K123353 · ArthroCare Corporation · GEI · Apr 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123353
Device Name
CIQ CONTROLLER, CIQ WANDS
Applicant
ArthroCare Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ArthroCare® Coblator IQTM ENT System is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: Adenoidectomy, Head, Neck, Oral, and Sinus Surgery, Myringotomy with Effective Hemorrhage Control, Nasopharyngeal/Laryngeal indications including Tracheal Procedures, Laryngeal Polypectomy, and Laryngeal Lesion Debulking, Papilloma Keloids, Submucosal Tissue Shrinkage, Traditional Uvulopalatoplasty (RAUP), Tissue in the Uvula/Soft Palate for the treatment of Snoring, Cysts, Mastoidectomy, Nasal Airway Obstruction by Reduction of Hypertrophic Nasal Turbinates, Neck Mass, Submucosal Palatal Shrinkage, Tonsillectomy (including palatine tonsils), Tumors.
Device Story
System comprises bipolar, single-use radiofrequency (RF) plasma wands and a controller unit. Controller functions as a high-frequency electrosurgical generator with an integrated peristaltic pump for saline delivery. During surgery, saline is delivered to the wand tip; RF energy creates a plasma field to ablate, resect, or coagulate soft tissue and achieve hemostasis. Used in ENT surgical settings by physicians. Output allows precise tissue removal and bleeding control, potentially reducing thermal damage to surrounding healthy tissue compared to traditional electrosurgery.
Clinical Evidence
No clinical data included. Evidence based on bench testing and pre-clinical animal studies comparing tissue effects to predicate devices.
Technological Characteristics
Bipolar, RF-based electrosurgical system. Components: single-use plasma wands and controller with integrated peristaltic pump for saline delivery. Connectivity: standalone. Sterilization: not specified. Software: minor updates to controller.
Indications for Use
Indicated for patients undergoing otorhinolaryngology (ENT) surgical procedures requiring soft tissue ablation, resection, coagulation, or hemostasis. Includes procedures such as tonsillectomy, adenoidectomy, turbinate reduction, and laryngeal/nasopharyngeal surgeries.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
ArthroCare Coblator IQ System (Controller) (K091674)
Submission Summary (Full Text)
{0}------------------------------------------------
# K123357
Page 1 of 2
## 510(k) Summary
ArthroCare® Corporation
COBLATOR IQTM ENT SYSTEM
APR 5 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | ArthroCare Corporation |
|-----------------|---------------------------------------------------------------------|
| Address: | 7000 West William Cannon Drive<br>Austin, TX 78735 |
| Contact Person: | Mitchell Dhority<br>Vice President, Clinical and Regulatory Affairs |
| Date Prepared: | October 29, 2012 |
### Device System Names/Components
| Proprietary: | Coblator IQTM ENT System |
|-----------------|--------------------------|
| Common: | ENT System |
| Classification: | Class II |
| Product Code: | GEI |
| CFR Section: | 21 CFR 878.4400 |
#### Predicate Device
| ArthroCare ENT Plasma Wands | K070374 (April 25, 2007) |
|-----------------------------|----------------------------|
| | K063538 (December 1, 2006) |
| | K033257 (October 29, 2003) |
| | K014290 (March 28, 2002) |
ArthroCare Coblator IQ System (Controller)
K091674 (January 15, 2010)
## Description
The ArthroCare Coblator IQ ENT System (CIQ System) is comprised of eight ArthroCare Coblator IQ ENT Plasma Wands (CIQ Wands) and the ArthroCare Coblator IQ Controller (CIQ Controller).
The CIQ Wands are bipolar, single use, radiofrequency-based electrosurgical devices designed for use with the CIQ Controller for specific indications in otorhinolaryngology (ENT) procedures.
The CIQ Controller is a bipolar, high frequency electrosurgical generator with an integrated peristaltic pump for saline delivery.
{1}------------------------------------------------
## Intended Use/Indications For Use
The ArthroCare® Coblator IQTM ENT System is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: 1 ... ..
- . Adenoidectomy
- 1 Head, Neck, Oral, and Sinus Surgery
- . Myringotomy with Effective Hemorrhage Control
- . Nasopharyngeal/Laryngeal indications including Tracheal Procedures, Laryngeal Polypectomy, and Laryngeal Lesion Debulking
-
- 1 Papilloma Keloids
- Submucosal Tissue Shrinkage
- 1 Traditional Uvulopalatoplasty (RAUP)
- . Tissue in the Uvula/Soft Palate for the treatment of Snoring
#### Performance Testing - Bench
- Cysts
123353
- Mastoidectomy 트
- l Nasal Airway Obstruction by Reduction of Hypertrophic Nasal Turbinates
- l Neck Mass
- Submucosal Palatal Shrinkage 트
- Tonsillectomy (including palatine tonsils)
- Tumors
Bench testing was conducted to evaluate the performance of the Coblator IQ ENT System compared to the predicate devices. The test results demonstrate that the Coblator IQ ENT System meets all design and performance specifications.
#### Performance Testing - Animal
Pre-Clinical animal studies were conducted to compare the tissue effects using the Coblator IQ ENT System as compared to the predicate devices. Based on the test results, the proposed devices are substantially equivalent to the predicate devices.
#### Performance Testing - Clinical
No clinical data are included in this submission.
#### Summary
The ArthroCare Coblator IQ ENT System (CIQ System) is comprised of eight ArthroCare Coblator IQ : ENT Wands (CIQ Wands) and the ArthroCare Coblator IQ Controller (CIQ Controller).
The modified CIQ Wands are substantially equivalent to the predicate ArthroCare ENT Plasma Wands. The modification to performance specifications, materials, and labeling are minor changes, and do not affect the safety or efficacy of the devices.
There have also been minor software changes to the CIQ Controller cleared in 2010. However, these changes do not affect the safety or efficacy of the device.
Page 2 of 2
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Letter dated: April 5, 2013
Arthrocare Corporation % Mr. Mitchell Dhority Vice President, Clinical and Regulatory Affairs 7000 West William Cannon Drive Austin, Texas 78735
Re: K123353 Trade/Device Name: ArthroCare® Coblator IQ™ ENT System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 12, 2013 Received: March 27, 2013
Dear Mr. Dhority:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Mitchell Dhority
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
# Mark N.Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K123353 510(k) Number (if known):
Device Name: ArthroCare® Coblator IQTM ENT System
#### Indications for Use:
The ArthroCare® Coblator IQTM ENT System is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including:
- Adenoidectomy
- Head, Neck, Oral, and Sinus Surgery
- Myringotomy with Effective Hemorrhage Control
- 다 Nasopharyngeal/Laryngeal indications including Tracheal Procedures, Laryngeal Polypectomy, and Laryngeal Lesion Debulking
- Papilloma Keloids
- Submucosal Tissue Shrinkage
- Traditional Uvulopalatoplasty (RAUP)
- 트 Tissue in the Uvula/Soft Palate for the treatment of Snoring
- . Cysts
- Mastoidectomy 미
- l Nasal Airway Obstruction by Reduction of Hypertrophic Nasal Turbinates
- 0 Neck Mass
- 트 Submucosal Palatal Shrinkage
- 이 Tonsillectomy (including palatine tonsils)
- 트 Tumors
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH. Office of Device Evaluation (ODE)
Joshua C.
Nipper -S
For
(Division Sign-Off) Division of Surgical Devices
K123353 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.