DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is indicated for use in superficial soft tissue approximation (i.e., mucosa and skin) where only short term wound support (7-10 days) is required. DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is not intended for use in ligation, cardiovascular, ophthalmic or neurological procedures.
Device Story
DemeQUICK is a sterile, braided, absorbable surgical suture composed of 90% glycolide and 10% L-lactide copolymer, coated with a mixture of 30% glycolide, 70% L-lactide, and calcium stearate. It functions as a temporary wound support device for superficial soft tissue approximation. The suture is absorbed via hydrolysis, with tensile strength loss occurring over 7-10 days and mass absorption complete in 30-40 days. It is used by clinicians in surgical settings to close skin or mucosal wounds. The device is provided in various lengths and diameters, with or without attached needles, and is sterilized via gamma irradiation. It provides mechanical support to tissues during the initial healing phase, reducing the need for long-term suture presence.
Clinical Evidence
Bench testing only. Testing included USP performance requirements (suture diameter, length, knot pull tensile strength, needle attachment strength, extractable color, sterility) and biocompatibility testing per ISO 10993-1. In-vitro and in-vivo resorption studies were performed to confirm degradation profile.
Indicated for superficial soft tissue approximation (mucosa and skin) requiring short-term wound support (7-10 days). Contraindicated for ligation, cardiovascular, ophthalmic, or neurological procedures.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "DemeTECH". The logo consists of a stylized triangle inside of a circle on the left, followed by the company name in a bold, italicized font. The "CH" in "DemeTECH" is slightly smaller and raised, with a registered trademark symbol next to it.
1 of 3
K123940
**MAY 30 2014**
### 510K Summary
This summary is submitted in accordance with the Safe Medical Device Act (SMDA) of 1990 and Title 21 CFR § 807.92. This summary demonstrates the equivalence of Demetech Sutures to those of the legally marked devices listed. ,
| A. | Applicant: | Demetech Corporation,<br>14175 NW 60th Ave.<br>Miami Lakes FL. 33014 |
|----|----------------------|----------------------------------------------------------------------------------|
| B. | Contact Person: | Luis Arguello<br>Email: luis@demetech.us<br>Phone # 305-824-1048 Ext 113 |
| | Alternate Contact | Anthony J Dimercurio, tony@demetech.us |
| C. | Date Prepared: | May 29, 2014 |
| D. | Trade Name: | DemeQUICK (Rapid Absorbable) Surgical<br>Suture |
| | Common Name: | PGLA Rapid Absorbable Surgical<br>Suture |
| | Classification Name: | Absorbable poly (glycolide/l-lactide) Surgical<br>Suture braided or monofilament |
- Predicate Devices: DemeQUICK (Rapid Absorbable) Surgical Suture is substantially E. equivalent to these following two predicate devices:
- Riverpoint's Vilet Quick® Synthetic Absorbable sufure, reference 510k number, . K120556, Portland, OR.
- . Ethicon's Vicryl Rapide® Synthetic Absorbable sufure, reference 510K number, K033746, Ethicon Inc. Somerville NJ.
- Device Description: F.
DemeQUICK Absorbable suture is an absorbable sterile surgical suture composed of copolymers made from 90% alvcolide and 10% L-lactide. DemeQUICK Absorbable suture is coated with a copolymer made from 30% glycolide 70% L-lactide and calcium stearate, it is available un-dved from sizes: USP 6-0 to USP 2. The DemeQUICK Absorbable suture complies with the requirements established by the United States Pharmacopoeia (USP) for absorbable suture, with tensile strength meeting the collagen table.
- ું. Intended Use:
DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is indicated for use in general soft tissue approximation (i.e., mucosa and skin) where only short term wound support (7-10 days) is required. DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is not intended for use in ligation, cardiovascular, ophthalmic or neurological procedures.
