OPHTACATH KIT

K123831 · France Chirurgae Instrumentation · OKS · Jul 29, 2013 · OP

Device Facts

Record IDK123831
Device NameOPHTACATH KIT
ApplicantFrance Chirurgae Instrumentation
Product CodeOKS · OP
Decision DateJul 29, 2013
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Pediatric

Indications for Use

The OphtaCath® Unilateral Kits (2-mm and 3-mm) are intended for use during dilation of the obstructed nasolacrimal duct to treat epiphora. For patients older than 12 months but under 30 months, the recommended catheter is the 2 mm balloon diameter. For patients 30 months and older, the recommended catheter is the 3 mm balloon diameter.

Device Story

OphtaCath Unilateral Kit is a sterile, single-use lacrimal balloon catheter system. Device consists of a semi-flexible 304 stainless steel stylet, PBX tubing, and a nylon balloon covered by a polyethylene sleeve. Kit includes a Flamingo Inflation Device for balloon dilation, pressure monitoring, and deflation. Used by clinicians to treat nasolacrimal duct obstruction and epiphora. Clinician inserts catheter into the lacrimal duct; inflates balloon to a nominal pressure of 8 atm to dilate the obstruction; monitors pressure via inflation device; deflates balloon post-procedure. Provides mechanical dilation of the duct to restore tear drainage, potentially relieving epiphora symptoms.

Clinical Evidence

Bench testing only. Validated manufacturing process, stylet flexibility, deflation time, compliance, burst pressure, and leak resistance. Biocompatibility testing met standards. Sterilization and package integrity validated. Assembly compatibility with inflation device confirmed.

Technological Characteristics

Materials: Nylon balloon, 304 stainless steel stylet, PBX tubing, polyethylene sleeve. Nominal fill pressure: 8 atm. Dimensions: 240 mm length; 2 mm or 3 mm inflated diameter. Sterilization: Ethylene oxide. Packaging: Tyvek peelable pouch. Accessory: Flamingo Inflation Device for manual pressure control.

Indications for Use

Indicated for dilation of obstructed nasolacrimal duct to treat epiphora in patients 12 months of age and older. 2 mm balloon recommended for patients 12-30 months; 3 mm balloon recommended for patients 30 months and older.

