K123859 · Lensar, Inc. · OOE · Mar 27, 2013 · Ophthalmic
Device Facts
Record ID
K123859
Device Name
TBD
Applicant
Lensar, Inc.
Product Code
OOE · Ophthalmic
Decision Date
Mar 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K123859 · Mar 27, 2013
TBD
Lensar, Inc.
Human donor globes
Human donor globes were used to compare the accuracy of laser-generated partial thickness corneal incisions (PTIs) against porcine ex vivo eye model results to ensure performance consistency.
Human donor globes; Corneal incision accuracy; Performance validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Comparison of laser PTIs in porcine eyes vs. human donor globes; Comparative analysis of incision depth accuracy
Human donor globes; Sample Size: Not specified (described as a 'smaller number')
Porcine ex vivo eye model
Mean error in target accuracy and variance of error in incision depth
Indications for Use
The LENSAR Laser System - fs 3D (LLS-fs 3D) is intended for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, laser phacofragmentation, and the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Device Story
Ophthalmic surgical laser system; uses femtosecond laser to perform anterior capsulotomy, phacofragmentation, and corneal incisions. Input: biometric data from 3D optical surface model of eye. Operation: system uses biometric camera and laser to perform 'Localized Imaging'—placing low-energy mark in corneal stroma to measure position relative to anterior/posterior surfaces; system calculates offsets/tilt to ensure accurate laser beam placement despite eye movement. Output: laser-generated corneal incisions and lens fragmentation. Used in clinical setting by ophthalmologists. Benefits: improved accuracy/precision of corneal incision depth compared to manual diamond blade techniques; reduces risk of incorrect placement.
Clinical Evidence
Bench testing only. Porcine ex vivo eye model compared laser-generated partial thickness incisions (PTIs) to manual diamond blade incisions. Results: laser PTIs showed higher depth accuracy (closer to 600um target) and precision than manual. Human donor globe testing confirmed similar mean error and variance. Endothelial cell density evaluation showed no loss when residual corneal bed >85um.
Technological Characteristics
Femtosecond laser system; pulse energy control/monitoring; biometric system with 3D optical surface modeling; localized imaging feedback loop. Connectivity: integrated system. Software: modified for custom shot patterns and self-correcting feedback. Sterilization: not specified.
Indications for Use
Indicated for patients undergoing cataract surgery for crystalline lens removal, including anterior capsulotomy, laser phacofragmentation, and creation of full/partial thickness single-plane and multi-plane corneal arc cuts/incisions.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
# 510(K) SUMMARY
MAR 2 7 2013
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(a).
### SUBMITTER'S NAME. ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED
a. Applicant:
D
LENSAR, Inc. 2800 Discovery Drive, Suite 100 Orlando, FL 32826 Tel: (888) 536-7271 Fax: (407) 386-7228
b. Contact Person:
S.K. McGarvey Regulatory Consultant skmcgg@aol.com Cell: (702) 524-3701 Fax: (407) 386-7228
#### DATE SUMMARY PREPARED: December 12, 2012
#### NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME
- LENSAR Laser System fs 3D (LLS-fs 3D) a. Trade/Proprietary Name:
- b. Common/Usual Name: LENSAR Laser System - fs 3D (LLS-fs 3D)
- Ophthalmic Laser, Phacofragmentation System C. Classification Name:
- 21 CFR 886.4390 OOE; 21 CFR 886.4670 HOC d. Classification Code(s):
#### PREDICATE DEVICES
| 510(K) # | TRADE NAME | MANUFACTURER |
|----------|-----------------------------|----------------------------|
| K122829 | LENSAR Laser System - fs 3D | LENSAR, Inc. |
| K120214 | LENSAR Laser System - fs 3D | LENSAR, Inc. |
| K101626 | LenSx 550 Laser System | LenSx Lasers, Inc. (Alcon) |
#### DEVICE DESCRIPTION
The predicate LENSAR Laser System is an ophthalmic surgical laser that has been cleared by the Agency for use in anterior capsulotomy and laser phaco fragmentation in cataract surgery performed individually or consecutively during the same surgery under K120214 (LENSAR Laser System - fs 3D). In K122829, the laser expands the Indication for Use to single-plane and multi-plane clear corneal cuts/incisions. This expansion of the
{1}------------------------------------------------
Indication for Use is for partial as well as full thickness single-plane and multi-plane arc cuts/incisions in the cornea.
The current LLS-fs 3D uses the same laser and the same beam guidance system to deliver laser pulses to the eye as the predicate LENSAR device cleared via 510(k) K120214 and K 122829. Also, the patient interface device (PID), controlled force docking mechanism and moveable optical head to dock the laser to the stationary patient are unchanged from that 510(k) K120214 and K122829.
The LLS-fs 3D biometric system, which measures and constructs a three dimensional model of the optical surfaces within the eye, is unchanged from the predicate device except for software modifications to allow the system to analyze the shape and position of the peripheral cornea. To improve the accuracy of incision placement in the cornea, a new functionality called "Localized Imaging" has been added to the system. Localized Imaging is a process whereby the biometric camera is positioned to observe the corneal incision site and a very small laser mark (at low energy) is made at the center of the stroma. By imaging and measuring the position of the mark relative to the cornea anterior and posterior surfaces, the system mitigates any potential of incorrect laser beam placement due to eve movement. The corneal incision surgery is performed immediately after Localized Imaging and results in accurate incision depth in the cornea (<75 microns error).
The Indication for Use was expanded to include creation of the corneal incisions for entry to the eye in cataract surgery in K122829 and is now further expanded to include partial thickness corneal arc cuts/incisions.
