PRIMALIF TM LLIF UNITARY PEEK LATERAL LUMBAR INTERBODY FUSION SYSTEM
K123207 · Osteomed LP · MAX · Nov 29, 2012 · Orthopedic
Device Facts
Record ID
K123207
Device Name
PRIMALIF TM LLIF UNITARY PEEK LATERAL LUMBAR INTERBODY FUSION SYSTEM
Applicant
Osteomed LP
Product Code
MAX · Orthopedic
Decision Date
Nov 29, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PrimaLIFTM LLIF Unitary PEEK Lateral Lumbar Interbody Fusion System (PrimaLIFTM LLIF) is indicated for intervertebral body fusion of the lumbar spine to be used with autogenous bone graft, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) and these patients may have up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation systems such as PrimaLOK™ SP and PrimaLOK™ FF which have been cleared for use in the lumbar spine. The PrimaLIFTM LLIF implants are intended for single use only.
Device Story
Unitary PEEK interbody cage system for lateral lumbar interbody fusion (LLIF). Device provides anterior column stability; features wedge-shaped nose, serrated endplate surfaces, and central cavity for bone graft. Includes instrumentation (osteotomes, trials, inserters) for direct lateral approach. Used by surgeons in clinical settings to treat degenerative disc disease. Implants are single-use; intended to facilitate fusion when used with autogenous bone graft and supplemental fixation systems.
Clinical Evidence
Bench testing only. Static and dynamic axial compression testing per ASTM F2077 and static subsidence testing per ASTM F2267 performed. No clinical data required.
Technological Characteristics
Unitary interbody cages made of PEEK OPTIMA LT1 (ASTM F 2026-08) with Tantalum markers (ASTM F560-08). Instrumentation made of chrome-coated stainless steel, anodized aluminum, and medical-grade silicone. Design features wedge-shaped nose, serrated endplate surfaces, and central bone graft cavity. Direct lateral approach.
Indications for Use
Indicated for intervertebral body fusion of lumbar spine (L2-S1) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis or retrolisthesis, following 6 months of failed non-operative treatment. Requires use with autogenous bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "OSTEOMED" in bold, black letters. The word is inside of a black oval shape. The oval shape is slightly tilted upwards.
## 510(k) Summary
| Submitter Information: | OsteoMed<br>3885 Arapaho Road<br>Addison, Texas 75001<br>Phone: (972) 677-4600<br>Fax: (972) 677-4601 | NOV 2 9 2012 |
|----------------------------------------|-------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Mrs. Piedad Peña | |
| Date Prepared: | November 21, 2012 | |
| Device Information: | | |
| Proposed Trade Name: | OsteoMed Spine PrimaLIF™ LLIF Unitary<br>PEEK Lateral Lumbar Interbody Fusion System | |
| Classification Name: | Intervertebral body fusion device | |
| Regulation Number: | 21 CFR 888.3080 | |
| Device Product Code: | MAX | |
| Device Class: | Class II | |
| Predicate Devices: | | |
| Lanx Fusion System - Lateral (K103666) | | |
Classification Name: Orthosis, spinal intervertebral fusion (21 CFR 888.3080 Product Code: MAX)
Device Class: II
#### DePuy Lumbar I/F Cage with VSP Spine System (P960025)
Classification Name: Intervertebral Fusion Device with Bone Graft, lumbar (21 CFR 888.3080 Product Code: MAX)
Device Class: II
#### Synthes Oracle Spacer (K072791)
Classification Name: Spinal Intervertebral Body Fusion (21 CFR 888.3080 Product Code: MAX)
Device Class: II
#### Device Description:
The PrimaLIFTM LLIF Unitary PEEK Lateral Lumbar Interbody Fusion System is a comprehensive interbody cage system that provides structural stability of the anterior column. It is comprised of unitary PEEK interbody cages of various sizes. The instruments provided with the system include osteoteomes, implant trials, and an inserter to facilitate placement of the interbody device.
#### Materials:
The device is made of PEEK (poly-ether-ketone) manufactured from Invibio PEEK OPTIMA® LT1 as described by ASTM F 2026-08 with Tantalum markers inserted for visibility with imaging as described by ASTM F560-08.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 (972) 677-4600 FAX: (800) 390-2620 Customer Service: (800) 456-7779
COLSON ASSOCIATE
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "OSTEOMED" in bold, black letters. The word is surrounded by a black, oval shape that is thicker on the sides and thinner on the top and bottom. The logo is simple and modern, with a focus on the company name.
