely on patient test results for clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Stability and shelf-life were validated using three lots of each level, measured on ABL735 analyzers over 36 months. Value assignment was performed using 144 measurements across multiple validated ABL7xx series devices to establish mean ±2SD target ranges. Traceability to NIST and IUPAC standards was documented.
Technological Characteristics
Aqueous solution containing biological buffer, salts, glucose, lactate, dyes, and preservative. Equilibrated with CO2 and O2. Provided in 2mL ampoules. Three-level system. Traceable to NIST SRM and IUPAC pH scale. Storage: 2-8°C. In-use: 18-32°C.
Indications for Use
Indicated for use as an assayed quality control system for evaluating the accuracy and precision of blood gas, oximetry, electrolyte, metabolite, and hemoglobin parameters on specified Radiometer ABL analyzers.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
QUALICHECK5+ (k980135)
Submission Summary (Full Text)
{1}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130236
B. Purpose for Submission:
New Device
C. Measurand:
Quality control material for blood gases
D. Type of Test:
Not applicable
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
Range+ QUALICHECK, Level 1
Range+ QUALICHECK, Level 2
Range+ QUALICHECK, Level 3
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660, Quality Control Material (Assayed and Unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJS, Controls for Blood-Gases
{2}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges.
Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not applicable
I. Device Description:
The Range+ QUALICHECK is an assayed quality control system, which can be used for quality control of the ABL700 Series, ABL800/8x7, ABL77, ABL80 and ABL90 analyzers from Radiometer. The Range+ QUALICHECK is a three level quality control system. Each level consists of 30 2mL ampoules of solution per box. The quality control solution is an aqueous solution containing a biological buffer, salts, glucose, lactate, dyes and a preservative, and it is equilibrated with carbon dioxide and oxygen. Range + QUALICHECK measures the following parameters: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil.
J. Substantial Equivalence Information:
1. Predicate device name(s):
QUALICHECK5+
2. Predicate 510(k) number(s):
k980135
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Indications for Use | For In Vitro Diagnostic Use. It is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. | Same |
| Matrix | Aqueous solution | Same |
| Blood gas measurement | pH, pO2, pCO2 | Same |
| Oximetry Measurement | ctHb, sO2, FO2Hb, FCOHb, FMetHb | Same |
| Electrolyte Measurement | cK+, cNa+, cCa2+, cCl- | Same |
| Metabolite Measurement | cGlucose, cLactate | Same |
| Hemoglobin Measurement | ctBil | Same |
| Compatible analyzers | ·ABL700/705/710/715/720/725/730/735 ·ABL805/810/815/820/825/830/835 ·ABL80 - OSM SW ·ABL90 FLEX | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Storage | 2°C to 8°C until expiration date 5 hours between 18-32°C | 2°C to 25°C until expiration date, including up to a total of 15 days at up to 32°C |
| Oximetry Measurement | FHbf in Level 1 | FHbf in all four levels |
| Levels | Three levels | Four levels |
K. Standard/Guidance Document Referenced (if applicable):
None
{4}
# L. Test Principle:
Not applicable
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The materials are traceable to the following standard materials:
| Parameter | Unit | Traceable to |
| --- | --- | --- |
| pH | - | The IUPAC pH scale. Primary pH standards are certified by the Danish Primary Laboratory for Electrochemistry (DPLEC) at the Danish Institute of Fundamental Metrology (DFM) under DANAK accreditation no. 255. The system is validated by comparison with SRM produced by National Institute of Standards and Technology (NIST) |
| pCO2 | mmHg | NIST Standard Reference Material (SRM) gas 1674b and SRM 2625a |
| pO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cK+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 999b |
| cNa+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 919b |
| cCa2+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 915 |
| cCl- | mmol/L (37°C) | NIST Standard Reference Material (SRM) 999b |
| cGlucose | mmol/L (37°C) | NIST Standard Reference Material (SRM) 917b |
| cLactate | mmol/L (37°C) | L-Lactic Acid Lithium Salt. SIGMA L2250 |
| cMalt | mmol/L (37°C) | L-Lactic Acid Lithium Salt. SIGMA L2250 |
| pH | - | NIST Standard Reference Material (SRM) gas 1674b and SRM 2625a |
| pCO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| pO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cK+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cNa+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cCa2+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cCl- | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cGlucose | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
{5}
| ctHb | g/dL | NIST SRM (absorbance, wavelength). Hemoglobin-cyanide standard. J.T. Baker (Product No. 3061) |
| --- | --- | --- |
| sO₂ | % | NIST SRM (absorbance, wavelength). NIST SRM gas, whole blood, pH=7.4, ctHb=15 g%, sO₂=100% |
## Stability:
Shelf-life characteristics for the Range+ QUALICHECK control solutions were determined using three different lots of each level at 32°C for 10 days followed by 8°C for 36 months. The products were placed at 32°C for 10 days to assess the potential short term temperature deviations during transportation and handling. Shelf-life and in-use stability was tested by measurements on 2 ABL735 analyzers, 6 replicates at each time point (0, 12, 25, and 36 months). Protocols and acceptance criteria have been reviewed and found to be adequate.
The control solutions are stable for 2 years at 2-8°C.
In-use Stability: Ampoules should be conditioned for at least 5 hours at a constant temperature between 18°C-32°C before use. The contents should be used immediately after opening.
## Value Assignment
The sponsor provided assigned values and control ranges for each level. To determine the assigned values and control ranges for Range+ QUALICHECK, six (6) trays of 1000 ampoules are sampled randomly from the Range+ QUALICHECK batch. Twelve ampoules are sampled from each of the trays and 72 ampoules are sampled from the reference batch. The samples are conditioned and shaken at 25°C in a water bath for 6 hours. Measurements are performed on a minimum of 3 validated ABL7xx series devices with data collection. Measurement of each parameter is performed alternately on the reference ampoule and the sample ampoule and repeated 12 times on each ABL7xx for a total of 144 measurements. Target ranges are calculated based on the mean ±2SD.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
{6}
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Representative target values for the control solutions have been provided in the labeling.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.