#### STATEMENT OF INTENDED USE
The LENSAR Laser System - fs 3D (LLS-fs 3D) is intended for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, laser phacofragmentation, and the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON
The femtosecond laser system, including pulse energy control and monitoring, used in the current LLS-fs 3D is the same as that used in the predicate device cleared in K120214 and K122829. The primary difference between the current LENSAR Laser System - fs 3D device and the predicate device is the addition of software for the indication of partial ' thickness single-plane and multi-plane arc cuts/incisions in the cornea for cataract surgery.
In the predicate device, the biometric system measured the shape and position of the central cornea for use in the ray tracing necessary to correctly place laser incisions within the crystalline lens. In the current device, corneal measurements are also used to generate custom shot patterns for the corneal incisions. In addition, the current device implements
{2}------------------------------------------------
Localized Imaging to make the corneal incisions more accurate, the system now images before and after making laser mark(s) and analyzes the images to apply any offsets or tilt in the intended mark position(s) to the subsequent corneal incisions. Both systems are now integrated to provide self-correcting feedback.
| 510(k) Number | Clearance Date | Device Description |
|---------------|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| K122829 | 12/03/2012 | LENSAR Laser System - fs 3D (LLS-fs 3D) for creation<br>of single-plane and multi-plane cuts/ incisions in the<br>cornea. |
| K120214 | 06/08/2012 | LENSAR Laser System - fs 3D (LLS-fs 3D) for anterior<br>capsulotomy and laser phacofragmentation |
| K101626 | 10/18/2010 | LenSx Laser System for anterior capsulotomy,<br>phacofragmentation, and creation of single-plane and<br>multi-plane arc cuts/incisions in the cornea |
The LLS-fs 3D is of comparable type and is substantially equivalent to the following predicate devices:
- . The activities used to evaluate the LENSAR Laser System - fs 3D (LLS-fs 3D) and the information and reports provided in this 510(k) submission do not identify any new issues of safety or effectiveness. The optical radiation hazard analysis confirms the continuing ocular safety and the equivalence to the predicate device detailed in 510(k) K120214 and K122829.
- . The LLS-fs 3D laser technology and mechanism of laser-tissue interaction are unchanged from that of the femtosecond laser cleared under K120214 and K122829.
- The Indication for Use statement for anterior capsulotomy and laser . phacofragmentation for the LLS-fs 3D is the same as that of the predicate 510(k)cleared LENSAR Laser System - fs 3D (K120214) and the indication for use statement for corneal cuts/incisions in the cornea is the same as that for the LENSAR Laser System - fs 3D (K122829). The indication for use statement for corneal are cuts/incisions in the cornea is the same as that for the LenSx Laser (K101626).
- . The differences between the modified LLS-fs 3D and the predicate devices are insignificant and do not affect the safety or effectiveness of the device.
## BRIEF SUMMARY OF PERFORMANCE TEST RESULTS
The performance data supporting substantial equivalence of the LENSAR Laser System fs 3D to the predicate LENSAR device is summarized as follows:
- . An analysis of the optical radiation hazard to non-target tissue demonstrated that the current LLS-fs 3D femtosecond laser, biometric system scanning diode light source, and patient eye illumination (light emitting diodes) are eye safe under all normal operating and known fault conditions.
{3}------------------------------------------------
- A hazard analysis of all potential hazards to the patient, surgeon and other system . operators was performed to consider all changes between the current LLS-fs 3D and predicate LLS-fs 3D device. The hazard analysis demonstrates that all potential hazards have acceptable levels of probability/severity characteristics.
- . The accuracy of the depth of conventional manual versus laser-generated partial thickness corneal incisions (PTIs) was tested using a porcine ex vivo eye model. The depths of the incisions were measured using an optical coherence tomographer. A comparison of the depth accuracy of manual and laser PTIs demonstrated that the laser PTIs had measured depths much closer to the target 600 um depth than the manually-created PTIs.
- . The laser PTIs in porcine eyes were compared to a smaller number of laser PTIs in human donor globes. The results showed that the mean errors in target accuracy were not different and the variance of the error in incision depth was also about the same.
- . Overall, the testing showed that PT's created by the LLS-fs 3D were more accurate and precise than those made using a diamond blade with the conventional manual technique.
- · Evaluation of the effect of the laser partial thickness incisions on endothelial cells of ex vivo eyes showed that the laser method resulted in no loss of endothelial cell density when the residual corneal bed is maintained at >85um.
- The addition of partial thickness arc cuts/incisions did not result in additional . hazards although the probabilities and severities of some of the hazards identified for the predicate device were changed.
Based on the above supportive documentation, the LENSAR Laser System - fs 3D with modified software to permit partial thickness corneal arc cuts/incisions in cataract surgery is substantially equivalent to the LENSAR Laser for anterior capsulotomy and cataract fragmentation (K120214), and full thickness single-plane and multi-plane clear corneal incisions (K122829) with respect to technological characteristics, and to the LenSx Laser (K101626) for the indication for use in corneal arc cuts/incisions.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
March 27, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
LensAR, Inc. % Ms. S.K. McGarvey Regulatory Consultant 2800 Discovery Drive, Suite 100 Orlando, FL 32826
Re: K123859
Trade/Device Name: Lensar Laser System- fs 3D (LLS-fs 3D) Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Femtosecond Laser Regulatory Class: Class II Product Code: OOE, HOF Dated: February 8, 2013 Received: February 14, 2013
Dear Ms. McGarvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. S.K. McGarvey
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia YAlexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123859
Device Name: LENSAR Laser System - fs 3D (LLS-fs 3D)
Indications For Use:
The LENSAR Laser System - fs 3D (LLS-fs 3D) is intended for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, laser phacofragmentation, and the creation of full and partial thickness single-plane and multi-plane arc cuts/ incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Prescription Use الااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااااا (part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Leonid Livshitz 2013.03.21 16:50:40-04'00'
(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number: K 123859
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.