K123708 · Usa Paramed Medical Systems, Inc. · MOS · Apr 10, 2013 · Radiology
Device Facts
Record ID
K123708
Device Name
SPINE COIL
Applicant
Usa Paramed Medical Systems, Inc.
Product Code
MOS · Radiology
Decision Date
Apr 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the SPINE COIL is to enable the MrOpen K101295 systems designed by Paramed to perform MR scan of the human spine either in the conventional recumbent position or with the patient sitting or standing (weight-bearing). The SPINE COIL inherits the same limitations of the Tomograph. The SPINE COIL enables the MRI System to which it is applied to acquire images of the Spine in transverse, sagittal, coronal and oblique cross-sectional directions. The images which are produced correspond to spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon MR parameters (spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
Device Story
Spine Coil 03-2016 is a flexible surface coil accessory for Paramed MrOpen MRI systems; enables spine imaging in recumbent, sitting, or standing (weight-bearing) positions. Device functions as an RF receiver coil to acquire proton density signals; images processed by the MRI system to display T1, T2, flow velocity, and chemical shift data. Used in clinical settings by trained MRI technicians/physicians. Output interpreted by medical experts for diagnostic purposes. Flexible design enhances patient comfort during weight-bearing exams. No patient contact with skin is intended; materials include polycarbonate and ethylene vinyl acetate copolymer foam.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence supported by comparison of technical specifications (SNR, channel count) and material composition to the predicate device.
Indicated for patients requiring MR imaging of the spine in recumbent, sitting, or standing (weight-bearing) positions using the Paramed MrOpen MRI system.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "PARAMed MEDICAL SYSTEMS". The word "PARAMed" is in large, bold, sans-serif font, with "PARAMed" in a cursive font. Below "PARAMed" is the text "MEDICAL SYSTEMS" in a smaller, sans-serif font.
. .
. ﺑ
. .
52/109
# 8. 510(K) Summary [807.92]
APR 1 0 2013
The following Special 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92.
#### Manufacturer Information
| Name & Address | Paramed Srl |
|----------------------|----------------------------------|
| | Corso Perrone 73R |
| | 16152 Genova, Italy |
| Submitted by: | Luca Vescovo, Authorized officer |
| Establishment Reg. # | 3004994584 |
807.92(a)(1)_The submitter's name, address, telephone number, a contact person, and the date the summary was prepared;
#### Submitter Information
Correspondent:
Contact Person
Date
Rim Vesan
29/11/2012
Richard Olson, Correspondent USA Paramed Medical Systems, Inc 6204 W. Oakton Street Morton Grove, IL 60053 Toll Free: 1 866 840-7565 T 1 847 470-0580nF 1 847 470-0612 Luisella De Benedetti Paramed S.r.l. Corso F.M. Perrone 73R 16152 Genova +39 010 6489 358 luisella.debenedetti@paramed.it
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "Paramed Medical Systems". The word "Paramed" is in large, bold, sans-serif font. Below it, in a smaller font, are the words "Medical Systems". The word "Paramed" is in all caps, while the words "Medical Systems" are in small caps.
807.92(a)(2) The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known;
| Trade Name: | SPINE COIL |
|-------------------------------------|----------------------------------------------------|
| Common Name: | Accessory for Magnetic resonance diagnostic system |
| Classification Name(s): | System, Nuclear Magnetic Resonance Imaging |
| Classification and class of device: | 21 CFR 892.1000, class II |
| Classification Number: | 90MOS |
| | |
807.92(a)(3) An identification of the legally marketed device to which the submitter claims equivalence. A legally marketed device to which a new device may be compared for a determination regarding substantial equivalence is a device that was legally marketed prior to May 28, 1976, or a device which has been reclassified from class III to class II or I (the predicate), or a device which has been found to be substantially equivalent through the 510(k) premarket notification process;
#### Predicate Devices
Paramed
MrOpen 0.5T
K101295
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Paramed Medical Systems. The word "Paramed" is in large, bold, sans-serif font. Below that, the words "Medical Systems" are in a smaller, sans-serif font. The "d" in "Paramed" has a stylized, cursive extension.
807.92(a)(4) A description of the device that is the subject of the premarket notification submission, such as might be found in the labeling or noctional material for the device, including an explanation of how the promocronal macorais the scientific concepts that form the basis for the device, and the significant physical and performance characteristics of the device, such as device design, material used, and physical properties;
The proposed SPINE COIL is a surface coil aimed to the Spine district. It is characterized by the fact of being flexible. This is done to increase patient comfort.
Below the pictures of the Flexible coil (code 03-2015) and the proposed SPINE COIL are shown to give an idea of the difference.
Image /page/2/Picture/4 description: The image contains two separate pictures of what appears to be a CPAP machine. The picture on the left shows the machine with a white box on top of it and a tube connected to the side. The picture on the right shows the side of the machine with a tube connected to it.
Figure 1: Flexible coil 03-2015 on the left and SPINE COLL 03-2016 on the right
Thanks to the flexible structure the proposed and the predicate coils grant patient comfort when employed in weight bearing examinations. Both coils can be used in recumbent position just as the cleared 03-2015 coil.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Paramed Medical Systems". The word "Paramed" is in bold, block letters, and the words "Medical Systems" are in smaller block letters underneath. The word "Paramed" is followed by a stylized script font.
