K123814 · Fx Solutions · KWT · Aug 13, 2013 · Orthopedic
Device Facts
Record ID
K123814
Device Name
HUMELOCK II CEMENTED SHOULDER SYSTEM
Applicant
Fx Solutions
Product Code
KWT · Orthopedic
Decision Date
Aug 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Humelock II Cemented Shoulder System is indicated for use in total and hemi-shoulder replacement to treat: 1. Proximal humeral fractures 2. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis; 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail). The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.
Device Story
Humelock II Cemented Shoulder System is a modular total and hemi-shoulder prosthesis; components include humeral stem, humeral head, double taper connector, and glenoid component. System used in orthopedic surgery for shoulder replacement. Humeral stem (Ti-6Al-4V) connects to humeral head (Co-Cr-Mo) via double taper connector (Ti-6Al-4V). Glenoid component (UHMWPE) features two pegs for cemented fixation. Offset taper design allows head rotation relative to humeral cut surface for optimal bone coverage. Surgeon implants device to restore joint function and alleviate pain in patients with arthritis or fractures. Device relies on cemented fixation for stem and glenoid components. Clinical benefit includes joint replacement for patients with severe shoulder pathology.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing, including mechanical stability testing of the modular connection and characterization of UHMWPE material.
Technological Characteristics
Humeral stem: Ti-6Al-4V alloy (ISO 5832-3). Humeral head: Co-Cr-Mo alloy (ISO 5832-12). Glenoid: UHMWPE (ISO 5834-2) with tantalum radiopaque markers (ASTM F560). Modular design with double taper connector. Cemented fixation only.
Indications for Use
Indicated for patients requiring total or hemi-shoulder replacement due to proximal humeral fractures, severe pain/disability from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or other conditions where arthrodesis/resection are not acceptable (e.g., revision of prior implants).
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Fx Solutions Humelock Cemented Shoulder Prosthesis (K111097)
{0}------------------------------------------------
### K123814
| 3. 510(k) Summary | |
|--------------------------------------|---------------------------------------------------------------------------------------------------------|
| Prepared On: | December 10, 2012 |
| Applicant / Sponsor: | Fx Solutions<br>1663 rue de Majornas<br>01440 Viriat<br>France |
| Manufacturer: | Compagnie Financière & Médicale<br>13 Bd Victor Hugo<br>01000 Bourg En Bresse<br>France<br>AUG 1 3 2013 |
| Contact: | Jean-Jacques Martin<br>+33 4 74 55 35 55<br>www.fxsolutions.fr |
| Proprietary Name: | Humelock II Cemented Shoulder System |
| Common Name: | Total and Hemi-Shoulder Replacement System |
| Classification Names: | 21 CFR 888.3650: Shoulder joint metal/polymer non<br>constrained cemented prothesis, Class II |
| | 21 CFR 888.3690: Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis, Class II |
| Product Codes: | KWT, HSD |
| Substantially<br>Equivalent Devices: | K111097 - Fx Solutions Humelock Cemented Shoulder<br>Prosthesis |
| | K060874 - DePuy Global AP Shoulder System |
Device Description:
The Humelock II Cemented Shoulder System is a total and hemi-shoulder prosthesis consisting of a humeral stern, a humeral head, a double taper connector and, when used for total shoulder replacement, a glenoid component.
The humeral stem is manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3 and is available in diameters of 6-15mm. The distal end of the humeral stern is cylindrical with a polished surface. The proximal portion of the humeral stem has a grit blasted surface.
8
{1}------------------------------------------------
#### K123814
The humeral stem incorporates a female taper for attachment of the double taper connector, which connects to the humeral head.
The double taper connector is manufactured from Ti-6A1-4V alloy conforming to ISO 5832-3. One size is available and is compatible with all sizes of humeral stems and humeral heads. The double taper connector has a male taper on each end is used to connect the humeral head to the humeral stem. An impactor / extractor hole is incorporated into the proximal end of the taper.
The humeral head is manufactured from wrought Co-Cr-Mo alloy conforming to ISO 5832-12 and is available in diameters of 39 - 54mm in centered and offset styles. The offset of the taper allows the head to be rotated relative to the cut surface of the humerus to provide optimal coverage of the bone. A female taper allows attachment to the double taper connector, which connects to the humeral stem.
The glenoid component is manufactured from ultra high molecular weight polyethylene (UHMWPE) conforming to ISO 5834-2. It is available in sizes extra small, small, medium, large and extra large. The glenoid component features two pegs for cemented fixation to the glenoid bone. Each peg contains a radiopaque marker manufactured from tantalum conforming to ASTM F560.
#### Intended Use / Indications:
The Humelock II Cemented Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- 1. Proximal humeral fractures
- 2. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).
The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.
{2}------------------------------------------------
#### Summary of Technologies/Substantial Equivalence:
Substantial equivalence of the Humelock II Cemented Shoulder Prosthesis to the predicate devices is based on a comparison of indications, intended use, materials, design and sizing, and mechanical testing.
#### Non-Clinical Testing:
Mechanical testing was conducted to demonstrate the stability of the modular connection between the double taper connector and the humeral stem. Pull off testing of the humeral head from the taper was conducted previously. Characterization data was provided for the UHMWPE used to manufacture the glenoids.
#### Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence between the Humelock II Cemented Shoulder System and the predicate shoulder systems.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with stylized wings and tail feathers.
r
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2013
Mr. Jean-Jacques Martin Managing Director Fx Solutions 1663 rue de Majornas Viriat, 014440 France
Re: K123814
Trade/Device Name: Humelock II Cemented Shoulder System Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, HSD Dated: July 1, 2013 Received: July 2, 2013
Dear Mr. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mr. Jean-Jacques Martin
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Erin 阿狄eith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 2. Indications for Use
510(k) Number (if known): _K123814_
Device Name: Humclock II Cemented Shoulder System
#### Indications for Use:
The Humelock II Cemented Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- 1. Proximal humeral fractures
- 2. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (c.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).
The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.
Prescription Use -- X --(Part 21 CFR 801 Subpart D)
,
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 4 of 4
## Elizabeth 断房rânk -S
Division of Orthopedic Devices
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.