← Product Code [NHA](/productcode/NHA) · K130183

# SFI BAR ABUTMENTS (K130183)

_Sterngold Dental, LLC · NHA · Jul 7, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130183

## Device Facts

- **Applicant:** Sterngold Dental, LLC
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Jul 7, 2013
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.

## Device Story

SFI Bar® Implant Abutments serve as prosthetic interfaces between dental implants and removable overdentures. Abutments are screwed into existing dental implants; a stress-free bar is then attached to the abutment to support the prosthesis. Used in dental clinics by clinicians to restore chewing function in edentulous patients. Device provides mechanical fixation; no software or electronic components involved.

## Clinical Evidence

No clinical data. Bench testing only, including torque tests, application, and functional testing. Torque tests demonstrated safety margins exceeding manufacturer recommendations. Design validation performed per 21 CFR 820.30.

## Technological Characteristics

Material: Wrought Titanium 6 Aluminum-4 Vanadium ELI Alloy (ASTM F136). Seven platform configurations ([A], [B], [C], [S], [X], [X], [Z]) for compatibility with various dental implant systems. Non-sterile supply. Mechanical connection via screw fixation.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- SFI Bar® System Complete 2-Implant and 4-Implant ([K083876](/device/K083876.md))
- SFI Bar® Implant Adapter Straumann, SFI Bar® Implant Adapter Neo- ([K102382](/device/K102382.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 13, 2013 SFI Bar® Implant Abutments

K130183

# 510(k) Summary

| Sponsor:                                                                     | Sterngold Dental, LLC<br>23 Frank Mossberg Drive<br>Attleboro, MA 02703                                        |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Contact:                                                                     | Maria Rao, QA/RA Director<br>Ph: 508-226-5660 ext 1206                                                         |
| Trade Name:                                                                  | SFI Bar® Implant Abutments                                                                                     |
| Common Name:                                                                 | Implant Abutments                                                                                              |
| Classification Name:                                                         | Endosseous Dental Implant Abutment                                                                             |
| Classification:                                                              | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II |
| Product Code:                                                                | NHA (21CFR 872.3630)                                                                                           |
| Legally Marketed Device to which Equivalence is claimed (Predicate Devices): | Predicate Device(s): K083876                                                                                   |

K102382

| K083876 | SFI Bar® System Complete 2-Implant and 4-Implant                  |
|---------|-------------------------------------------------------------------|
| K102382 | SFI Bar® Implant Adapter Straumann, SFI Bar® Implant Adapter Neo- |

#### Description of Device:

Device Description: The SFI Bar® Implant Abutments provide the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI Bar® Implant Abutments consist of an abutment, which is attached to a stress free bar for the fixation of removable overdentures. The implant abutment is screwed into the dental abutment. The implant abutments fit Nobel Biocare Brånemark System®. Nobel Biocare (Steri-Oss®), Keystone (Lifecore), 3i Implant Inovations®, Sterngold-ImplaMed®, Interpore IMZ™, Osstem, Zimmer (Paragon, Centerpulse), OIC, IMTEC Corporation®, MIS Implants, Minimatic/Stryker, Bud, Straumann, Biolok International, INNOVA, Implant Direct, Zimmer (Calcitek® Centerpulse), BioHorizons®. See table 1 for platform compatibility.

There are seven (7) different platforms and each platform is compatible with one or more implant types. The platforms of the abutments are [A], [B], [C], [S], [X], [X], [Z]. Table 1 specifies which dental implants are compatible with these platforms. The difference between each platform is the internal connection with the specific implant. The devices are supplied non-sterile, therefore there is no shelf life.

{1}------------------------------------------------

### Abutment Insertion

Choose the abutment with the proper cuff height that fits on the existing implant. Abutment platforms should be 1 to 2 mm above the gingival level and approximately parallel to the occlusal plane. However, to allow the subsequent placement of the sections of SFI-Bar® on top of these abutments so that the bars are approximately parallel to the occlusal plane, it may be necessary to choose some abutments with different heights. Abutments are screwed into each implant.

# Intended Use of the Device:

The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.

| Implant Brand                  | Model                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nobel Biocare Brånemark System | 3.3 Fixture, 3.75 Fixture, 4.0 Fixture, 5.0 Fixture (Old Version), 3.75 MkII Self-tapping Fixture, 4.0 MkII Self-tapping Fixture                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Sterngold-ImplaMed             | 3.3 Hex Cylinder, 4.0 Hex Cylinder, 3.75 Standard Hex Screw, 3.75 Self-tapping Hex Screw, 3.75 Self-tapping "SST" Hex Screw, 4.0 Standard Hex Screw, 4.0 Self-tapping Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 5.0 RP "SST" Hex Screw, 3.75 RP Acid Etched, 4.0 RP Acid Etched, 5.0 RP Acid Etched, 4.1 Stern IC (4.8 head), 3.3 Stern IC (4.8 head)                                                                                                                                                              |
| Nobel Biocare (Steri-Oss®)     | 3.8 HL Cylinder, 3.8 HL Threaded, 4.5 HL Threaded, 3.5 NobelReplace™, Replace® Select (NP), 4.0 NobelReplace Straight, (RP), 4.3Replace®Select&NobelReplace™ (RP)                                                                                                                                                                                                                                                                                                                                                                                       |
| Keystone (Lifecore)            | 3.75 Restore® Self-tapping Screw, 4.0 Restore® Self-tapping Screw, 3.75 Restore® External Hex Screw, 4.0 Restore® External Hex Screw, 4.0 Restore® External Hex Cylinder, 4.2 Sustain® External Hex Cylinder, 3.75 Sustain® External Hex Screw, 4.0 Sustain® External Hex Screw, 4.2 Sustain® External Hex MC Cylinder, Stage-1TM (3.3 and 4.0 fixtures)                                                                                                                                                                                                |
| 3i Implant Inovations          | 3.25 External Hex Miniplant®, 3.25 ICE™ Miniplant®, 3.25 OSSEOTITE® Miniplant®, 3.3 Cylinder Miniplant®, 3.3 External Hex Cylinder, 3.75 ICE™ Self-tapping, 3.75 OSSEOTITE®, 3.75 Self-tapping Threaded, 3.75 Standard Threaded, 4.0 External Hex Cylinder, 4.0 ICE™ Self-tapping, 4.0 OSSEOTITE®, 4.0 Standard Threaded, 4.25 External Hex Cylinder, TG OSSEOTITE® (4.8 Platform), 4.0 OSSEOTITE® Certain™, 4.0 OSSEOTITE® NTCertain™, 4.0 OSSEOTITE®CERTAIN PREVAIL, 5.0 Osseotite® Certain, 5.0 Osseotite® NT Certain, 5.0 Osseotite®Certain Prevail |
| IMTEC Corporation®             | 3.3 Universal Flare Cylinder, 3.75 Universal Self-tapping, 3.75 Universal Self-tapping Coated, 4.0 Spike Cylinder, 4.0 Universal Cylinder                                                                                                                                                                                                                                                                                                                                                                                                               |
| Interpore IMZ™                 | 3.3 Hex Cylinder, 3.75 Self-tapping Threaded, 4.0 Hex Cylinder, 4.0 Self-tapping Threaded, 4.25 Hex Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Osstem                         | 4.1 US II, III, II Plus, III Plus, SS II, III (4.8 head)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

The SFI-Bar® Implant Abutments are compatible with the following implant systems:

