U&U HYDROGEN PEROXIDE GAS PLASMA STERILIZATION POUCH/ROLL
K123162 · U&U Medical Technology Co, Ltd. · FRG · Feb 26, 2014 · General Hospital
Device Facts
Record ID
K123162
Device Name
U&U HYDROGEN PEROXIDE GAS PLASMA STERILIZATION POUCH/ROLL
Applicant
U&U Medical Technology Co, Ltd.
Product Code
FRG · General Hospital
Decision Date
Feb 26, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll are intended to be used to enclose medical devices that are to be terminally sterilized in the STERRAD 100NX Hydrogen Peroxide Gas Plasma sterilizer.
Device Story
Sterilization packaging system consisting of Tyvek and medical plastic film pouches/rolls; heat-sealed or adhesive-sealed by healthcare personnel. Used to enclose medical instruments prior to terminal sterilization in STERRAD 100NX Hydrogen Peroxide Gas Plasma sterilizer. Packaging maintains sterility of contents for 30 days post-sterilization. Protects sterilized instruments during storage and transport until point of use. Benefits patient by ensuring maintenance of sterile barrier for surgical/medical devices.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included sterilant penetration, package integrity, maintenance of package integrity, material compatibility, shelf-life, and biocompatibility. All tests passed.
Technological Characteristics
Materials: Tyvek and laminated medical plastic film. Construction: Heat-sealed or adhesive-sealed on three sides. Form factor: Various pouch sizes and rolls. Sterilization: Compatible with STERRAD 100NX Hydrogen Peroxide Gas Plasma. Shelf life: 2 years. Sterile barrier maintenance: 30 days.
Indications for Use
Indicated for enclosing general medical instruments (metal/nonmetal, including hinged), stainless steel single-channel lumens (ID ≥3mm, length ≤150mm), and polyethylene/Teflon lumen tubing (ID ≥3mm, length ≤500mm) for terminal sterilization in STERRAD 100NX Standard cycle. Limit 20 tubing pieces per cycle; one device per pouch.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Tyvek Pouch/Roll with STERRAD Chemical Indicator (K103210)
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 2 6 2014
| 510(k) Submission |
|-------------------|
|-------------------|
U&U ChangZhou Modical Packaging Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Madical Technology Co., Limited RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [Gas Plasma Sterilization Pouch / Roli]
21/02/14 Rev 1.02
## Sec 005_510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92
Date Prepared: 2014-01-05 Submission Numbers for Pre-Submission: K123162
### 1. Submitter Name and Address:
| Name: | U&U Medical Technology Co., Ltd |
|------------------|-----------------------------------------------------------|
| Address: | Dongzhou Village, Hengshanqiao, Changzhou, Jiangsu, China |
| Contactor Name: | Garfield Wang |
| TEL: | +86-13902471751 |
| E-mail: | Wangxuebo_11@hotmail.com |
| US Agent: | Pan Angels Corp: |
| Address: | 3330 Fowler Street, Los Angeles, CA 90063, U.S.A |
| TEL: | (323)422-8581 |
| Contact person : | Mr. Michael Kim |
#### 2. Submission Devices Information:
Trade/Proprietary Name: U&J Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll Common Name: Plasma Sterilization Pouch / Roll Classification name: Sterilization Wrap Class: Il FDA review panel code: INCB Product code: FRG - sterilization wrap CFR Regulation Number: 21CFR 880.6850
#### 3. Predicate Devices Information:
Trade Name: Tyvek Pouch/Roll with STERRAD Chemical Indicator 510(K) Number: K103210 Manufacturer: Advanced Sterilization Products Company.
