K123289 · Cardinal Health-Medical Products and Services · FRG · Feb 28, 2013 · General Hospital
Device Facts
Record ID
K123289
Device Name
DURABLUE STERILIZATION WRAP
Applicant
Cardinal Health-Medical Products and Services
Product Code
FRG · General Hospital
Decision Date
Feb 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider using the following modalities: - Pre-vacuum steam at 270°F/132°C for 4 minutes. - 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). Maintenance of package sterility was validated with real-time aging testing for a duration of 180 days for each indicated modality. For pre-vacuum steam sterilization, the wrap has been validated for dry times for Models CH100 and CH200 and 30 minutes for Models CH300, CH400, CH500, and CH600. Models CH400, CH500, and CH600 have been validated for pre-vacuum steam sterilization of two lumens 3 mm in diameter or larger and 400 mm in length or less. For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55°C. Models CH400, CH500, and CH600 have been validated for sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less for EO sterilization.
Device Story
DuraBlue™ Sterilization Wrap consists of double-layer 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. Used by healthcare providers in clinical settings to wrap medical instruments prior to sterilization. Design supports simultaneous double-wrapping and aseptic opening. Device acts as a microbial barrier to maintain sterility of enclosed contents for 180 days post-sterilization. Six models (CH100-CH600) accommodate varying load weights (3-25 lbs) and specific lumen configurations. Healthcare providers use the wrap to ensure instrument sterility until point of use, reducing risk of contamination.
Clinical Evidence
Bench testing only. Validation included sterilization efficacy, microbial barrier properties (aerosol challenge), event-related shelf life (180 days), material compatibility, and biocompatibility per ISO 10993. Testing confirmed sterility maintenance for 180 days post-sterilization for all indicated modalities and models.
Technological Characteristics
Material: 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. Six models (CH100-CH600) vary by basis weight. Sterilization modalities: Pre-vacuum steam (270°F/132°C) and 100% Ethylene Oxide (55°C). Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization via pre-vacuum steam (270°F/132°C, 4 min) or 100% ethylene oxide (725-735 mg/L, 131°F/55°C, 40-80% RH, 60 min). Maintains sterility of enclosed contents for 180 days.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
# 1-123289
Image /page/0/Picture/1 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic of three curved lines above the text "CardinalHealth". The text is in a bold, sans-serif font, with "Cardinal" in a darker shade than "Health".
1430 Waukegan Road McGraw Park, IL 60085
www.cardinal.com
| 510(k) SUMMARY<br>DuraBlue™ Sterilization Wrap | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Manufacturer: | Cardinal Health 200, LLC<br>1430 Waukegan Road<br>McGaw Park, IL 60085 |
| Regulatory Affairs Contact: | Lavenia Ford<br>1430 Waukegan Road<br>McGaw Park, IL 60085 |
| Telephone Number: | (847) 887-3323 |
| Date summary Prepared: | February 12, 2013 |
| Trade Name: | DuraBlue™ Sterilization Wrap |
| Classification: | Class II per 21 CFR § 880.6850 |
| Classification Name: | Sterilization Wrap |
| Common Name: | Sterilization Wrap |
| Product Code: | FRG |
| Predicate Device: | K112211 - DuraBlue™ Sterilization Wrap - Pre-vacuum Steam (4<br>min/270°C) |
| | K120542 - DuraBlue™ Sterilization Wrap - Ethylene Oxide(EO) (725-<br>735 mg/L at 55°C/40%-80%RH/60 min) |
#### Description:
Cardinal Health DuraBlue™ Sterilization Wraps are double layer sterilization wraps made from 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. They are intended to be used to enclose another medical device that is to be sterilized by a health care provider pre-vacuum steam at 270°F/132°C for 4 minutes, or 100% ethylene oxide (EO) with a concentration of 725-735 mgL at 131°F/55°C and 40%-80% relative humidity for 60 minutes. This wrap design allows for use of the simultaneous double-wrapping technique and also allows for a sterilized pack to be opened aseptically.
