Intended for femoral venous and arterial access during cardiopulmonary bypass (CPB).
Device Story
Wire-reinforced thin-wall polymer cannulae; designed for femoral venous and arterial access during cardiopulmonary bypass (CPB). Features include tapered tips for insertion, incremental depth markings, and optional radiopaque barium stripes for visualization; some versions include suture rings. Available with or without Duraflo heparin coating. Used in clinical settings by physicians to drain blood flow during short-term (≤ 6 hours) CPB procedures. Device provides physical conduit for extracorporeal circulation; performance relies on mechanical integrity (kink resistance, crush resistance, leak prevention).
Clinical Evidence
Bench testing only. Testing included sterility (ISO 11135-1), ethylene oxide residuals (ISO 10993-7), biocompatibility (ISO 10993-1), dimensional inspection, wire encapsulation, leak testing, kink resistance, crush resistance, collapse resistance, tensile strength, cap removal force, and corrosion resistance. All performance testing met pre-determined acceptance criteria.
Technological Characteristics
Wire-reinforced thin-wall polymer cannulae. Materials include stainless steel and plastic tips. Features include tapered tips, depth markings, and optional radiopaque barium stripes. Available with Duraflo heparin coating. Sterilized via ethylene oxide per ISO 11135-1. Complies with ISO 10993-1 (biocompatibility), ISO 10993-7 (EO residuals), and ISO 14971 (risk management).
Indications for Use
Indicated for rapid femoral venous and arterial access for short-term (≤ 6 hours) cardiopulmonary bypass in pediatric and adult populations. Use depends on flow rate requirements and patient anatomy. Duraflo-coated components indicated for cardiopulmonary surgery requiring heparin-coated blood paths.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}------------------------------------------------
K123303
| 510(k) Summary | |
|----------------|--|
|----------------|--|
| Submitter: | Edwards Lifesciences LLC |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dannette Crooms, Manager, Regulatory Affairs<br>12050 Lone Peak Pkwy<br>Draper, UT 84020<br>(801) 565-6209 |
| Date Prepared: | October 22, 2012 |
| Trade Name: | Edwards Lifesciences® Femoral Access Venous Cannulae |
| Classification Name: | cardiopulmonary bypass vascular catheter, cannula or tubing<br>Class II (21 CFR §870.4210). |
Predicate Devices
- K891576, Cleared October 5, 1989 .
- K905312, Cleared January 15, 1991 .
- K033464, Cleared June 22, 2004 .
## Device Description:
The Edwards Femoral Access Venous Cannulae with or without Duraflo® coating are wire-reinforced thin-wall polymer cannulae. The wire reinforcement is intended to prevent kinking during use.
The clear proximal section of the cannula is unreinforced for clamping and terminates in either a barbed connector for 1/4" or 3/8" tubing connection.
The cannulae are available in various sizes and lengths. Edwards Femoral Access Venous Cannulae are available in a range of sizes with a variety of tip and hole configurations, including stainless steel and plastic tips. The cannulae tips are tapered for easy insertion.
Some cannulae feature incremental depth markings to aid in proper placement and positioning. As an additional aid to placement, the clear tip of some versions of the
# MAR 22 2013
{1}------------------------------------------------
cannula contains two radiopaque barium stripes for visualization. Some product codes may also include a suture ring.
Edwards Femoral Access Venous Cannulae are intended to provide a means of draining the blood flow of a patient during cardiopulmonary bypass procedures.
Edwards Femoral Access Venous Cannulae are also available with Durafto® coating.
#### Intended Use:
Intended for femoral venous and arterial access during cardiopulmonary bypass (CPB).
#### Indications for Use:
Edwards Lifesciences Femoral Access Cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (≤ 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician.
Femoral Access Cannula may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy.. Please consult labeling to determine pressure drop related to flow rates.
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
#### Comparative Analysis:
The basis for this submission is a clarification of the Indications for Use statement. No physical changes are being made to these devices. The subject devices have the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate devices. It has been demonstrated that the subject Femoral Access Venous Cannulae are comparable to the predicate devices in fundamental scientific technology, material types, principles of operation, and functional performance evaluations. No new issues of safety or efficacy have been raised.
{2}------------------------------------------------
## Functional/Safety Testing:
Bench and laboratory testing was performed and assures that the product meets its specifications per the table below. The performance testing met the acceptance criteria.
| Testing | Criteria |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Sterility | Per ISO11135-1, Sterilization of health care products - Ethylene oxide - Part 1: |
| Ethylene oxide<br>sterilization residuals | ISO 10993-7, Biological evaluation of medical devices - Part 7: |
| Biocompatibility | Per ISO 10993-1 for External communicating device, direct<br>circulating blood path, duration ≤ 24 hours. |
| Dimensional Inspection | Confirmation of critical dimensions such as ID, OD and length. |
| Wire Encapsulation | Cannula body wire reinforcement shall be fully encapsulated. |
| Cannula Leak | Pressure drop must meet minimum requirement. |
| Kink | The cannulae shall not kink at a pre-determined diameter. |
| Cannula Crush | The cannulae must not crush at a pre-determined compression. |
| Cannula Collapse | The cannulae must not collapse at a pre-determined vacuum<br>pressure. |
| Tensile | Confirmation of the bond strength of the catheter assembly must<br>meet pre-determined loads. |
| Cap removal force | Confirm cap removal force is within pre-determined limits. |
| Corrosion | Metallic components shall show no signs of corrosion. |
The Femoral Access Arterial Cannulae conform to the following standards:
- · ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing.
- · ISO 10993-7, Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- · ISO11135-1, Sterilization of health care products Ethylene oxide Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- · ISO14971, Medical devices Application of risk management to medical devices
#### Conclusion:
The Femoral Access Venous Cannulae are substantially equivalent to the cited predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines extending from its body, representing the department's commitment to health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2013
Edwards Lifesciences LLC C/O Dannette Crooms 12050 Lone Peak Pkwy. Draper, UT 84020
Re: K123303
Trade/Device Name: Femoral Access Venous Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: February 15, 2013 Received: February 19, 2013
Dear Ms. Crooms:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Dannette Crooms
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sonna M-Patel-raman -S
- Bram D. Zuckerman, M.D. for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Statement of Indications for Use
510(k) Number (if known): K123303
Edwards Lifesciences Femoral Access Cannulae Device Name:
Indications for Use:
Edwards Lifesciences Femoral Access Cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (≤ 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician.
Femoral Access Cannula may be used in pediatic populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates.
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Prescription Use × (Per 21 CFR 801.109)
OR Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
Sonna MEPatél Traman -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.