← Product Code [KNW](/productcode/KNW) · K123606

# CASSI II ROTATIONAL CORE BIOPSY SYSTEM; CASSI QUADPOINT DISPOSABLE (K123606)

_Scion Medical Technologies, LLC · KNW · Mar 1, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123606

## Device Facts

- **Applicant:** Scion Medical Technologies, LLC
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Mar 1, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The device is indicated for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.  It is not intended for use in bone. The device is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities.  It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.  The extent of histologic abnormality cannot be reliably determined from its mammographic appearance.  Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy).  When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

## Device Story

CASSI II Rotational Core Biopsy System is a biopsy instrument for soft tissue and breast abnormalities. System components include a control unit handle, valve block, battery, cryogen, biopsy needle (CASSI QuadPoint), and sample collection tray. Operation involves a sticking needle that uses cold temperatures at its tip to engage target tissue; a coaxially mounted cutting cannula/piston assembly then cores the tissue specimen. Used by clinicians in a clinical setting to obtain tissue for histologic examination. Provides diagnostic samples; assists in partial or complete removal of imaged abnormalities. Benefits include precise tissue acquisition for pathology.

## Clinical Evidence

No clinical data provided; submission relies on substantial equivalence to a previously cleared predicate device.

## Technological Characteristics

System includes control unit handle, valve block, battery, cryogen, and biopsy needle (sticking needle and cutting cannula/piston). Employs cryo-assisted tissue engagement and mechanical rotational coring. No software or electronic algorithm described.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Sanarus CASSI™ II Rotational Core Biopsy System ([K051581](/device/K051581.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123606
Page 1 of 2

MAR 1 2013

| 510(k) Summary                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's<br>Name and<br>Address: | Scion Medical Technologies, LLC<br>90 Oak Street<br>Newton, MA 02464<br>U.S.A.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Contact Name<br>and Information:    | Joseph Ostendorf<br>Regulatory Affairs Consultant<br>Scion Medical Technologies<br><br>Address: 23879 Blue Spruce Road<br>Sauk Centre, MN 56378<br>U.S.A.<br><br>Telephone: (503) 784-6756<br>Fax: (888) 582 - 6211<br>E-mail: jeostendorf@gmail.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Date Prepared:                      | 20 November 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Proprietary<br>Name(s):             | CASSI™ II Rotational Core Biopsy System, with the<br>CASSI QuadPoint™ Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Common Name:                        | Biopsy instruments and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code:                       | KNW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification of<br>Device:        | Class II, 21 CFR 876.1075                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate<br>Device:                | Sanarus CASSI™ II<br>Rotational Core Biopsy<br>System K051581 July 27, 2005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description:              | The proposed CASSI II Rotational Core Biopsy System consists<br>of the following primary components: a fully integrated control unit<br>handle (the CASSI II handle), a valve block, battery, cryogen, a<br>biopsy needle (CASSI QuadPoint comprised of a sticking needle<br>and cutting cannula/piston assembly), and a sample collection<br>tray. The sticking needle is operated by the control unit and uses<br>cold temperatures at its tip to engage the tissue to be sampled.<br>The cutting cannula is coaxially mounted around the sticking<br>needle and is used to core the tissue specimen. The cutting                                                                                                                  |
| Indications for<br>Use:             | The device is indicated for use in obtaining biopsies from soft<br>tissues such as liver, kidney, prostate, spleen, lymph nodes and<br>various soft tissue tumors.  It is not intended for use in bone.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                     | The device is also indicated to provide breast tissue samples for<br>diagnostic sampling of breast abnormalities.  It is designed to<br>provide breast tissue for histologic examination with partial or<br>complete removal of the imaged abnormality.  The extent of<br>histologic abnormality cannot be reliably determined from its<br>mammographic appearance.  Therefore, the extent of removal of<br>the imaged evidence of an abnormality does not predict the extent<br>of removal of a histologic abnormality (e.g., malignancy).  When<br>the sampled abnormality is not histologically benign, it is essential<br>that the tissue margins be examined for completeness of removal<br>using standard surgical procedures. |
| Technological<br>Characteristics:   | The purpose of this premarket notification is to seek clearance for<br>an updated branding and packaging configuration to the Sanarus<br>CASSI II Rotational Core Biopsy System (K051581) and to<br>summarize minor modifications made to the device since the last<br>submission.  The modified device will be marketed under the<br>trade names CASSI II Rotational Core Biopsy System and the<br>CASSI QuadPoint disposables.                                                                                                                                                                                                                                                                                                     |
| Conclusion:                         | In summary, Scion Medical Technologies, LLC believes that the<br>proposed CASSI II Rotational Core Biopsy System and the<br>CASSI QuadPoint disposables, as described in this submission,<br>do not raise any new or significant questions of safety and<br>efficacy and are substantially equivalent to the predicate Sanarus<br>CASSI II Rotational Core Biopsy System, which was determined<br>to be substantially equivalent and cleared on July 27, 2005<br>(K051581).                                                                                                                                                                                                                                                          |

540/1

{1}------------------------------------------------

K123606
Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains a logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 1. 2013

Scion Medical Technologies, LLC % Mr. Joseph Ostendorf Regulatory Affairs Consultant Ostendorf Consulting 23879 Blue Spruce Road SAUK CENTRE MN 56378

Re: K123606

Trade/Device Name: CASSI™ II Rotational Core Biopsy System with the CASSI QuadPoint™ Disposable Regulation Number: 21 CFR& 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: January 25, 2013 Received: January 31, 2013

## Dear Mr. Ostendorf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name:

CASSI™ II Rotational Core Biopsy System with the CASSI QuadPoint™ Disposable

Indications For Use:

The device is indicated for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.

The device is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.q., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -S

Page 1 of 1

luctive, Gastr

---

**Source:** [https://fda.innolitics.com/device/K123606](https://fda.innolitics.com/device/K123606)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
