Systematic literature review; Complaint data analysis
The sponsor utilized a systematic literature review and an analysis of postmarket complaint data to evaluate the safety of the MTS System and its leads when used in patients with concomitant cardiac devices, supporting the update of contraindications to warning statements.
Systematic literature review; Complaint data; Postmarket surveillance; Safety evaluation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic literature review and complaint data analysis; Retrospective analysis of literature and postmarket surveillance data
Patients with spinal cord stimulation (SCS) systems and concomitant cardiac devices
Not applicable for this study
Safety and risk of interference with cardiac device functionality
Indications for Use
The Multiprogram Trial Stimulator System is indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach. The Multiprogram Trial Stimulator is intended to be used with St. Jude Medical Neuromodulation Division percutaneous trial leads and external test extensions to deliver paresthesia to single or multiple painful sites for trial stimulation either intraoperatively or postoperatively for up to a maximum of 30 days. The St. Jude Medical Neuromodulation Division neurostimulation systems are indicated for spinal cord stimulation (SCS) in the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. St. Jude Medical Neuromodulation Division percutaneous leads model number 3066, 3163, 3166, 3169, extension model numbers 3382, 3383, 3341, 3342, 3343, and receiver model number 3408, transmitter model number 3508, and antenna model numbers 1220 and 1230 are also indicated to stimulate electrically peripheral nerves to relieve severe intractable pain.
Device Story
MTS System delivers low-intensity electrical impulses to spinal cord or peripheral nerves to treat chronic pain; system includes battery-powered external stimulator, trial cables, and percutaneous leads. Leads feature platinum-iridium electrodes and contact bands with polyurethane or silicone rubber sheaths; placed via epidural needle, guide wire, and stylet. Stimulator programmed to deliver pulses via sixteen output polarities; communicates via RS232 port with Rapid Programmer. Used intraoperatively or postoperatively by clinicians; patient wears stimulator during trial period (max 30 days). Output provides targeted paresthesia to painful sites; helps clinicians evaluate patient response to stimulation before permanent implant. Modification replaces demand-type pacemaker contraindication with warning to verify lack of interference and avoid unipolar programming.
Clinical Evidence
Clinical evaluation included prospective clinical evidence, systematic literature review, and complaint data analysis. Evidence supports replacing the demand-type pacemaker contraindication with a warning and clinical mitigation instructions to reduce risk of SCS interference with cardiac devices.
Technological Characteristics
System consists of external battery-powered stimulator and percutaneous leads. Leads constructed of platinum-iridium electrodes and contact bands with polyurethane or silicone rubber insulation. Leads feature 4 cylindrical electrodes. Connectivity via RS232 port to Rapid Programmer. System delivers electrical pulses via sixteen output polarities.
Indications for Use
Indicated for patients with chronic pain of the trunk and limbs or severe intractable peripheral nerve pain. Used as sole therapy or adjunct in multidisciplinary approach. Trial stimulation limited to 30 days.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
MTS™ Multiprogram Trial Stimulator System (K033757)
{0}------------------------------------------------
K123299
JAN 1 8 2013
St. JUDE MEDICAL
MORE CONTROL. LESS RISK
St. Jude Medical Neuromodulation Division 1 Preston Road
510(k) Summary Contact Information Owner/Operator: St. Jude Medical, Inc., Neuromodulation Division Terrina Wilder, RAC Submitter's Name: 6901 Preston Rd., Plano, TX 75024 Address: (972) 309-8115 (Office) (972) 309-8265 (Fax) Telephone Number: October 19, 2012 Date:
Device Names
6.0
Trade Names:
Common or Usual Names:
Classification:
Classification Names:
Class II
CFR 882.5870)
MTS™ System
MTS™ System
Stimulator, Spinal Cord, Implanted for Pain Relief (21 CFR 882.5880) Stimulator, Peripheral Nerve, Implanted for Pain Relief (21
Wide-Spaced Quattrode™ Percutaneous Lead Wide-Spaced Quattrode™ Percutaneous Trial Lead
Wide-Spaced Quattrode™ Percutaneous Lead Wide-Spaced Quattrode™ Percutaneous Trial Lead
Predicate Device Names :
MTS™ Multiprogram Trial Stimulator System (K033757) Wide-Spaced Quattrode™ Leads (K072462)
Establishment Registration Number: 1627487
# Device Description
MTS™ System
The MTS™ System (also called the Multiprogram Trial Stimulator) is a multi-programmable device designed to deliver low intensity electrical impulses to nerve structures in the dorsal aspect of the
Page 12 of 30
{1}------------------------------------------------
spinal cord. The system consists of an external stimulator, one or two trial cables, and a single or multiple percutaneous trial lead.
