K123421 · Olympus Medical Systems Corp. · NEZ · Apr 29, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K123421
Device Name
OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM
Applicant
Olympus Medical Systems Corp.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
Apr 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K123421 · Apr 29, 2013
OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM
Olympus Medical Systems Corp.
Retrospective analysis of recorded image data from a clinical study
Retrospective evaluation of antenna reception performance using recorded image data from a clinical study of healthy volunteers.
Retrospective analysis; Antenna performance; Clinical study data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of clinical study data
Healthy volunteers; Sample Size: 29 subjects
Predicate device
Video gap occurrence as an indicator of reception performance
Indications for Use
The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.
Device Story
System captures small intestine images via ingestible capsule (EC-Y0006); capsule contains objective lens, image sensor, 4 white LEDs, radio transmitter (315MHz), and silver oxide batteries. Patient wears antenna unit (MAJ-Y0163) and recorder (RE-Y0002) to receive/store image data. Physician uses workstation (PC with ENDOCAPSULE software) to download, view, and analyze images. Software features include 3D tracking (capsule position estimation via antenna signal strength), red color detection (highlighting suspected blood/lesions), and various playback modes (Express viewing, Overview). System aids clinicians in identifying small intestine mucosal abnormalities; patient benefits from non-invasive visualization of small bowel.
Clinical Evidence
Clinical study with 30 healthy volunteers evaluated 3D Track function accuracy and image quality. 3D Track accuracy verified against radiographic ground truth. Image quality assessed via investigator questionnaire, showing superiority to predicate. Antenna performance evaluated retrospectively via video gap occurrence, confirming equivalence to predicate.
Technological Characteristics
Capsule (11mm x 26mm) made of biocompatible polymer composites. Sensing: CMOS image sensor with 4 white LEDs. Transmission: 315MHz radio. Power: Silver oxide batteries. Connectivity: Antenna unit to wearable recorder; data transfer to PC workstation via cradle. Software: Server-client architecture on commercial PC. Sterilization: Not specified.
Indications for Use
Indicated for visualization of small intestine mucosa in patients requiring endoscopic examination. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
123421 117
### 510(k) SUMMARY
# APR 2 9 2013
#### OLYMPUS INTESTINAL ENDOSCOPE SMALL CAPSULE SYSTEM
November 5, 2012
#### 1 General Information
를 Applicant:
B Official Correspondent:
1 Manufacturer: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No .: 8010047
Daphney Germain-Kolawole Regulatory Affairs Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-5691 FAX: 484-896-7128 Email: daphney.germain-kolawole@olympus.com
SHIRAKAWA OLYMPUS CO., LTD. 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima, 961-8061, Japan Establishment Registration No .: 3002808148
OLYMPUS MEDICAL SYSTEMS CORP. HINODE PLANT 34-3 Hirai, Hinode-machi, Nishitama-Gun, Tokyo, 190-0182, Japan Establishment Registration Number: 3003637092
#### 2 Device Identification
Device Trade Name:
트 Common Name:
l Regulation Number:
. Regulation Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM
Capsule Imaging System
21 CFR 876.1300
Ingestible telemetric gastrointestinal capsule imaging system
{1}------------------------------------------------
| | Regulatory Class: | II (Special Controls) |
|--|-----------------------|--------------------------|
| | Classification Panel: | Gastroenterology/Urology |
| | Product Code: | NEZ |
#### 3 Legally Marketed Device to which Substantial Equivalence is Claimed
The following table shows the device to which we claim substantial equivalence (predicate device).
## Table 18-1 Predicate Device of the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM
| Predicate Device | 510(k)<br>Number |
|----------------------------------|------------------|
| OLYMPUS CAPSULE ENDOSCOPE SYSTEM | K090210 |
## 4. Device Description
OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is a capsule imaging system used for visualization of the small intestine mucosa.
The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM consists of a CAPSULE ENDOSCOPE (EC-Y0006), an ACTIVATOR (MAJ-1478), a RECORDER (RE-Y0002), a CRADLE (MAJ-Y0160), a BATTERY PACK (MAJ-Y016), a RECORDER HOLDER (MAJ-Y0162), an ANTENNA UNIT (MAJ-Y0163), an ANTENNA HOLDER (MAJ-Y0164), an ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) and an ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166).
ENDOCAPSULE SMALL INTESTINAL CAPSULE ENDOSCOPE SET (MAJ-Y0176) is a package containing five CAPSULE ENDOSCOPEs (EC-Y0006).
ENDOCAPSULE RECORDER SET (MAJ-Y0177) is a package containing the ACTIVATOR (MAJ-1478), the RECORDER (RE-Y0002), the CRADLE (MAJ-Y0160), the BATTERY PACK (MAJ-Y0161), the RECORDER HOLDER (MAJ-Y0162), the ANTENNA UNIT (MAJ-Y0163), and the ANTENNA HOLDER (MAJ-Y0164).
