AXON-BUS PROSTHETIC SYSTEM

K123795 · Otto Bock Austria · GXY · Jun 3, 2014 · Neurology

Device Facts

Record IDK123795
Device NameAXON-BUS PROSTHETIC SYSTEM
ApplicantOtto Bock Austria
Product CodeGXY · Neurology
Decision DateJun 3, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Axon-Bus Prosthetic System is to be used exclusively for exoprosthetic fittings of the upper limbs.

Device Story

The Axon-Bus Prosthetic System is a powered upper limb prosthesis for unilateral amputees. It utilizes cutaneous electrodes to detect EMG signals from the user's residual limb muscles. These signals are processed by the AxonMaster control unit, which drives the Michelangelo Hand (terminal device) and manages passive/active elbow and wrist components. The system offers multiple control strategies (proportional or digital) and switching methods (e.g., co-contraction, pulse, automatic) to accommodate individual patient needs. A prosthetist configures the system via AxonSoft software using Bluetooth data transfer. The device provides various grip patterns (power, three-point, two-point) and thumb positions to assist with activities of daily living. It is intended for use by a single patient under the guidance of an authorized prosthetist. The system benefits patients by restoring functional grasping and manipulation capabilities, with electrical thumb positioning and elbow locking enhancing control efficiency.

Clinical Evidence

No clinical studies were performed. Substantial equivalence was demonstrated through bench testing, including software validation, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility testing (ISO 10993-1, -5, -10) for all patient-contacting materials. Internal performance testing confirmed the device meets design specifications for grip force, speed, mechanical strength, and component compatibility.

Technological Characteristics

Powered upper limb prosthesis with Li-Ion battery (11.1V). Components: Michelangelo Hand, AxonMaster control unit, electrodes (Titanium Grade 1 contact area, ASA housing, silicone pad), and passive wrist/elbow adapters. Connectivity: Bluetooth for PC-based adjustment via AxonSoft. Sensing: EMG electrodes (90-450 Hz). Materials: ASA, Titanium, Silicone, Cyanacrylate. Standards: ISO 22523, IEC 60601-1, ISO 10993. Software-controlled via microprocessor.

Indications for Use

Indicated for unilateral amputees requiring upper limb exoprosthetic fitting, including transradial, transhumeral, or dysmelia-related forearm/upper arm cases. Contraindicated for unusual activities involving excessive strain/shocks (e.g., contact sports, mountain biking, free climbing) and operation of motor vehicles or heavy industrial machinery.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # ottobock. 510(k) Summary Axon-Bus Prosthetic System Project document Nr. PB-PP00120100A-641 Page 1 of 13 # K123795 # 510(k) Summary # 1 Submitter Information Manufacturer: Otto Bock Healthcare Product GmbH Establishment Registration: 9615892 Kaiserstrasse 39 Adress: 1070 Vienna Austria Contact Person: Reinhard Wolkerstorfer Regulatory Affairs Manager reinhard.wolkerstorfer@ottobock.com Telephone: +43 1 523 37 86 676 Fax Number: +43 1 523 22 64 Date Prepared: October 8, 2012 Date Revised: May 30, 2014 # 2 Device Information | . Trade Name: | Axon-Bus Prosthetic System | |------------------------------|------------------------------------------------| | Common or Usual Name: | Powered, External Upper Limb Prosthetic System | | Classification Name: | Cutaneous Electrode (21 CFR 882.1320) | | Classification Product Code: | GXY (Electrode, Cutaneous) | | Subsequent Product Code: | IQZ (Hand, External Limb Component, Powered) | # 3 Identification of Legally Marketed Predicate Devices | Name: | MyoSystem with Customizing | |---------------|------------------------------------| | Manufacturer: | Otto Bock HealthCare Products GmbH | | 510k Number: | K032833 | | Date Cleared: | June 3rd, 2010 | | Name: | i-limb System | |---------------|---------------------| | Manufacturer: | Touch Bionics, Inc. | | 510k Number: | - | | Date Cleared: | - | # JUN 0 3 2014 {2}------------------------------------------------ | Name: | Pro Hand System | |---------------|---------------------| | Manufacturer: | Motion Control, Inc | | 510k Number: | - | | Date Cleared: | - | | | | | Name: | Bebionic System | | Manufacturer: | RSL Steeper | | 510k Number: | - | | Date Cleared: | - | # 4 Description of the Device ## 4.1 General Description The Axon-Bus Prosthetic