BECKMAN COULTER IMMAGE IMMUNOCHEMISTRY SYSTEM LOW CONCENTRATION IMMUNOGLOBULIN, BECKMAN COULTER
K130122 · Beckman Coulter, Inc. · CFN · Jan 7, 2014 · Immunology
Device Facts
Record ID
K130122
Device Name
BECKMAN COULTER IMMAGE IMMUNOCHEMISTRY SYSTEM LOW CONCENTRATION IMMUNOGLOBULIN, BECKMAN COULTER
Applicant
Beckman Coulter, Inc.
Product Code
CFN · Immunology
Decision Date
Jan 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Indications for Use
IMMAGE® Immunochemistry Systems IGMLC Reagent, when used in conjunction with IMMAGE® Immunochemistry Systems and Cerebrospinal Fluid Protein Calibrator, is intended for quantitative determination of immunoglobulin M in human serum or cerebrospinal fluid (CSF) by rate nephelometry.
Device Story
IMMAGE Immunochemistry System uses rate nephelometry to measure human IgM concentration in serum and CSF. System automatically dilutes and delivers samples and reagents to reaction cuvettes. Particle-bound anti-IgM antibodies bind to IgM, forming insoluble complexes that increase light scatter. 670 nm laser light source; detector at 90° angle measures scatter intensity increase. System calculates rate of change of scatter signal, which is directly proportional to IgM concentration. Used in clinical laboratories by trained personnel. Output provides quantitative IgM levels to assist clinicians in diagnosing CNS conditions related to blood-CSF barrier permeability and intrathecal synthesis.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2 (n=80 per sample pool); total %CV ranged 2.85-15.64% for CSF and 2.85-10.84% for serum. Linearity demonstrated per CLSI EP06-A (r² ≥ 0.997). Method comparison (n=80) against predicate showed high correlation (R² ≥ 0.994 for CSF, R² ≥ 0.995 for serum). LoB, LoD, and LoQ established per CLSI EP17-A.
Indicated for quantitative determination of Immunoglobulin M (IgM) in human serum or cerebrospinal fluid (CSF) for clinical laboratory use. Prescription use only.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
Beckman Coulter IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent (K993547)
Beckman Coulter Cerebrospinal Fluid Protein Calibrator (CSF CAL) (K993547)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130122
B. Purpose for Submission:
Modification to cleared device (k993547)
C. Measurand:
Immunoglobulin M
D. Type of Test:
Quantitative, Nephelometric
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent
IMMAGE® Immunochemistry Systems CSF-CAL Cerebrospinal Fluid Protein Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5510, Immunoglobulins A, G, M, D, and E immunological test system
21 CFR §862.1150, Calibrator
2. Classification:
Class 2
3. Product code:
CFN – Method, Nephelometric, Immunoglobulins (G, A, M)
JIX – Calibrator, Multi-Analyte Mixture
{1}
4. Panel:
Immunology (82)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent:
IMMAGE® Immunochemistry Systems IGMLC Reagent, when used in conjunction with IMMAGE® Immunochemistry Systems and Cerebrospinal Fluid Protein Calibrator, is intended for quantitative determination of Immunoglobulin M (IGMLC) in human serum or cerebrospinal fluid (CSF) by rate nephelometry.
IMMAGE® Immunochemistry Systems CSF-CAL Cerebrospinal Fluid Protein Calibrator:
CSF CAL (Cerebrospinal Fluid Protein Calibrator), when used in conjunction with Beckman Coulter Low Concentration Immunoglobulin A (IGALC) and Low Concentration Immunoglobulin M (IGMLC) reagents is intended for use on IMMAGE for the calibration of these reagents.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
IMMAGE Immunochemistry Systems
I. Device Description:
The IMMAGE® Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent kit includes (a) one IGMLC Cartridge containing IGMLC Particle Reagent (particle bound anti-IgM), Antigen Excess Solution (processed diluted human serum), sodium azide (used as a preservative) and bovine serum albumin and non-reactive chemicals, (b) two Evaporation Caps, and (c) one IGMLC Reagent Bar Code Card.
2
{2}
The Beckman Coulter Cerebrospinal Fluid Protein Calibrator (CSF CAL) components include (a) Cerebrospinal Fluid Protein Calibrator, (b) Cerebrospinal Fluid Protein Calibrator Bar Code Card, (c) Cerebrospinal Fluid Protein Calibrator Bar Code Strips, and (d) Value Assignment Sheet.
