← Product Code [KDI](/productcode/KDI) · K123288

# HEMOCONCENTRATOR (K123288)

_Maquet Cardiopulmonary, AG · KDI · Jun 26, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123288

## Device Facts

- **Applicant:** Maquet Cardiopulmonary, AG
- **Product Code:** [KDI](/productcode/KDI.md)
- **Decision Date:** Jun 26, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The MAQUET Hemoconcentrators are used to remove excess fluid from the blood or to concentrate highly diluted blood solutions. They are only intended for use for blood concentration during and/or following cardiopulmonary bypass procedures. The maximum duration of use is 6 hours. The physician in charge of treatment has sole responsibility for decisions concerning use of the hemoconcentrator.

## Device Story

Hemoconcentrators (BC 60 plus, BC 140 plus) are extracorporeal devices used during or after cardiopulmonary bypass. They function by removing excess fluid from blood or concentrating diluted blood solutions. The device is a membrane-based filter; it requires filling and venting prior to use. It is operated by clinical staff in a surgical or intensive care setting. The physician determines the necessity of use based on the patient's fluid status. The device benefits the patient by managing blood volume and concentration during complex cardiac procedures.

## Clinical Evidence

Bench testing only. Testing included device integrity, performance, biocompatibility per ISO 10993-1, and compliance with ISO 8637 for cardiovascular implants and extracorporeal systems.

## Technological Characteristics

Hemoconcentrator utilizing a membrane filter. Materials evaluated for biocompatibility per ISO 10993-1. Performance tested per ISO 8637. Device is ready for use, requiring filling and venting; membrane contains no stabilizers. Mechanical device; no software or energy source.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Hemocor HPH® 700 ([K983085](/device/K983085.md))
- Hemocor HPH® 1400 ([K923139](/device/K923139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123288

JUN 2 6 2013

## MAQUET GETINGE GROUP

### 510(k) Summary

[as required by 21 CFR 807.92(c)]

| Submitter            | MAQUET Cardiopulmonary AG<br>Kehler Strasse 31<br>76437 Rastatt<br>Germany                                         |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| Contact Person       | Dr. Ingrid Richter<br>Phone: 011 49 7478 921 337<br>Fax: 011 49 7478 921 8667<br>e-mail: ingrid.richter@maquet.com |
| Date Prepared        | October 19, 2012                                                                                                   |
| Device Trade Name    | Hemoconcentrators : BC 60 plus, BC 140 plus                                                                        |
| Common/Usual Name    | Hemoconcentrator                                                                                                   |
| Classification Names | High Permeability Hemodialysis System<br>Hemoconcentrator<br>(21 CFR 876.5860 - Product Code: KDI)                 |
| Predicate Devices    | Hemocor HPH® 700, Minntech Corp. USA,<br>K983085<br>Hemocor HPH® 1400, Minntech Corp. USA,                         |

#### Device Description

BC 60 plus, BC 140 plus Hemoconcentrators are used to remove excess fluid from the blood during and/or following cardiopulmonary bypass procedures. Hemoconcentrators are ready for use after they have been filled and vented as the membrane contains no stabilizers. The type of Hemoconcentrator used is determined by the protocol used.

Page 1 of 2

{1}------------------------------------------------

# MAQUET GETINGE GROUP

#### Indications for Use

The MAQUET Hemoconcentrators are used to remove excess fluid from the blood or to concentrate highly diluted blood solutions. They are only intended for use for blood concentration during and/or following cardiopulmonary bypass procedures.

The maximum duration of use is 6 hours.

The physician in charge of treatment has sole responsibility for decisions concerning use of the hemoconcentrator.

#### Statement of Technical Comparison

The Hemoconcentrators, BC 60 plus, BC 140 plus are comparable to the Hemocor HPH® 700 K983085, Hemocor HPH® 1400 K923139 from Minntech Corp. USA, reqarding the principals of operation, performance and indications for use.

#### Non-clinical Testing

The Hemoconcentrators BC 60 plus and BC 140 plus have been tested to and met the requirements of ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing as well as the requirements of ISO 8637 Cardiovascular implants and extracomoreal systems - Haemodialysers. haemodiafilters, haemofilters and haemoconcentrators.

#### Determination of Substantial Equivalence

Evaluation and testing on safety and effectiveness was executed to demonstrate that the Hemoconcentrators BC 60 plus, BC 140 plus described in this submission are substantially equivalent to the Hemocor HPH® 700 (K983085), Hemocor HPH® 1400 (K923139) from Minntech Corp., USA.

The following areas have been tested and / or evaluated:

- Integrity .
- Performance -
- Biocompatibility -
- Sterility .

#### Conclusion

The data given demonstrate that the Hemoconcentrator BC 60 plus and BC 140 plus are substantially equivalent to the named predicate devices which have FDA market clearance.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized waves or lines.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

#### June 26, 2013

MAQUET Cardiopulmonary AG % Ms. Ingrid Richter Regulatory Affairs Manager Kehler Strasse 31 RASTATT 76437 GERMANY

Re: K123288

Trade/Device Name: BC 60 plus, BC 140 plus Hemoconcentrator Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: May 17, 2013 Received: May 20, 2013

Dear Ms. Richter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Ms. Ingrid Richter

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): __K123288

Device Name:

#### BC 60 plus, BC 140 plus Hemoconcentrator

Indications for Use:

The MAQUET Hemoconcentrators are used to remove excess fluid from the blood or to concentrate highly diluted blood solutions. They are only intended for use for blood concentration during and/or following cardiopulmonary bypass procedures. The maximum duration of use is 6 hours.

The physician in charge of treatment has sole responsibility for decisions concerning use of the hemoconcentrator.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Herbert P. Lerner -S

K123288

30

---

**Source:** [https://fda.innolitics.com/device/K123288](https://fda.innolitics.com/device/K123288)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