System is to be used exclusively for upper limb exoprosthetic fitting. The Axon-Bus Prosthetic System is suitable for unilateral amputations starting with the transradial/transhumeral amputation level or, in case of dysmelia, for forearm or upper arm fittings. The Axon-Bus prosthetic system was developed for everyday use and must not be used for unusual activities. These unusual activities include, for example, sports with excessive strain and/or shocks to the wrist unit (pushups, downhill, mountain biking) or extreme sports (free climbing, paragliding, etc.). Furthermore, the Axon-Bus prosthethic system should not be used for the operation of motor vehicles, heavy equipment (e.g. construction machines), industrial machines or motor-driven equipment. The prosthesis is intended exclusively for use on one patient. Use of the product by another person is not approved by the manufacturer. Fitting a patient with the Axon-Bus prosthetic system may only be carried out by a prosthetist who has been authorized by Ottobock after completion of a corresponding training course. System Components: - Michelangelo Hand (terminal device) . - AxonFlexion Adapter (passive flexion) . - . AxonRotation Adapter (passive rotation) - AxonArm (passive elbow joint with mechanical and/or electrical lock) . - AxonEnergy Integral (battery) . - AxonCharge (charger) � - AxonMaster (control unit) . - Electrode (detecting EMG Input signals) . - AxonSoft (adjustment software) . - ♥ AxonSkin (prosthetic glove) ## 4.2 Device Functions The components of the Axon-Bus Prosthetic System are assembled by a prosthetist according to the individual needs of the amputee. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "ottobock." in a bold, sans-serif font. The text is black against a white background. The word is followed by a period. Adjustments to the prosthesis components can be performed through Bluetooth® data transfer using the AxonSoft software. The Bluetooth module is integrated into the control unit. The program is selected for the respective user situation from among four control option. The adjustment software running on a personal computer allows the prosthetist to adjust the settings of the system by selecting the right program from among four control option. The hand component contains two drives. The main drive of the hand component is responsible for the gripping movements and gripping force. Actively driven elements are the thumb, index finger and middle finger while the ring finger and little finger passively follow the other fingers. The thumb drive permits electronic positioning. Rotating the thumb outward creates a wide open palm, so that additional movement options are possible. Flexion and extension (bending and stretching) are based on the relaxed wrist (flexible mode). Pronation and supination (inward and outward rotation) can be passively performed by the user. The elbow component allows passive flexion and extension. Locking and unlocking is carried out mechanically (e.g., by means of body harnesses) or electrically (e.g., by means of electrode). | Feature | OPEN | CLOSE | Indication | |----------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Channel | Input signal no. 2 | Input signal no. 1 | For patients with 2<br>(strong) electrode<br>signals | | Needed signals | Sustained electrode<br>signal | Sustained electrode<br>signal. | | | Application of gripping<br>force | N.A. (no gripping force<br>applied during opening<br>of the hand) | Proportional, according<br>to strength of the<br>electrode signal<br>(resulting from the<br>muscle contraction). | The neutral position of<br>the Michelangelo Hand<br>is the initial position for<br>Opposition Mode and<br>Lateral<br>Mode. | | Application of gripping<br>speed | Proportional, according to strength of the electrode<br>signal (resulting from the muscle contraction). | | | | Stopping | No electrode signal | | | ## 4.3 Description and Comparison of the Control Options #### Comparison of this control option with the predicate devices: No differences (all predicate devices and the Axon-Bus Prosthetic System (2-Multi) are controlled proportional (gripping force and gripping speed) with 2 electrodes. {4}------------------------------------------------ Page 4 of 13 | Control Option 2 - "2-ELE" | | | | |----------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Feature | OPEN | CLOSE | Indication | | Channel | Input 2 | Input 1 | For patients with 2<br>(weak) electrode<br>signals. | | Needed signals | Sustained