Regulatory Classification

Identification

Lacrimal stents and intubations sets are intended to repair the lacrimal drainage system. Lacrimal stents and intubations sets are indicated for, but limited to, the treatment of epiphora in infants and adults; canalicular pathologies such as stenosis; obstruction or laceration; and conditions requiring dacryocystorhinostomy (conventional or laser); or imperforation of the nasolacrimal duct in an infant.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 2 9 2013 # 5. 510(k) Summary as required by 21 CFR§807.92(c) 510(k) Owner: FCI SAS (France Chirurgie Instrumentation) 20-22 rue Louis Armand 75015 Paris, France Telephone: +33 1 53 98 98 98 Facsimile: Contact person: Barbara S. Fant, Pharm.D. Clinical Research Consultants, Inc. 3308 Jefferson Avenue Upper Level Cincinnati, OH 45220 Phone: (513) 961-8200 Facsimile: (513) 961-2858 Date: July 15, 2013 Trade Name: OphtaCath® Unilateral Kit Common name: Lacrimal balloon catheter Classification Name: Lacrimal Stents and Intubation Sets Product Code: OKS #### Identification of a Legally Marketed Predicate Device The OphtaCath® lacrimal balloon catheter is substantially equivalent to the LacriCATH lacrimal balloon catheter marketed by Quest Medical, Inc. 510(k) Premarket Notification Number: K113867, FDA Product Code OKS. The OphtaCath® is only available in kit form; whereas. the LacriCath is available as a procedure kit, or the kit components can be purchased separately from Quest Medical, Inc. The OphtaCath® Unilateral Kit is substantially equivalent to the LacriCath Unilateral Kit in contents and accessory devices. The OphtaCath® and LacriCath unilateral kits each contain one Jacrimal balloon catheter and an inflation device for the balloon catheter. The inflation devices contained in the OphtaCath® and LacriCath kits are Class I, 510(k) exempt devices when used with ophthalmic lacrimal balloon catheters; thus, a substantial equivalence determination is unnecessary since these devices are exempt from premarket notification. The Flamingo Inflation Device that is used with the OphtaCath® was originally cleared for marketing for use as an inflation device with cardiovascular balloon catheters as the Sedat Flamingo, 510(k) Premarket Notification Number: K082755, FDA Product Code MAV, Class II, {1}------------------------------------------------ Subsequently, the trade name was modified to Flamingo Inflation Device, and the product line was extended to add the indication for use as an inflation device for ophthalmic lacrimal duct catheters under Product Code HNW, Class I. 510(k) exempt. The Flamingo Inflation Device for use with cardiovascular balloon catheters (product reference # 0218ND) is the exact same device as the Flamingo Inflation Device for ophthalmic lacrimal balloon catheters (product reference # 0275ND), and the devices in each product reference number are identical in every respect, except that the instructions for use contained in each package contains only the respective indication for use to assure maximum safe use of the device. Similarly, the Atrion QL® Balloon Catheter Inflation Device was originally cleared for marketing under 510(k) Premarket Notification Number: K972964, FDA Product Code KOE, Subsequent Product Code HNW. The ophthalmic indication for use with lacrimal balloon catheters was added to the device. and FDA subsequently assigned Product Code HNW to the Atrion QL® device and rendered it Class I, 510(k) exempt status for the ophthalmic indication. ## General Description The OphtaCath® lacrimal duct catheter is a sterile and single-use balloon catheter consisting of a semi-flexible stainless steel stylet covered by PBX tubing that ends with a nylon balloon. The balloon is covered by a protective polyethylene sleeve. The balloon is designed to be inflated to a known diameter and length at the specified pressure. Markings are placed at 10 mm and 15 mm · proximal to the working portion of the balloon. The overall length of the catheter is approximately 240 mm long. The catheter is available in a 2 mm inflated diameter. The 2 mm balloon has a length of 13 mm and a diameter of 0.90 mm before inflation. The 3 mm balloon has a length of 15 mm and a diameter of 1.0 mm before inflation. The Flamingo Inflation Device is an accessory to the OphtaCath®, and is a sterile, single-use inflation device for dilating the balloon catheter, monitoring balloon pressure, and deflating the balloon postoperatively. The OphtaCath® is sold as a unilateral kit that contains one 2 mm or 3 mm balloon catheter and an inflation device (OphtaCath® Unilateral Kits, 2-mm and 3-mm). #### Intended Use/Indications for Use The intended use for the OphtaCath® is the same as for the LacriCath predicate device. The OphtaCath® and LacriCath devices (2-mm and 3-mm) are both indicated for use for the treatment of nasolacrimal duct obstruction; and, the LacriCath is also indicated for use for the treatment of dacryocystocele. These differences in the indications for use have no impact on the safety and effectiveness of the device itself. Unlike the LacriCath, the OphtaCath® indications for use statement contains the recommended