← Product Code [GEX](/productcode/GEX) · K123573

# YOULASER CO2 LASER (K123573)

_Quanta System Spa · GEX · Apr 15, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123573

## Device Facts

- **Applicant:** Quanta System Spa
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Apr 15, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for: incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.

## Device Story

YOULASER CO2 is a 30-watt, 10.6 µm CO2 laser system. It consists of a power supply, laser source with air cooling, optical bench, articulated arm with CO2 scanner, and control electronics. Operated by clinicians in medical settings, the device delivers laser energy to soft tissues for surgical procedures or skin resurfacing. The system features a touch screen interface for parameter control. The microcontroller-based electronics manage power supply and laser emission. The device provides therapeutic benefits through precise tissue ablation and coagulation, facilitating various surgical and aesthetic interventions.

## Clinical Evidence

Bench testing only. Testing evaluated ablation, coagulation depth, and skin healing evolution at typical settings to support substantial equivalence.

## Technological Characteristics

30W, 10.6 µm CO2 laser; air-cooled; articulated arm with CO2 scanner; metallic shell with polyurethane panel; touch screen interface; microcontroller-based control electronics; footswitch and remote interlock connectivity.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Lutronic Corp - MOSAIC e CO2 ([K080496](/device/K080496.md))
- Lutronic Corp - MOSAIC e CO2 ([K091115](/device/K091115.md))
- Lasering Srl - SLIM Evolution Family ([K063001](/device/K063001.md))
- Lasering Srl - SLIM Evolution II Family ([K110984](/device/K110984.md))
- El.En S.p.A - Smart CO2 Laser System ([K072159](/device/K072159.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## K 123573

## 510(K) SUMMARY

## APR 1 5 2013

## Introduction:

This document contains the 510(k) Summary for the YOULASER CO2 laser system.

The content of this summary is based on the requirements of 21 CFR 807.92(c).

| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058           |
|--------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Contact Person:                           | Maurizio Bianchi<br>Regulatory Affairs Manager<br>Quanta System SPA                        |
|                                                  | Email: Maurizio.bianchi@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared:                                   | November 16th, 2012                                                                        |
| Device Name:                                     | YOULASER CO2                                                                               |
| Classification:                                  | Class II                                                                                   |
| Classification Name:                             | Laser surgical instrument for use in general<br>and plastic surgery and in dermatology     |
| Regulation Number:                               | 21 CFR 878.4810                                                                            |
| Product Code:                                    | GEX                                                                                        |

#### Predicate Devices:

The YOULASER CO2 Laser System is claimed to be substantially equivalent to the following legally marketed predicate devices:

| Lutronic Corp - MOSAIC e CO2 (K080496)            |
|---------------------------------------------------|
| Lutronic Corp - MOSAIC e CO2 (K091115)            |
| Lasering Srl - SLIM Evolution Family (K063001)    |
| Lasering Srl - SLIM Evolution II Family (K110984) |
| El.En S.p.A - Smart CO2 Laser System (K072159)    |

{1}------------------------------------------------

#### Performance Standards:

There are no mandatory performance standards for this device.

A specific investigation to evaluate the safety and the effectiveness has been planned and performed :

A specific non clinical testing was conducted on the proposed YOULASER CO2 device to support a determination of substantial equivalence

The testing has evaluated the ablation, coagulation depth and the healing evolution of the skin at the typical settings of use of Youlaser CO2.

In conclusion the results of these tests provided reasonable assurance that the YOULASER CO2 is as safe as effective as the predicated devices and support a determination of substantial equivalence.

#### General Device Description:

The YOULASER CO2 Laser System includes a single model named YOULASER CO2 emitting a maximum power of 30 Watt at 10.6 µm.

The YOULASER CO2 Laser is composed externally of a metallic shell with a frontal polyurethane panel containing the touch screen display. On this panel the key switch, emergency red push button and the operation led are inserted too. On the rear panel the footswitch connector, the remote interlock, the power switch are located.

The laser system is composed of power supply, CO2 laser source with air cooling system, optical bench, articulated arm with CO2 scanner, the control electronics.

The electronic, based on a microcontroller, manages the voltage power supply and the CO2 laser source.

#### Indications for Use:

YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for:

incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.

YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.

#### Comparison of Technological Characteristics:

The laser specifications for the YOULASER CO2 laser system are substantially equivalent to the laser specifications for its identified predicate device with respect to the laser source, wavelengths, maximum energy, spot size, fluence,

{2}------------------------------------------------

pulse width, repetition rate, beam delivery, power monitor, actuator, and aiming beam.

## Comparison of Intended Use:

The intended use of the YOULASER CO2 laser system is the same as the intended use of its previously cleared devices.

#### Substantial Equivalence:

The Quanta System YOULASER CO2 laser system is as safe and effective as the predicate devices.

The YOULASER CO2 laser system has the same intended use and similar technological characteristics and principles of operation as its predicate devices. The minor technological differences between the YOULASER CO2 laser system and its predicate devices raise no new issues of substantial equivalence or safety and effectiveness.

Thus, the YOULASER CO2 laser system is substantially equivalent to its identified predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is centered horizontally. The words are all capitalized and evenly spaced.

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 15, 2013

Quanta System, S.P.A. % Mr. Maurizio Bianchi Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058

Re: K123573 Trade/Device Name: YOULASER CO2 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 8, 2013 Received: March 11, 2013

Dear Mr. Bianchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

## Page 2 – Mr. Maurizio Bianchi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, FOR

# Peter D. Ruimm -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known): K123573

Device Name: YOULASER CO2

### Indications for Use:

YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for: incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.

YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.

X ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use AND/OR (Part 21 C.F.R. 801 Subpart D)

Over-The-Counter Use (21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden 2013.04.105 09:32:07 -04'00'

(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K123573

{6}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known): K123573

YOULASER CO2 Device Name:

#### Indications for Use:

YOULASER CO2 laser when used in traditional, non fractionated mode is indicated for:

incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery.

YOULASER CO2 laser when used in fractionated mode (dot scanner) is indicated only for ablative skin resurfacing.

Prescription Use X Use (Part 21 C.F.R. 801 Subpart D)

(21 C.F.R. 807 Subpart C)

Over-The-Counter

(Please Do Not WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

FOR

Peter D. Rumm, 2013.04.15 -04'00'

- (Division Sign-Off) for MXM Neil R. Ogden, M.S. Chief, General Surgery Devices Branch I Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

---

**Source:** [https://fda.innolitics.com/device/K123573](https://fda.innolitics.com/device/K123573)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