electrode<br>signal | Sustained electrode<br>signal. | | | Application of gripping<br>force | N.A. (no gripping force<br>applied during opening<br>of the hand) | Proportional, according<br>to strength of the<br>electrode signal<br>(resulting from the<br>muscle contraction). | The open position of the<br>Michelangelo Hand is<br>the initial position for<br>Lateral Mode,<br>Opposition<br>Mode and Neutral<br>Mode. | | Application of gripping<br>speed | Proportional, according to strength of the electrode<br>signal (resulting from the muscle contraction). | | | | Stopping | No electrode signal | | | #### Comparison of this control option with the predicate devices: Both, Pro Hand System and the Axon-Bus Prosthetic System (2-ELE) can be controlled with reduced proportionality (gripping force and gripping speed) with 2 electrodes. The MyoSystem with Customizing offers reduced proportionality for only 1 electrode. No detailed information about other predicate devices. {5}------------------------------------------------ Page 5 of 13 | Establishing digital control option (using and adjustment of "2-ELE") | | | | |-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | OPEN | CLOSE | Indication | | Channel | Input 2 | Input 1 | For patients with 2 very<br>weak electrode signals. | | Needed signals | Sustained electrode<br>signal | Sustained electrode<br>signal. | | | Application of gripping<br>force | N.A. (no gripping force<br>applied during opening<br>of the hand) | Time proportional: The<br>level of the gripping<br>force is determined by<br>the duration of the<br>electrode signal | Adjust working range<br>according to the weak<br>electrode signals -> if<br>selected threshold is<br>reached, constant<br>gripping force and<br>constant gripping speed<br>are applied (digital<br>behavior). | | Application of gripping<br>speed | constant speed by<br>means of a muscle<br>signal at any level<br>exceeding the adjusted<br>threshold | constant speed by<br>means of a muscle<br>signal at any level<br>exceeding the adjusted<br>threshold | | | Stopping | No electrode signal | | The open position of the<br>Michelangelo Hand is<br>the initial position for<br>Lateral Mode,<br>Opposition<br>Mode and Neutral<br>Mode. | ## Comparison of this control option with the predicate devices: Both, MyoSystem with Customizing and the Axon-Bus Prosthetic System (using and adjustment of control option "2-ELE") are controlled with constant speed and time proportional gripping force with 2 electrodes. No detailed information about other predicate devices. | Control Option 3 - "1-Vario" | | | | |----------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------| | Feature | OPEN | CLOSE | Indication | | Channel | Input 1 | | For patients with 1 | | Needed signals | Increasing signal<br>through muscle<br>contraction | Declining signal through<br>muscle relaxation | (strong) electrode signal<br>The open position of the<br>Michelangelo Hand is | | Application of gripping<br>force | N.A. (no gripping force<br>applied during opening<br>of the hand) | Proportional to the<br>decline of the electrode<br>signal | the initial position for<br>Lateral Mode,<br>Opposition | | Application of gripping<br>speed | Proportional:<br>The speed is determined by the speed and strength<br>of the muscle contraction/relaxation. | | Mode and Neutral<br>Mode. | | Stopping | Through very slow muscle relaxation via the<br>electrode. | | | ## Comparison of this control option with the predicate devices: Pro Hand System, MyoSystem with Customizing and the Axon-Bus Prosthetic System (1-Vario) are controlled with increasing/declining signals with 1 electrode. No detailed information about other predicate devices. {6}------------------------------------------------ | Control Option 4 - "1-Double" | | | | |-------------------------------|-------------------------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | Feature | OPEN | CLOSE | Indication | | Channel | Input 1 | | For patients with 1 strong electrode signal | | Needed signals | with a fast and strong signal | with a slow, gentle signal | The open position of the Michelangelo Hand is the initial position for Lateral Mode,<br>Opposition Mode and Neutral Mode. | | Application of gripping force | N.A. (no gripping force applied during opening of the hand) | determined by the duration of the electrode signal | | | Application of gripping speed | Constant | | | | Stopping | No electrode signal | | | ### Comparison of this control option with the predicate