K123174 · GE Healthcare · LLZ · Nov 16, 2012 · Radiology
Device Facts
Record ID
K123174
Device Name
CENTRICITY PACS-IW WITH UNIVERSAL VIEWER
Applicant
GE Healthcare
Product Code
LLZ · Radiology
Decision Date
Nov 16, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.
Device Story
Centricity PACS-IW with Universal Viewer is a software-only PACS component for medical image management. It receives DICOM-compliant images from modalities (CT, MR, US, NM, CR, MG, DX, PET/PT, XA, IO, RF, SC, VL). Deployed over TCP/IP networks on off-the-shelf hardware, it provides a unified workspace for image review, manipulation, and interpretation. Features include 'smart' hanging protocols based on DICOM tags, bookmarking of display states, and measurement tools (Cardio Thoracic Ratio, Ellipse). An optional Zero Footprint (ZFP) viewer allows web-based access for clinicians without local installation. The system does not generate images but facilitates clinical decision-making by providing diagnostic-quality visualization. It supports integration with GE's IDI Mammo and Advantage Workstation (AW) Server. The ZFP viewer is not intended for primary diagnosis.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including risk analysis, requirements/design reviews, usability analysis, and verification testing (unit, integration, performance, regression, system) and validation (simulated use).
Technological Characteristics
Software-only device running on off-the-shelf computer hardware. Utilizes TCP/IP networking. Supports DICOM imaging standards. Features include ZFP web-based viewer, hanging protocols, and measurement tools. Software developed under moderate level of concern. No specific materials or energy sources as it is a software-only PACS application.
Indications for Use
Indicated for trained professionals (radiologists, physicians, nurses, technicians, assistants) to view, communicate, process, and display medical images and data from various imaging sources. Includes mammograms. Contraindicated for primary diagnosis using lossy compressed mammographic images or digitized film screen images; mammographic interpretation requires FDA-approved 5MP resolution monitors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date written in a bold, sans-serif font. The date is 'NOV 16 2012', indicating November 16th, 2012. The letters and numbers are clearly legible against a white background. The text appears to be extracted from a document or label.
25
![GE logo]
GE Healthcare 510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: October 8, 2012
Submitter: GE Healthcare 540 West Northwest Highway Barrington, IL 60010
Primary Contact Person:
Cheryl Bork Regulatory Affairs Manager GE Healthcare Phone: 847-277-6038 Fax: 847-277-4506
Secondary Contact Person:
Jeme Wallace Regulatory Affairs Director GE Healthcare Phone: 847-277-4468 Fax: 847-939-1446
Device Trade Name: Centricity PACS-IW with Universal Viewer
Common/Usual Name: Picture Archiving and Communication System
Classification Names: 21 CFR 892.2050, System, Image Processing, Radiological
Product Code: LLZ
Predicate Device: K121387 - GE Healthcare Centricity PACS-IW
Device Description:
Centricity PACS-IW with Universal Viewer is an Internet based medical image display and interpretation software product that is part of a picture archiving and communications system. It provides users with capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images (including digital mammograms).
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters and the circle are black, creating a strong contrast.
#### 510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary .
Device Description: (cont.) Centricity PACS-IW with Universal Viewer includes features to access and manage medical imaging studies from Computed Tomography (CT), Magnetic Resonance (MR), Ultrasound (US), Nuclear medicine (NM). Computerized radiography (CR). Diaital mammography (MG),Digital x-ray (DX), Positron Emission Tomography (PET/PT), X-Ray Angiography (XA), Digital Intra-oral X-Ray (IO), Radiofluoroscopic x-ray (RF), Secondary Capture Images (SC), Visible Light (VL) Endoscopic, Microscopic and Photographic Image Storage, Slide Coordinates Microscopic Image Storage, Presentation States (PS), Key Image Notes (KIN) and other DICOM imaging modalities.
> Centricity PACS-IW with Universal Viewer is designed to be deployed over conventional TCP/IP networking infrastructure available in in most healthcare organizations and utilizes commercially available computer platforms and operating systems.
> The system does not produce any origing| medical images. All images located on the Centricity PACS, PACS-IW, and Enterprise Archive systems have been received from DICOM compliant modalities and/or image acquisition systems.
> The Universal Viewer Zero Foot Print (ZFP) clinical viewer is an optional viewer which allows trained professionals to display and manipulate images stored in Centricity Enterprise Archive or other DICOM archive devices. These trained professionals include but are not limited to physicians, radiologists, nurses, medical technicians, and assistants.
