K123319 · Olympus Surgical Technologies America · KNS · Dec 14, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K123319
Device Name
GYRUS ACMI BICOAG HEMOSTASIS PROBE
Applicant
Olympus Surgical Technologies America
Product Code
KNS · Gastroenterology, Urology
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gyrus ACMI BiCOAG Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures.
Device Story
Single-use disposable bipolar RF accessory; used with external bipolar generator (up to 500Vp-p) and endoscope. Device passed through endoscope working channel (10 Fr or 7 Fr). Features central irrigation port for sterile water delivery to clear blood/debris from surgical site. Bipolar electrodes spiral down tip to deliver desiccating RF current to bleeding lesions. Operated by physicians during endoscopic procedures in clinical settings. Provides endoscopic control and hemostasis via bipolar energy application. Benefits patient by enabling visualization and coagulation of GI bleeding sites.
Clinical Evidence
Bench testing only. Verification testing included physical dimensions, flow rates, pull testing, leak testing, electrical performance, packaging, shelf life, sterilization, and biocompatibility. Non-clinical bench study on porcine tissue compared coagulation performance of proposed probe against predicate and other market competitors. All tests met acceptance criteria.
Technological Characteristics
Bipolar electrosurgical probe; 10 Fr (3.3mm) and 7 Fr (2.3mm) diameters. Materials: gold conductive traces, zirconia tip, conductive epoxy, tin-silver solder. Energy source: external bipolar generator (up to 500Vp-p). Features central irrigation port. Single-use, disposable. Sterilization method not specified.
Indications for Use
Indicated for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures in patients requiring endoscopic hemostasis.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{1}------------------------------------------------
Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772 BiCOAG Hemostasis Probe
## 510(K) SUMMARY
General Information
Manufacturer:
DEC 1 4 2012
大12331
rf
Gyrus ACMI 6845 Wedgewood Rd. Maple Grove, MN 55311 ERN: 3005975494
Gyrus ACMI 136 Turnpike Rd. Southborough, MA 01772-2104 ERN: 3003790304
Neil Kelly (508) 804-2690 neil.kelly@Olympus-OSTA.com
BiCOAG Hemostasis Probe
BiCOAG, BiCOAG Probe
K092571
October 26, 2012
Unit, Electrosurgical, Endoscopic (With or Without Accessories) Gastroenterology/Urology KNS 21 CFR 8876.4300 Class II
Trade Name:
Generic/Common Name:
Predicate Device
Gyrus ACMI BiCOAG Hemostasis Probe
Indications for Use
The Gyrus ACMI BiCOAG Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures.
#### Product Description
The BiCOAG Hemostasis Probe is a single use disposable high frequency RF accessory to be used in conjunction with a generator with bipolar frequency energy rated up to 500Vp-p. The BiCOAG Hemostasis Probe is intended to be used for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract. It contains a central irrigation port. The probe is to be available in 10 Fr and 7 Fr (3.3mm and 2.3mm respectively) outer diameter configurations to allow its use through a minimum 3.7mm and 2.8mm endoscope working channel, respectively. The device is intended to be passed through the working channel of an endoscope to provide hemostasis throughout the gastrointestinal tract.
Page 10
P. 1/3 22
510(k) Submitter:
Contact Person:
Date Prepared:
Classification Name:
{2}------------------------------------------------
The BiCOAG Hemostasis Probe enables irrigating fluid (sterile water) to be delivered at the electrode tip. This fluid enables the clearing of blood and other substances that often impede vision and coagulation of the surgical site.