balloon diameters based on age, which aids in proper device selection. - . The OphtaCath® Unilateral Kits (2-mm and 3-mm) are intended for use during dilation of the obstructed nasolacrimal duct to treat epiphora in patients 12 months of age and older. For patients older than 12 months but under 30 months, the recommended catheter is the 2 mm balloon diameter. For other patients 30 months or older, the recommended catheter is the 3 mm balloon diameter. {2}------------------------------------------------ #### Comparison of Technological Characteristics The OphtaCath® and LacriCath balloon catheters are sterile, single-use balloon catheters. As shown in the table below, the OphtaCath® (2-mm and 3-mm) has the same, or similar, technological characteristics, as the LacriCath (2-mm and 3-mm) predicate device. | | LacriCath<br>Predicate Device (K113867) | | OphtaCath® | | |-----------------------------|-----------------------------------------|----------|----------------------------|----------| | Manufacturer | Quest Medical, Inc. | | FCI SAS | | | Balloon Material | Nylon | | Nylon | | | Stainless Steel Core/Stylet | 304 Stainless steel core | | 304 Stainless steel stylet | | | Nominal Fill Pressure | 8 atm | | 8 atm | | | Balloon Size | Diameter | Length | Diameter | Length | | | 2 mm | 13 mm | 2 mm | 13 mm | | | 3 mm | 15 mm | 3 mm | 15 mm | | Balloon Diameter | Deflated | Inflated | Deflated | Inflated | | | 0.8 mm | 2 mm | 0.9 mm | 2 mm | | | 0.9 mm | 3 mm | 1.0 mm | 3 mm | | Catheter Length | 200 mm | | 240 mm | | | Sterilization Method | Ethylene Oxide | | Ethylene Oxide | | | Sterile Packaging | Tyvek peelable pouch | | Tyvek peelable pouch | | | Unilateral Kit Accessory | Inflation device | | Inflation device | | The predicate device is also available in a 5 mm balloon diameter, whereas, the OphtaCath® is only available in the 2 mm and 3 mm sizes. # Brief Summary of Non-Clinical Tests and Results The manufacturing process was validated, and demonstrated the capacity of FCI to manufacture the OphtaCath® lacrimal balloon catheter. Bench top testing was performed on samples before and after sterilization to validate the flexibility of the stylet, deflation time, compliance, and burst of the catheter and its components. All nonclinical test results met the established specifications for the device. Test results demonstrate that the OphtaCath® catheter did not contain any leakages while being inflated and all welds were of sufficient tensile strength to resist breakage. The biocompatibility of the balloon raw materials and finished, sterilized device was tested to the applicable standards and met required specifications. Ethylene oxide sterilization validation studies and package integrity studies were performed according to the applicable standards; and, the test results support the shelf-life and storage conditions for the device. The assembly of the balloon catheter with the inflation device was tested and validated to demonstrate the connectivity of the kit components and the fluid fill volume of the inflation device. Test results demonstrate the compatibility of the components. From the testing, it can be concluded that the OphtaCath® lacrimal balloon catheter performs as intended without leakage or breakage; and that the assembled device performs as intended. {3}------------------------------------------------ # Basis of Substantial Equivalence The OphtaCath® lacrimal balloon catheter is substantially equivalent to the LacriCath lacrimal balloon catheter in material, intended use, basic design concept, dimensions, sterilization methods, performance testing, biocompatibility and clinical use. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 29, 2013 FCI SAS (France Chirurgie Instrumentation) % Barbara S. Fant, Pharm. D. President Clinical Research Consultants, Inc. 3308 Jefferson Ave. Upper Level Cincinnati, OH 45220 Re: K12383 i Trade/Device Name: OphtaCath® Unilateral Kit Regulation Number: None Regulation Name: Lacrimal Stents and Intubation Sets Regulatory Class: Unclassified Product Code: OKS Dated: June 19, 2013 Received: June 19, 2013 Dear Dr. Fant: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {5}------------------------------------------------ Page 2 - Dr. Barbara S. Fant You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); Iabeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kesia Y. Alexander -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## 4. Indications for Use Statement 510(k) Number (if known): K123831 · Device Name: OphtaCath Unilateral Kit : Indications for Use: The OphtaCath® Unilateral Kits (2-mm and 3-mm) are intended for use during dilation of the obstructed nasolacrimal duct to treat epiphora. For patients older than 12 months but under 30 months, the recommended catheter is the 2 mm balloon diameter. For patients 30 months and older, the recommended catheter is the 3 mm balloon diameter. ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X OR Over-The-Counter Use James P. Bertram -S 2013.07.24 10:51:01 -04'00' (Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number_ K123831_ ______________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...