{2}------------------------------------------------

| Zimmer Dental                                | 3.5 Bio-Vent® X             TM, 3.75 Swede-VentTM Conical Neck CST, 3.75 Swede-<br>VentTM Standard, 4.0 Swede-VentTM Standard, 4.0 Bio-Vent® XTM, 3.25 Micro-<br>Vent® (3.5 head), 3.3 Screw-Vent® (3.5 head), 3.5 Bio-Vent® (3.5<br>head), 3.7 Screw-Vent® (3.5 head), 3.75 Screw-Vent® (3.5 head), 4.3 Core-<br>Vent® (3.5 head), 4.25 Micro-Vent® (4.5 head), 4.5 Bio-<br>Vent® (4.5 head), 4.7 Screw-Vent® (4.5 head), 5.3 Core-Vent® (4.5 head),<br>3.7 Tapered Swiss PlusTM (4.8 platform), 4.8 Tapered Swiss PlusTM 4.1 Straight<br>Swiss PlusTM, 4.8 Straight Swiss PlusTM |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Zimmer (Calcitek®, Centerpulse)<br>Straumann | 3.75 ThreadLocTM<br>ITI TETM 3.3 (4.8 head), ITI 3.3 Std & Std Plus (4.8 head), ITI TETM 4.1 (4.8 head),<br>ITI 4.1 Std. & Std. Plus (4.8 head), ITI, 4.8 Std. & Std. Plus (4.8 head)                                                                                                                                                                                                                                                                                                                                                                                              |
| Biolok International                         | 4.5 Silhouette Screw, 4.0 Micro-Lok Screw, 4.0 Micro-Lok Cylinder, 3.75 Micro-Lok Screw, 3.3 Micro-Lok Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Bud                                          | 3.25 Bud Screwvent, 3.75 Bud Screwvent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| INNOVA                                       | 4.1 ENDOPORE® External Connection, 4.0 ENTEGRATM External Connection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| OIC                                          | 3.0 Osteo Standard ST, 3.25 Osteo Standard ST, 3.75 Osteo Standard ST                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| MIS IMPLANTS                                 | 3.3mm Internal Hex, 3.75mm Internal Hex, 4.20mm Internal Hex, 5.0mm Internal Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| BioHorizons®                                 | 3.5 Internal, 4.0 Internal, 4.5 Internal, 3.5 Single Stage, 4.5 Single Stage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Implant Direct                               | Legacy 3.5mm, Legacy 4.5mm, RePlantTM 4.3mm, RePlantTM 3.5mm, 3.7mm ScrewPlant, 4.7mm ScrewPlant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Minimatic/Stryker                            | 3.3 External Hex Cylinder, 3.75 External Hex Screw, 4.0 External Hex Cylinder,<br>4.0 External Hex Screw, 4.75 External Hex Screw, 5.0 External Hex Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                       |

# Summary Technological Characteristics:

The proposed implant abutments are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.

The material of the implant abutments conform to ASTM F136, Wrought Titanium 6 Aluminum-4 Vanadium ELI Alloy.

# Comparison/Compatibility

# Substantial Equivalence:

The proposed implant abutments are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.

To ensure compatibility the following process was carried out: The implant abutments were designed and developed, and manufactured according to manufacturer's specifications and controlled procedures. A validation protocol was done in accordance with Design Control requirements per FDA CFR820.30.

Table 2 summarizes the substantial equivalence comparison to the predicate devices.

{3}------------------------------------------------

# Performance Data:

Torque tests, application and functional testing have been conducted to evaluate the performance characteristics of the SFI Bar® Implant Abutments. The test methods used were the same as in predicate devices. Testing has shown that the SFI Bar® Implant Abutments included in this application are equivalent in performance characteristics to the predicate SFI Barto. The acceptance criteria were met.

Torque test results indicated that there is sufficiently large safety margin for fracture of the SF1 Bar® Implant Abutments specified on this submission to occur when tightening. Test samples torqued above 50 Ncm, well above the implant manufacturer's recommended torque value. There is no risk or possibility of harm to the patient.

# Summary of Testing to Demonstrate

### Safety and Effectiveness / Conclusion:

Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. The non-clinical testing consisted of analysis of platforms to identify worst case test samples. Fatigue testing was not done as the basic design is the same as the predicate devices. The evaluation was based on FDA guidance "Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments." Torque tests, application and functional tests have been carried out.

The summary of technological characteristics as well as the torque test, application and functional testing indicate that the device is safe and effective for its intended use and performs as well or better than the predicate devices.