The probe is bipolar with electrodes that spiral down the tip in a uniform manner. These electrodes carry desiccating RF current to a site where bleeding is occurring and allows for coagulation at the target site. The hole at the end of the probe allows for irrigating water to clean the surgical site as required when used in conjunction with a bipolar generator and endoscope. The system thus provides endoscopic control and hemostasis of upper and lower GI bleeding by the application of bipolar energy.
| Change | Predicate Device K092571 (From) | Proposed Device (To) | Reason |
|-------------------------------------|---------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conductive Trace Material | Nickel-plated Palladium | Gold | Improves manufacturability. New material found to be biocompatible, and performed substantially equivalent to the predicate material. (Material) |
| Tip to Wire Electrical Joint | Tin-lead solder | Conductive epoxy | Improves manufacturability of electrical joint with gold trace material and proposed conductive epoxy is RoHS compliant. (Material, Component) |
| Wire to Power Cord Electrical Joint | Tin-lead solder | Tin-silver solder | Proposed solder is intended for electrical applications and is RoHS compliant. (Material) |
| 10Fr Tip Material | Machined alumina | Molded zirconia | Higher mechanical strength and improves manufacturability. Material already in use on the 7Fr predicate device. (Material) |
| Bifurcation Joint | Overmolded | Two-piece Assembly (potting cup and lid) | Improves reliability and resistance to leaks with improved flow. (Component) |
| Coaxial Plug | Manufactured | Purchased | Coaxial plug is an industry standard design and was readily available from an existing vendor. Removes steps from current manufacturing process. (Component) |
| Luer Adhesive Change | UV-cure adhesive | UV-cure adhesive with tint | Original adhesive has become obsolete by the vendor. New adhesive performs substantially equivalent to the original but changes color from blue to clear when cured which aids in assembly. (Material) |
| Tip to Body Stock Adhesive Change | Moisture-cure adhesive | Epoxy | Improved manufacturability and epoxy is RoHS compliant. (Material, Component) |
## Proposed to Predicate Comparison
Page 11
P.2/3
23
{3}------------------------------------------------
These device modifications have been tested against the predicate device and are found to perform substantially equivalent. No new issues of safety or effectiveness are raised by these changes.
#### Performance Data. Technological features, and Substantial Equivalence
The Gyrus ACMI BiCOAG Hemostasis Probe, as described in this submission, is substantially equivalent to the predicates in intended use, principles of operation, performance, and fundamental scientific technology.
Design verification testing was carried out on the bench to ensure the performance and reliability of the proposed device all with passing results. Testing such as physical dimensions, flow rates, pull testing, leak testing, performance, electrical, packaging, shelf life, sterilization, and Biocompatibility were all carried out and met acceptance criteria. In addition, a non-clinical bench study was carried out by a trained user on porcine tissue to evaluate the coagulation of the proposed probe against the predicate, and a few other market competitors. All specifications of the BiCOAG Hemostasis Probe that required verification and/or validation to ensure the proposed device is substantially equivalent to the predicate BiCOAG Hemostasis Probe have successfully achieved the acceptance criteria and raised no new issues of safety or effectiveness.
#### Summary
The proposed Gyrus ACMI BiCOAG Hemostasis Probe, as described in this submission, raises no new issues of safety or effectiveness as compared to the predicate device and has been found substantially equivalent to the currently marketed predicate Gyrus ACMI BiCOAG Hemostasis Probe.
Page 12
p.3/3 2
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized symbol resembling a bird or abstract human figure, with three curved lines representing the body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Olympus Surgical Technologies America % Mr. Neil Kelly Regulatory Affairs Specialist 136 Turnpike Road Southborough, Massachusetts 01772
December 14, 2012
Re: K123319
Trade/Device Name: Gyrus ACMI® BiCOAG Hemostatis Probe Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KNS Dated: November 21, 2012 Received: November 21, 2012
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Gyrus ACMJ、Inc. 136 Turnpike Road Southborough, MA 01772
### Indications for Use Statement
510(k) Number: K123319
Device Name: Gyrus ACMI® BiCOAG Hemostasis Probe
Indications for Use:
The Gyrus ACMI BiCOAG Hemostasis Probe is intended for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures.
OR
Prescription Use:_X_
Over-the-Counter Use: ________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dwight Yen 2012.12.12 07:57:38 -05'00'
(Division Sign-Off)
Division of Surgical Devices
510(k) Number _K123319
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.