K123958 · Wuxi Jiajian Medical Instrument Co., Ltd. · GZJ · Oct 28, 2013 · Neurology
Device Facts
Record ID
K123958
Device Name
JIAJIAN POINTOSELECT DIGITAL
Applicant
Wuxi Jiajian Medical Instrument Co., Ltd.
Product Code
GZJ · Neurology
Decision Date
Oct 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Jiajian® Pointoselect Digital is intended for use in the symptomatic relief of chronic intractable pain, postoperative pain, and acute pain.
Device Story
Jiajian Pointoselect Digital is a handheld, 9V battery-powered transcutaneous electrical nerve stimulator (TENS). Device consists of a console with an LCD, a handheld probe, a hand-grip electrode, and lead wires. Operation requires the patient to hold the hand-grip electrode to ground the device while the probe is applied to specific treatment sites. The device delivers an asymmetric biphasic square wave electrical pulse to the skin. The LCD displays selected operation mode, intensity, frequency, pulse width, and battery level. It is used for pain management; the healthcare provider or patient adjusts intensity and frequency settings to provide symptomatic relief. The device is intended for the relief of chronic intractable, postoperative, and acute pain.
Clinical Evidence
Bench testing only. The device was tested for compliance with IEC 60601-1 (general safety), IEC 60601-2-10 (nerve/muscle stimulators), and IEC 60601-1-2 (EMC). Biocompatibility of the probe was verified per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/hypersensitivity). Performance testing confirmed that stimulus parameters remain stable (within ±10%) under reduced battery conditions.
Technological Characteristics
Handheld TENS device; 9V battery powered (6F22 or 6LR61). Output: single channel, asymmetric biphasic square wave. Frequency: 2-18Hz; Pulse duration: 60-120μS. Materials: Handheld probe and lead wires tested for biocompatibility (ISO 10993-5, ISO 10993-10). Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2. Connectivity: None (standalone).
Indications for Use
Indicated for symptomatic relief of chronic intractable pain, postoperative pain, and acute pain in patients requiring transcutaneous electrical nerve stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2013
Wuxi Jiaiian Medical Instrument Co., Ltd. c/o Ms. Doris Dong, Manager Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songiiang Area Shanghai 201600 China
Re: K123958
Trade/Device Name: Jiajian Pointoselect Digital Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ Dated: September 17, 2013 Received: September 25, 2013
Dear Ms. Dong:
This letter corrects our substantially equivalent letter of October 28, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Doris Dong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 Indications for Use
510(k) Number (if known): K123958
Device Name: Jiajian® Pointoselect Digital
### Indications for Use:
The Jiajian® Pointoselect Digital is intended for use in the symptomatic relief of chronic intractable pain, postoperative pain, and acute pain.
> Prescription Use _ (Part 2) CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of __ l
{3}------------------------------------------------
Wuxi Jiajian Medical Instrument Co., Ltd Qinghong Rd., Ehu Town, Xishan District, Wuxi. Jiangsu, China 214116
K123958
### Section 5 510(k) Summary [As required by 21 CFR 807.92]
| 1. Submission Information: | |
|------------------------------|--|
| 510(k) Number: | |
| Date: | |
| Type of 510(k) Submission: | |
| Basis for 510(k) Submission: | |
| Submitter/Manufacturer: | |
| | |
e of 510(k) Submission:
sis for 510(k) Submission:
omitter/Manufacturer:
Contactor:
OCT 2 8 2013 March 13th, 2013 Traditional New device Wuxi Jiajian Medical Instrument Co., Ltd Qinghong Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China 214116 Doris Dong, Consultant Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, China 201600 E-mail: doris_d@126.com / Url: www.ceve.org.cn Tel: 86 21-31261348 / Fax: 86 21-37824346
2. Device Description:
Proprietary Name: Jiajian® Pointoselect Digital Common Name: Transcutaneous electrical nerve stimulator Classification Name: Transcutaneous electrical nerve stimulator for pain relief Regulation Number: 882.5890 Product Code: GZI Device Class: ll Review Panel: Neurology Indications for use: Jiajian® Pointoselect Digital is intended for use in the symptomatic relief of chronic intractable pain, postoperative pain, and acute pain. Device Description:
Jiajian® Pointoselect Digital is a newly designed and easy to operate hand held stimulator. It is 9V battery powered device, with an LCD in the console. which can display the selected operation mode, intensity, frequency, pulse width, and battery level of the device, and so on.
The device composes of a console, a hand held probe, a hand grip electrode, and lead wires.
The hand probe can be used for treatment at individual sites. When using the hand probe, the patient must hold the hand grip electrode in order to electrically ground the device.
Standards:
IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2
ISO 10993-5, ISO 10993-10;
## 3. Predicate Device Identification
| 510k Number: | K060517 |
|---------------|-----------------|
| Product Code: | GZJ |
| Device Name: | Pointer Excel |
| Manufacturer: | LHASA OMS, INC. |
#### 4. Substantial Equivalence:
Detailed comparison data is included in the section of "Substantial Equivalence Discussion" of this 510(k) submission.