#### 4. Devices Description:
The U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll are manufactured from TYVEK and medical plastic film that are heat sealed on three sides. The fourth side has an adhesive tape that is used to seal the pouch or heat-sealed by the heat-seal machine. The TYVEK conforms to recognized material standards and can be sterilized by STERRAD 100 NX Hydrogen Peroxide Gas Plasma. The Sterilization Pouch has the same intended use, Essential Component, Raw material, Sterilization method, manufacturing methods and same technological characteristics as these predicate devices. Substantial equivalent to the predicate device was established by physical testing of the TYVEK (pressure drop vs. flow and filtration efficiency) and film (thickness, tensile strength and elongation) from non-sterile, sterilized finished devices, as well as, performance of these finished devices (seal strength, package burst, dye migration).
The pouches are manufactured from TYVEK that is thermally sealed to laminated film on the left, right, and bottom of pouch. The fourth side has an adhesive tape that is used to that is
[PMN-510(K) SUBMISSION]
{1}------------------------------------------------
510(k) Submission
U&U ChangZhou Medical Packaging Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [Gas Plasma Sterillzation Pouch / Roll]
Rev 1.02 21/02/14
used to seal the pouch or heat-sealed by the heat-seal machine prior to sterilization of the enclosed medical device.
#### 5. Model numbers of the pouches:
| Ref | Model | Description | Size |
|---------|--------|-------------------------------------|--------------|
| Number | Number | | |
| UUP0001 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 3" X 8" |
| UUP0002 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 4" X 10.25" |
| UUP0003 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 6" X 12.5" |
| UUP0004 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 4" X 14" |
| UUP0005 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 8" X 16" |
| UUP0006 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 6" X 16.5" |
| UUP0007 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 10" X 19" |
| UUP0008 | UUP | TYVEK Pouch-Heat-Seal and Self-Seal | 12.5" X 22" |
| UUR0001 | UUR | TYVEK Roll-Heat-seal | 3" X 228' |
| UUR0002 | UUR | TYVEK Roll-Heat-seal | 4" X 228' |
| UUR0003 | UUR | TYVEK Roll-Heat-seal | 6" X 228' |
| UUR0004 | UUR | TYVEK Roll-Heat-seal | 14" X 228' |
| UUR0005 | UUR | TYVEK Roll-Heat-seal | 16.5" X 228' |
| UUR0006 | UUR | TYVEK Roll-Heat-seal | 20" X 228' |
#### 5. Intended Use:
U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll are intended to be used to enclose medical devices that are to be terminally sterilized in the STERRAD 100NX Hydrogen Peroxide Gas Plasma sterilizer.
## The recommended Hydrogen Peroxide Gas Plasma cycle parameter: STERRAD 100NX STANDARD Cycle
#### Sterlization load claim:
## STERRAD 100NX Sterilizer Standard cycle:
· General medical instruments (metal and nonmetal, including hinged devices)
· Instruments with single-channel stainless steel lumens with an Internal diameter 3 mm or larger and length 150 mm or shorter.
• Polyethylene and Teflon lumen tubing with an internal diameter of 3 mm or larger and length of 500 mm or shorter.
· These items must be sterilized without any additional load items
• Limit of 20 pieces of tubing per cycle
Load condition: One device per Sterilization Pouch.
Load chamber: Two shelves (total weight: 21.4 Ibs or 9.7 kg)
Load Pouch: Leave enough material beyond the seal for the opener to easily grasp (usually 1- 1 % inches).
SHELF LIFE: It is recommended that the products are put to their end use within 2 years of manufacture. The recommended "best before" date and the manufacturing date are stated on the label. This device was demonstrated to maintain sterility of the contents for 30 days after Sterilization.
[PMN-510(K) SUBMISSION]
{2}------------------------------------------------
| U&U ChangZhou Medical Packaging Technology Co., Ltd<br>Dongzhou Village, Hengshanglao, Changzhou, Jiangsu, China<br>U&U (HONGKONG) Medical Technology Co., Limited<br>RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG<br>[Gas Plasma Sterilization Pouch / Roll] | | 510(k) Submission<br> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|-----------------------|
| | | |
| | | |
| | Rev 1.02 | 21/02/14 |
## 6. Technological Characteristics:
Performance testing was conducted to show that the U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll maintain sterility until the seal of the Pouch/Roll is opened.