> The Universal Viewer Zero Foot Print option is not intended for primary diagnosis.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. There are three small decorative elements resembling stylized leaves or flourishes placed around the circle's perimeter.
510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary
Intended Use: Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations.
> Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA.
Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.
Technology: The Centricity PACS-IW with Universal Viewer device employs the same fundamental scientific technology as its predicate device, Centricity PACS-IW cleared under K121387, with the following modifications:
- . Modified viewer to present a common unified workspace for radiologists and clinicians to perform the review, manipulation and diagnostic interpretation of images and other information generated by acquisition.
- . Provides integration to Centricity PACS backend. The PACS-IW with Universal Viewer can read image data and other patient information from Centricity PACS backend.
- Implementation of a "Bookmark" feature for . image display, that enables the user to capture the Display State of an exam context as it exists at any particular time, and then restores this same state at a later time.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The logo is black and white.
- Enhanced hanging protocols ("smart" hanging . protocols) for image display, which will assess information provided in the DICOM tags to evaluate the type of study that is being displayed to determine the most appropriate method for the study to be hung.
- Enhanced image annotations and . measurements to include the ability to calculate Cardio Thoracic Ratio (CTR) (web client only) and Ellipse (ZFP only).
- Supports interface with GE Healthcare's IDI . Mammo product, in place of Cedara Mammo.
- . Supports interface with GE Healthcare's Advantage Workstation (AW) Server advanced visualization tools.
### Technology (cont.)
- Hardware minimum specifications were . modified as a result of technology advancements and obsolescence issues.
- The optional ZFP (Zero Footprint) Clinician Viewer . is a non-diagnostic DICOM image and results viewer. Zero footprint:
- image enables an EMR or other similar o applications.
- displays image data from GE O Healthcare's Centricity Enterprise Archive.
- can be accessed on a client device o without requiring the user to override security settings or initiate an installation.
Centricity PACS-IW with Universal Viewer receives medical images and other information from various data sources. The information can be stored, communicated, processed and displayed within the
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a circular frame. The letters and the frame are solid black, contrasting with the white background.
510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary
system or across computer networks at distributed locations, the same as its predicate devices. Centricity PACS-IW with Universal Viewer is a software-only device that runs on commercially available off-the-shelf computer hardware platforms.
The Centricity PACS-IW with Universal Viewer device will continue to have an intended use and functionality fitting within the definition of 21 CFR 892.2050. Picture Archiving and Communication Systems, Product Code LLZ.
#### Determination of Substantial Equivalence:
#### Summary of Non-Clinical Tests
The software documentation was provided at a moderate level of concern following the FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Centricity PACS-IW with Universal Viewer complies with voluntary standards as detailed in this premarket notification submission. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Requirements Reviews .
- Design Reviews .
- Usability Analysis .
- . Testing on unit level (Verification)
- Integration testing (Verification) .
- . Performance testing (Verification)
- . Regression testing (Verification)
- System testing (Verification) ♥
- Simulated use testing (Validation) .
#### Summary of Clinical Tests:
The subject of this premarket notification submission, Centricity PACS-IW with Universal Viewer, did not require clinical studies to support substantial equivalence.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a circular border. The letters are stylized and connected, creating a unique and recognizable design. The logo is in black and white.
510(k) Premarket Notification Submission Centricity PACS-IW with Universal Viewer Section 5: 510(k) Summary
Comparison of the Intended Uses, the technological Conclusion: characteristics, and performance specifications demonstrate the functional equivalence of the subject device to the predicate device. Verification and Validation testing results demonstrate that no adverse effects have been introduced by these differences.
> Information provided in this premarket notification submission supports the Centricity PACS-IW with Universal Viewer medical device to be as safe, as effective and substantially equivalent to its predicate device.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its head facing left and its wings spread. The eagle is composed of three thick, curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
November 16, 2012
Ms. Cheryl Bork Regulatory Affairs Manager GE Healthcare-HCIT 540 W. Northwest Highway BARRINGTON IL 60010
Re: K123174
Trade/Device Name: Centricity PACS-IW with Universal Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 8, 2012 Received: October 25, 2012
Dear Ms. Bork:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Ms. Bork
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quarty of would provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Michael D. O'Hara
Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K123174
Device Name: Centricity PACS-IW with Universal Viewer
Indications for Use:
Lentricity PACS-IW with Universal viewer is a software only product that is part of a Picture Archiving and Communication System (PACS) medical device.
Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA.
Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.
Prescription Use Yes (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K123174
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.