← Product Code [MQB](/productcode/MQB) · K123499

# AEROPILOT (K123499)

_Konica Minolta Medical & Graphic, Inc. · MQB · Dec 20, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123499

## Device Facts

- **Applicant:** Konica Minolta Medical & Graphic, Inc.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Dec 20, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

AeroPilot is a software device that is used in conjunction with REGIUS Unitea to control Konica Minolta Digital Radiography systems. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.

## Device Story

AeroPilot is software installed on an off-the-shelf PC (operation console) used with REGIUS Unitea; functions as an interface between X-ray generators and Konica Minolta Digital Radiography systems. It acquires X-ray images; replaces traditional film/screen systems. Operated by clinical staff in a diagnostic setting. Output is digital radiographic images for clinical review and decision-making. Benefits include transition to digital imaging workflow.

## Clinical Evidence

Bench testing only. Performance testing demonstrated that image quality is equivalent to the predicate devices.

## Technological Characteristics

Software-based device; operates on off-the-shelf PC hardware. Functions as an interface for X-ray image acquisition and system control. Complies with ISO 14971 for risk management.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- AeroDR SYSTEM ([K102349](/device/K102349.md))
- AeroSync for AeroDR SYSTEM ([K102349](/device/K102349.md))
- AeroDR SYSTEM with P-21 ([K113248](/device/K113248.md))

## Reference Devices

- REGIUS Unitea ([K071436](/device/K071436.md))
- Konica Minolta Digital Radiography systems ([K120477](/device/K120477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KI 234 9

## 510(k) Summary

as required by 807.92

DEC 2 0 2012

## 1. Company Identification

Konica Minolta Medical & Graphic, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675

### 2. Submitter's Name and Address

Shiqeyuki Kojima Manager Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053

### 3. Date of Submission

November 9, 2012

- 4. Device Trade Name AeroPilot
#### 5. Common Name

Picture Archiving Communications System

### 6. Classification, Product Code

Class II, 90LLZ

#### 7. Predicate Device

AeroDR SYSTEM, 510(k) number K102349 AeroSync for AeroDR SYSTEM, 510(k) number K102349 AeroDR SYSTEM with P-21, 510(k) number K113248 (Konica Minolta Digital Radiography system)

#### 8. Indications for Use

AeroPilot is a software device that is used in conjunction with REGIUS Unitea Unitea to control Konica Minolta Digital Radiography system. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/ screen systems in general-purpose diagnostic procedures.

This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.

5 - 1

{1}------------------------------------------------

## 9. Device Description

The AeroPilot is a software device that is used in conjunction with our REGIUS Unitea. K071436, to control Konica Minolta Digital Radiography systems. The AeroPilot is the software designed to be installed in Off-the-shelf PC (operation console) which is one of component of REGIUS Unitea and works with Konica Minolta Digital Radiography systems to be an interface with X-ray generator or between REGIUS Unitea and the specified Konica Minolta Digital Radiography systems, and to acquire X-ray images like the specified Konica Minolta Digital Radiography systems do.

### 10. Risk Analysis

The Risk Analysis for the AeroPilot (software) was conducted on the basis of ISO14971, "Medical devices - Application of risk management to medical devices". As a result of risk control measures, the risk associated with all of the identified hazards was reduced to an acceptable level or ALARP (As low as reasonably practicable).

Please refer to the Section 13, Software, Appendix 13-G

#### 11. Substantial Equivalence to Predicate Device

The predicate devices of AeroPilot is our Konica Minolta Digital Radiography systems (K102349, K120477, K113248).

A comparison of the Indications for Use, Configuration, Specifications, and Principal of Operation of this proposed device and the predicate devices has shown in Section 9. Substantial Equivalence Comparison Table.

The result of the Section 9, above mentioned the Risk Analysis showed that there is no new safety and efficacy issue of the proposed device. Also, the results of performance testing show that the image quality of proposed device is equivalent to the predicate device. Please refer to the section 15 and Section 14(as reference data)

Therefore, the function of AeroPilot is substantially equivalent to the specified function that AeroDR SYSTEMS have.

#### 12. Conclusion

Comprehensively, we conclude that the AeroPilot has the same technological characteristics as the predicate devices. This 510(k) has demonstrated substantial equivalence as the predicate devices.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 20, 2012

Konica Minolta Medical & Graphic, Inc. % Mr. Russell Munves Storch, Amini & Munves, P.C. Official Correspondent 140 East 45th Street, 25th floor Two Grand Tower NEW YORK, NY 10017

Re: K123499

Trade/Device Name: AeroPilot Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified flouroscopic x-ray system Regulatory Class: II Product Code: MOB Dated: November 9, 2012 Received: November 13, 2012

Dear Mr. Munves:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2-Mr. Munves

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Janine M. Morris -S

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

#### 510(k) Number: K123499

Device Name: Aero Pilot

Indications for Use:

AeroPilot is a software device that is used in conjunction with REGIUS Unitea to control Konica Minolta Digital Radiography systems. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.

Prescription Use __ Yes_______________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

# Janine M. Morris -S

(Division Sign Off)

Division of Radiological Health

Office of In Vitro Diagnostic and Radiological Health

510(k)_

Page 1 of

---

**Source:** [https://fda.innolitics.com/device/K123499](https://fda.innolitics.com/device/K123499)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
