← Product Code [EIH](/productcode/EIH) · K123664

# INCORIS ZI (K123664)

_Sirona Dental Systems GmbH · EIH · Apr 8, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123664

## Device Facts

- **Applicant:** Sirona Dental Systems GmbH
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Apr 8, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Classic sintering - Framework and reduced crowns in the anterior and posterior tooth region . - Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics . - Crown caps in the anterior and posterior region . - Cone and telescoping crowns . Speed sintering - Framework and reduced crowns in the anterior and posterior tooth region . - Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 units - Crown caps in the anterior and posterior region . - Cone and telescoping crowns Super speed sintering - Framework and reduced crowns with a maximum wall-thickness of 2 mm .

## Device Story

inCoris ZI consists of zirconia ceramic (ZrO2) blocks used for custom dental restorations. Blocks are mounted to metal carriers and processed via Sirona CAD/CAM grinding machines. Following grinding, restorations undergo sintering (classic, speed, or super speed) to achieve final material properties. Used in dental clinics/labs by trained professionals. Output is a custom-fit dental restoration (crowns, bridges, caps). Benefits include biocompatible, high-strength ceramic restorations tailored to patient anatomy.

## Clinical Evidence

No clinical data. Bench testing performed, including sintering tests, mechanical property evaluation, crack damage inspection, restoration fit, color suitability, and usability testing.

## Technological Characteristics

Material: Yttria-stabilized zirconia ceramics (ZrO2). Standards: ISO 6872:2008, ISO 13356:2008, ISO 10993-1:2009. Physical properties: Density ≥ 6.05 g/cm3, Bending strength > 1100 MPa. Form factor: Block with metal carrier for CAD/CAM spindle mounting. Processing: CAD/CAM grinding followed by sintering.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- inCoris ZI ([K062509](/device/K062509.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K.123.669

# APR 0 8 2013

## 510(k) Summary for Sirona Dental Systems inCoris ZI

### 1. Sponsor

Sirona Dental Systems GmbH

Fabrikstrasse 31.

D-64625 Bensheim

Germany

Contact Person: Fritz Kolle

Telephone: +49 6251 16 3294

Date Prepared: March 05, 2013

#### Device Name 2.

Proprietary Name: inCoris ZI

Common/Usual Name: Powder, Porcelain

Classification Name: Porcelain powder for clinical use

#### 3. Predicate Devices

inCoris ZI (K062509)

#### Intended Use 4.

Classic sintering

- · Framework and reduced crowns in the anterior and posterior tooth region
- . Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics
- . Crown caps in the anterior and posterior region
- Cone and telescoping crowns .

### Speed sintering

- . Framework and reduced crowns in the anterior and posterior tooth region
- · Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 . units
- · Crown caps in the anterior and posterior region
- · Cone and telescoping crowns

Sirona Dental Systems 510(k) inCoris ZI

Page 1

APPENDIX F

{1}------------------------------------------------

Super speed sintering

- Framework and reduced crowns with a maximum wall-thickness of 2 mm
#### న. Device Description and Function

The inCoris ZI are blocks of various sizes from which custom made dental restorations are grinded using Sirona CAD/CAM system. inCoris ZI ceramics constitute blocks comprised of zirconia ceramics (ZrO2). The blocks are initially manufactured in a partially sintered state; then, they are o individually processed to specification, and finally, densely sintered. One end plane of a block is mounted to a metal carrier that is inserted in the spindle's clamping chuck of the grinding machine. The blocks are available in different colors.

#### 6. Scientific Concept

The underlying scientific concept is

- . Processing dental restorations by Sirona Dental CAD/CAM System
- . Restorations are grinded from an inCoris ZI block by a Sirona CAM machine
- . Different sintering time to gain appropriate material properties

### 7. Physical and Performance Characteristics

### 7.1. Design

The design of the inCoris ZI is described in section 5, Device Description and Function.

### 7.2. Material Used

inCoris ZI ceramics constitute blocks comprised of zirconia ceramics (ZrO2). One end plane of a block is mounted to a metal carrier that is inserted in the spindle's clamping chuck of the grinding machine. The material is biocompatible according to ISO 10993-1: 2009, "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process".

### 7.3. Physical Properties

Final technical data of densely sintered inCoris ZI.

| Density:                                        | $\geq$ 6.05 g cm-3 |
|-------------------------------------------------|--------------------|
| Thermal expansion<br>coefficient (20 - 500 °C): | 11.0 10-6 K-1      |
| Bending strength:                               | > 1100 MPa         |

Sirona Dental Systems 510(k) inCoris ZI

Page 2

APPENDIX F

7.4. Chemical Properties

{2}------------------------------------------------

| Component      | inCoris ZI |
|----------------|------------|
| ZrO2+HfO2+Y2O3 | ≥ 99.0%    |
| Y2O3           | 5.2%       |
| HfO2           | 2%         |
| Al2O3          | < 0.35%    |
| Fe2O3          | < 0.3%     |

#### 8. Summary of the technological characteristics

Both, Proposed and Predicate Sirona inCoris ZI are made of zirconia ceramics(ZrO2) and block shaped. Both devices meet ISO 6872: 2008, "Dentistry -- Ceramic materials" and ISO 13356: 2008, "Implants for surgery, Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP)". Physical and chemical properties are similar.

### Nonclinical Testing 9.

Sirona performed a series of tests to assess whether the device is appropriate for the indications for use. Sintering tests coupled with bench mechanical testing highlight that the mechanical properties are appropriate. Furthermore, crack damage inspection, restoration fit, color suitability, and overall usability tests were conducted.

#### 10. Clinical Testing

Clinical tests have not been performed.

### 11. Conclusion

Based on the comparison of intended use, indications, contra-indications, material properties and processing/fabrication, Sirona Dental Systems believes that the Proposed and Predicate (K062509) Sirona inCoris ZI blocks are substantially equivalent.

Sirona Dental Systems 510(k) inCoris ZI

APPENDIX F

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human figures in a row, representing health and human services. The figures are depicted in a dark color, contrasting with the background.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

### April 8, 2013

Mr. Fritz Kolle Sirona Dental Systems GmbH Fabrikstrasse 31 Bensheim. Germany D-64625

Re: K123664

Trade/Device Name: inCoris ZI Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: March 5, 2013 Received: March 7, 2013

Dear Mr. Kolle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Kolle

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kwame O. Ulmer -S for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and

Dental Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K 12 3664

Device Name: ____ inCoris ZI

Indications for Use:

Classic sintering

- Framework and reduced crowns in the anterior and posterior tooth region .
- Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics .
- Crown caps in the anterior and posterior region �
- Cone and telescoping crowns ●

Speed sintering

- Framework and reduced crowns in the anterior and posterior tooth region ●
- . Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 units
- Crown caps in the anterior and posterior region .
- . Cone and telescoping crowns

Super speed sintering

- Framework and reduced crowns with a maximum wall-thickness of 2 mm .
Prescription Use X (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Sirona Dental Systems 510(k) | March 05, 2013 |
|------------------------------|----------------|
| inCoris ZI                   |                |

|  | Mary S. Runner -S                            |
|--|----------------------------------------------|
|  | Susan Runner, DDS, MA                        |
|  | 2013.04.04 10:34:20                          |
|  | (Division Sign-Off)                          |
|  | Division of Anesthesiology, General Hospital |
|  | Infection Control, Dental Devices            |

| 510(k) Number: | K123664 |
|----------------|---------|
|----------------|---------|

510(k) Number:_

---

**Source:** [https://fda.innolitics.com/device/K123664](https://fda.innolitics.com/device/K123664)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