The trial stimulator is a battery-powered device that can be worn by the patient and is intended to deliver targeted paresthesia to single or multiple painful stimulation either intraoperatively or post operatively for a maximum of 30 days. The device contains electronic circuitry that combines the function of a receiver and a transmitter and allows the stimulator to be easily programmed to deliver electrical pulses via sixteen output polarities through the trial cable to the implanted lead. The Trial Stimulator communicates via an RS232 port with the Rapid Programmer.
# Wide-Spaced Quattrode™ Trial and Permanent Leads
The wide-spaced Quattrode trial and permanent leads are for use in spinal cord stimulation and perioheral nerve stimulation. Each lead consists of a variety of platinum iridium electrodes on the distal end connected by individually insulated wires to platinum iridium contact bands on the proximal end. The insulated wires are covered by a biocompatible polyurethane or silicone rubber sheath. The lead assembly consists of 4 cylindrical electrodes spaced at precise intervals. These leads are designed for introduction into the epidural space using an epidural needle, a guide wire, the optional Introde-AK (Introde) lead introducer, and a stylet to aid in positioning. Also, the permanent wide-spaced Quattrode leads (model #s: 3161, 3163, 3166 and 3169) are designed to be placed directly on or adjacent to a peripheral nerve. The exception is found with the widespaced Quattrode trial leads (model #3066) in that they are used during a trial implantation period not to exceed 30 days.
### Indications for Use Statements
### MTS™ System
The Multiprogram Trial Stimulator System is indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach.
The Multiprogram Trial Stimulator is intended to be used with St. Jude Medical Neuromodulation Division percutaneous trial leads and external test extensions to deliver paresthesia to single or multiple painful sites for trial stimulation either intraoperatively or postoperatively for up to a maximum of 30 days.
# Wide-Spaced Quattrode™ Percutaneous Leads
The St. Jude Medical Neuromodulation Division neurostimulation systems are indicated for spinal cord stimulation (SCS) in the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach.
St. Jude Medical Neuromodulation Division percutaneous leads model number 3066, 3163, 3166, 3169, extension model numbers 3382, 3383, 3341, 3342, 3343, and receiver model number 3408, transmitter model number 3508, and antenna model numbers 1220 and 1230 are also indicated to stimulate electrically peripheral nerves to relieve severe intractable pain.
The intended uses for the MTS™ System and the Wide-Spaced Quattrode™ Percutaneous Leads are the same as the predicate devices. The proposed change to replace the demand-type pacemaker. contraindication does not affect the safe and effective use of the medical devices.
### Technological Characteristics
Summary of Non-clinical Studies
{2}------------------------------------------------
Non-clinical data performed to support previous submissions would still be applicable and would support this submission as well. As an effort to evaluate the continued safety and effectiveness of the medical devices mentioned in this submission, a risk evaluation was conducted. Identified risks were mitigated by replacing the contraindication for demand-type pacemakers with a warning statement instructing clinicians to verify that there is no interference between the implantable devices intraoperatively and to avoid a unipolar programming mode.
# Summary of Clinical Studies
A clinical evaluation was performed on the MTS™ System which includes associated accessory devices such as leads. This evaluation provides evidence to support the replacement of the contraindication for demand-type cardiac pacemakers with a warning that provides the clinician with mitigators to reduce the risk of spinal cord stimulation (SCS) interference with cardiac device functionality in patients with concomitant systems. The evidence to support this change includes prospective clinical evidence, a systematic literature review, and complaint data analysis.
