CAPNO MODULE, 92517

K130112 · Spacelabs Healthcare · CCK · Mar 21, 2013 · Anesthesiology

Device Facts

Record IDK130112
Device NameCAPNO MODULE, 92517
ApplicantSpacelabs Healthcare
Product CodeCCK · Anesthesiology
Decision DateMar 21, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.1400
Device ClassClass 2
AttributesPediatric

Intended Use

The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits. The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel.

Device Story

Modular sidestream/mainstream CO2 analyzer; inserts into Spacelabs Healthcare Ultraview SL or XPREZZON monitors. Uses infrared sensor technology to measure CO2 concentration and respiratory rate. Monitor provides user interface, numeric displays, and capnograph waveform. Used primarily in operating room environments by qualified medical personnel. Alerts clinicians when parameters fall outside user-defined limits; facilitates patient monitoring and clinical decision-making regarding respiratory status.

Clinical Evidence

Bench testing only. Performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, and ISO 21647 standards. Software verification and validation conducted per IEC 62304 and FDA guidance. No clinical data presented.

Technological Characteristics

Infrared sensor-based gas analyzer. Supports both sidestream and mainstream sampling. Modular form factor for integration with Spacelabs monitors. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 21647 (respiratory gas monitors).

Indications for Use

Indicated for monitoring carbon dioxide (CO2) and respiration rate in adult, pediatric, and neonate patients. Used under the direction of qualified medical personnel to alert clinicians when concentrations exceed user-defined limits.

Regulatory Classification

Identification

A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## んに30112 # 510(k) Summary ﺎ ## MAR 2 1 2013 · t; | Submission Date: | 08 January 2013 | | | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------------------------------| | Submitter: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | | | Submitter Contact: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029<br>Mr. David J. Geraghty<br>Spacelabs Healthcare<br>Phone: +1 (425) 657-7200, ext 5889<br>Fax: +1 (425) 657-7210<br>Email: david.geraghty@spacelabs.com | | | | Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 | | | | Manufacturing Site: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | | | Trade Name: | Spacelabs Healthcare Capno Module, 92517 | | | | Common Name: | CO2 monitor | | | | Classification Name: | Carbon dioxide gas analyzer | | | | Classification<br>Regulation: | 21 CFR §868.1400 | | | | Product Code: | CCK | | | | Substantially<br>Equivalent Devices: | New Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model | | | Spacelabs Healthcare<br>Capno Module, 92517 | K121017 | Spacelabs Healthcare<br>Capnography Pod<br>(92516) | ﮨﮯ-ﺎ {1}------------------------------------------------ ## 510(k) Summary The Spacelabs Healthcare Capno Module, 92517 (92517) is an easy-to-Device Description: use modular unit used with Spacelabs Healthcare Ultraview SL or XPREZZON monitors. The 92517 is inserted into the bay within the monitors. which is then used to control the 92517, and provide the user interface for the 92517. > The 92517 is a sidestream or mainstream analyzer intended to provide a measurement of the following parameters: carbon dioxide (CO2); and respiratory rate. The monitor provides a number display for CO2 and respiratory rate, and a capnograph waveform. The 92517 is intended to be used primarily in the operating room environment. Intended Use: The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits. > The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel. Technology Comparison: . The 92517 employs the same technological characteristics as the predicate device. | Characteristic | Predicate Device | Proposed Device | |---------------------------|------------------------------------------------|---------------------------| | Parameters | Carbon dioxide (CO2); and<br>respiratory rate. | Same | | EtCO2 | Yes | Same | | FiCO2 | Yes | Same | | Measurement<br>Technology | Infrared Sensor | Same | | Sampling<br>Technique | Sidestream | Sidestream and mainstream | {2}------------------------------------------------ ## 510(k) Summary #### Summary of Performance Testing: The 92517 was tested for performance in accordance with the following Electrical Safety Standards: IEC 60601-1: 1988, Am1: 1991, Am2: 1995, Medical electrical . equipment - Part 1. General requirements for safety; and UL 60601-1: 2003, Medical electrical equipment - Part 1. General . requirements for safety. Test results indicated that the 92517 complies with the Standards. The 92517 was tested for performance in accordance with the following Electromagnetic Compatibility (EMC) Standard: Testing IEC 60601-1-2: 2007, Medical electrical equipment -- Part 1-2: . General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility -- Requirements and tests. Test results indicated that the 92517 complies with the Standards. Software device modifications made to the 92517 were designed and Software Testing developed according to a robust software development process, and were rigorously verified and validated. Software information is provided in accordance with internal documentation and the following Standards and guidance documents: . . FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; FDA guidance: Off-the-shelf software use in medical devices, 09 . Sep 99; . FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; IEC 60601-1-4: 2000, Medical electrical equipment Medical . electrical equipment - Part 1-4: General requirements for safety -Collateral Standard: Programmable electrical medical systems; and . IEC 62304: 2006, Medical device software -- Software life cycle processes. Test results indicate that the 92517 complies with its predetermined specification and the Standards and guidance documents. {3}------------------------------------------------ ## 510(k) Summary #### Performance Testing The 92517 was tested for performance in accordance with internal documentation and the following Standards: - IEC 60601-1-8: 2006, Medical electrical equipment Part 1-8: . General requirements for basic safety and essential performance -Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems; and - . ISO 21647: 2004, Medical electrical equipment – Particular requirements for the basic safety and essential performance of respiratory gas monitors. Test results indicated that the 92517 complies with its predetermined specification and with the applicable Standards. Conclusion Verification and validation activities were conducted to establish the performance and safety characteristics of the software device modifications made to the 92517. The results of these activities demonstrate that the 92517 is safe and effective when used in accordance with its intended use and labeling. Therefore, the 92517 is considered substantially equivalent to the predicate device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is centered and appears to be the heading of a document or publication. The words are stacked on a single line. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### March 21, 2013 Spacelabs Healthcare C/O Mr. Thomas Kroenke Principal Consultant Speed To Market, Incorporated P.O. Box 3018 NEDERLAND CO 80466 Re: K130112 Trade/Device Name: Spacelabs Healthcare Capno Module, 92517 Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: February 14, 2013 Received: February 22, 2013 Dear Mr. Kroenke: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Image /page/5/Figure/7 description: The image shows the text "Kwame O. Ulmer for". The text is in black and white, and the font is simple and easy to read. The letters "O.U" are stylized with a geometric pattern. The word "for" is smaller than the rest of the text. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Indications for Use 510(k) Number (if known): Device Name: Indications for Use: K 130 112 Spacelabs Healthcare Capno Module, 92517 The Capno Module, 92517 (92517) is intended to provide a means of monitoring carbon dioxide and respiration rate and alert clinical personnel when the concentration moves outside of user-defined limits. The 92517 is intended to be used with and controlled by a Spacelabs Healthcare monitors. The 92517 is intended to be used for monitoring adult, pediatric and neonate patients, under the direction of qualified medical personnel. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Lester W. Schultheis 2013.03.12 12:11:55 =04'00' (Division Sign-Off) (Division Sign-On) Division of Anestheslology, General Hospital Infection Control, Dental Devices 1130112 510(k) Number. Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%