The only modification to the predicate devices involving the clarification of the sterlity maintenance information provided in the Indication for Use portion of the Instructions for Use Labeling.
FEB 2 8 2013
{1}------------------------------------------------
- . Predicate Statement:
- "The wrap is intended to allow sterilization of the enclosed medical device(s) and o maintain sterility of the enclosed device(s) for 30 days."
- . Proposed Statement:
- o "The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). Maintenance of package sterility was validated with real-time aging testing for a duration of 180 days for each indicated modality."
This modification is based on Event-Related Shelf Life testing that demonstrates that the DuraBlue™ Sterilization Wrap maintains sterility of the enclosed contents for 180 days following sterilization by prevacuum steam or EO.
This submission covers six different models of Cardinal Heatth DuraBlue™ Sterilization Wrap. Each model is made from material of a different basis weight, though all models utilize the same material technology.
{2}------------------------------------------------
| Element of | PROPOSED | PREDICATE | PREDICATE |
|-----------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Comparison | Cardinal Health DuraBlue™ | Cardinal Health DuraBlue™ | Cardinal Health DuraBlue™ |
| | Sterilization Wrap - Pre-vacuum | Sterilization Wrap - Pre- | Sterilization Wrap - Ethylene |
| | Steam or EO | vacuum Steam<br>(K112211) | Oxide (EO)<br>(K120542) |
| Key Claims | 180 day Maintenance of Sterility | • Indications for use with<br>pre-vacuum sterilization | • Indications for use with<br>Ethylene Oxide<br>sterilization |
| | Testing for all modalities | | |
| Product | 100% polypropylene spunbond-<br>meltblown-spunbond (SMS) fabric | Same | Same |
| Properties | | | |
| Intended | Cardinal Health DuraBlue™ | Cardinal Health DuraBlue™ | Cardinal Health DuraBlue™ |
| Use | Sterilization Wrap is intended to | Sterilization Wrap is intended | Sterilization Wrap is intended |
| | enclose another medical device that is | to be used to enclose another | to be used to enclose another |
| | to be sterilized by a health care | medical device that is to be | medical device that is to be |
| | provider using the following modalities: | sterilized by a health care | sterilized by a health care |
| | • pre-vacuum steam at 270°F/132°C | provider by pre-vacuum steam<br>at 270°F/132°C for 4 minutes.<br>The wrap is not indicated for | provider by 100% ethylene<br>oxide (EO) with a |
| | for 4 minutes | use with gravity steam | concentration of 725-735 mg/L |
| | • 100% ethylene oxide (EO) with a | sterilization. | at 131°F/55°C and 40%-80% |
| | concentration of 725-735 mg/L at | | relative humidity for 60 |
| | 131°F/55°C and 40%-80% relative | | minutes. |
| | humidity for 60 minutes | The wrap is intended to allow | |
| | The wrap is intended to allow | sterilization of the enclosed | The wrap is intended to allow |
| | sterilization of the enclosed medical | medical device(s) and to | sterilization of the enclosed |
| | device(s) and to maintain sterility of the | maintain sterility of the<br>enclosed devices for 30 days. | medical device(s) and to |
| | enclosed device(s). Maintenance of<br>package sterility was validated with | | maintain sterility of the<br>enclosed devices for 30 days. |
| | real-time aging testing for a duration of | The wrap has been validated | The wrap has been validated |
| | 180 days following each indicated | for dry times for pre-vacuum | for an aeration time of 8 hours |
| | modality. | steam sterilization of 20 | at 55°C. |
| | | minutes for Models CH100 | |
| | For pre-vacuum steam sterilization, | and CH200, and for 30 | Models CH400, CH500, and |
| | the wrap has been validated for dry | minutes for Models CH300, | CH600 have been validated |
| | times of 20 minutes for Models CH100 | CH400, CH500 and CH600. | for sterilization of two lumens |
| | and CH200 and 30 minutes for Models | | of 3 mm diameter or larger and |
| | CH300, CH400, CH500, and CH600, | | 400 mm in length or less for<br>EO sterilization. |
| | | | |
| | For EO sterilization, the wrap has<br>been validated for an aeration time of | | |
| | 8 hours at 55°C. | | |