### A. Basic technological characteristics, New device VS. Predicate device:
| | REAMPIEFS | 9.000 b.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Canada Banks<br>100 - 100 - 20 - 40 -<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 510/b) Nimb<br> | 1°199000<br><br>----------------------------------- | K060517 | |
1-3
{4}------------------------------------------------
Wuxi Jiajian Medical Instrument Co., Ltd ..... .. .
Image /page/4/Picture/1 description: The image shows a logo with the text "JIA JIAN" in bold, black letters. A checkmark symbol is placed above the letters "IA", and a circled "R" registration mark is located to the right of the word "JIAN". Below the text, there are some Chinese characters.
| | | Qinghong Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China 214116 | |
|-----|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Marketing clearance date: | | Jun 19th, 2006 |
| 3. | Device Name | Jiajian Pointoselect Digital | Pointer Excel |
| 4. | Manufacturer | Wuxi Jiajian Medical Instrument Co., Ltd | LIIASA OMS, INC. |
| 5. | Accessories for stimulation | I pcs. of hand probe for stimulation with | I pcs. of hand probe for stimulation; |
| | mode | lead wire;<br>1 pcs. of hand electrode with lead wire | I pcs. of hand electrode with lead wire |
| 6. | Intended use | Jiajian® Pointoselect Digital is intended<br>for use in the symptomatic relief of<br>chronic intractable pain, postoperative<br>pain, and acute pain. | Pointer Excel is intended for use in the<br>symptomatic relief of chronic intractable<br>pain, postoperative pain, and acute pain. |
| 7. | Power Source(s) | 9 Volt battery type 6F22 (Carbon-zinc), or<br>9 Volt battery type 6LR61 (Alkaline) | DC 9V battery, Type 6F22 |
| | - Method of Line Current<br>Isolation | N/A for DC current | N/A for DC current |
| | - Patient Leakage Current | - | - |
| | - Normal Condition | 20μA | Not Stated in the manual |
| | - Single Fault Condition | N/A | Not Stated in the manual |
| 8. | Number of Output Mode: | 1 | 1 |
| 9. | Number of Output channels: | 1 | 1 |
| 10: | Compliance with Voluntary<br>Standards? | ISO 10993-5, ISO 10993-10:<br>IEC 60601-1, IEC 60601-2-10, IEC<br>60601-1-2 | ISO 10993-5, ISO 10993-10:<br>IEC 60601-1, IEC 60601-2-10, IEC<br>60601-1-2 |
| 11. | Waveform | Biphasic | Biphasic |
| 12. | Shape | Asymmetric biphasic square wave | Asymmetric biphasic square wave |
| 13. | Maximum Output Voltage | $6.6V ±15% @500Ω$ | $11V±15% @500Ω$ |
| 14. | Maximum Output Current | $13.2mA ±15% @500Ω$ | $22mA ±15% @500Ω$ |
| 15. | Pulse Duration | 60-120μS | 220μS |
| 16. | Frequency | 2~18Hz | 1-16Hz |
| 17. | Net Charge | 0μC @500Ω, + and - pulses cancel | 0μC@500Ω |
| 18. | Maximum Phase Charge | 2.4μC@500Ω | 4.8μC@500Ω |
| 19. | Maximum Current Density | 12.08mA/cm²@500Ω | 10.35mA/cm²@500Ω |
| | (r.m.s.) | | |
# B. Substantial Equivalence Discussion
| Similarities between New device<br>and Predicate Device: | Intended use, DC power source, biphasic square waveform, Net charge, frequency<br>range, maximum current density, maximum power density, adjustable frequencies and<br>intensity, standards |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Differences between New device<br>and Predicate Device: | Weight, dimensions: Maximum output voltage, Maximum output current, Output pulse<br>duration, Maximum phase charge |
| Conclusion: | The new device has same intended use, biphasic square waveform, complied standards,<br>adjustable output intensity and frequency, and similar maximum output current density<br>and power density.<br>Though there are differences between the new devices and the predicate devices, such as<br>output intensity values and pulse width, the differences would not raise new safety<br>concerns.<br>To sum up, the new device Jiajian Pointoselect Digital is substantially equivalent to<br>Predicate devices of Pointer excel (K060517). |
{5}------------------------------------------------
Wuxi Jiajian Medical Instrument Co., Ltd Qinghong Rd., Ehu Town, Xishan District, Wuxi, Jiangsu. China 214116
### 5. Safety and Effectiveness of the device
Jiajian® Pointoselect Digital was tested and found to meet the safety standards of:
* IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995;
* IEC 60601-2-10 Edition 2.0 2012-06, Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators; and
* IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2: General Requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests
The pointer probe was tested and found to comply with the biocompatibility standards of:
· ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Viro cytotoxicity, and * ISO 10993-10:2002, Biological evaluation of medical devices - Part 10: Tests for irritation and
delayed-type hypersensitivity
The output lead wires used for Jiajian® Pointoselect Digital were tested and found to comply with the safety standards of:
* IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988: Amendment I, 1991-11, Amendment 2, 1995
The device was also tested basis on reduced battery level, and was found that the stimulus parameters were not significantly affected (less than ±10%).
The conclusion drawn from the testing is that the device is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.