The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits. The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel.
Device Story
Modular sidestream/mainstream CO2 analyzer; inserts into Spacelabs Healthcare Ultraview SL or XPREZZON monitors. Uses infrared sensor technology to measure CO2 concentration and respiratory rate. Monitor provides user interface, numeric displays, and capnograph waveform. Used primarily in operating room environments by qualified medical personnel. Alerts clinicians when parameters fall outside user-defined limits; facilitates patient monitoring and clinical decision-making regarding respiratory status.
Clinical Evidence
Bench testing only. Performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, and ISO 21647 standards. Software verification and validation conducted per IEC 62304 and FDA guidance. No clinical data presented.
Technological Characteristics
Infrared sensor-based gas analyzer. Supports both sidestream and mainstream sampling. Modular form factor for integration with Spacelabs monitors. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 21647 (respiratory gas monitors).
Indications for Use
Indicated for monitoring carbon dioxide (CO2) and respiration rate in adult, pediatric, and neonate patients. Used under the direction of qualified medical personnel to alert clinicians when concentrations exceed user-defined limits.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
Spacelabs Healthcare Capnography Pod (92516) (K121017)
Submission Summary (Full Text)
{0}------------------------------------------------
## んに30112
# 510(k) Summary
ﺎ
## MAR 2 1 2013
·
t;
| Submission Date: | 08 January 2013 | | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------------------------------|
| Submitter: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | |
| Submitter Contact: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029<br>Mr. David J. Geraghty<br>Spacelabs Healthcare<br>Phone: +1 (425) 657-7200, ext 5889<br>Fax: +1 (425) 657-7210<br>Email: david.geraghty@spacelabs.com | | |
| Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 | | |
| Manufacturing Site: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | |
| Trade Name: | Spacelabs Healthcare Capno Module, 92517 | | |
| Common Name: | CO2 monitor | | |
| Classification Name: | Carbon dioxide gas analyzer | | |
| Classification<br>Regulation: | 21 CFR §868.1400 | | |
| Product Code: | CCK | | |
| Substantially<br>Equivalent Devices: | New Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
| | Spacelabs Healthcare<br>Capno Module, 92517 | K121017 | Spacelabs Healthcare<br>Capnography Pod<br>(92516) |
ﮨﮯ-ﺎ
{1}------------------------------------------------
## 510(k) Summary
The Spacelabs Healthcare Capno Module, 92517 (92517) is an easy-to-Device Description: use modular unit used with Spacelabs Healthcare Ultraview SL or XPREZZON monitors. The 92517 is inserted into the bay within the monitors. which is then used to control the 92517, and provide the user interface for the 92517.
> The 92517 is a sidestream or mainstream analyzer intended to provide a measurement of the following parameters: carbon dioxide (CO2); and respiratory rate.
The monitor provides a number display for CO2 and respiratory rate, and a capnograph waveform. The 92517 is intended to be used primarily in the operating room environment.
Intended Use: The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits.
> The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel.
Technology Comparison: . The 92517 employs the same technological characteristics as the predicate device.
| Characteristic | Predicate Device | Proposed Device |
|---------------------------|------------------------------------------------|---------------------------|
| Parameters | Carbon dioxide (CO2); and<br>respiratory rate. | Same |
| EtCO2 | Yes | Same |
| FiCO2 | Yes | Same |
| Measurement<br>Technology | Infrared Sensor | Same |
| Sampling<br>Technique | Sidestream | Sidestream and mainstream |
{2}------------------------------------------------
## 510(k) Summary
#### Summary of Performance Testing:
The 92517 was tested for performance in accordance with the following Electrical Safety Standards: IEC 60601-1: 1988, Am1: 1991, Am2: 1995, Medical electrical . equipment - Part 1. General requirements for safety; and UL 60601-1: 2003, Medical electrical equipment - Part 1. General . requirements for safety. Test results indicated that the 92517 complies with the Standards. The 92517 was tested for performance in accordance with the following Electromagnetic Compatibility (EMC) Standard: Testing IEC 60601-1-2: 2007, Medical electrical equipment -- Part 1-2: . General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility -- Requirements and tests. Test results indicated that the 92517 complies with the Standards. Software device modifications made to the 92517 were designed and Software Testing developed according to a robust software development process, and were rigorously verified and validated. Software information is provided in accordance with internal documentation and the following Standards and guidance documents: . . FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; FDA guidance: Off-the-shelf software use in medical devices, 09 . Sep 99; . FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; IEC 60601-1-4: 2000, Medical electrical equipment Medical . electrical equipment - Part 1-4: General requirements for safety -Collateral Standard: Programmable electrical medical systems; and . IEC 62304: 2006, Medical device software -- Software life cycle processes. Test results indicate that the 92517 complies with its predetermined
specification and the Standards and guidance documents.
{3}------------------------------------------------
## 510(k) Summary
#### Performance Testing
The 92517 was tested for performance in accordance with internal documentation and the following Standards:
- IEC 60601-1-8: 2006, Medical electrical equipment Part 1-8: . General requirements for basic safety and essential performance -Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems; and
- . ISO 21647: 2004, Medical electrical equipment – Particular requirements for the basic safety and essential performance of respiratory gas monitors.
Test results indicated that the 92517 complies with its predetermined specification and with the applicable Standards.
Conclusion
Verification and validation activities were conducted to establish the performance and safety characteristics of the software device modifications made to the 92517. The results of these activities demonstrate that the 92517 is safe and effective when used in accordance with its intended use and labeling.
Therefore, the 92517 is considered substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is centered and appears to be the heading of a document or publication. The words are stacked on a single line.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### March 21, 2013
Spacelabs Healthcare C/O Mr. Thomas Kroenke Principal Consultant Speed To Market, Incorporated P.O. Box 3018 NEDERLAND CO 80466
Re: K130112
Trade/Device Name: Spacelabs Healthcare Capno Module, 92517 Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: February 14, 2013 Received: February 22, 2013
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/5/Figure/7 description: The image shows the text "Kwame O. Ulmer for". The text is in black and white, and the font is simple and easy to read. The letters "O.U" are stylized with a geometric pattern. The word "for" is smaller than the rest of the text.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name:
Indications for Use:
K 130 112
Spacelabs Healthcare Capno Module, 92517
The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits.
The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis 2013.03.12 12:11:55 =04'00'
(Division Sign-Off) (Division Sign-On)
Division of Anestheslology, General Hospital Infection Control, Dental Devices
1130112 510(k) Number.
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.