JUNG RIM SINGLE USE HYPODERMIC SYRINGE WITH NEEDLE

K124037 · Jung Rim Medical Industrial Co., Ltd. · FMF · Sep 12, 2013 · General Hospital

Device Facts

Record IDK124037
Device NameJUNG RIM SINGLE USE HYPODERMIC SYRINGE WITH NEEDLE
ApplicantJung Rim Medical Industrial Co., Ltd.
Product CodeFMF · General Hospital
Decision DateSep 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Intended Use

Jung Rim single use hypodermic syringe with Needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.

Device Story

Single-use, sterile, disposable hypodermic syringe with needle; consists of barrel, piston, gasket, needle, and needle cap. Available in various sizes (1-50 mL) and needle gauges (16G-31G). Used by clinicians or patients for fluid injection or aspiration. Operation involves manual filling by pulling the plunger, removing air bubbles by tapping the barrel, and depressing the plunger to deliver the dose. Benefits include sterile, single-use delivery of fluids. No electronic or software components.

Clinical Evidence

No clinical data. Bench testing only. Device performance validated against ISO 7886-1, ISO 7864, ISO 9626, ISO 6009, and ISO 594-1. Biocompatibility testing per ISO 10993-1. Sterilization validated per ISO 11135.

Technological Characteristics

Piston-style syringe with needle. Materials: lubricated polypropylene barrel, polypropylene plunger, styrene ethylene/butylene copolymer gasket, stainless steel (STS304) needle, polypropylene hub/sheath. Bonding: Epoxy. Lubricant: Dimethylpolysiloxane. Sterilization: Ethylene oxide (SAL 10⁻⁶). Connectivity: None. Software: None.