## Conclusion
As a result of the non-clinical and clinical data demonstrated that the MTS™ System and the Wide-Spaced Quattrode™ Leads are as safe, effective and perform as well as the legally marketed devices identified in this 510(k) summary.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign.
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle or bird-like figure with three distinct head profiles, suggesting a sense of community or collaboration. The bird is positioned to the right of the text, which is arranged in a circular pattern around the left side of the image. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
January 18, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
St. Jude Medical, Inc., Neuromodulation Division Terrina Wilder, RAC Senior Regulatory Affairs Specialist 6901 Preston Rd. Plano, TX 75024
Re: K123299
Trade/Device Name: MTS Wide-Spaced Quattrode Percutaneous Leads Regulation Numbers: 21 CFR 882.5880 Regulation Name: Stimulator, Spinal Cord, Implanted for Pain Relief Stimulator, Peripheral Nerve, Implanted for Pain Reliefs Regulatory Class: Class II Product Code: GZB, GZF Dated: October 19, 2012 Received: October 23, 2012
Dear Ms. Terrina Wilder:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave and have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use surfor to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of The general bonnote proving practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Terrina Wilder
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse events) (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you uestle specific advice tor your do need career and offices/ucm115809.htm for go to mtp.//www.lua.gov/rtoout Brical Health's (CDRH's) Office of Compliance. Also, please the Center for Devices and Radional by reference to premarket notification" (21CFR Part note the regulation entined, "Wholenants of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the 1 ou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (000) 050 2611 01 (0 MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joyce M. Whang
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for St. Jude Medical. The logo consists of a 3x3 grid of squares on the left, followed by the text "St. Jude MEDICAL" in a bold, sans-serif font. Below the text, there is a tagline that reads "MORE CONTROL LESS RISK" in a smaller font.
St. Jude Medica Neuromodulation Division 6901 Praston Road Plano TX 75024 USA Tel 972 309 8000 Fax 972 309 8150
Indications for Use 5.0
510(k) Number (if known):
Device Name:
MTS™ System
Indications for Use:
The Multiprogram Trial Stimulator System is indicated for the treatment of chronic pain of the trunk and f the Multiplogian Than Stilliantial agent or as an adjunct to other modes of therapy used in a multidisciplinary approach.
The Multiprogram Trial Stimulator System is intended to be used with St. Jude Medical Neuromodulation The Maniprogram That Scimal test extensions to deliver targeted paresthesia to single or Division perculuneous that lease and ther intraoperatively or postoperatively for up to a maximum of 30 days.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joyce M. Whang
Page 1 of
Division of Neurological and Physical Medicine Devices 510(k) Number: K123299
Page 10 of 30 .
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for St. Jude Medical. The logo consists of a grid of nine squares on the left, followed by the text "ST. JUDE MEDICAL" in a serif font. Below the text, in smaller letters, is the phrase "MORE CONTROL LESS RISK".
St. Jude Medical Neuromodulation Division 6901 Preston Road Plano, TX 75024 USA Tel 972 309 8000 Fax 972 309 8150
#### 5.1 Indications for Use
510(k) Number (if known):
Device Name:
Wide-Spaced Quattrode™ Leads
Indications for Use:
The St. Jude Medical Neuromodulation Division neurostimulation systems are indicated for spiral cord rrie 3t. Jude Medical Nedromedation of the trunk and limbs, either as the sole mitigating Stimulution (800) in the vo other modes of therapy used in a multidisciplinary approach.
St. Jude Medical Neuromodulation Division percutaneous leads model numbers 3066, 3161, 3163, 3166, 31. Jude Medical Nedromobilers 3382, 3383, 3341, 3342, 3343, and receiver model number 3408, 5105, Excellsion friodel Numbers 1220 and antenna model numbers 1220 and 1230 are also indicated to stimulate electrically peripheral nerves to relieve severe intractable pain.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joyce M. Whang
Page 1 of _
Page 11 of 30
Division of Neurological and Physical Medicine Devices 510(k) Number: K123299
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.