{1}------------------------------------------------
#### Non-Clinical Tests Performed: H.
Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, blocompatibility testing in accordance to ISO 10993-1 and in-vitro and in-vivo resorption to further demonstrate substantial equivalence to the predicate devices. Physical properties and functionality testing assured that the device conformed with suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterlity to methods outlined in USP 35,
| COMPARISON TABLE DEMEQUICK ABSORBABLE TO PREDICATE DEVICES | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|-------------------------------------------------|-------------------------------------|
| Comparison Items | DemeQUICK<br>Absorbable<br>Suture | Riverpoint<br>Medical Vilet<br>Quick®<br>Suture | Ethicon<br>Vicryl Rapide®<br>Suture |
| Suture meets or exceeds the performance<br>requirements for "Absorbable Surgical<br>Suture" as defined in the Official Monograph<br>of the United States Pharmacopeia. | Same | Same | Same |
| Suture Materials meet the performance<br>requirements for Diameter as defined in the<br>United States and European Pharmacopeias | Same | Same | Same |
| Suture meets or exceeds the performance<br>requirements defined in the United States<br>Pharmacopeia for "Tensile Strength" < 881 ><br>collagen | Same | Same | Same |
| Suture Material meets or exceeds the<br>performance requirements defined in the<br>United States Pharmacopeia and the current<br>edition USP for "Needle Attachment" < 871 > | Same | Same | Same |
| Suture meets or exceeds the performance<br>requirements defined in the United States<br>Pharmacopeia for "Suture Length<br>Requirement" (95% of stated label length) | Same | Same | Same |
| Suture is packaged in a same or equivalent<br>manner with sterile single or double package<br>having labeling conforming to 21 CFR and<br>current edition of USP. | Same | Same | Same |
| Suture Material is a composition of<br>absorbable flexible, braided thread prepared<br>from 90% glycolide and 10% L-lactide | Same | Same | Same |
{2}------------------------------------------------
| 3 | of | 3 |
|---|----|---|
|---|----|---|
| Suture material is offered un-dyed | Same | Same | Same |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|------|
| Suture material is supplied coated | Same | Same | Same |
| Suture is sterilized by Gamma Irradiation | Same | Same | Same |
| Suture Material is designed being a sterile,<br>flexible, braided thread offered in a variety of<br>lengths and a range of diameters with or<br>without various needles attached. | Same | Same | Same |
| Progressive loss of tensile strength and<br>eventual absorption of DemeQUICK (Rapid<br>Absorbable) Surgical Suture occurs by<br>means of hydrolysis. Absorption begins as a<br>loss of tensile strength followed by a loss of<br>mass. Absorption is essentially complete in<br>approximately 30 to 40 days. | Same | Same | Same |
| DemeQUICK (Rapid Absorbable) Synthetic<br>Surgical Suture is indicated for use in<br>general soft tissue approximation (i.e.,<br>mucosa and skin) where only short term<br>wound support (7-10 days) is required.<br>DemeQUICK (Rapid Absorbable) Synthetic<br>Surgical Suture is not intended for use in<br>ligation, cardiovascular, ophthalmic or<br>neurological procedures. | Same | Same | Same |
DemeQUICK (Rapid Absorbable) Surgical Suture is composed of the same material as are the predicated devices using the same design' being a sterile, flexible, multifilament threads meeting the requirements of the United States Pharmacopeia. DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is manufactured in the same manner as the predicate devices, being produced from 90% glycolide and 10% L-lactide manufactured in operations considered standard in the fiber industry to form the finished suture fiber in bulk. The manufacturer supplies the identical bulk fibers to Demetech as it sells to other suture manufacturers including those with approved 510K submissions.
A summary of recent USP performance testing results are presented below on gamma sterilized finish sutures, together with the gamma sterilized Biocompatibility documentation already submitted to FDA, demonstrates the substantial equivalence of DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture to that of the predicated devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Flampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
May 30, 2014
DemeTECH Corporation % Mr. John O Brien AJW Technology Consultants Incorporated 445 Apollo Beach Boulevard Apollo Beach, Florida 33572
Re: K123940
Trade/Device Name: DemeQUICK (Rapid Absorbable) Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/l-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: May 7, 2014 Received: May 28, 2014
Dear Mr. O Brien:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. John O'Brien
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DemeTECH. The logo consists of a stylized mountain peak to the left of the company name. The company name is written in a bold, sans-serif font, with the letters "TECH" slightly smaller and underlined.
# Section 2 - Indications for Use Statement
Indication for use
(assigned by FDA Reviewer) 510K Number: K123940
Device Name: DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture
Indication for Use:
DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is indicated for use in superficial soft tissue approximation (i.e., mucosa and skin) where only short term wound support (7-10 days) is required. DemeQUICK (Rapid Absorbable) Synthetic Surgical Suture is not intended for use in ligation, cardiovascular, ophthalmic or neurological procedures.
| Prescription Use | "X" | And/Or | Over the-Counter |
|-------------------------------------|-----|--------|-------------------------|
| Use<br>(Part 21 CFR 801; Subpart D) | | | (21 CFR 801; Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Peter L. Hudson -5 2014.05.30 10:56:41 -04'00'
2-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.