| Length of | Pre-vacuum steam: 180 days | Pre-vacuum steam: 30 days | Ethylene oxide: 30 days |
| Maintenance | | | |
| of Sterility | Ethylene oxide: 180 days | | |
| Testing | | | |
| Lumen<br>Claims | Pre-vacuum steam: Models CH400,<br>CH500 and CH600 have been | Pre-vacuum Steam: Models | Ethylene Oxide (EO): Models |
| | validated for sterilization of two lumens | CH400, CH500 and CH600<br>have been validated for | CH400, CH500, and CH600 |
| | of 3mm diameter or larger and 400mm | sterilization of lumens of 3mm | have been validated for<br>sterilization of two lumens of |
| | in length or less. | diameter or larger and 400mm | 3mm diameter or larger and |
| | | in length or less. | 400mm in length or less. |
| | Ethylene Oxide (EO): Same as | | |
| | K120542 | | |
# Table 1: Overall Comparison to Predicate Devices
{3}------------------------------------------------
Indications for Use:
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider using the following modalities:
- Pre-vacuum steam at 270°F/132°C for 4 minutes ●
- . 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes
The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). Maintenance of package sterility was validated with real-time aging testing for a duration of 180 days for each indicated modality.
For pre-vacuum steam sterilization, the wrap has been validated for dry times of 20 minutes for Models CH100 and CH200 and 30 minutes for Models CH300, CH400, CH500, and CH600. Models CH400, CH500, and CH600 have been validated for pre-vacuum steam sterilization of two lumens 3 mm in diameter or larger and 400 mm in length or less.
For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55℃. Models CH400, CH500, and CH600 have been validated for sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less for EO sterilization.
| Sterilization<br>Wrap Model | Intended Load | Maximum Recommended<br>Wrapped Package Content<br>Weight |
|-----------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------|
| | | Pre-Vacuum Steam and EO |
| CH100 | Very light weight package<br>(for example: batteries) | 3 lbs |
| CH200 | Light weight package<br>(for example: telescope with light cord) | 6 lbs |
| CH300 | Light to moderate weight package<br>(for example: general use medical instruments) | 9 lbs |
| CH400 | Moderate to heavy weight package<br>(for example: general use medical instruments) | 13 lbs |
| CH500 | Heavy weight package<br>(for example: general use medical instruments) | 17 lbs |
| CH600 | Very heavy weight package<br>(for example: general use medical instruments) | 25 lbs |
#### Table 2 - Wrap Model Recommendations
{4}------------------------------------------------
The following loads were used in the pre-vacuum steam Sterility Validation Studies:
- · CH100: 16 huck towels (17 in. x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.).
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm 1. D x 400 mm), and 8 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
• CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 12 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
• CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 20 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. trav.
The following loads were used in the EO Sterility Validation Studies:
· CH100: 16 huck towels (17 in, x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 2 fluid-resistant drapes (108 in. x 88 in.), 2.5 lbs of metal mass.
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass,
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 7.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm 1. D x 400 mm), and 11.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 19.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. trav.
Note: The loads used in the Sterlify Validation Study corresponded to the maximum wrapped package content weights in Table 2.
#### Substantial Equivalence
The proposed DuraBlue" Sterilization Wrap is substantially equivalent to the predicate devices. Both devices:
- � Have the same intended use
- Have the same material composition .
- Have the same physical and chemical properties ●
- Have the same configurations/dimensions .
- Are indicated for the same sterilization parameters .
- Are indicated for the same Maximum Wrapped Package Content Weights .