## 6.1 Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll Performance Tests
| Element | Results |
|----------------------------------|---------|
| Sterilant Penetration | PASSED |
| Package Integrity | PASSED |
| Maintenance of Package Integrity | PASSED |
| Material Compatibility | PASSED |
| Shelf-Life | PASSED |
| Biocompatibility | PASSED |
| Performances | PASSED |
| Biocompatibility | PASSED |
## 6.2 Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll Comparison Table
| Element of Comparison | Submission Device | Predicate Device<br>K103210 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Pouch Construction:<br>Tyvek | YES | YES |
| Film | YES | YES |
| User Heat Seal | YES | YES |
| User Self Seal | YES | YES |
| Single Use Device | YES | YES |
| Shelf-Life | 2 Years, This device was<br>demonstrated to<br>maintain sterility of the<br>contents for 30 days<br>after Sterilization. | 2 Years |
| Labeling | Meet the requirements of<br>21 CFR Part 801 | Meet the requirements of<br>21 CFR Part 801 |
## 7. Conclusion:
The materials, intended use, performance, function, and operational features of both the submitted device and the predicate device are substantially equivalent.
END
โคพท-240(K) ลักษณะระเวณ์
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized human figure embracing a sphere, symbolizing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 26, 2014
U&U Medical Technology Company, Limited Mr. Garfield Wang General Manager Dongshou Village, Hengshangiao, Changzhou, Jiengsu 213119 CHINA
Re: K123162
Trade/Device Name: U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: January 23, 2014 Received: January 29, 2014
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page-2 - Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/4/Picture/7 description: The image shows a signature and some text. The text includes "Sincerely yours," "Tejashri Purohit-Sheth, M.D.," "Clinical Deputy Director," and "DAGRID." There is also a large, stylized signature to the left of the text. The word "FOR" is at the bottom right.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
610(k) Submission
U&U ChangZhou Modical Packaging Tochnology Co., Ltd
Dongzhou Village,Hengshanqfao,Changzhou,Jiangsu,China U&U (HONGKONG) Modical Technology Co., Limited RM C1-D &F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [Gas Plasma Sterllization Pouch / Roll]
Rev 1.02 21/02/14
## Sec 004 Indications for Use
510(k) Number (if known): K123162
## Device Name: U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roll
## Indications for Use
U&U Hydrogen Peroxide Gas Plasma Sterillzation Pouch / Roll are intended to be used to enclose medical devices that are to be terminally sterilized in the STERRAD 100NX Hydrogen Peroxide Gas Plasma sterilizer.
### The recommended Hydrogen Peroxide Gas Plasma cycle parameter: STERRAD 100NX STANDARD Cycle
#### Sterilization load claim:
#### STERRAD 100NX Sterilizer Standard cycle:
· General medical instruments (metal and nonmetal, including hinged devices)
· Instruments with single-channel stalniess steel lumens with an Internal diameter 3 mm or larger and length 150 mm or shorter.
· Polyethylene and Teflon lumentubing with an internal diameter of 3 mm or larger and length of 500 mm or shorter.
· These items must be sterilized without any additional load items
· Limit of 20 pieces of tubing per cycle
Load condition: One device per Sterilization Pouch.
Load chamber: Two shelves (total weight: 21.4 Ibs or 9.7 kg)
Load Pouch: Leave enough material beyond the seal for the opener to easily grasp (usually 1- 1 % inches).
SHELF LIFE: It is recommended that the products are put to their end use within 2 years of manyfacture. The recommended "best before" date and the manufacturing date are stated on the label. This device was demonstrated to maintain sterility of the contents for 30 days after Sterilization.
> AND/OR Over-The-Counter Use Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Elizabeth F. Gave 2014.02.24 17:35:14 =05'00'
[PMN-510{K) SUBMISSION]
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.