{4}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure with its wings spread, symbolizing protection and care.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

July 3, 2013

Sterngold Dental, Limited Liability Company C/O Ms. Maria Rao Director of Regulatory Affairs 23 Frank Mossberg Drive ATTLEBORO MA 02703

Re: K130183

Trade/Device Name: SFI Bar® Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 30, 2013 Received: June 3, 2013

Dear Ms. Rao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Rao

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.hum for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Lester W. Schultheis Jr 2013.07.03 13:53:09 -04'00'

Acting for Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Sterngaid Dental, LLC
Abbreviated SIO(k) Premarket Notiflention

March 13, 2013 8F1 Barto Impiant Abutments

510(k) Number (if known):

Device Name: SFI Bar@ Abutments

#### Indications for Use:

The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.

The SFI-Bar® Implant Abutments are compatible with the following implant systems:

| Implant Brand                  | Model                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nobel Biocare Branemark System | 3.3 Fixture, 3.75 Fixture, 4.0 Fixture, 5.0 Fixture (Old Version), 3.75 Mkll Self-<br>tapping Fixture, 4.0 Mkll Self-tapping Fixture                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sterngold-ImplaMed             | 3.3 Hex Cylinder, 4.0 Hex Cylinder, 3.75 Standard Hex Screw, 3.75 Self-tapping<br>Hex Screw, 3.75 Self-tapping "SST" Hex Screw, 4.0 Standard Hex Screw, 4.0<br>Self-tapping Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 4.0 Self-tapping<br>"SST" Hex Screw, 5.0 RP "SST" Hex Screw, 3.75 RP Acid Etched, 4.0 RP Acid<br>Etched, 5.0 RP Acid Etched, 4.1 Stern IC (4.8 head), 3.3 Stern IC (4.8 head)                                                                                                                                                                       |
| Nobel Biocare (Steri-Oss®)     | 3.8 HL Cylinder, 3.8 HL Threaded, 4.5 HL Threaded, 3.5 NobelReplace™,<br>Replace® Select (NP), 4.0 NobelReplace Straight, (RP),<br>4.3Replace®Select&NobelReplace™ (RP)                                                                                                                                                                                                                                                                                                                                                                                                      |
| Keystone (Lifecore)            | 3.75 Restore® Self-tapping Screw, 4.0 Restore® Self-tapping Screw, 3.75<br>Restore® External Hex Screw, 4.0 Restore® External Hex Screw, 4.0 Restore®<br>External Hex Cylinder, 4.2 Sustain® External Hex Cylinder, 3.75 Sustain®<br>External Hex Screw, 4.0 Sustain® External Hex Screw, 4.2 Sustain® External<br>Hex MC Cylinder, Stage-1TM (3.3 and 4.0 fixtures)                                                                                                                                                                                                         |
| 3i Implant Inovations          | 3.25 External Hex Miniplant®, 3.25 ICE™ Miniplant®, 3.25 OSSEOTITE®<br>Miniplant®, 3.3 Cylinder Miniplant®, 3.3 External Hex Cylinder, 3.75 ICE™<br>Self-tapping, 3.75 OSSEOTITE®, 3.75 Self-tapping Threaded, 3.75 Standard<br>Threaded, 4.0 External Hex Cylinder, 4.0 ICE™ Self-tapping, 4.0<br>OSSEOTITE®, 4.0 Standard Threaded, 4.25 External Hex Cylinder, TG<br>OSSEOTITE® (4.8 Platform), 4.0 OSSEOTITE® Certain™, 4.0<br>OSSEOTITE® NTCertain™, 4.0 OSSEOTITE®CERTAIN PREVAIL, 5.0<br>Osseotite® Certain, 5.0 Osseotite® NT Certain, 5.0 Osseotite®Certain Prevail |
| IMTEC Corporation®             | 3.3 Universal Flare Cylinder, 3.75 Universal Self-tapping, 3.75 Universal Self-<br>tapping Coated, 4.0 Spike Cylinder, 4.0 Universal Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Interpore IMZ™                 | 3.3 Hex Cylinder, 3.75 Self-tapping Threaded, 4.0 Hex Cylinder, 4.0 Self-<br>tapping Threaded, 4.25 Hex Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Osstem                         | 4.1 US II, III, II Plus, III Plus, SS II, III (4.8 head)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{7}------------------------------------------------

