K123681 · Wipro GE Healthcare Private, Ltd. · MOS · Feb 22, 2013 · Radiology
Device Facts
Record ID
K123681
Device Name
1.5T 4CH LARGE FLEX COIL, 1.5 4CH SMALL FLEX COIL
Applicant
Wipro GE Healthcare Private, Ltd.
Product Code
MOS · Radiology
Decision Date
Feb 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Flex Coils are designed to produce diagnostic images of human anatomies and are used as general purpose coils. The Flex Coils are receive-only coils designed to be used with 1.5T MRI systems manufactured by GE Healthcare to provide images of various parts of human body including shoulder, elbow, wrist, knee, hip and ankle. Large flex coil can be used for scanning large anatomy and small flex coil can be used for scanning smaller anatomy. The flexible coil is wrapped around the anatomy of interest, such as an elbow or knee or can be used for planar imaging.
Device Story
1.5T 4CH Flex Coils are receive-only, flexible MRI coils; wrapped around anatomy (shoulder, elbow, wrist, knee, hip, ankle) or used for planar imaging. Device contains 4 elements with integrated preamplifiers; tuned to proton frequency of 63.86MHz. Operates as phased array to combine images from 4 channels. Used in clinical settings with GE 1.5T MR systems; operated by trained MRI technicians/radiologists. Output is raw signal data processed by host MRI system into diagnostic images; assists clinicians in anatomical visualization and diagnosis.
Clinical Evidence
No clinical data. Bench testing only; internal scans performed to obtain sample images.
Technological Characteristics
4-element phased array receive-only coil; integrated preamplifiers; tuned to 63.86MHz. Flexible form factor for wrapping around anatomy. Complies with IEC60601-1, IEC60601-2-33, IEC60601-1-2, ISO14971, and ISO10993-1. No software algorithm; hardware-based signal acquisition.
Indications for Use
Indicated for patients requiring diagnostic MRI imaging of human anatomy, including shoulder, elbow, wrist, knee, hip, and ankle, using GE 1.5T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
K123681 Page 1 of 3
## GE Healthcare 510(k) Premarket Notification Submission
# FEB 2 2 2013
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | 26-Nov-2012 |
| Submitter: | GE Healthcare, (Wipro GE Healthcare Pvt Ltd)<br>122 (Part 1), John F. Welch Technology Centre,<br>Export Promotional Industrial Park, Whitefield,<br>Bangalore, INDIA 560066 |
| Primary Contact Person: | Shashidhar C S<br>Regulatory Affairs Leader<br>GE Healthcare, (Wipro GE Healthcare Pvt Ltd)<br>122 (Part 1), John F. Welch Technology Centre,<br>Export Promotional Industrial Park, Whitefield,<br>Bangalore, INDIA 560066<br>Telephone: +91 80 40883613<br>Fax: +91 80 28411645 |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director - MR<br>GE Healthcare,(GE Medical Systems LLC)<br>3200N Grandview Blvd., Mail Coil-W-827<br>Waukesha, WI 53188, USA<br>Phone: (262) 521-6848<br>Fax: (262) 364-2785 |
| Device: | |
| Trade Name: | 1.5T 4CH FLEX COIL |
| Common/Usual Name: | Coil, magnetic resonance, specialty |
| Classification Names: | 21CFR 892.1000, Magnetic resonance diagnostic device |
| Product Code: | MOS |
| Predicate Device(s): | K030953, GE 3.0T General Purpose Flex Coil |
| Device Description: | The 1.5T 4CH Flex Coils are receive-only coils designed<br>to provide images of various parts of human body. The<br>flexible coil is wrapped around the anatomy of interest,<br>such as an elbow or knee or can be used for planar<br>imaging. |
The 1.5T 4CH Flex coils are available in two models –
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The letters are connected and have a flowing, elegant appearance. The logo is in black and white.
K123681 Page 2 of 3
GE Healthcare 510(k) Premarket Notification Submission
1.5T 4CH Large Flex coil and 1.5T 4CH Small Flex Coil, and are designed for use with GE 1.5T MR systems.
Intended Use: The Flex Coils are designed to produce diagnostic images of human anatomies and are used as general purpose coils. The Flex Coils are receive-only coils designed to be used with 1.5T MRI systems manufactured by GE Healthcare to provide images of various parts of human body including shoulder, elbow, wrist, knee, hip and ankle. Large flex coil can be used for scanning large anatomy and small flex coil can be used for scanning smaller anatomy. The flexible coil is wrapped around the anatomy of interest, such as an elbow or knee or can be used for planar imaging.
Technology: 1.5T 4CH Flex Coils are flexible and can be wrapped around the anatomy of interest, which is similar to its predicate device.
> 1.5T 4CH Flex coils are General Purpose receive only coils with 4 elements and integrated preamplifiers.
> 1.5T 4CH Flex coils are based on phased array technique for combining the images from 4 different channels. Coils are tuned to the proton frequency of 63.86MHz.
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
**Substantial Equivalence:**
The 1.5T 4CH Flex Coil complies with voluntary standards IEC60601-1, IEC60601-2-33, IEC60601-1-2, ISO14971, and ISO10993-1.
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Requirements Reviews ●
- Design Reviews �
- . Performance testing
- Integration testing .
- Safety testing .
{3}------------------------------------------------
K123681 Page 3 of 3
GE Healthcare 510(k) Premarket Notification Submission
- Maximum B1 Peak test -- This test is to verify the . coil's ability to withstand maximum B1 peak fields and high B1 field energy concentrations without posing a risk to safety through arcing or voltage breakdown
- Blocking Network analysis This test determines . the effectiveness of the blocking networks(s) for transmit decoupling to ensure safety and to minimize B1 distortion
- Surface temperature test under normal condition .
- Surface temperature test under unplugged condition .
#### Summary of Clinical Tests:
Internal scans within GE Healthcare facility were performed to obtain sample clinical images. The subject of this premarket submission, 1.5T 4CH Flex Coil did not require external clinical studies to support substantial equivalence
Conclusion: GE Healthcare considers the 1.5T 4CH Flex Coil to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Section 05
Image /page/2/Picture/11 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are white against a black background, and the circle is outlined in white.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2013
Shashidhar C S Regulatory Affairs Leader WIPRO GE HEALTHCARE PRIVATE, LTD. 122 (Part 1) John F Welch Trading Centre, EPIP, Whitefield Rd BANGALORE, INDIA 560066
Re: K123681
Trade/Device Name: 1.5T 4CH FLEX COIL Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 26, 2012 Received: November 30, 2012
Dear Mr. Shashidhar C S:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic . product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Shashidhar C S
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael Hara
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
K123681
Device Name:
1.5T 4CH FLEX COIL
Indications For Use:
The Flex Coils are designed to produce diagnostic images of human anatomies and are used as general purpose coils. The Flex Coils are receive-only coils designed to be used with 1.5T MR1 systems manufactured by GE Healthcare to provide images of various parts of human body including shoulder, elbow, wrist, knee, hip and ankle. Large flex coil can be used for scanning large anatomy and small flex coil can be used for scanning smaller anatomy. The flexible coil is wrapped around the anatomy of interest, such as an elbow or knee or can be used for planar imaging.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Michael FDA Hara
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.