devices: Pro Hand System, MyoSystem with Customizing and the Axon-Bus Prosthetic System (1-Double) are controlled with time proportional gripping force and constant gripping speed with 1 electrode. A high or strong muscle signal opens the hand, and a low signal closes the hand. No detailed information about other predicate devices. ## 4.4 Description and Comparison of Switching Methodes The AxonMaster offers five switching methodes. The prosthetist selects the switching methodes according to the user needs. The switching events are triggered by muscle signals of the user. As a result of a triggered switching event the user can switch between different joints and gripping patterns. | Method | Co-<br>Contraction | Four-Channel<br>Control | Pulse | Long Open | Automatic<br>Switching after<br>Neutral Position is<br>Reached | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| | Description | | | | | | | Required<br>signals | 2 | 2 | 1 | 1 | 0 (no signal<br>required, switching<br>after neutral position<br>is reached) | | Indication | For patients<br>with co-<br>contraction<br>ability | For patients with 2<br>fast HI signals | For<br>patients<br>with one<br>HI signal | For patients who<br>can maintain the<br>open hand<br>position with a<br>high signal. | For patients who<br>have a preferred<br>type of grip | | Execution | Both<br>electrode<br>signals<br>must<br>exceed the<br>co-<br>contraction<br>thresholds<br>within a<br>certain time,<br>and then<br>decline. | Quick and strong<br>signals from either<br>electrode controls<br>one joint or gripping<br>pattern. Slow and<br>weak signals from<br>either electrode<br>controls another<br>joint. | The<br>electrode<br>signal<br>must<br>exceed<br>the pulse<br>threshold<br>and then<br>decline. | The electrode<br>signal for<br>opening must be<br>held at the value<br>for the maximum<br>speed for a<br>certain period of<br>time when the<br>hand is already<br>fully open. | Switching is<br>triggered after the<br>neutral position has<br>been reached and a<br>certain delay time<br>has passed. | {7}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the word "ottobock." The word is written in a bold, sans-serif font. The text is black and the background is white. There is a period at the end of the word. ## Comparison of the switching methods with the predicate devices: #### Co-Contraction: No differences, all predicate devices and the Axon-Bus Prosthetic System offer Co-Contraction switching mode ## Four-Channel Control: The MyoSystem with Customizing and the Axon-Bus Prosthetic System offer Four-Channel Control for switching between the joints. ### Pulse: The i-limb System uses double and triple impulse for switching. The Axon-Bus Prosthetic System uses single impulse for switching. #### Long Open: The Bebionic System and the Axon-Bus Prosthetic System offer the long opening method for switching between the grip patterns. #### Switching when Neutral Position is reached: No predicate device uses this switching method. ## 4.5 Technological and Performance Characteristics | System | | | |--------------------------------------------|------------------------------------------|--------------------| | Power Source | rechargeable battery | | | PC Software for adjustment | Yes | | | Software/Firmware/ Microprocessor Control? | Yes | | | Hand | | | | Operating temperature [°C] | -10 to +60 | | | Weight [g] | 600 (incl. Passive flexion and rotation) | | | max. Gripping force (opposition mode) [N] | 70 | | | max. Gripping force (lateral mode) [N] | 60 | | | max. Gripping force (neutral mode) [N] | 15 | | | max. Grip Speed [mm/s] | 325 | | | max. Opening Width [mm] | 120 | | | Battery | | | | Chemistry | Li Ion | | | Number of cells | 3 | | | Battery capacity [mAh] | various (1150 / 1500) | | | Nominal voltage [V] | 11,1 | | | Charging time [hrs] | max. 3,5 | | | Battery weight [g] | 142 | | | Battery dimensions (LxWxH) [mm] | various (75x60x21 / 55x35x23) | | | Installation | integrated | | | Electrode | | | | Electrode Temp. Range | -15°C to 60°C | | | Material<br>of Main | Housing | Plastics (ASA) | | | Contact Area | Titanium (Grade 1) | | | Pad | Silicone | | | Bonding Agent | Cyanacrylate | | Frequency Bandwidth [Hz] | 90 - 450 | | | Adjustment | Potentiometer 1-7 | | | Installation | suspension arms / suction socket | | | Dimensions [mm] | 27 x 18 x 9,5 | | | Elbow | | | | Elbow Temp. Range [°C] | 0 to 60 | | | Automatic Forearm Balance | Yes | | | Lock | Mechanical electric | | {8}------------------------------------------------ | Maximum vertical load with locked elbow joint and<br>a forearm length of 305 mm / 12 inch [kg] | 23 | |------------------------------------------------------------------------------------------------|----------| | Weight [g] (without hand, wrist and battery) | 750 | | Max. forearm length [mm] | 305 | | Flexion angle [°] | 15 - 145 | The Axon-Bus Prosthetic System offers the following functions: - Multiple control strategies . - O Different Control Options available for individual patient needs - Different grip patterns . - Power grip o - Three-Point grip O - Two-Point grip 0 - Rest position 0 - o Two Thumb positions - Passive wrist functions . - Mechanical and electrical elbow lock . ## 4.6 Comparison to Predicate Devices The Axon-Bus Prosthetic System has similar intended use and technical features compared to the predicate devices listed above. #### 4.6.1 Similarities The subject device shows the following similarities in functional performance: #### Available control options: | Similarity: | Various control options available for individual patient needs | |--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Description: | All devices allow the selection of various control strategies. Regarding application of<br>gripping force and gripping speed all devices allow the selection of proportional control<br>methods. The Axon-Bus Prosthetic System and the MyoSystem with Customizing also allow<br>the selection of digital control methods (time proportional application of gripping force and<br>constant gripping speed).<br>Regarding number of input channels all devices allow the control via 2 individual electrode<br>signals. Pro Hand System, MyoSystem with Customizing and the Axon-Bus Prosthetic<br>System allow also the control via 1 electrode signal. | #### Available switching methodes: Various switching methodes available for individual patient needs Similarity: All devices allow the patient to switch between various joints and gripping patterns. All Description: devices offer the Co-Contraction switching method. The MyoSystem with Customizing and the Axon-Bus Prosthetic System offer Four-Channel Control for switching between the joints. The Bebionic System and the Axon-Bus Prosthetic System offer the long opening method for switching between the grip patterns. #### Available grip patterns: Similarity: Various grip patterns available for individual patient needs All devices allow the patient to use various grip patterns. All devices allow power grip which Description: allows holding objects with large diameters. In addition all devices allow a two-point grip, whereby objects can be fixated between two fingers. The Axon-Bus prosthetic System also allows a three point grip (also supported by i-limb System and Bebionic System), which allows the user to fixate small objects. The Axon-Bus Prosthetic System, the i-limb System and the Bebionic System allow the user to switch between 2 individual thumb positions, whereby the users can switch between opposition and lateral grip. The Axon-Bus Prosthetic {9}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "ottobock." in bold black font. The word is lowercase and followed by a period. The font is sans-serif and the letters are closely spaced. System, the i-limb System and the Bebionic System allow the user to use a rest position, which gives the hand a natural and physiological appearance. Passive wrist functions Similarity: Passive rotation and flexion of the wrist available All devices allow the patient to rotate the wrist joint passive. The Pro Hand System and the Description: Axon-Bus Prosthetic System also allow passive flexion of the wrist joint. Due to wrist rotation pro- and supination of the hand is possible. Mechanical and electrical elbow lock Similarity: Mechanical and electrical locking mechanism available for the elbow joint Description: The MyoSystem with Customizing and the Axon-Bus Prosthetic System allow mechanical and electrical lock of the elbow joint. The Bebionic System allows electrical lock of the elbow joint. ### 4.6.2 Differences and Comparison The subject device shows the following differences: | Weight of Hand: | Description<br>& Comparison: | The Axon-Bus Prosthetic System's hand is a little bit heavier in weight compared to the<br>predicate devices. | |------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Discussion: | This is because the Axon-Bus Prosthetic System supports also passive flexion for the wrist.. | | Grip Forces: | Description<br>& Comparison: | Grip forces vary for the different devices. | | | Discussion: | The Axon-Bus Prosthetic System's grip forces for the different types of grips are at least<br>higher than compared to one or more predicate devices. The deviations in grip force do not<br>result in any risk for the user. | | Speed: | Description<br>& Comparison: | Grip Speed varies for the different devices. | | | Discussion: | The Axon-Bus Prosthetic System's grip speed is higher than compared to all predicate<br>devices and therefore it does not result in any risk for the user. | | Battery Characteristics: | Description<br>& Comparison: | The Axon-Bus Prosthetic System has a rechargeable Li-Ion Battery Pack with 11,1 V. Most<br>of the predicated devices use rechargeable 7V Li-Ion Battery Packs. | | | Discussion: | The higher capacity and nominal voltage is needed to provide the performance related to<br>grip force and grip speed. This also leads to varieties in charging time, weight and<br>dimensions. None of these deviations do result in any risk for the user. | | | Material of Electrode Contact Area: | | | Description<br>& Comparison: | The contact area of the Axon-Bus Prosthetic System electrode is made up of titanium (grade<br>1). The electrode of the predicate device (MyoSystem with Customizing) uses stainless steel | | | Discussion: | Titan is known as a material with good biocompatibility properties, and is therefore very<br>common e.g. for use in implants. Nevertheless, the used material (titanium grade 1) is tested<br>according to biocompatibility requirements of ISO 10993. | | | | Material of other Parts Contacting Patient Skin (Electrode Housing, Silicone Pad): | | | Description<br>& Comparison: | The Housing of the Axon-Bus Prosthetic System Electrode is made up of ASA, whereas the<br>Housing of the predicate device (MyoSystem with Customizing) uses ABS. The Axon-Bus<br>Prosthetic System Electrode uses also a Silicone Pad. | | | Discussion: | The used Silicone Pad is certified according to USP Class VI. In addition, all used materials<br>(ASA, Silicone Pad and Bonding Agent) are tested according to biocompatibility<br>requirements of ISO 10993. | | | Index-Point Grip: | | | | Difference: | The Axon-Bus Prosthetic System does not support an index-point grip. | | | Description<br>& Comparison: | The I-limb System offers an Index-Point Grip: thumb, little, ring and middle fingers close and<br>switch off. Only the index finger will move. This option is particularly useful when operating a<br>computer keyboard, cell phone or elevator button.<br>The Bebionic System also offers an index-point grip, where middle, ring and small fingers<br>close against the palm and the thumb is driven against the middle finger. Once this position<br>is selected, typing on a keyboard or input pad, pressing a bell or a button can be achieved.<br>The other predicate devices do not offer this type of grip. | | | Discussion: | This additional available option of the predicate devices is considered only for special<br>activities of daily living (e.g. typing on a computer keyboard, etc.). Therefore the difference in<br>the available grip options does not raise any questions with respect to safety and<br>effectiveness in comparison to the predicate device. | | | Switching Method - "Pulse": | | | | Difference: | The Axon-Bus Prosthetic System offers uses single impulse for switching, whereas the i-limb<br>System uses double and triple impulse for switching. | | | Description<br>& Comparison: | A single impulse (single muscle signal generated by the user) can be selected as switching<br>method for the Axon-Bus Prosthetic System, and as a result the user can switch between<br>different joints and gripping patterns. The i-limb System also allows the selection of a<br>switching method triggered via impulses, but to trigger such a switching the user has to<br>generate double impulses (two uninterrupted muscle signals) or triple impulses (three<br>uninterrupted muscle signals).<br>The other predicate devices do not offer this type of switching method. | | | Discussion: | This switching method needs only 1 impulse instead of 2 or 3 impulses of the muscle signal.