QLAB QUANTIFICATION SOFTWARE HEART MODEL

K130159 · Philips Health Care · LLZ · May 13, 2013 · Radiology

Device Facts

Record IDK130159
Device NameQLAB QUANTIFICATION SOFTWARE HEART MODEL
ApplicantPhilips Health Care
Product CodeLLZ · Radiology
Decision DateMay 13, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Left Atrial Volume—————External clinical evaluation of Heart Model functionality and performance>1 (clinicians)
Left Ventricular Volume—————External clinical evaluation of Heart Model functionality and performance>1 (clinicians)

Indications for Use

QLAB Quantification Software is a software application package. It is designed to view and quantify image data acquired on Philips Healthcare ultrasound products.

Device Story

QLAB Quantification Software with Heart Model is a software application for ultrasound image review and quantification. It functions as a standalone PC application, on a dedicated workstation, or integrated into Philips ultrasound systems. The Heart Model application performs automatic 3D anatomical segmentation of heart chambers for End Diastole (ED) and End Systole (ES) phases. It identifies chambers and draws borders for standard ASE apical and short-axis views. The user reviews these borders and chooses to accept or reject the automated output. The device generates LA and LV volumes, streamlining cardiac 3D quantification. Clinicians use these measurements to support cardiac assessment. The software benefits patients by providing standardized, efficient chamber quantification, with performance validated against cardiac MRI.

Clinical Evidence

Clinical evaluation involved external validation by clinicians to assess functionality and performance. The study confirmed the device met its intended use and that the automated LA and LV chamber measurements showed correlation to cardiac MRI. No specific statistical metrics (e.g., p-values, CIs) were provided in the summary.

Technological Characteristics

Software-based image processing and quantification application. Operates on standard PCs, dedicated workstations, or on-board ultrasound systems. Utilizes an automated 3D anatomical segmentation algorithm (reused/modified from CT). Supports ASE/ESE standard view displays. Connectivity via standard ultrasound system interfaces. No specific hardware materials or sterilization requirements as it is a software-only package.

Indications for Use

Indicated for clinicians requiring review and quantification of ultrasound image data for cardiac chamber analysis, specifically for obtaining 3D quantitative measurements of the left atrium (LA) and left ventricle (LV) in adult patients.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ QLAB Quantification Software with Heart # MAY 1 3 2013 #### 3. 510(k) Summary · This summary of safety and effectiveness is provided as part of the Premarket Notification in compliance with 21CFR. Part 807 Subpart E, Section 807.92. 1 > Submitter's name, address, telephone number, contact person Philips Ultrasound, Inc. 3000 Minuteman Road Andover, MA Penny Greco, Regulatory Affairs Specialist Telephone: (978) 659-4615 Facsimile: (978) 975-7324 E-mail: penny.greco@philips.com Date prepared: January 21, 2013 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known: | Common/Usual Name: | Picture Archiving and Communications Systems Workstation | |----------------------|----------------------------------------------------------------------------------| | Proprietary Name: | QLAB Quantification Software with Heart Model | | Classification Name: | CFR 892.2050, system, image processing, radiological, Product code LLZ, Class II | ## 3) Substantially Equivalent Devices Philips Ultrasound believes that the OLAB software with the Heart Model application is substantially equivalent to other commercially available products, including Philips QLAB Cardiac 3DQ and 3DQA (K040227 & K042540), Philips Brilliance CT (K042293), and Siemens syngo US Workplace (K091286). #### 4) Device Description The OLAB software application is available either as a stand-alone product that can function on a standard PC, a dedicated workstation, and on-board Philips' ultrasound systems. It can be used for the on-line and off-line review and quantification of ultrasound studies. {1}------------------------------------------------ K130159 Page 2 of 2 1 QLAB Quantification software now includes the Heart Model application. # Heart Model The Heart Model application provides automatic 3D anatomical segmentation and identification of the heart chambers for the End Diastole (ED) and End Systole (ES) cardiac phases. The Heart Model segmentation algorithm draws segmented borders for select standard American Society of Echocardiology (ASE) apical and short axis views. This provides a streamlined workflow for obtaining cardiac 3D quantitative heart chamber measurements and calculated result values. ## 5) Intended Use QLAB Quantification software is a software application package. It is designed to view and quantify image data acquired on Philips Healthcare ultrasound products. # 6) Technological comparison to predicate devices QLAB Heart Model algorithm is reused from Philips Brilliance CT (K042293), but modified for ultrasound. It is similar to QLAB 3DQ/3DQA and the Brilliance CT Comprehensive Cardiac Analysis (CCA) in that the user is able to evaluate the image in the ASE/ESE American Society of Echocardiography and European Society of Echocardiography standard views display. While 3DQ requires the user to place reference points to finish a trace, QLAB Heart Model finds the chambers of the Heart and then displays the Heart Model determined chamber borders in the ASE/ESE views for the user to Accept or Reject. Similar to Siemens syngo US Workplace, QLAB Heart Model automatically generates left atrial (LA) and left ventricular (LV) volumes. #### 7) Non-clinical performance data No performance standards for PACS systems or components have been issued under the authority of Section 514. The Heart Model application was tested in accordance with Philips verification and validation processes. Quality assurance measures were applied to the system design and development, including: - Risk Analysis . - Product Specifications ● - Design Reviews . - Software Verification & Validation ● #### Summary of Clinical Tests: The subject of this premarket submission, QLAB Heart Model, was clinically evaluated and accepted during external validation and determined to be as safe and effective as predicate devices. External clinicians evaluated the functionality and performance of Heart Model. The external evaluation confirmed that Heart Model met its intended use and that the LA and LV chamber measurements correlated to cardiac MRI. 8 {2}------------------------------------------------ 0 # 8) General Safety and Effectiveness Concerns The device labeling contains operating instructions for the safe and effective use of the QLAB software with Heart Model. #### 9) Conclusions The QLAB Quantification Software with the Heart Model application incorporates components common to all image viewing systems for the display and quantification tasks within a clinical setting. Software development for the QLAB software follows documented processes for software design, verification and validation testing. A risk assessment was completed to identify potential design hazards that could cause an error or injury based on the use of the quantification results. Appropriate steps have been taken to control all identified risks for this type of image display and quantification product. Verification and validation activities of the QLAB Quantification software with the Heart Model application indicate that it meets its intended use and introduces no new issues of safety and effectiveness. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 13, 2013 Philips Ultrasound, Inc. % Ms. Penny Greco Regulatory Affairs Specialist Philips Healthcare 3000 Minuteman Road ANDOVER MA 01810 Re: K130159 Trade/Device Name: QLAB Quantification Software Heart Model Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 25, 2013 Received: March 28, 2013 Dear Ms. Greco: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Smh?) for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K130159 Device Name: QLAB Quantification Software Indications for Use: QLAB Quantification Software is a software application package. It is designed to view and quantify image data acquired on Philips Healthcare ultrasound products. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Smh. D) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K130159 510(k)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...