K123813 · Vasonova, Inc. · OBJ · Jan 18, 2013 · Cardiovascular
Device Facts
Record ID
K123813
Device Name
VASCULAR POSITIONING SYSTEM (VPS) G4 CONSOLE
Applicant
Vasonova, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Jan 18, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The intended use of the VPS Stylet and Console (VPS System) is to quickly and accurately guide market available central catheters to the desired location which is the lower third of the SVC or at the cavo-atrial junction.
Device Story
VPS G4 Console provides real-time guidance for central venous catheter tip placement. System inputs include electrical cardiac signals (ivECG) and blood flow data (Doppler ultrasound) via a polyimide stylet with a distal Doppler sensor and stainless steel ivECG wire. Console processes these signals to provide visual guidance via a touchscreen tablet display. Used in clinical settings by healthcare providers to navigate catheters to the lower third of the SVC or cavo-atrial junction. Output consists of colored icons and waveform data; a 'Blue Bullseye' indicates the catheter tip has reached the desired location. System assists clinicians in confirming tip placement without immediate reliance on fluoroscopy or chest X-ray, potentially reducing radiation exposure and procedure time.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including software verification and validation, and compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-37 (ultrasonic diagnostic equipment safety).
Technological Characteristics
System comprises a console with tablet display and a polyimide stylet. Sensing utilizes a distal Doppler sensor and stainless steel ivECG wire. Connectivity is via extension cable to the console. Compliant with IEC 60601-1, IEC 60601-1-4, and IEC 60601-2-37. Software-driven data acquisition and processing.
Indications for Use
Indicated for guidance and tip positioning of central venous catheters in adult patients as an alternative to fluoroscopy or chest x-ray. Contraindicated/limited in patients with altered cardiac rhythms (atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm) or femoral/saphenous vein access, where P-wave presentation is altered; these patients require additional confirmation methods.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
## 12. 510(k) Summary
.
| Submitter | VasoNova, Inc., a wholly owned subsidiary of Teleflex, Inc.<br>155 Jefferson Drive, Suite 100<br>Menlo Park, CA 94025 | | |
|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--|
| Name of contact person | Christine Ford, RAC | | |
| | Office phone: 610-378-0131, ext. 3338<br>Mobile phone: 484-797-7110 | | |
| | Fax: 610-478-3179<br>Email: christine.ford@teleflex.com | | |
| Date summary prepared | December 7, 2012 | | |
| Device trade or<br>proprietary name | Vascular Positioning System™ (VPSTM) G4 Console | | |
| Common or usual name | Catheter, Ultrasound, Intravascular | | |
| Classification name | Diagnostic intravascular catheter, 21 CFR §870.1200, Class II, product code OBJ | | |
| Classification panel | Cardiovascular | | |
| Legally marketed device(s)<br>to which equivalence is<br>claimed | K103260 VasoNova Vascular Positioning System™ (VPSTM) Console | | |
| Device description | The VPS System consists of a VPS Console with tablet display and a VPS<br>Stylet.<br>The VPS G4 Console consists of two key integrated software-driven<br>components: a data acquisition module and an embedded computer with<br>processing capabilities. The data acquisition module within the console<br>transmits and receives ultrasound data while receiving electrical signals from<br>the heart through sensors mounted at the tip of the VPS Stylet. Using signals<br>gathered by the data acquisition module, the graphical user interface provides<br>the user with guidance for tip positioning of central venous access devices.<br>The separately provided VPS Stylet is a polyimide tube containing a Doppler<br>sensor on a coax cable and an intravascular electrocardiogram (ivECG) signal<br>sensing stainless steel wire. The Doppler sensor and the exposed portion of the<br>ivECG are located at the distal end of the stylet and are used to detect and<br>transmit physiological information to the VPS Console for analysis.<br>The<br>proximal end of the stylet connects to an extension cable that connects to the<br>VPS Console. | | |
| Intended use of the<br>device | The intended use of the VPS Stylet and Console (VPS System) is to quickly and<br>accurately guide market available central catheters to the desired location which<br>is the lower third of the SVC or at the cavo-atrial junction. | | |