# J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
Beckman Coulter IMMAGE Immunochemistry System Low Concentration Immunoglobulin M (IGMLC) Reagent and Beckman Coulter Cerebrospinal Fluid Protein Calibrator (CSF CAL); (k993547)
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device IGMLC Reagent and Calibrator (k130122) | Predicate IGMLC Reagent and Calibrator (k993547) |
| Intended Use: | IGMLC reagent is intended for the quantitative determination of immunoglobulin M (IGMLC) in human serum or cerebrospinal fluid (CSF) by rate nephelometry. CSF CAL is intended for use on IMMAGE for the calibration of these reagents. | Same |
| Instrument | IMMAGE Immunochemistry Systems | Same |
| Technology | Rate nephelometric method | Same |
| Result Output | Quantitative result based on calibration curve | Same |
| Calibrator Formulation | Prepared from human urine to which immunoglobulin A and immunoglobulin M have been added | Same |
| Specimen types | Serum, CSF | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device
IGMLC Reagent and Calibrator (k130122) | Predicate
IGMLC Reagent and Calibrator (k993547) |
| Analytical Sensitivity | CSF: 0.15 mg/L
Serum: 32.4 mg/L | CSF: 0.3 mg/L
Serum: 64.8 mg/L |
| Calibrator traceability | IFCC Reference preparation
ERM-DA470(k)/IFCC | IFCC Reference preparation CRM 470 |
| Analytical Range | CSF 0.15 - 10 mg/L
Serum 32.4 - 2160 mg/L | CSF 0.3 - 10 mg/L
Serum 64.8 - 2160 mg/L |
| Equivalency | Comparison to the Predicate assay | Concordance to IFE method |
| Internal Standards | Two additional internal standard levels were added to the internal standard curve at the low end | |
## K. Standard/Guidance Document Referenced (if applicable):
EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition
EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition
EP09-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition
EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantification, Approved Guideline
EN ISO 17511: In vitro diagnostic medical devices. Measurement of quantities in biological samples. Metrological traceability of values assigned to calibrators and control materials
## L. Test Principle:
The IGMLC test measures the rate of increase in light scattered from particles suspended in solution as a result of complexes formed during an antigen-antibody reaction. A beam of light is passed through the cuvette and, as the antigen-antibody reaction proceeds, the light passing through the cuvette is scattered increasingly as insoluble immune complexes are formed. The light scatter is monitored by measuring the light intensity at an angle away from incident light. The IGMLC Reagent and CSF Calibrator are designed for optimal performance on the IMMAGE Immunochemistry Systems. A curve fit model is applied to the assay result to create parameters that allow each reagent lot to have its own unique stored curve. For a CSF sample, the instrument will use a $50~\mu \mathrm{L}$ sample size and concentration will
{4}
read directly from the stored curve. For a serum sample, the instrument will make 1:39 dilution and pick up $8\mu \mathrm{L}$ of sample and then multiply result by 216.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision testing was performed in accordance with CLSI EP5-A2. Three separate serum sample pools representing low (32.4 - 43.4 mg/L; Serum-1), mid (432 - 1296 mg/L; Serum-2) and high (1728 - 2160 mg/L; Serum-3) range, and CSF samples pools representing low (0.15 - 0.20 mg/L; CSF-1), mid (2 - 6 mg/L; CSF-2) and high (8 - 10 mg/L; CSF-3) range were included in the study. The experimental design utilized duplicate sample analysis, twice daily, over the course of twenty working days $(n = 80)$ . The data were analyzed for within-run, between-run, between-day, and total precision and the mean U/mL and percent coefficient of variation (%CV) are summarized below.
| Sample | N | Mean) (mg/L) | Within-Run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CSF-1 | 80 | 0.196 | 0.03 | 16.55% | 0 | 0% | 0.01 | 4.57% | 0.03 | 15.64% |
| CSF-2 | 80 | 2.49 | 0.05 | 2.07% | 0.06 | 2.22% | 0 | 0% | 0.07 | 2.85% |
| CSF-3 | 80 | 8.81 | 0.29 | 3.25% | 0.29 | 3.32% | 0 | 0% | 0.38 | 4.30% |
| Serum-1 | 80 | 38.5 | 3.84 | 9.99% | 0.69 | 1.80% | 1.46 | 3.81% | 4.17 | 10.84% |
| Serum-2 | 80 | 1250.1 | 28.87 | 2.31% | 21.33 | 1.71% | 0 | 0% | 35.59 | 2.85% |
| Serum-3 | 80 | 1935.8 | 55.23 | 2.85% | 29.58 | 1.53% | 10.23 | 0.53% | 63.48 | 3.28% |
# b. Linearity/assay reportable range:
Linearity studies were conducted in accordance with CLSI Guideline EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition. Two sample sources, including internal standards traceable to ERM-DA470k/IFCC and patient sample pools, were used to demonstrate linearity. Each sample and dilution was evaluated in replicates of five.