The significant physical and performance characteristics of the SPINE COIL are here resumed:
# Examined anatomy:
Spine in all the sections
## Patient positions
Recumbent
Sitting or standing
### Device design
Paramed S.r.l. is a company specialized in the design of NMR Tomographs since the very beginning. The company is ISO 13485 certified since year 2006 and the Quality management System is laid down to grant overall process management including design. Written procedures are the guideline for designers who are highly specialized physicists. The Quality manager performs internal audits to to grant procedures' respect. Mechanical drawings/specifications document each component being manufactured or bought.
#### Materials used
Paramed S.r.l. has decided to cover its devices with fire preventing ULV0 covers to minimize fire risks or damages due to external fire.
Materials contacting patient or operator are Polycarbonate and Ethylene vinyl acetate copolymer foam which are the same already employed for previously cleared devices (in the specific to Paramed's K101295 MrOpen Generic Purpose 03-2010 coil.
Cleaning procedures are the same as per the other receiving coils and are described in the User Manual.
55/109
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Paramed Medical Systems. The word "Paramed" is written in a bold, sans-serif font on the top line. Below that, the words "Medical Systems" are written in a smaller, sans-serif font.
807.92(a)(5) A statement of the intended use of the device that is the subject of the premarket notification submission, including a general description of the diseases or conditions that the device will diagnose, treat, prevent, cure, or mitigate, including a description, where appropriate, of the patient population for which the device is intended. If the indication statements are different from those of the legally marketed device identified in paragraph
## Device Intended Use(s)
The intended use of the SPINE COIL is to enable the MrOpen K101295 systems designed by Paramed to perform MR scan of the human spine either in the conventional recumbent position or with the patient sitting or standing (weight-bearing). The SPINE COL inherits the same limitations of the Tomograph. The SPINE COIL enables the MRI System to which it is applied to acquire images of the Spine in transverse, sagittal, coronal and oblique cross-sectional directions. The images which are produced correspond to spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon MR parameters (spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "Paramed Medical Systems". The word "Paramed" is in a bold, sans-serif font, with the "d" having a cursive-like extension. Below "Paramed" is the text "MEDICAL SYSTEMS" in a smaller, sans-serif font. The logo is in black and white.
807.92 (a) (6) If the device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device identified in paragraph (a) (3) of this section, a summary of the technological characteristics of the new device in comparison to those of the predicate device. [omissis]
#### Technological Characteristics
The SPINE COIL 03-2016 is substantially equivalent to
- Paramed MrOpen K101295 Flexible COIL 03-2015 .
| Technological<br>Characteristic | SPINE COIL 03-2016 | Flexible coil 03-2015<br>predicate K101295 |
|----------------------------------|--------------------|--------------------------------------------|
| Channels | 4 | 1 |
| Minimum SNR | 55 | 65 |
| Minimum SNR for each<br>channels | 40 | n.a. |
| Examined anatomy | Spine | Spine |
- Paramed MrOpen K101295 General Purpose COL 03-2010
| Technological<br>Characteristic | SPINE COIL 03-2016 | General purpose coil 03-<br>2010 predicate K101295 |
|---------------------------------|------------------------------------------|----------------------------------------------------|
| Material [outer surface] | Ethylene vinyl acetate<br>copolymer foam | Ethylene vinyl acetate<br>copolymer foam |
| Manufacturer | Leidel & Kracht | Leidel & Kracht |
No new materials are in contact with the patient because:
- · The new coil is manufactured employing the same materials as the predicate devices' coils (no new material)
- · No contact with the skin is considered as normal use. The User Manual addresses the need to prevent coil contamination by minimizing contact of the coil with the patient's skin. No invasive use is intended, hence no contact with undamaged skin or body fluids.
Also in the event of a foreseeable misuse there is no change in the type and duration of contact with patient between the actual and predicate devices which are aimed to perform the same examinations according to the same procedures.
52/09
{7}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2013
LUISELLA DE BENEDETTI QUALITY MANAGER PARAMED S.R.L. CORSO F.M. PERRONE 73R, 16152 GENOVA ITALY
Re: K123708
Trade/Device Name: Spine Coil 03-2016 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 11, 2013 Received: March 21, 2013
Dear Ms. De Benedetti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll
{8}------------------------------------------------
Page 2-Ms. De Benedetti
free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mechael Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123708
Device Name: SPINE COIL 03-2016
Indications for Use:
The intended use of the SPINE COIL is to enable the MrOpen K101295 systems designed by Paramed to perform MR scan of the human spine either in the conventional recumbent position or with the patient sitting or standing (weight-bearing). The SPINE COIL inherits the same limitations of the Tomograph. The SPINE COIL enables the MRI System to which it is applied to acquire images of the Spine in transverse, sagittal, coronal and oblique cross-sectional directions. The images which are produced correspond to spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon MR parameters (spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael FOX 'Hara
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K123708
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.