K123531 · LeMaitre Vascular, Inc. · DQY · Dec 14, 2012 · Cardiovascular
Device Facts
Record ID
K123531
Device Name
UNBALLOON NON-OCCLUSIVE MODELING CATHETER
Applicant
LeMaitre Vascular, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UnBalloon Non-Occlusive Modeling Catheter is intended to assist in the modeling of self-expanding endoprostheses in large diameter vessels.
Device Story
UnBalloon Non-Occlusive Modeling Catheter is a percutaneous catheter used by vascular surgeons to assist in modeling self-expanding endoprostheses in large diameter vessels. Device features an expandable Nitinol mesh cage housed in a 14F retractable sheath; design allows mesh expansion without occluding blood flow. Radiopaque markers and Nitinol mesh provide visibility under fluoroscopy for precise positioning. Inner lumen accommodates 0.035 or 0.038 inch guidewires for over-the-wire access. Blue handle controls sheathing/unsheathing; clear handle/luer controls mesh expansion. Device is flushed via side ports. By expanding the mesh against the endoprosthesis, the device ensures proper apposition to the vessel wall, potentially improving clinical outcomes for endovascular procedures.
Clinical Evidence
Bench testing only. Performance evaluation included radial outward force, apposition length, and simulated anatomical use testing to verify the performance of the additional catheter sizes.
Technological Characteristics
Silicone surface-coated percutaneous catheter; expandable Nitinol mesh cage; 14F retractable sheath; radiopaque markers; over-the-wire guidewire compatibility (0.035/0.038 inch). Sterilization: Ethylene oxide (EO) per ANSI/AAMI/ISO 11135-1:2007. Biocompatibility: ISO 10993 compliant for limited contact (<24 hours) in circulating blood.
Indications for Use
Indicated for use by vascular surgeons to assist in the modeling of self-expanding endoprostheses within large diameter vessels.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K123531
## 510 (k) Summary
#### Submitter's information:
## DEC 1 4 2012
| Name: | LeMaitre Vascular, Inc. |
|----------|--------------------------|
| Address: | 63 Second Avenue |
| | Burlington, MA USA 01803 |
| Phone: | 781-425-1727 |
| Contact | Bryan Cowell, MSc., RAC |
| Person: | |
| Date of preparation: | 14 Nov 2012 |
|----------------------|-------------------------------------------|
| Device Name: | UnBalloon Non-Occlusive Modeling Catheter |
| Trade Name. | UnBalloon Non-Occlusive Modeling Catheter |
Common/ Classification Catheter, Percutaneous / Modeling Catheter Name: Classification Panel: 21CFR §870.1250 Class: II (2)
DQY
Product Code:
#### Establishment Registration: 1220948
Establishment: LeMaitre Vascular, Inc., 63 Second Avenue, Burlington, MA USA 01803
Owner/Operator: 1220948
#### Proposed Device Description:
The UnBalloon Non-Occlusive Catheter is a silicone surface coated (medical grade) modeling catheter with an expandable Nitinol mesh in a 14F retractable sheath. The Nitinol mesh design allows for expansion without occluding blood flow. The Nitinol mesh and radiopaque markers are highly visible under fluoroscopy and assist in the positioning of the device. The inner lumen allows for a 0.035 or 0.038 inch guidewire for over-the-wire access. Side ports and clear handle/luer allow the device and guidewire lumen to be flushed. The blue handle allows the device to be sheathed/unsheathed while the clear handle/luer controls the expansion of the Nitinol mesh. This submission modifies the current UnBalloon Non-Occlusive Modeling Catheter by expanding the product family line with additional nitinol mesh cage sizes and catheter lengths.
{1}------------------------------------------------
#### Proposed Intended Use:
The UnBalloon Non-Occlusive Modeling Catheter is intended to assist in the modeling of selfexpanding endoprostheses in large diameter vessels.
#### Predicate Device:
| 510(k): | K121839 |
|--------------------|-------------------------------------------|
| Device Name: | UnBalloon Non-Occlusive Modeling Catheter |
| SE Date: | 08/08/2012 |
| Regulation Number: | 870.1250 |
| Device Class Name: | Catheter, Percutaneous |
| Device Class: | 2 |
| 510(k): | K110891 |
|--------------------|-------------------------------------------|
| Device Name: | UnBalloon Non-Occlusive Modeling Catheter |
| SE Date: | 09/13/2011 |
| Regulation Number: | 870.1250 |
| Device Class Name: | Catheter, Percutaneous |
| Device Class: | 2 |
#### Substantial Equivalence:
#### Fundamental Scientific Technological Characteristics:
The UnBalloon Non-Occlusive Modeling Catheter is a silicone surface coated percutaneous/modeling catheter designed for vascular surgeons. This submission is to include additional product sizes.
#### Functional/ Safety testing:
The verification activities conducted indicate that UnBalloon Non-Occlusive Modeling Catheter device meets the product performance requirements of the device specifications and does not raise any additional safety issues.
#### Sterilization:
The device is validated for ethylene oxide (EO) sterilization according to ANSVAAMI/ISO 11135-1:2007, "Sterilization of Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization"
#### Biocompatibility:
All blood contact portions of the device were subjected to Biocompatibility testing according to ISO 10993 guidelines for an externally communicating device with limited contact duration (<24 hours), in circulating blood. There is no new material used for the additional catheter sizes as discussed in this submission.
{2}------------------------------------------------
### Summary of Product Testing:
The following tests have been completed to evaluate the performance of the catheter with shorter cage: .
ﻳ
Radial outward force Apposition Length Simulated Anatomical Use
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in bold, black font. The text is centered in the image. There is a logo to the left of the text.
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
LeMaitre Vascular, Inc. c/o Mr. Bryan Cowell Principal Regulatory Affairs Specialist 63 Second Avenue Burlington, MA 01803
DEC 1 4 2012
Re: K123531
Trade/Device Name: UnBalloon Non-Occlusive Modeling Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: DQY Dated: November 15, 2012 Received: November 16, 2012
Dear Mr. Cowell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Bryan Cowell
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to hittp://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
.
| 510(k) Number (if known) | K123531 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | UnBalloon Non-Occlusive Modeling Catheter |
| Indications for Use | The UnBalloon Non-Occlusive Modeling Catheter is intended to assist in the modeling of self-expanding endoprostheses in large diameter vessels. |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | X |
|---------------------------------------|-------------------------|
| | OR Over-The-Counter Use |
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K123531:
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :