K123509 · Hitachi Medical Systems America, Inc. · JAK · Mar 29, 2013 · Radiology
Device Facts
Record ID
K123509
Device Name
SCENARIA PHASE 2 WHOLE-BODY X-RAY CT SYSTEM
Applicant
Hitachi Medical Systems America, Inc.
Product Code
JAK · Radiology
Decision Date
Mar 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SCENARIA Phase 2 system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the SCENARIA Phase 2 can be post processed by the SCENARIA Phase 2 to provide additional information. Post processing capabilities included in the SCENARIA Phase 2 software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering. Volume datasets acquired by the SCENARIA Phase 2 can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
Device Story
Multi-slice CT system; uses X-ray tube and solid-state detector assembly on rotating gantry; collects up to 64 slices simultaneously. High-frequency generator produces fan beam X-ray output. System operates in axial, helical, gated, or dynamic modes. High-speed reconstruction sub-system processes raw data into cross-sectional images. Workstation (Windows-based PC) displays, stores, and archives images. Includes post-processing for CTA, MPR, and volume rendering. GuideShot option provides remote in-room display/controls for interventional procedures. Used in clinical settings by qualified physicians to aid diagnosis. DICOM interface allows data transfer to external devices.
Clinical Evidence
Clinical evaluation compared the SCENARIA Phase 2 to the predicate SCENARIA system (K101888). Bench testing evaluated dose profile, image noise, Modulation Transfer Function (MTF), slice thickness, sensitivity profile, slice plane location, and CT dose index. Results confirmed performance is comparable to the predicate.
Technological Characteristics
64-slice CT system; solid-state detector; high-frequency X-ray generator; fan beam output; slip-ring gantry rotation. PC-based workstation (Windows OS). DICOM connectivity. System components (gantry, tube, table) identical to predicate. Complies with IEC 60601-1, 60601-1-1, 60601-1-2, 60601-1-3, 60601-2-32, and 60601-2-44.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages. Used for axial, helical, gated, or dynamic acquisition modes. Supports interventional imaging via remote display/controls.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
K123509
Page 1 of 3
2-2
#### 510(k) Summary of Safety Effectiveness
MAR 29 2013
| Submission Information | |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371<br>ph: (330) 425-1313<br>fax: (330) 963-0749 |
| Contact: | Douglas J. Thistlethwaite |
| Date: | November 5, 2012 |
| Establishment Registration No. | 1528028 |
| Trade/Proprietary Name: | SCENARIA Phase 2 Whole-body X-ray CT System |
| Classification Name: | Computed Tomography X-ray System |
| Product Code | JAK |
| C.F.R. Section: | 892.1750 |
| Regulatory Class | Class II |
| Predicate Device(s): | SCENARIA Whole-body X-ray CT System (K101888) |
| Reason for Submission | New Technology |
## Device Intended Use
The SCENARIA Phase 2 system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes.
The volume datasets acquired by the SCENARIA Phase 2 can be post processed by the SCENARIA Phase 2 to provide additional information. Post processing capabilities included in the SCENARIA Phase 2 software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Volume datasets acquired by the SCENARIA Phase 2 can be transferred to external devices via a DICOM standard interface.
The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
## Device Description
#### Function
The SCENARIA Phase 2 is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles.
{2}------------------------------------------------
KB23509
Page 2 of 3
2-3
### Scientific Concepts
The SCENARIA Phase 2 system uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.
The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.
### Physical and Performance Characteristics
The SCENARIA Phase 2 system consists of a Gantry, Operator's Workstation, Patient Table, high-Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.
### Performance Comparison
A clinical evaluation comparison was conducted with the SCENARIA Phase 2 system and the original SCENARIA System (K101888) and found to be substantially equivalent as documented in Section 12 - Performance Testing - Clinical.
In addition, evaluations were conducted for dose profile, image noise, Modulation Transfer Function (MTF), slice thickness and sensitivity profile, slice plane location, and CT dose index as documented in Section 11 - Performance Testing - Bench.
The evaluation results confirm the performance characteristics of the SCENARIA Phase 2 are comparable to the predicate device and support our conclusion that the SCENARIA Phase 2 system is substantially equivalent.
## Device Technological Characteristics
The SCENARIA PHASE 2's Gantry, X-ray Tube, X-ray Generator, and Patient Table specifications are the same as the SCENARIA System (K101888). Likewise, the Helical Scanning, Image Processing, Display, Image Storage, and Performance features are equivalent to the predicate device. Lastly, the features related to dose which are available on the predicate device are also included on the SCENARIA PHASE 2. See Section 4 - Comparison to Predicate Device for additional information.
Therefore, the SCENARIA Phase 2 system is technologically equivalent in concept, function, and performance to the predicate device.
{3}------------------------------------------------
# Safety and Effectiveness
The SCENARIA Phase 2 is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820.
Additionally this system is in conformance with the applicable parts of:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, . 1988; Amendment 1, 1991-11, Amendment 2, 1995
- . IEC 60601-1-1:2000 Medical Electrical Equipment - Part 1-1: General Requirements for Safety - Collateral Standard; Safety Requirements for Medical Electrical Systems
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for . Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)}
- IEC 60601-1-3: 1994 Edition 1.0, Medical electrical equipment Part 1-3 Collateral . Standard: General Requirements for Radiation Protection in Diagnostic X-ray Equipment.
- IEC 60601-2-32: 1994 Edition 1.0, Medical electrical equipment Part 2: Particular . requirements for the safety of associated equipment of X-ray equipment
- IEC 60601-2-44 (2002-11) Edition 2.1, Medical electrical equipment Part 2-44: . Particular requirements for the safety of X-ray equipment for computed tomography
## Conclusions
The SCENARIA Phase 2 system has been developed and validated according to applicable standards. Testing has proven that the system is safe and effective for the indicated use. Risk and hazard analysis shows that there are no new safety issues associated with this system as compared with the predicate device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with healthcare, with three wavy lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2013
Hitachi Medical Systems America, Inc. % Mr. Doug Thistlethwaite Manager. Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087
Re: K123509
Trade/Device Name: SCENARIA Phase 2 Whole-body X-ray CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 25, 2013 Received: February 26, 2013
Dear Mr. Thistlethwaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Mr. Thistlethwaite
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123509
Device Name:
HITACHI SCENARIA Phase 2 Whole-body X-ray CT System
Indications for Use:
The SCENARIA Phase 2 system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes.
The volume datasets acquired by the SCENARIA Phase 2 can be post processed by the SCENARIA Phase 2 to provide additional information. Post processing capabilities included in the SCENARIA Phase 2 software include CT angiography (CTA). Multiplanar reconstruction (MPR) and volume rendering.
Volume datasets acquired by the SCENARIA Phase 2 can be transferred to external devices via a DICOM standard interface.
The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
> Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sm. 7).
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
K123509 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.