← Product Code [CAW](/productcode/CAW) · K123738

# MARK 5 NUVO LITE OCSI, MARK 5 NUVO LITE STD (K123738)

_Nidek Medical Products, Inc. · CAW · Feb 28, 2013 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123738

## Device Facts

- **Applicant:** Nidek Medical Products, Inc.
- **Product Code:** [CAW](/productcode/CAW.md)
- **Decision Date:** Feb 28, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5440
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The modified devices are intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life.

## Device Story

Portable oxygen concentrator; uses pressure swing adsorption (PSA) to separate oxygen from ambient air; draws air via piston-type compressor through molecular sieve adsorbent; nitrogen retained, oxygen enriched gas output (90% +6.5%/-3% at 5 l/min). Used in home/clinical settings; operated by patients or caregivers. Modifications include horizontal compressor orientation, two-part cabinet back, improved capacitor, and 5A circuit breaker. Provides oxygen-enriched gas via nasal cannula/tubing. Benefits patients by providing supplemental oxygen therapy. Status indicators (power, oxygen concentration/pressure) managed by model-specific PCB.

## Clinical Evidence

Bench testing only. No clinical data provided. Device tested to ISO 8359:1996, EN 60601-1-2:2001, IEC 60601-1:1988 + A1:1991 + A2:1995, UL 60601-1:2003, and CAN/CSA-C22.2 No 601.1-M90.

## Technological Characteristics

AC-powered oxygen concentrator; PSA process using synthetic zeolite. Piston-type compressor (horizontal axis). ABS thermoplastic enclosure (UL-94 V0). Filtration: 5um inlet air filter, 0.3um bacterial filter. Output: 90% (+6.5%/-3%) O2 at 5 l/min. Connectivity: Standalone. Power: 115V/230V. Circuit breaker: 5A. Capacitor: P2 protection per IEC 20252.

## Regulatory Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

## Predicate Devices

- Mark 5 Nuvo Lite OCSI ([K082566](/device/K082566.md))
- Mark 5 Nuvo Lite Std ([K082566](/device/K082566.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# FEB 2 8 2013

.

#### 510(k) Summary (Rev 12/21/12) б

| 510(k) Summary              | This summary of 510(k) safety and effectiveness information<br>is being submitted in accordance with the requirements of 21<br>CFR 807.92.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                       |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Submitter                   | Nidek Medical Products, Inc.<br>3949 Valley East Industrial Drive<br>Birmingham, AL 35217<br>Phone (205) 856-7200<br>Fax (205)856-0533                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                       |
| Submission<br>Correspondent | Van Muth<br>Regulatory Affairs<br>Nidek Medical Products<br>lmuth@nidekmedical.com<br>Phone (205) 856-7200 x242<br>Fax (205)856-0533                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                       |
| Date Prepared               | December 19, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                       |
| Trade Name                  | Modified Device #1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Mark 5 Nuvo Lite OCSI |
|                             | Modified Device #2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Mark 5 Nuvo Lite STD  |
| Common Name                 | Oxygen Concentrator or Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                       |
| Classification Name         | Generator, Oxygen Portable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                       |
| Regulation Number           | 21CFR 868.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                       |
| Class                       | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                       |
| Panel                       | Anesthesiology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                       |
| Product Code                | CAW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                       |
| Predicate Device #1         | Mark 5 Nuvo Lite OCSI oxygen concentrator cleared under<br>K082566 on December 3, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                       |
| Predicate Device #2         | Mark 5 Nuvo Lite Std oxygen concentrator cleared under<br>K082566 on December 3, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                       |
| Performance<br>Standards    | No applicable mandatory performance standards or special<br>controls have been established for this device under section<br>514 of the Federal Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                       |
| Indications for Use         | The modified devices are intended solely for medical use in<br>oxygen therapy programs under the supervision of a<br>physician. This device is available by prescription only and<br>is not intended to support or sustain life.<br>There has been no change to the intended use compared to<br>the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                       |
| Device Description          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                       |
| Device Description          | The modified devices are substantially equivalent to the predicate devices. Modifications were made to improve reliability, ease of assembly and serviceability The compressor was changed from a vertically oriented tension spring supported configuration to a horizontally oriented compression spring supported configuration to minimize damage from shipment and handling. The cabinet back was changed from a single part to two parts to simplify assembly and serviceability. Labeling was changed to increase type size for improved readability and to correct some clerical errors. The motor run capacitor was changed to a type that includes an internal protective device. The device circuit breaker was changed from a 10 A rating to a 5 A rating. The modified devices continue to be AC powered devices that provide a high level of inspired oxygen, using the same method as that of the predicate devices, by separating oxygen from ambient air with the pressure swing adsorption (PSA) process. Air is drawn into the device with a piston-type compressor and exposed to molecular sieve adsorbent that selectively retains nitrogen and other components that are released when the pressure is vented to the atmosphere. This cycle is controlled by an electronic valve and protected from over pressurization by the compressor's pressure relief valve. The device provides a nominal oxygen enriched gas concentration of 90% +6.5%/-3% at a flow rate of 5 l/min ±10%. The modified device is a durable, reusable, semi-portable unit weighing approximately 32 pounds [14.5 kg.]. The device is available in both 115V and 230V models. The device status indicators are the same as in the predicate devices: They are controlled by a printed circuit board which is device model specific. For the OCSI model, device status indicators are mains power and oxygen concentration. For the STD model, device status indicators are mains power and system pressure. The modified device is not life supporting, life sustaining or sterile. The modified devices may be used with one of the many legally marketed humidifiers, connecting tubing and nasal cannula as prescribed. Recommended accessory devices are described in the User's Guide. One of these devices may optionally be included with the device. |                       |

{1}------------------------------------------------

#### 6 510(k) SUMMARY (REV 12/21/12) CONTINUED

.

{2}------------------------------------------------

| Technological<br>Characteristics         | The modified devices are substantially equivalent to the<br>predicate devices as they utilize the same technology currently<br>used in the predicate Nuvo Lite concentrators.<br>The following table indicates the similarities and differences<br>between the modified devices and the predicates:        |                                                                                                                                                                                                               |  |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristic                           | Modified Devices<br>#1 & #2<br>Mark 5 Nuvo Lite                                                                                                                                                                                                                                                            | Predicate Devices<br>#1 & #2<br>Mark 5 Nuvo Lite                                                                                                                                                              |  |
| Intended Use                             | Same as Predicate                                                                                                                                                                                                                                                                                          | Intended solely for medical use in<br>oxygen therapy programs under the<br>supervision of a physician. This<br>device is available by prescription<br>only and is not intended to support<br>or sustain life. |  |
| Usage Environment                        | Same as Predicate                                                                                                                                                                                                                                                                                          | Home/Clinical                                                                                                                                                                                                 |  |
| Operating Life                           | 3 years                                                                                                                                                                                                                                                                                                    | 5 years                                                                                                                                                                                                       |  |
| Oxygen Concentration                     | Same as Predicate                                                                                                                                                                                                                                                                                          | 90% (±6.5/-3%)<br>0.125 to 5 Ipm                                                                                                                                                                              |  |
| Oxygen Concentration<br>Status Indicator | Same as Predicate                                                                                                                                                                                                                                                                                          | OSCI - Yes<br>STD - No                                                                                                                                                                                        |  |
| Operating<br>Ranges                      | Same as Predicate                                                                                                                                                                                                                                                                                          | 40°F - 100°F<br>(5°C - 40°C)<br>up to 95% RH                                                                                                                                                                  |  |
| Storage<br>Ranges                        | Same as Predicate                                                                                                                                                                                                                                                                                          | -4°F - 140°F<br>(-20°C - 60°C)<br>up to 95% RH                                                                                                                                                                |  |
| Over-Temperature<br>Protection           | Same as Predicate                                                                                                                                                                                                                                                                                          | UL 547 Class B<br>145 +/- 5°C                                                                                                                                                                                 |  |
| Avg. Power<br>Consumption                | Same as Predicate                                                                                                                                                                                                                                                                                          | 330 WATTS                                                                                                                                                                                                     |  |
| Input Power                              | Same as Predicate                                                                                                                                                                                                                                                                                          | 115V ~ 60 Hz                                                                                                                                                                                                  |  |
| Flow Regulation<br>Assembly              | Same as Predicate                                                                                                                                                                                                                                                                                          | Single-stage diaphragm pressure<br>reducer                                                                                                                                                                    |  |
| Flow Settings                            | Same as Predicate                                                                                                                                                                                                                                                                                          | 0.125 to 5 lpm                                                                                                                                                                                                |  |
| Flow Control                             | Same as Predicate                                                                                                                                                                                                                                                                                          | Flow Valve<br>(12 positionorifice)                                                                                                                                                                            |  |
| Noise Level                              | Same as Predicate                                                                                                                                                                                                                                                                                          | 45 dBA                                                                                                                                                                                                        |  |
| Compressor                               | Self-lubricated piston type;<br>Horizontal axis                                                                                                                                                                                                                                                            | Self-lubricated piston type<br>Vertical axis                                                                                                                                                                  |  |
| Compressor Free Flow                     | Same as Predicate                                                                                                                                                                                                                                                                                          | 2.7 scfm / 76 l/m                                                                                                                                                                                             |  |
| Compressor Support                       | Compression Steel Springs<br>Horizontal Axis                                                                                                                                                                                                                                                               | Tension Steel Springs<br>Vertical Axis                                                                                                                                                                        |  |
| Compressor Relief<br>Valve               | Same as Predicate                                                                                                                                                                                                                                                                                          | 45 psig/310kPa                                                                                                                                                                                                |  |
| Compressor Cooling                       | Same as Predicate                                                                                                                                                                                                                                                                                          | Tube-Axial Fan                                                                                                                                                                                                |  |
| Characteristic                           | Modified Devices<br>#1 & #2<br>Mark 5 Nuvo Lite                                                                                                                                                                                                                                                            | Predicate Devices<br>#1 <br>Mark 5 Nuvo Lite                                                                                                                                                                 |  |
