← Product Code [OWN](/productcode/OWN) · K123676

# CELLVIZIO 100 SERIES AND CELLVIZIO SYSTEMS WITH CONFOCAL MINIPROBES (K123676)

_Mauna Kea Technologies · OWN · Apr 18, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123676

## Device Facts

- **Applicant:** Mauna Kea Technologies
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Apr 18, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Cellvizio 100 Series System with Confocal Miniprobes is a confocal laser system with fiber optic probes that is intended to allow imaging of the internal microstructure of tissues in anatomical tracts, i.e., gastrointestinal. The AQ-Flex 19TM member of the GastroFlex M series of Confocal Miniprobes can be used within anatomical tracts, i.e., gastrointestinal, accessed by an endoscope or endoscopic accessories, including through EUS-FNA needles.

## Device Story

AQ-Flex 19 is a fiber-optic confocal miniprobe used with Cellvizio 100 Series/Cellvizio Systems. Device captures real-time optical images of internal tissue microstructure. Operates via confocal laser scanning through fiber optics. Used by physicians during endoscopic procedures; compatible with EUS-FNA needles via a locking accessory replacing the stylet. Output displayed for clinician visualization to assist in tissue assessment. Benefits include enhanced visualization of internal microstructure during minimally invasive procedures.

## Clinical Evidence

Bench testing only. Verification and validation testing demonstrated compatibility of the AQ-Flex 19 probe with endoscopes and endoscopic accessories, including EUS-FNA needles.

## Technological Characteristics

Confocal laser imaging system using fiber optic probes. Includes a locking accessory for secure placement within EUS-FNA needles. Sterilization instructions provided. System architecture remains consistent with previously cleared GastroFlex M series.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Cellvizio 100 Series System and Cellvizio System with Confocal Miniprobes ([K120208](/device/K120208.md))
- Expect Endoscopic Aspiration Needle ([K112198](/device/K112198.md))
- EchoTip Ultra Ultrasound Needle System ([K083330](/device/K083330.md))
- EchoTip Ultra High Definition Ultrasound Access Needle ([K092359](/device/K092359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Mauna Kea Technologies

Page ① of ↷

Traditional 510(k)

## Section 8.Premarket Notification 510(k) Summary

APR 1 8 2013

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

| 510(k) Number: | K123676 |
|----------------|---------|
|----------------|---------|

#### Applicant Information:

Mauna Kea Technologies Name: Address: 9 rue d'Enghien F-75010 Paris, France

Phone: +33 1 48 24 03 45 Fax: +33 1 48 24 12 18

Contact Person:

Michael A Daniel, Consultant

| Phone Number:     | (415) 407-0223 |
|-------------------|----------------|
| Office:           | (925) 254-5228 |
| Facsimile Number: | (925) 254-5187 |

#### Device Information: ·

| Device Trade Name:        | Cellvizio® 100 Series System and Cellvizio® System with Confocal<br>Miniprobes™ |
|---------------------------|---------------------------------------------------------------------------------|
| Common Name:              | Endoscope and Accessories                                                       |
| Classification Name(s):   | Confocal Optical Imaging                                                        |
| Product Code/ Regulation: | OWN / 21 CFR 876.1500                                                           |
| Classification:           | Class II                                                                        |

#### Predicate Device:

- The AQ-Flex 19 Confocal Miniprobes are identical in all respects to the Confocal Miniprobe . GastroFlex M most recently cleared with Cellvizio 100 Series System and Cellvizio System with Confocal Miniprobes, K120208.
- Endoscopic UltraSound Needles : Expect Endoscopic Aspiration Needle (K112198 from Boston . Scientific), EchoTip Ultra Ultrasound Needle System (K083330 from Cook Endoscopy) and EchoTip Ultra High Definition Ultrasound Access Needle (K092359 from Cook Ireland Ltd)

#### Device Description:

The subject device, "AQ-Flex 19" is a member of the GastroFlex M family of devices cleared via K122042 and K120208. It is identical in design to the GastroFlex M devices described in K120208. There are no differences in design, materials, or function between the subject AQ-Flex 19 device and the recently cleared GastroFlex M devices.

{1}------------------------------------------------

#### Mauna Kea Technologies

page (2) of (2)

Traditional 510(k)

A locking accessory is being added to the package as an aid to secure placement inside a EUS-FNA needle in place of a Stylet. Instructions for Sterilization have also been added.

Indications for Use:

The Cellvizio 100 Series System with Confocal Miniprobes is a confocal laser system with fiber optic probes that is intended to allow imaging of the internal microstructure of tissues in anatomical tracts, i.e., gastrointestinal.

The AQ-Flex 19TM member of the GastroFlex M series of Confocal Miniprobes can be used within anatomical tracts, i.e., gastrointestinal, accessed by an endoscope or endoscopic accessories, including through EUS-FNA needles.

#### Comparison to Predicate Device:

No change is being made to the subject/predicate device. The subject device is identical to the neviously cleared device (K120208). The indication for use is being expanded to include selected needle-based endoscopic procedures.

Verification and validation testing has shown that the AQ-Flex 19 Confocal Miniprobes are compatible with endoscopes or endoscopic accessories designed and commonly used in these procedures (such as endoscopic needles).

#### Summary:

The AQ-Flex 19 M Confocal Miniprobes, when used as part of the Cellvizio 100 Series and the Cellvizio Systems, have been shown to be substantially equivalent to cleared predicate devices and can be used as intended to image the internal microstructure of tissues in a variety of anatomical locations. These previously cleared probes have been verified to be compatible with endoscopes and endoscopic accessories designed to be used in these applications.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem consists of a stylized eagle or bird-like figure with three overlapping, curved lines representing its wings or body. The text is arranged in a circular fashion around the emblem.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

April 18, 2013

Mauna Kea Technologies % Daniel and Daniel Consulting Mr. Michael Daniel 8 Snowberry Court Orinda, California 94563

Re: K123676

Trade/Device Name: Cellvizio 100 Series and Cellvizio Systems with Confocal Miniprobes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OWN Dated: April 01, 2013 Received: April 02, 2013

Dear Mr. Daniel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Michael Daniel

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, FOR

# Peter Di Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Section 7.Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Cellvizio® 100 Series System and Cellvizio® System with Confocal Miniprobes™

## Indications for Use:

The Cellvizio 100 Series System with Confocal Miniprobes is a confocal laser system with fiber optic probes that is intended to allow imaging of the internal microstructure of tissues in anatomical tracts, i.e., gastrointestinal.

The AQ-Flex 19TM member of the GastroFlex M series.of Confocal Miniprobes can be used within anatomical tracts, i.e., gastrointestinal. accessed by an endoscope or endoscopic accessories, including through EUS-FNA needles.

Prescription Use **X** AND/OR Over-The-Counter-Use *_*
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE

Neil R Oge 2013.04.17

(Division Sign-Off) for MXM

Division of Surgical Devices

510(k) Number K123676

---

**Source:** [https://fda.innolitics.com/device/K123676](https://fda.innolitics.com/device/K123676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
