ViZion + DR is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion + allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
Device Story
ViZion + DR is a digital radiography system consisting of a flat panel digital detector (FPD) and image acquisition software. It captures X-ray signals to produce digital radiographic images, replacing conventional film-screen systems. The device is used in clinical settings by healthcare professionals to perform general radiographic examinations. The system integrates with PACS for image viewing and storage. The output is a DICOM-compliant digital image, which clinicians use to support diagnostic decision-making. The device benefits patients by providing the advantages of digital radiography, such as improved workflow and filmless environment, while maintaining image quality comparable to or better than previous systems.
Clinical Evidence
Clinical evaluation consisted of a qualitative assessment of clinical images by a board-certified radiologist. The study compared images acquired with the new panel against those from the predicate panel. The radiologist concluded that the image quality of the ViZion + DR was as good as or better than the predicate device.
Technological Characteristics
Digital flat panel X-ray detector system. Available in 14" x 17" and 17" x 17" sizes. Pixel sizes: 139 μm (17"x17") and 150 μm (14"x17"). Connectivity: DICOM output. Power: AC line. Electrical safety: IEC 60601-1. EMC: IEC 60601-1-2. Software: OPAL-RAD PACS interface.
Indications for Use
Indicated for general radiographic imaging of human anatomy, including skull, chest, shoulders, spine, abdomen, pelvis, and extremities, as a replacement for conventional screen-film systems. Contraindicated for fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
### 510(k) Summary 510(k) Number K12 364 Y
JAN 1 1 2013
#### Viztek, LLC. 6491 Powers Avenue Jacksonville, FL 32217 Phone: 800.366.5343 Fax: 904.448.9936 Date Prepared: November 9, 2012 Bruce Ashby, Sales and Marketing Manager Contact:
- 1. Identification of the Device: Proprietary-Trade Name: ViZion + DR Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) 90 MQB, Common/Usual Name: Digital X-Ray Receptor Panel
- 2. Equivalent legally marketed device: Viztek ViZion DR, K112661 and Atlaim ATAL 8, K113812.
- 3. Indications for Use (intended use) ViZion + DR is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion + allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
- 4. Description of the Device: The ViZion + DR system represents the straightforward integration of a new digital x-ray receptor panel and our previously cleared software. This is a MODIFICATION of our clearance K112661 wherein we have changed the supplier of the panel.
- ViZion + DR is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD) ViZion + is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and the new receptor panel were not changed. Our main predicate is ViZion DR, K112661, wherein we combined our OPAL-RAD software with a new digital panel. We now also offer two sizes of panels: 14" x 17" and 17" x 17" panels.
- 5. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate devices. Clinical images collected demonstrate equal or better image quality as compared to our predicates.
{1}------------------------------------------------
## 6. Substantial Equivalence Chart
| Characteristic | Viztek ViZion DR | Atlaim ATAL 8, K113812. | Viztek ViZion + DR |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K112661 | | |
| Intended Use: | ViZion DR is intended for<br>digital image capture use in<br>general radiographic<br>examinations, wherever<br>conventional screen-film<br>systems may be used,<br>excluding fluoroscopy,<br>angiography and<br>mammography. ViZion<br>allows imaging of the skull,<br>chest, shoulders, spine,<br>abdomen, pelvis, and<br>extremities. | The ATAL 8 and ATAL<br>8c is indicated for use 'in<br>general radiographic<br>images of human anatomy.<br>It is intended to replace<br>radiographic film/screen<br>systems in all general-<br>purpose diagnostic<br>procedures,<br>excluding fluoroscopic,<br>angiographic, and<br>mammographic<br>applications | ViZion + DR is intended<br>for digital image capture<br>use in general radiographic<br>examinations, wherever<br>conventional screen-film<br>systems may be used,<br>excluding fluoroscopy,<br>angiography and<br>mammography. ViZion +<br>allows imaging of the<br>skull, chest, shoulders,<br>spine, abdomen, pelvis,<br>and extremities. |
| Configuration | This submission is for the<br>Digital Panel and Software<br>only, no generator or stand<br>provided. | This submission is for the<br>Digital Panel and Software<br>only, no generator or stand<br>provided. | This submission is for the<br>Digital Panel and Software<br>only, no generator or stand<br>provided. |
| Digital Panel | Samsung LTX240AA01-A<br>(K090742)<br>Pixel size 143 μm<br>3072 x 3072 pixels 9<br>million pixels. | Pixel size 139 μm<br>3072 X 3072 pixels (9<br>million pixels | iRay Technology<br>(Shanghai) Ltd.<br>For the 17" x 17" panel:<br>Pixel size 139 μm<br>3064×3072 pixels (9<br>million pixels)<br>For the 14" x 17" panel:<br>Pixel size 150 μm<br>2288×2800 pixels (6.4<br>million pixels) |
| Software | Employs OPAL-RAD<br>PACS image viewing and<br>acquire interface software<br>technology, K063337 | The Atal-8 can be used<br>with any FDA-cleared off-<br>the-shelf radiographic<br>image acquisition software<br>package that outputs a<br>suitable DICOM image<br>like OmniVision software<br>by OmniVision, cleared<br>under K110040. | SAME as K112661, outputs a DICOM image. |
| DICOM | Yes | Option | Yes |
| Power source | AC Line | AC Line | AC Line |
| Electrical<br>safety and<br>EMC | Electrical Safety per IEC-<br>60601. UL listed | Electrical Safety per IEC<br>60601-1 and EMC per IEC<br>60601-1-2. | Electrical Safety per IEC<br>60601-1 and EMC per IEC<br>60601-1-2. |
- 7. Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety and Electromagnetic Compatibility. MTF and DQE measurements, Risk Analysis and Software validation was conducted in accordance with FDA guidance documents. .
{2}------------------------------------------------
- 8. Summary of Clinical Testing: Clinical images were acquired and evaluated by a board certified radiologist who concluded the images from the new panel are as good as or better than the images acquired with the predicate panel.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Viztek Inc that the Viztek ViZion + DR is as safe and effective as the predicate device, have few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 11, 2013
Viztek LLC % Mr. Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114
Re: K123644
Trade/Device Name: ViZion + DR. Digital Flat Panel X-Ray Detector System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-ray System Regulatory Class: II Product Code: MQB Dated: November 15, 2012 Received: November 28, 2012
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2-Mr. Kamm
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Sean M. Boyd -S
Janine M. Morris · Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use 510(k) Number (if known): K12:3644
Device Name: ViZion + DR, Digital Flat Panel X-Ray Detector System
#### Indications For Use:
ViZion + DR is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion + allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Sean M.Boyd -S.
(Division Sign Off)
Division of Radiological Health
Office of In Vitro Diagnostics and Radiological
51000 K123644
Page 1 of 1 .
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.