The CAPSULE ENDOSCOPE (EC-Y0006) has an outer diameter of 11mm, a total length of 26mm, and is designed and manufactured with polymer composites that are biocompatible and provide sufficient tolerance to exposure to body fluids (digestive tract juice) and external compression forces. The clear top cover contains a compact objective lens in front of the image sensor. Four white light emitting diodes (LEDs) are allocated around the objective lens. The exterior casing contains operational circuits for the image sensor, LEDs, radio transmitter, and antenna for radio transmission in 315MHz. The CAPSULE ENDOSCOPE (EC-Y0006) has two silver oxide batteries to power the circuit within the capsule.
{2}------------------------------------------------
The RECORDER (RE-Y0002) is used in combination with the ANTENNA UNIT (MAJ-Y0163) for receiving signals from the CAPSULE ENDOSCOPE (EC-Y0006) and for recording image data captured by the CAPSULE ENDOSCOPE (EC-Y0006). The RECORDER (RE-Y0002) is powered by the BATTERY PACK (MAJ-Y0161). It is capable of continuously receiving and recording image data from the CAPSULE ENDOSCOPE (EC-Y0006). The RECORDER has a color LCD display panel for displaying received images, recorded images, patient ID, battery pack level, etc. The RECORDER (RE-Y0002) is inserted into a pouch on the RECORDER HOLDER (MAJ-Y0162) enabling the patient to move freely during the examination. The BATTERY PACK (MAJ-Y0161) is charged when the RECORDER (RE-Y0002) is attached to the CRADLE (MAJ-Y0160). Unlike the predicate device that includes Battery charger (MAJ-1478), no independent battery charger is required.
The ANTENNA UNIT (MAJ-Y0163) is an integrated antenna sheet containing eight antenna elements and is stored in the ANTENNA HOLDER (MAJ-Y0164). The ANTENNA HOLDER is attached covering the patient's waist.
A workstation for observation of the image data captured by the CAPSULE ENDOSCOPE (EC-Y0006) is constructed by installing the ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) or the ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166) to a commercially available personal computer. Functions of each software are shown in the table below.
| Model | Functions |
|--------------------------------------------|-------------------------------|
| ENDOCAPSULE SOFTWARE | Observation of the image data |
| SERVER-CLIENT (MAJ-Y0165) | Storage of the image data |
| ENDOCAPSULE SOFTWARE<br>CLIENT (MAJ-Y0166) | Observation of the image data |
When the user uses more than or equal to two workstations in the network environment, the user can select the ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166) as the workstation from the 2nd workstation. The user can share data between two workstations or more.
The ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) includes the ENDOCAPSULE SOFTWARE LIGHT which can be installed on commercially available personal computer for viewing the image data captured by the CAPSULE ENDOSCOPE (EC-Y0006). The ENDOCAPSULE SOFTWARE LIGHT is provided for user convenience.
The workstation facilitates the downloading of endoscopic image data recorded in the RECORDER (RE-Y0002) to the workstation via the CRADLE (MAJ-Y0160) for observation of patient image data. The workstation function sends patient information data to the RECORDER (RE-Y0002) from the workstation via the CRADLE (MAJ-Y0160).
The workstation also provides the following functions:
- Image display whose maximum rate of the image renewal is based upon the user setting:
- Display / store images selected by the user as thumbnail data;
- Development and printing of diagnostic reports with bookmarked images;
- Image / video clip export;
- Data storage to a network server.
{3}------------------------------------------------
In addition, the image processing function offers the following user selectable functions:
- . 3D Track function (new function) The signal reception strength of multiple antennas is calculated to estimate the capsule position, and the capsule track is displayed.
- . Red Color Detection Function A software feature that highlights frames suspicious for blood or red lesions based upon analysis of red pixels within the image.
- . Average Color Bar Calculates the average RGB colors of an image and displays the averaged color composite as a color line.
- . Adjust mode Automatically adjusts the speed of frame display based upon both image movement and user setting.
- . Express viewing function A software feature that allows for image playback and that has two playback modes. The Express-Sclected mode plays back images whose mucosal area is dissimilar from that of previous image. The Express-Skipped mode plays back the images skipped in the Express-Selected mode.
- 사 Overview function
A software feature that displays images whose mucosal area is dissimilar from that of previous images as an outline of the complete case.
### 5 Indications for Use
The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.
#### Comparison of Technological Characteristics б
The technological characteristics of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM are identical or similar to the predicate device identified in "3 Legally Marketed Device to which Substantial Equivalence is Claimed" above. A detailed comparison of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM with the predicate device is provided in the premarket notification.