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE COMPARED TO THE<br>PREDICATE DEVICE | | | |
| Characteristic | Predicate VPS Console | Subject VPS G4 Console | |
| Guidance indicator | Colored icons, waveform data | Colored icons, waveform data | |
| Guidance signal type(s) | ivECG, Doppler ultrasound | ivECG, Doppler ultrasound | |
| Signal Conductor | Two conductor stylet wires<br>(coax cable and ivECG wire) | Two conductor stylet wires<br>(coax cable and ivECG wire) | |
| Acoustic Output | One setting; not controlled by the user | One setting; not controlled by the user | |
| User Interface | PC-based system with keyboard and touch screen | PC-based system connected to a touchscreen tablet display | |
| Software Interface | Custom GUI | Custom GUI | |
| PERFORMANCE TESTS | | | |
| Characteristic | Standard/Test | Results Summary | |
| Electromagnetic Compatibility | IEC 60601-1-2: 2007-03 - Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance:<br>Collateral standard: Electromagnetic compatibility - Requirements<br>and tests | Compliant | |
| Electrical Safety and Acoustic<br>Safety | IEC 60601-1:1988 (A1:1991 + A2:1995) - Medical electrical<br>equipment - Part 1: General requirements for safety IEC 60601-1-4: 1996 (A1:1999)/EN 60601-1-4:2000- Medical<br>electrical equipment - Part 1-4: Collateral Standard: Programmable<br>electrical medical systems IEC 60601-2-37:2001 (A1:2004 + A2:2005) - Medical electrical<br>equipment - Part 2-37: Particular requirements for the safety of<br>ultrasonic medical diagnostic and monitoring equipment | Compliant | |
| Software verification and<br>validation | VasoNova Software and Algorithm Verification and Validation testing | Compliant | |
| Conclusions: | The intended use, technological characteristics, method of collecting and<br>interpreting physiological data, and performance of the predicate VasoNova<br>VPS Console and the VPS G4 Console along with testing to voluntary standards<br>and applicable FDA guidance documents provide evidence that the subject VPS<br>G4 console is substantially equivalent to the predicate VPS Console in terms of<br>safety, efficacy, and performance.<br>The described physical, software, and hardware modifications are enhancements<br>to various aspects of the predicate VPS Console. They do not alter the<br>fundamental scientific technology of the device; they do not alter the method in<br>which physiological data is collected from the patient; they do not alter the<br>method in which physiological data is interpreted by the software algorithm to<br>determine the symbols used to indicate catheter migration through the<br>vasculature and final tip placement. Any differences between the subject VPS<br>G4 Console and the predicate VPS Console do not raise new issues of safety or<br>effectiveness. | | |
{1}------------------------------------------------
.
:
.
.
.
.
.
1 : 1 : 1 : 1 :
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
JAN 1 8 2013
Vasonova Inc. c/o: Christine Ford Regulatory Affairs Manager 155 Jefferson Drive, Suite 100 Menlo Park, CA 94025
Re: K123813
Trade Name: Vascular Positioning System (VPSTM) G4 Console Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: OBJ Dated: December 7, 2012 Received: December 19, 2012
Dear Ms. Ford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{3}------------------------------------------------
Page 2 -- Ms. Christine Ford
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
**Owen P. Faris -S**
for
Bram D. Zuckerman, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): __ K123813
## Vascular Positioning System (VPS™) G4 Console Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
The VPS Stylet and Console are indicated for guidance and tip positioning for central The Console is intended for use with the stylet supplied by venous catheters. VasoNova. The Vascular Positioning System consists of a Console and associated power supply and cabling and a stylet. The VPS Console, when used with the VPS Stylet, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the Vascular Positioning System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location. The Vascular Positioning System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. NOTE: If a steady blue bull's eye is not obtained, standard hospital practice should be followed to confirm catheter tip location.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Prescription 'Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR) |
|-----------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------|
(Division Sign-Off)
Division of Cardiovascular Devices
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.