Sterngold Deztal, LLC Abbrevisted 510(k) Premarket Notification

#### March 13, 2013 SFI Bar® Implant Abutments

| Zimmer Dental                   | 3.5 Bio-Vent® XTM, 3.75 Swede-VentTM Conical Neck CST, 3.75 Swede-<br>VentTM Standard, 4.0 Swede-VentTM Standard, 4.0 Bio-Vent® XTM, 3.25 Micro-<br>Vent® (3.5 head), 3.3 Screw-Vent® (3.5 head), 3.5 Bio-Vent® (3.5<br>head), 3.7 Screw-Vent® (3.5 head), 3.75 Screw-Vent® (3.5 head), 4.3 Core-<br>Vent® (3.5 head), 4.25 Micro-Vent® (4.5 head), 4.5 Bio-<br>Vent® (4.5 head), 4.7 Screw-Vent® (4.5 head), 5.3 Core-Vent® (4.5 head),<br>3.7 Tapered Swiss PlusTM (4.8 platform), 4.8 Tapered Swiss PlusTM 4.1 Straight<br>Swiss PlusTM, 4.8 Straight Swiss PlusTM |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Zimmer (Calcitek®, Centerpulse) | 3.75 ThreadLocTM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Straumann                       | ITI TETM 3.3 (4.8 head), ITI 3.3 Std & Std Plus (4.8 head), ITI TETM 4.1 (4.8 head),<br>ITI 4.1 Std. & Std. Plus (4.8 head), ITI, 4.8 Std. & Std. Plus (4.8 head)                                                                                                                                                                                                                                                                                                                                                                                                     |
| Biolok International            | 4.5 Silhouette Screw, 4.0 Micro-Lok Screw, 4.0 Micro-Lok Cylinder, 3.75 Micro-Lok Screw, 3.3 Micro-Lok Cylinder                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Bud                             | 3.25 Bud Screwvent, 3.75 Bud Screwvent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| INNOVA                          | 4.1 ENDOPORE® External Connection, 4.0 ENTEGRATM External Connection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| OIC                             | 3.0 Osteo Standard ST, 3.25 Osteo Standard ST, 3.75 Osteo Standard ST                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| MIS IMPLANTS                    | 3.3mm Internal Hex, 3.75mm Internal Hex, 4.20mm Internal Hex, 5.0mm Internal Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| BioHorizons®                    | 3.5 Internal, 4.0 Internal, 4.5 Internal, 3.5 Single Stage, 4.5 Single Stage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Implant Direct                  | Legacy 3.5mm, Legacy 4.5mm, RePlantTM 4.3mm, RePlantTM 3.5mm,<br>3.7mm ScrewPlant, 4.7mm ScrewPlant                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Minimatic/Stryker               | 3.3 External Hex Cylinder, 3.75 External Hex Screw, 4.0 External Hex Cylinder,<br>4.0 External Hex Screw, 4.75 External Hex Screw, 5.0 External Hex Cylinder                                                                                                                                                                                                                                                                                                                                                                                                          |

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Part 21 CFR 801 Subparts D)

AND/OR

Over-the -Counter Use (21 CFR 807 Subpart D)

Sheena A. Green -5 2013.07.03 08:48:41 -04/00'

for M. Susan Runner, DDS, MA

(Division Sign-Off) (Division Sign-Off)
Division of Anesthestlology, Genèral Hospital
Division of Anesthestology, Genèral Devices Olvision of Antosthosionology.
Infection Control, Dental Devices k130183

510(k) Number:_

---

**Source:** [https://fda.innolitics.com/device/K130183](https://fda.innolitics.com/device/K130183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