The instrumentation is made from various grades of chrome coated stainless steel, anodized aluminum, and/or medical grade silicone.
#### Intended Use:
The OSTEOMED PrimaLIFTM LLIF is indicated for intervertebral body fusion of the lumbar spine to be used with autogenous bone graft, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) and these patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation systems such as PrimaLOKTM SP and PrimaLOKTM FF which have been cleared for use in the lumbar spine.
The PrimaLIFTM LLIF implants are intended for single use only.
#### Technological Characteristics:
The predicate devices consist of interverterbal body fusion devices designed with wedge shaped noses, serrations or teeth at endplate contacting surfaces, and central cavities to hold bone graft along with instrumentation for facilitating placement of the device using a direct lateral approach. The OsteoMed PrimaLIFTM LLIF device also is designed with wedge shaped noses, serrations or teeth at endplate contacting surfaces, and central cavities to hold bone graft along with instrumentation for facilitating placement of the device using a direct lateral approach.
Material used for the PrimaLIF™ LLIF is the same as the Lanx Lateral and Synthes Oracle Spacer predicate devices, PEEK OPTIMA® LT1 as described by ASTM F 2026-08. This material is biocompatible.
#### Performance/Clinical Data:
The PrimaLIF™ LLIF device has been tested in static and dynamic axial compression per ASTM F2077 and static subsidence per ASTM F2267. The intended use of the device is the same as the predicate devices. Clinical testing is not required to support substantial equivalence.
In conclusion, the device is safe and effective and performs as well as the predicate devices.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 (973) 677-4600 FAX: (800) 390-2620 Customer Service: (800) 456-7779
Image /page/1/Picture/13 description: The image shows a logo with the word "Colson" in a stylized font on the left, and the words "A COLSON ASSOCIATE" in a smaller, sans-serif font on the right. The word "Colson" is enclosed in a circular shape. The logo appears to be for a company or organization called "Colson Associate".
{2}------------------------------------------------
K123207 Page 3 of 3
Image /page/2/Picture/1 description: The image shows the logo for Osteomed. The logo consists of the word "OSTEOMED" in bold, sans-serif font, with the letters slightly slanted to the right. The word is enclosed within a horizontally stretched oval shape, with the top and bottom lines of the oval curving around the word. There is a small dot above and to the right of the letter "D".
### Substantial Equivalence:
The basis of substantial equivalence for this device is based on similarities in intended use, anatomic location, mechanical properties and method of stabilization to the predicate devices including the Lanx Lateral (K103666), the DePuy Lumbar I/F Cage (P960025), and the Synthes Oracle Spacer (K072791), based on their promotional materials, labeling and FDA clearance letters. Also, the basis of substantial equivalence for this device is based on similarities in material of manufacture to the Lanx Lateral (K103666) and the Synthes Oracle Spacer (K072791) based on their promotional materials, labeling and FDA clearance letters.
Due to the similarity of materials and design to the predicate devices, OsteoMed believes that the PrimaLIFTM LLIF device does not raise any new safety or effectiveness issues and performs as well as the predicate devices.
OsteoMed 3885 Arapaho Road Addison, Texas 75001 974) 677-4600 FAX: (800) 390-2620 Customer Service: (800) 456-7779
Image /page/2/Picture/6 description: The image shows the logo for A Colson Associate. The logo consists of the word "Colson" in a stylized font, with a vertical line extending from the top of the "l". To the right of the word "Colson" is the text "A COLSON ASSOCIATE" in a smaller, sans-serif font. The logo is simple and professional, and it is likely used to represent a business or organization.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 29, 2012
OsteoMed LP % Piedad Peña, M.S. Manager, Regulatory Affairs 3885 Arapaho Road Addison, Texas 75001
Re: K123207
Trade/Device Name: PrimaLIFTM LLIF Unitary PEEK Lateral Lumbar Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: October 11, 2012 Received: October 12, 2012
Dear Ms. Peña:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Piedad Peña, M.S.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known): K123207
Device Name: PrimaLIF™ LLIF Unitary PEEK Lateral Lumbar Interbody Fusion System Indications for Use:
The PrimaLIFTM LLIF Unitary PEEK Lateral Lumbar Interbody Fusion System (PrimaLIFTM LLIF) is indicated for intervertebral body fusion of the lumbar spine to be used with autogenous bone graft, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) and these patients may have up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation systems such as PrimaLOK™ SP and PrimaLOK™ FF which have been cleared for use in the lumbar spine.
The PrimaLIFTM LLIF implants are intended for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Ronald P. Jean
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123207
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.