| Electric Motor                           | Same as Predicate                                                                                                                                                                                                                                                                                          | Inductive type                                                                                                                                                                                                |  |
| Capacitor                                | P2 Protection per IEC20252                                                                                                                                                                                                                                                                                 | P0 Class Protection per IEC 20252                                                                                                                                                                             |  |
| Process Timing                           | Same as Predicate                                                                                                                                                                                                                                                                                          | Timer Printed Circuit Board                                                                                                                                                                                   |  |
| Air Separation Module                    | Same as Predicate                                                                                                                                                                                                                                                                                          | Two beds                                                                                                                                                                                                      |  |
| Sieve Material                           | Same as Predicate                                                                                                                                                                                                                                                                                          | Zeochem synthetic zeolite                                                                                                                                                                                     |  |
| Outlet Pressure                          | Same as Predicate                                                                                                                                                                                                                                                                                          | 7 psig/48kPa                                                                                                                                                                                                  |  |
| Filtration                               | Same as Predicate                                                                                                                                                                                                                                                                                          | Polyurethane foam 10 ppi cells,<br>nominal 5 um inlet air filter, and<br>0.3 µm bacterial filter                                                                                                              |  |
| Protective Enclosure                     | Same as Predicate                                                                                                                                                                                                                                                                                          | ABS thermoplastic<br>UL-94 V0                                                                                                                                                                                 |  |
| Cabinet Dimensions                       | Same as Predicate                                                                                                                                                                                                                                                                                          | 14"W x 9"D x 23"H                                                                                                                                                                                             |  |
| Weight                                   | Same as Predicate                                                                                                                                                                                                                                                                                          | 32 lbs./14.5kg                                                                                                                                                                                                |  |
| Wiring                                   | Same as Predicate                                                                                                                                                                                                                                                                                          | Power cord, power on/off switch,<br>& inter connecting lead wires                                                                                                                                             |  |
| Power Cord                               | Same as Predicate                                                                                                                                                                                                                                                                                          | 16 Ga/1.3mm PVC Sheath, Integral                                                                                                                                                                              |  |
| Power Plug                               | Same as Predicate                                                                                                                                                                                                                                                                                          | Two Blade, Polarized                                                                                                                                                                                          |  |
| Circuit Breaker                          | 5 Amp                                                                                                                                                                                                                                                                                                      | 10 Amp                                                                                                                                                                                                        |  |
| Tubing<br>Materials                      | Same as Predicate                                                                                                                                                                                                                                                                                          | PVC, Aluminum, Silicone, &<br>Polyurethane                                                                                                                                                                    |  |
| Labeling                                 | Same as Predicate                                                                                                                                                                                                                                                                                          | English Text & Symbols                                                                                                                                                                                        |  |
| Optional Accessories                     | Same as Predicate                                                                                                                                                                                                                                                                                          | Legally marketed nasal cannula,<br>connecting tubing, & humidifiers                                                                                                                                           |  |
| Labeling                                 | Same as Predicate                                                                                                                                                                                                                                                                                          | English Text & Symbols                                                                                                                                                                                        |  |
| Nonclinical<br>Performance               | The device was tested to applicable requirements ISO<br>8359:1996, EN 60601-1-2:2001, IEC 60601-1:1988 + A1:1991<br>+ A2:1995, UL 60601-1:2003, CAN/CSA-C22.2 No 601.1-<br>M90, and FDA Reviewer Guidance document "Excerpts<br>Related to EMI from November 1993" as appropriate to the<br>area of usage. |                                                                                                                                                                                                               |  |
| Conclusion                               | The modified devices include all the same or similar materials,<br>operate with the same technology and demonstrate the same<br>performance so are substantially equivalent to the legally<br>marketed predicate devices.                                                                                  |                                                                                                                                                                                                               |  |

.

:

{3}------------------------------------------------

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms, representing health, services, and people. The figure is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## February 28, 2013

Mr. Van Muth Regulatory Affairs Nidek Medical Products, Incorporated 3949 Valley East Industrial Drive BIRMINGHAM AL 35217

Re: K123738

Trade/Device Name: Mark 5 Nuvo Lite OCSI, Mark 5 Nuvo Lite STD Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: January 21, 2013 Received: January 31, 2013

### Dear Mr. Muth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Kwame O. Ulmer

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

#### 5 STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): Device Names:

Mark 5 Nuvo Lite OCSI Mark 5 Nuvo Lite STD

Indications for Use:

The modified devices are intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life.

k123738

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

## Lester V 2013.02.28

(Division Sign-Off) Division of Anesthesiology, General Hospi Infection Control, Dem

510(k) Number:

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K123738](https://fda.innolitics.com/device/K123738)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