<br>For the user these switching methods differ only in the number of impulses which must be | | {10}------------------------------------------------ {11}------------------------------------------------ | ottobock. | | | |-----------|--|--| |-----------|--|--| 510(k) Summary Axon-Bus Prosthetic System generated. Therefore this difference in the available switching method does not raise any questions with respect to safety and effectiveness in comparison to the predicate device. | Switching Method - "Automatic Switching after Neutral Position is Reached": | | | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|--|--|--|--| | Difference: | No predicate device uses this switching method. | | | | | | | | | | | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | | | | # Description & Comparison: This switching method does not require a muscle signal for triggering Switching is triagered after the neutral position of the hand has been reached, and a certain delay time has passed. This switching method is for patients who have a preferred type of grip. So it such a Discussion: preferred grip type and this switching method are selected, this helps the user to start the next movement always with the preferred type of grip. Therefore this allows the user to control the prosthetic device more efficiently, but it does not raise any questions with respect to safety and effectiveness in comparison to the predicate device. #### Positioning of thumb: Difference: Electrical rotation of the thumb instead of passive rotation #### Description & Comparison: The Axon-Bus Prosthetic System allows the patient to move the thumb lateral to the index fingers so that the user can grip items from the side. This is established electronically, whereas the i-limb System and the Bebionic System allow this lateral movement of the thumb only in a mechanically way. The MyoSystem with Customizing and the Pro Hand System do not offer this lateral movement of the thumb (neither electronically nor mechanically). Discussion: The electrical positioning of the thumb does not require the other hand of the patient for moving the thumb. Therefore this feature helps the patient to control the prosthetic device more efficiently, but it does not raise any questions with respect to safety and effectiveness in comparison to the predicate device. #### 4.6.3 Conclusion of Comparison to Predicate Devices: All patient contacting materials of the Axon-Bus Prosthetic System are equivalent to those of the predicates and are in compliance with ISO 10993-1. The subject device is also substantially equivalent to its predicates based on comparison of functional and performance characteristics. The differences of the subject device don't raise any questions with respect to safety and effectiveness of the device. # 5 Indications for Use The Axon-Bus Prosthetic System is to be used exclusively for exoprosthetic fittings of the upper limbs. {12}------------------------------------------------ Page 12 of 13 # 6 Summary of Performance Testing The following Performance Standards were used for performance testing of the Axon-Bus Prosthetic System: | No. | Title | Version | Comments | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------| | ISO 22523 | External limb prostheses and external orthoses<br>- Requirements and test methods | 2006-<br>10-01 | | | IEC 60601-1 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance | 2005 | Protection against electrical<br>hazards from ME equipment:<br>Classification according to<br>chapter 6, testing according<br>to chapter 8. | | IEC 60601-1-2 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements<br>and tests | 2007 | Electromagnetic<br>Compatibility | | ISO 10993-1 | Biological evaluation of medical devices<br>Part 1: Evaluation and testing within a risk<br>management process | 2009 | Biocompatibility | | ISO 10993-5 | Biological evaluation of medical devices Part<br>5: Tests for in vitro cytotoxicity | 2009 | | | ISO 10993-10 | Biological evaluation of medical devices<br>Part 10: Tests for irritation and skin<br>sensitization | 2009 | | | EN 980 | Symbols for use in the labeling of medical<br>devices | 2008 | Labeling Symbols | | IEC 60721-3-2 | Classification of environmental conditions Part<br>3: Classification of groups of environmental<br>parameters and their severities - Section 2:<br>Transport | 1997 | | | IEC 60068-2-1 | Environmental testing Part 2-1: Tests - Test A:<br>Cold | 2007 | | | IEC 60068-2-2 | Environmental testing Part 2-2: Tests - Test B:<br>Dry heat | 2007 | | | IEC 60068-2-30 | Environmental testing Part 2-30: Tests - Test<br>Db: Damp heat, cyclic (12 + 12-hour cycle) | 2005 | | | IEC 60068-2-78 | Environmental testing Part 2-78: Tests - Test<br>Cab: Damp heat, steady state | 2001 | Shipping test | | IEC 60068-2-14 | Environmental testing Part 2: Tests - Test N:<br>Change