The linear regression parameters (slope, intercept and $\mathbf{r}^2$ ) of the observed values vs. predicted values are shown below.
| Sample type | Sample | Dilution range (mg/L) | Slope | Intercept | r2 |
| --- | --- | --- | --- | --- | --- |
| Serum | Low-1 | 29.70 – 297.01 | 0.977 | 6.918 | 0.999 |
| | Low-2 | 29.72 – 297.20 | 0.999 | 2.340 | 0.997 |
| | Low-3 | 30.51 – 305.12 | 0.970 | 5.538 | 0.999 |
| | High-1 | 341.40 – 1707.02 | 1.033 | 34.837 | 0.998 |
{5}
| Sample type | Sample | Dilution range (mg/L) | Slope | Intercept | r2 |
| --- | --- | --- | --- | --- | --- |
| | High-2 | 354.68 – 1773.39 | 1.019 | 22.365 | 0.999 |
| | High-3 | 386.35 – 1931.75 | 1.034 | 58.827 | 0.999 |
| CSF | Low-1 | 0.13 – 1.33 | 1.073 | -0.110 | 0.997 |
| | Low-2 | 0.15 – 1.52 | 1.012 | -0.030 | 0.999 |
| | Low-3 | 0.15 – 1.50 | 1.006 | -0.011 | 0.999 |
| | High-1 | 1.74 – 8.70 | 0.986 | -0.002 | 0.998 |
| | High-2 | 1.73 – 8.63 | 0.985 | 0.088 | 0.999 |
| | High-3 | 1.74 – 8.72 | 0.992 | 0.033 | 0.999 |
The claimed reportable Analytical Measuring range is shown in the table below:
| IgM Assay | Assay reportable range |
| --- | --- |
| Serum | 32.4 – 2160 mg/L |
| CSF | 0.15 – 10 mg/L |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The IMMAGE system uses a stored response curve that is created during manufacture by diluting single-level internal standard which is traceable to the IFCC reference preparation to plasma protein, material ERM-DA470k/IFCC. The traceability process is based on EN ISO 17511. The curve is established using multiple instruments. A curve fit model is applied to the assay result to create parameters that allow each reagent lot to have its own unique stored curve. The single point calibrator is also value assigned using the same internal standard that is used for the stored curve.
Reagent Stability: No change (referred to k993547).
d. Detection limit:
Limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were determined in accordance with CLSI EP17-A.
LoB: For serum, four patient sample pools created from IgM negative sera were assayed daily in replicates of 5 for a period of three days (total 60 data points) on two instruments. For CSF, CSF diluent was assayed daily in replicates of 5 for a period of three days (total 15 data points) on two instruments. Datasets were ranked from lowest to highest for these blank readings and the 95th percentile reading was determined from this ranking.
LoD: For serum, four pools created from a serum patient sample and IgM depleted human serum at a mean value $\leq 32.4\mathrm{mg / L}$ and assayed daily in replicates of 5 for a
{6}
period of three days (total 60 data points) on two instruments. For CSF, four pools patient sample pools at a mean value $\leq 0.15\mathrm{mg / L}$ created from a CSF patient sample pool and CSF diluent were assayed daily in replicates of 5 for a period of three days (total 60 data points) on two instruments.
LoQ: In order to establish LoQ for serum and CSF, three low-concentration IgM pools were run over 2 reagents lots and 2 instruments. For serum, there were a total of 45 replicates, 3 reps per pool, 3 pools, and 5 runs. For CSF, there were a total of 40 replicates from 3 pools: 3 reps for two of the pools and 2 reps for the third pool, and 5 total runs. The LoQ samples were created from an internal standard with a known value.
| IgM assay | LoB (mg/L) | LoD (mg/L) | LoQ (mg/L) |
| --- | --- | --- | --- |
| Serum | 7.46 | 18.05 | 32.4 |
| CSF | 0 | 0.0586 | 0.13 |
e. Analytical specificity:
Established in k993547.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison and bias estimation experiments were designed using CLSI guideline EP09-A2. The testing compared IGMLC pre-design change lots and IGMLC design changed lots. Eighty (80) CSF and serum specimens with IgM concentrations spanning the analytical ranges of the two sample types were obtained from commercial sources, including paired CSF and serum patient samples from clinical laboratories and commercial sources. About $5\%$ CSF samples (4 of 80 samples) and $12.5\%$ of the serum samples (10 of 80 samples) were contrived. The method comparison samples were tested internally over multiple days; one run per day with eight of each of the sample types (CSF and Serum) per day, over a period of 10 days. The IgM values in CSF samples ranged from 0.150 to $8.982\mathrm{mg / mL}$ while IgM values in serum samples ranged from 34.73 to $2,156.19\mathrm{mg / mL}$ (by the new method). Deming regression analysis was performed to obtain slope, $\mathbf{R}^2$, intercept and a scatter plot of data. Results are summarized in tables below.
{7}
CSF samples:
| n | Slope (95% CI) | Y-Intercept (95% CI) | R² |
| --- | --- | --- | --- |
| 80 | 0.963 (0.946 to 0.980) | -0.133 (-0.190 to -0.076) | 0.994 |
Serum samples:
| n | Slope (95% CI) | Y-Intercept (95% CI) | R² |
| --- | --- | --- | --- |
| 80 | 1.048 (1.031 to 1.065) | 5.695 (-12.7 to 24.117) | 0.995 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The reference interval values of 429 to 1893 mg/L for serum IGMLC were established using a population of 130 apparently healthy male and female adults from California. The reference interval of 0.19 – 0.29 mg/L for CSF was taken from published literature [Burtis, C. A., Ashwood, E. R., Tietz, Texbook of Clinical Chemistry, 3rd Edition, W. B. Saunders, Philadelphia, PA (1999)].
{8}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.