#### Summary of Testing
DuraBlue™ Sterilization Wrap performance has been tested in accordance with the applicable requirements recommended in the FDA's Guidance Document Premarket Notification 510(k) Submissions for Medical Sterilization Packaging System in Health Care Facilities; Draft Guidance for Industry and FDA (March 7, 2002) in the predicate Premarket Notifications. Testing included sterlization efficacy, physical properties, event related maintenance of package sterlity, and biocompatibility in compliance with the methods of ISO 10993 for pre-vacuum steam and EO. Data from testing demonstrates that the DuraBlue™ Sterilization Wrap maintains sterlity of the enclosed contents for 180 days following sterilization by pre-vacuum steam and EO.
{5}------------------------------------------------
Table 3: Performance Testing of Proposed DuraBlue Sterilization Wrap sterilized by Pre-vacuum steam or EO
| Performance Properties | | Results |
|---------------------------------------------------------------|-----------------------------|----------------|
| Sterilization Efficacy | | PASS |
| Microbial Barrier<br>Properties | Aerosol Challenge | PASS |
| | Event Related Shelf<br>Life | PASS- 180 days |
| Material Compatibility with Indicated<br>Sterilization Method | | Compatible |
| Biocompatibility | | PASS |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# February 28, 2013
Cardinal Health-Medical Products and Services C/O Mr. Ned Devine Underwriters Laboratories, Incorporated 333 Pfingsten Road NORTHBROOK IL 60062
Re: K123289
Trade/Device Name: Cardinal Health DuraBlue 114 Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: February 15, 2013 Received: February 19, 2013
# Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized graphic above the company name. The graphic is composed of several curved lines that resemble a bird in flight. The text "CardinalHealth" is written in a bold, sans-serif font.
Indications for Use
510(k) Number (if known): K123289
Device Name: Cardinal Health DuraBlue™ Sterilization Wrap
Indications for Use:
Cardinal Health DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider using the following modalities:
- Pre-vacuum steam at 270°F/132°C for 4 minutes .
- 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-. 80% relative humidity for 60 minutes
The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). Maintenance of package sterility was validated with real-time aging testing for a duration of 180 days for each indicated modality.
For pre-vacuum steam sterilization, the wrap has been validated for dry times for Models CH100 and CH200 and 30 minutes for Models CH300, CH400, CH500, and CH600. Models CH400, CH500, and CH600 have been validated for pre-vacuum steam sterilization of two lumens 3 mm in diameter or larger and 400 mm in length or less.
For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55°C. Models CH400, CH500, and CH600 have been validated for sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less for EO sterilization.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Devices Evaluation (ODE) Shani J. Smith =============================================================================================================================================================== 2013.02.28 15:52:20=05'00'
Division Sign-Off Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k)
{9}------------------------------------------------
Table 1 - Wrap Model Recommendations
| Sterilization<br>Wrap Model | Intended Load | Maximum Recommended<br>Wrapped Package Content<br>Weight |
|-----------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------|
| | | Pre-Vacuum Steam and EO |
| CH100 | Very light weight package<br>(for example: batteries) | 3 lbs |
| CH200 | Light weight package<br>(for example: telescope with light cord) | 6 lbs |
| CH300 | Light to moderate weight package<br>(for example: general use medical instruments) | 9 lbs |
| CH400 | Moderate to heavy weight package<br>(for example: general use medical instruments) | 13 lbs |
| CH500 | Heavy weight package<br>(for example: general use medical instruments) | 17 lbs |
| CH600 | Very heavy weight package<br>(for example: general use medical instruments) | 25 lbs |
The following loads were used in the pre-vacuum steam Sterility Validation Studies:
· CH100: 16 huck towels (17 in. x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.).
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm ¶. D x 400 mm), and 8 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
• CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 12 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
• CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 20 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
{10}------------------------------------------------
The following loads were used in the EO Sterility Validation Studies:
· CH100: 16 huck towels (17 in. x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 2 fluid-resistant drapes (108 in. x 88 in.), 2.5 lbs of metal mass.
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 7.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 11.5 Ibs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm 1. D x 400 mm), and 19.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
Note: The loads used in the Sterility Validation Study corresponded to the maximum wrapped package content weights in Table 1.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.