of temperature | 1986 | | | IEC 68-2-64<br>(EN 60068-2-64) | Environmental testing Part 2-64: Test methods<br>- Test Fh: Vibration, broad-band random<br>(digital control) and guidance | 1993 | | | IEC 68-2-27<br>(EN 60068-2-27) | Environmental testing Part 2-27: Tests - Test<br>Ea and guidance: Shock | 1987 | | | IEC 68-2-32<br>(EN 60068-2-32) | Basic environmental testing procedures Part 2:<br>Tests - Test Ed: Free fall | 1990 | | Table 1: Standards used for performance testing Extensive testing was performed (e.g. software testing and electrical safety tests including EMC). Also biocompatibility testing according to ISO 10993-1 was performed on all patient contacting materials. No clinical studies were performed. Since all samples tested met the acceptance criteria, substantial equivalence has been demonstrated through these tests. {13}------------------------------------------------ **ottobock.** In Addition, internal testing regarding performance characteristics of the hand, wrist and elbow has been conducted on the Axon-Bus Prosthetic System to ensure that the device meets design specifications, operates as it is intended, and ensures that safety functions and features operate as they are intended. | Component | Performance Attribute | Test Result | |------------|-------------------------------------------------------------------------------------------------|-------------| | Hand | Weight of the prosthesis | Passed | | | Compatibility of components | Passed | | | Activating and deactivating the system | Passed | | | Performing different types of grips | Passed | | | Gripping different sizes of objects, max. opening distance | Passed | | | Grip force (strength, increase and decrease, maintaining grip force when<br>engine not running) | Passed | | | Positioning speed | Passed | | | Emergency opening | Passed | | | Mechanical strength (static and dynamic testing) | Passed | | | | | | Wrist | Positions of flexion and rotation unit | Passed | | | Mechanical strength (static and dynamic testing, support a specific<br>weight) | Passed | | | Connecting/ disconnecting of flexion adapter to/from rotation adapter | Passed | | Elbow | Mechanical strength (static and dynamic testing) | Passed | | | Maintaining position (Maximum holding force): Overload Protection | Passed | | Tube Valve | Function of Tube Valve for Suction Socket | Passed | Table 2: Overview of internal testing regarding performance characteristics All tests that has been conducted on the Axon-Bus Prosthetic System to ensure that the device meets design specifications, operates as it is intended, and ensures that safety functions and features operate as they are intended were passed. # 7 Conclusion The Axon-Bus Prosthetic System is safe and effective for its intended use. The Axon-Bus Prosthetic System is substantially equivalent to the predicate devices. {14}------------------------------------------------ Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with outstretched arms, possibly representing care and protection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 3, 2014 Otto Bock Health Care Products Reinhard Wolkerstorfer Regulatory Affairs Manager Kaiserstrasse 39 1070 Vienna AUSTRIA Re: K123795 Trade Name: Axon-Bus Prosthetic System Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY, IQZ Dated: May 20, 2014 Received: May 23, 2014 Dear Mr. Wolkerstorfer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of I he general condons proving practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease be advised that I Dri 3 1558anted or or device complies with other requirements of the Act that I DA has inade a decommiations administered by other Federal agencies. You must {15}------------------------------------------------ ## Page 2 - Mr. Reinhard Wolkerstorfer comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Felipe Aquel -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {16}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K123795 Device Name Axon-Bus Prosthetic System Indications for Use (Describe) The Axon-Bus Prosthetic System is to be used exclusively for exoprosthetic fittings of the upper limbs. Type of Use (Select one or both, as applicable) [ Prescription Use (Part 21 CFR 801 Subpart D) O Over-The-Counter Use (21 CFR 801 Subpart C) ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # ല --- Date: 2014.06.03 21:19:23 -04'00' Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. {17}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...