Indications for Use

Indicated for medical purposes to inject fluid into or withdraw fluid from the body in patients requiring hypodermic injection or aspiration.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUNG RIM MEDICAL # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS ### 1. Submitted by : September /02 / 2013 Kyu Lee / Overseas sales manager Jung Rim Medical Industrial Co., Ltd. #593-81, Sagok-Ri, Iwol-Myeon, Jincheon-Gun, Chungbuk, 365-820, Korea Phone : +82 2 404 7986 Fax : + 82 2 404 7989 #### Device Name : 2. Trade Name Jung Rim single use hypodermic Syringe with Needle . - � Piston syringe - Classification : Class II - Classification Name Syringe, Piston - Product Code FMF .. - Regulation Number 21 CFR 880.5860 : : - Review Panel General Hospital : SEP 12 2013 - � Hypodermic single lumen needle | - Classification | : Class II | |-----------------------|----------------------------------| | - Classification Name | : Hypodermic single lumen needle | | - Product Code | : FMI | | - Regulation Number | : 21 CFR 880.5870 | | - Review Panel | : General Hospital | ### 510(k) Summary {1}------------------------------------------------ ### 3. Predicate Device : ### a) K number : K113091 - Manufacturer : Jiangyin Caina Technology Co., Ltd. I ) Proposed Device Name: Syringes with or without needles Classification: Class II Product Code: FMF Regulation Number: 21 CFR 880.5860 2) Proposed Device Name: Needles Classification: Class II Product Code: FMI Regulation Numiber: 21 CFR 880.5570 ### b) K number : K072739 - Manufacturer : ShanDong WeiGao Group Medical Polymer Products Co., LTD 1) Proposed Device Name: Sterile Hypodermic Syringe for single use, with/without needle (1,2,3,5,10,20,30,50,100(ml)) Classification: Class II Product Code: FMF Regulation Number: 21 CFR 880.5860 2) Proposed Device Name: Sterile Hypodermic Needle for single use (16G,18G,19G,20G,21G,22G,23G,24G,25G,26G,27G,29G) Classification: Class H Product Code: FMI Regulation Numiber: 21 CFR 880.5570 {2}------------------------------------------------ ### 4. Device Description : The Single use hypodermic Syringe with Needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. The product consists of a Barrel, Piston, Gasket, Needle Needle cap. The Single use hypodermic Syringe with Needle are offered various sizes (Luer slip (1.3.5,10,20,30,50 (ml)), and Needle size (16G, 17G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G, 31G) Proposed syringes will be provided along with a proposed needle in one single package. The combinations of sizes are various upon the request of the users. These are sterile (Eo gas sterilization). The devices are disposable, single use devices. ### 5. Intended For Use : 1) The single us hypodermic syringe The Single use hypodermic Syringe with needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. ### 2) The Hypodermic needle The Sterile hypodermic needle for single use is intended for use with syringes for general purpose fluid injection/aspiration Proposed syringes will be provided along with a proposed needle in one single package. The combinations of sizes are various upon the request of the users. JUNG RIM MEDICAL {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the text "JUNG RIM MEDICAL" in a bold, sans-serif font. The text is black, and there is a horizontal line underneath it. To the left of the text is a logo that is partially visible. ## 6. Technological Characteristics : The Single use hypodermic Syringe with needle and the predicate device have the identical technological characteristics and perform equivalently. | Device Name | Subject Device | Predicate Device #1 | Predicate Device #2 | | | |-------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | Manufacturer | Jung Rim Medical Industrial Co.,<br>Ltd. | Jiangyin Caina Technology Co.,<br>Ltd. | ShanDong WeiGao Group Medical<br>Polymer Products Co., LTD | | | | 510(k) Number | NEW | K113091 | K072739 | | | | Classification | Class II | Class II | Class II | | | | Product code | FMF | FMF | FMF | | | | Regulation Number | 21 CFR 880.5860 | 21 CFR 880.5860 | 21 CFR 880.5860 | | | | Classification | Class II | Class II | Class II | | | | Product code | FMI | FMI | FMI | | | | Regulation Number | 21 CFR 880.5870 | 21 CFR 880.5870 | 21 CFR 880.5870 | | | | Intended for use | Syringe | The Single use hypodermic Syringe<br>with needle is intended to be used<br>for medical purposes to inject fluid<br>into or withdraw fluid from body. | Syringes with or without needle<br>are intended to inject fluids into or<br>withdraw fluids from the body. | The sterile Hypodermic syringe for<br>single Use With/without needle is<br>intended to be used for medical<br>purposes to inject fluid into or<br>withdraw fluid from body. | | | | Needles | The Sterile hypodermic needle for<br>single use is intended for use with<br>syringes for general purpose fluid<br>injection/aspiration. | Needles are intended to be used<br>with a legally marketed syringe<br>intend to inject fluids into or<br>withdraw fluids from the body. | The Sterile Hypodermic needle for<br>single use is intended for use with<br>syringes and injection devices for<br>general purpose fluid<br>injection/aspiration. | | | Syringe | Syringe type | Piston syringe | Piston syringe | Piston syringe | | | | Tip type | Luer slip | Luer slip and Luer lock | Luer slip and Luer lock | | | | Volume | 1ml 3ml 5ml 10ml 20ml 30ml 50ml | 1,2,3,5,10,20,30,50 (ml) | 1,2,3,5,10,20,30,50,100(ml) | | | Needle | Gauge | 16G, 17G, 18G, 19G, 20G, 21G,<br>22G,23G, 24G, 25G, 26G, 27G, 28G,<br>29G, 30G, 31G | 16G,18G,19G,20G,21G,22G,23G,<br>24G,25G,26G,27G,29G | 16G,18G,19G,20G,21G,22G,23G,<br>24G,25G,26G,27G,29G | | | | Tip<br>configuration | Image: tip configuration | same | same | | | | Hub color | According to ISO 6009 | According to ISO 6009 | According to ISO 6009 | | | | | Device Name | Subject Device | Predicate Device #1 | Predicate Device #2 | | | | Manufacturer | Jung Rim Medical Industrial Co.,<br>Ltd. | Jiangyin Caina Technology Co.,<br>Ltd. | ShanDong WeiGao Group Medical<br>Polymer Products Co., LTD | | | | 510(k) Number | NEW | K113091 | K072739 | | | Syringe | Barrel | lubricated polypropylene | lubricated polypropylene | lubricated polypropylene | | | | Plunger | polypropylene | polypropylene | polypropylene | | | | Gasket | Styrene ethylene / Butylene<br>copolyme | Resin stopper | • Resin stopper | | | | Tubing | lubricated stainless (STS304) | lubricated stainless (STS304) | lubricated stainless (STS304) | | | Needle | Hub | polypropylene | polypropylene | polypropylene | | | | Sheath | polypropylene | polypropylene | polypropylene | | | | Bonding | Epoxy bond | Epoxy bond | Epoxy bond | | | | Lubricant | Dimethylpolysiloxane | Dimethylpolysiloxane | Dimethylpolysiloxane | | | | Principle of<br>operation | 1. Check that the syringe package is<br>undamaged and unopened.<br>Discard if damaged or opened.<br>2. Peel open the end of the package<br>and remove the syringe.<br>3. Wipe the top of the vial with<br>alcohol and allow it to air dry.<br>4. Remove the shield from the<br>needle.<br>Do not move the plunger, and do<br>not try to inject air into the vial, as<br>this will disable the syringe.<br>5. Pick up the vial and invert it.<br>Insert needle into the vial.<br>6. Keep the needle tip in the fluid at<br>all times.<br>Do not draw air into the syringe.<br>It may lead to an inaccurate dose.<br>To fill the syringe, slowly pull<br>the plunger<br>To remove air bubbles, remove<br>the needle from the vial.<br>7. Holding the syringe upright, tap<br>the barrel to bring the bubbles<br>towards the tip of the syringe.<br>8. Then carefully push the plunger<br>to the dose mark, thus expelling<br>the air bubble.<br>9. Administer the injection<br>10. Depress the plunger completely<br>to deliver the dose. | Same | Same | | | | Biocompatibility | ISO10993-1 | ISO10993-1 | ISO10993-1 | | | | Sterilization | Ethylene oxide gas (SAL 10⁻⁶) | Ethylene oxide gas (SAL 10⁻⁶) | Ethylene oxide gas (SAL 10⁻⁶) | Page 4 of 6 {4}------------------------------------------------ ## Jung Rim Syringe 510(K) Submission . ・ 7 - 1 JUNG RIM MEDICAL 510(k) Summary · Page 5 of 6 י {5}------------------------------------------------ # JUNG RIM MEDICAL ### Performance Testing The Single use hypodermic Syringe with needle have been designed and successfully tested to meet the applicable requirements outlined in ISO7886-1, ISO7864, ISO9626, ISO6009 and ISO594-1. ### Biocompatibility Testing The material of the Single use hypodernic Syringe with needle have successfully passed testing as outlined in ISO10993-1 for devices categorized as External communicating devices, Limited exposure. ### Sterilization and Shelf-life Testing Sterilization of the Single use hypodermic Syringe with needle has been validated using the half-cycle method as outlined in ISO11135. The maximum levels of residues of ethylene oxide and ethylene chlorohydrins will not exceed the limits presented in ISO10993-7. Shelf-life testing supports a shelf-life of 5-years after sterilization ### Clinical Data No prospective clinical trials were conducted in support of this Traditional 510(k) #### 8. Conclusion Based on the information provided in this premarket notification of Jung Rim Medical Industrial Co., Ltd. Concludes that The Sterile Hypodermic Syringe for Single Use with needle is substantially equivalent to predicate devices Page 6 of 6 {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(19 Silver Spring, MI) 20993-0002 ### September 12, 2013 Jung Rim Medical Industrial Company Limited C/O Mr. Kyu Lec 593-81. Sagok-Ri. Iwol-Myeon, Jincheon-Gun Chungbuk 365-820 Korea Re: K124037 Trade/Device Name: Jung Rim Single Use Hypodermic Syringe with Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF, FMF, FMI Dated: August 14, 2013 Received: August 15. 2013 Dear Mr. Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it mav be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {7}------------------------------------------------ Page 2 - Mr. Lee or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mary Si Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indication for use Statement S10(k) Number ( If known ) : _ < 1 2 4 0 37 Device Name : Jung Rim single use hypodermic syringe with Needle Indication for use : Jung Rim single use hypodermic syringe with Needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 2) CFR 801 Subpart D) AND/OR Over-The-Counter (Part 21 CFR 801 Subpart C) JUNG RIM MEDICAL (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | Richard C. Chapman | |--|--------------------| | | 2013.09.12 | | | 13:10:16 -04'00' | | | Division Sign-Off) | |--|----------------------------------------------| | | Division of Anesthesiology, General Hospital | | | fection Control, Dental Devices | | 10(k) Number: | K124037 | |---------------|---------| |---------------|---------| Indication for use statement Page 1 of 2
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%