← Product Code [ONF](/productcode/ONF) · K130015

# SHASER V-MINI HAIR REMOVAL SYSTEM (K130015)

_Shaser, Inc. · ONF · Jun 13, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130015

## Device Facts

- **Applicant:** Shaser, Inc.
- **Product Code:** [ONF](/productcode/ONF.md)
- **Decision Date:** Jun 13, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The SHASER HRS3 hair removal system is an over the counter device intended for removal of unwanted hair. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

## Device Story

Shaser V-MINI is a cordless, rechargeable, light-based hair removal system for OTC home use. Device utilizes pulsed light energy to target hair follicles for permanent reduction. Operated by consumer via fingerswitch; includes limited-life treatment head and battery charger. Output is light energy pulse; clinical benefit is long-term stable reduction of regrowing hair. Safety and efficacy are managed by user adherence to instructions for use.

## Clinical Evidence

No clinical data provided. Evidence consists of human factors/usability testing: 150 subjects for label comprehension and 123 subjects for usability, including low-literacy populations. Results confirmed sufficient comprehension and safe, appropriate use in a simulated OTC home environment.

## Technological Characteristics

Powered light-based non-laser surgical instrument (Product Code ONF). Cordless, rechargeable, 0.45 kg, 2.1 x 0.6 x 0.8 cm. Electrical requirement: 115 VAC, 15A, 50-60 Hz charger. Fingerswitch activation. Limited-life treatment head.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Shaser HRS2 Hair Removal System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

. .

## 510(K) Summary

・・

- .

| Shaser, Inc.<br>10 Maguire Road<br>JUN 1 3 2013<br>Building I<br>Lexington, MA 02421                                                                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Anthony Burns<br>Sr. Director of Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| December 31, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Shaser V-MINI Hair Removal System                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Light based hair removal device                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Powered Light Based Non-Laser Surgical Instrument with Thermal<br>Effect. Product Code ONF                                                                                                                                                                                                                                                                                                                                                                                                           |
| Shaser HRS2 Hair Removal System                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Over-The-Counter, Cordless, Rechargable, Personal Light-Based Hair<br>Removal System For Permanent Hair Reduction                                                                                                                                                                                                                                                                                                                                                                                    |
| Emission activation is by fingerswitch. Includes limited life treatment<br>head and battery charger. Overall weight of the device is 0.45 Kg, and<br>the size is 2.1 x 0.6 x 0.8 cm (HxWxD).                                                                                                                                                                                                                                                                                                         |
| Charger Electrical Requirement is 115 VAC, 15A, 50-60 Hz, single<br>phase.                                                                                                                                                                                                                                                                                                                                                                                                                           |
| The SHASER V-MINI hair removal system is an over the counter<br>device intended for removal of unwanted hair. It is also intended for<br>permanent reduction in unwanted hair. Permanent hair reduction is<br>defined as the long-term stable reduction in the number of hairs<br>regrowing when measured at 6, 9, and 12 months after the completion<br>of a treatment regimen.                                                                                                                     |
| The Shaser V-MINI Hair Removal System has the same indication for<br>uses, the same principle of operation, the same pulse energy range and<br>wavelength range as the predicate device.                                                                                                                                                                                                                                                                                                             |
| Nonclinical Performance Data:<br>none                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Label comprehension and usability test of consumers' ability to<br>understand the instructions for use and to evaluate their ability to use<br>the device safely in a simulated OTC home-use environment.<br>· 150 study subject were tested for label comprehension and 123<br>study subjects tested for usability. Both test populations included<br>low literacy subjects.<br>The results of the two tests confirms sufficient label comprehension<br>and safe and appropriate use of the device. |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{1}------------------------------------------------

1. 1. 1. 1.

Conclusion:

The Shaser V-MINI Hair Removal System is a safe and effective device for the intended uses.

:

.

Additional Information:

None

{2}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

June 13, 2013

Shaser, Inc. % Ms. Sarah Hackett 10 Maguire Road, Suite 120 Building One Lexington, Massachusetts 02421

Re: K130015

Trade/Device Name: Shaser V-Mini Hair Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and Plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: May 06, 2013 Received: May 08, 2013

Dear Ms. Hackett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Sarah Hackett

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours,

## David Krause -S

- for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: _ Shaser V-MINI Hair Removal System

Indications For Use:

The SHASER HRS3 hair removal system is an over the counter device intended for removal of unwanted hair. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

Prescription Use (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use V (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Neil R Ogden 2013.06.11 16:24:42 -04'00'

(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K130015](https://fda.innolitics.com/device/K130015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