{4}------------------------------------------------
#### 7 Substantially Equivalent Discussion
The subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is a modification, to the previously cleared OLYMPUS CAPSULE ENDOSCOPE SYSTEM (K090210). The indications for use, principles of operation and fundamental technology of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM are similar or identical to the predicate device. The indicated patient population and anatomical site are also identical to the predicate The subject device also meets the guidance entitled "Class II Special device. Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA, Document issued on: November 28, 2001." The proposed changes in this submission do not raise new safety and effectiveness issues related to device design.
Based upon the results of bench testing and clinical testing performed, we believe that the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is substantially equivalent to the identified predicate device as demonstrated by this submission.
#### Summary of non-clinical testing કિ
The following non-clinical testing was performed.
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
The following bench testing were performed.
1) Specification testing including "PH resistance testing", "Mechanical and structural integrity testing (including bite test)", "Battery life testing", "Field of view, depth of focus, distortion, color/illumination uniform, sensitivity of imaging sensor, color reproduction, resolution, and image noise testing", "Shock, vibration, temperature, and after the shelf life)".
2) Side-by-side bench testing between the subject device and the predicate device including "Image quality (including displayed lines and villus, field of view, optimum working distance)" and "Antenna reception performance".
3) Evaluation of 3D track function including "Accuracy evaluation" and "Display evaluation"
4) Stability testing
The testing results showed the subject device has sufficient performance or equivalent performance to the predicate device.
{5}------------------------------------------------
The following biocompatibility testing were performed for the patient contacting parts of the EC-Y0006 and testing results met the biocompatibility requirement.
- Cytotoxicity
- Sensitization
- Acute Intracutaneous Reactivity
- Acute Oral Toxicity
- Acute Systemic Toxicity
- Implantation instead of sub-chronic toxicity
The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a "Moderate Level of Concern."
The following standards were used during the design and validation of the subject devices:
-ISO 14971: 2007 -IEC 60601-1: 1988, Amendment 1: 1991, Amendment 2: 1995 -IEC 60601-1-1: 2000 -IEC 60601-1-2: 2001, Amendment 1: 2004 -IEC 60601-2-18: 1996, Amendment 1: 2000 -IEC 60950-I : 2005 -180 11135-1: 2007 -ISO 10993-1: 2009 -ISO 10993-5: 2009 -ISO 10993-6: 2007 -ISO 10993-10: 2010 -1SO 10993-11: 2006
#### ਹੈ Summary of clinical testing
A clinical study was performed and thirty healthy volunteers were involved in this study. The objective of the study was to evaluate the accuracy of the 3D Track function and image quality. Additionally, the performance of the new antenna was retrospectively evaluated utilizing the recorded image data in the clinical study above. The summary of the clinical study results is described below.
### 3D Track
The accuracy of the 3D Track function was evaluated. Healthy volunteers swallowed the capsule endoscope and underwent x-ray photography. Based upon the radiographic image data, positions of the capsule endoscope were derived. Then the positions of the capsule endoscope derived from the radiographic image and the estimated positions of the capsule endoscope at the time of the radiographic image Three subjects were excluded, and therefore twenty-seven shooting were compared. subjects were eligible. The test results met the predefined acceptance criteria. Therefore, it was verified that the accuracy of the 3D Track function was maintained as we presumed.
{6}------------------------------------------------
### Image quality
Image quality of the subject device was assessed by the principal investigator questionnarie regarding resolution, brightness including evenness of illumination, color tone, and field of view. The principal investigator's comments showed that the subject device is greatly superior to the predicate device in regard to image quality.
### Antenna Performance
The performance of the new antenna was evaluated since the shape and type of the antennas of the subject device is different from those of the predicate device. The evaluation was made using video gap occurrence as the indicator of the reception performance. One subject was excluded, and therefore twenty-nine subjects were eligible. The evaluation results met the predefined acceptance criteria. Thus, the reception performance of the subject device is equivalent to that of the predicate device.
#### 10 Conclusion
When compared to the predicate device, the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2013
OLYMPUS MEDICAL SYSTEMS CORP. % Ms. Daphney Germain-Kolawole Regulatory Affairs Project Manager Olympus America, Inc. 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034-0610
Re: K123421
Trade/Device Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Regulation Number: 21 CFR& 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: March 15, 2013 Received: March 18, 2013
Dear Ms. Germain-Kolawole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
# Page 2 - Ms. Daphney Germain-Kolawolė
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
- http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
1/1
# Indications for Use
510(k) Number (if known): K123421
Device Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Indications For Use:
The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.
Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher S
2013.04.29 15:44:16 -04'00'
Page 1 of __ l________________________________________________________________________________________________________________________________________________________________
K123421
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.