SMARTGUARD NIGHT GUARD, SMARTGUARD ORIGINAL, SMARTGUARD NIGHT GUARD ORIGINAL, SMARTGUARD ELITE
K123161 · Smartguard, Inc. · MQC · Mar 14, 2013 · DE
Device Facts
| Record ID | K123161 |
| Device Name | SMARTGUARD NIGHT GUARD, SMARTGUARD ORIGINAL, SMARTGUARD NIGHT GUARD ORIGINAL, SMARTGUARD ELITE |
| Applicant | Smartguard, Inc. |
| Product Code | MQC · DE |
| Decision Date | Mar 14, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
OTC: - Protection against bruxism or night time teeth grinding - Reduce damage to the teeth and reduce the noise associated with bruxing and/or grinding. Rx: - Protection of teeth grinding, bruxism, and jaw clenching. - Protection of restorations from injury due to bruxism or clenching. - Relief of bruxism related headaches and pains. - Short-term pain relief from muscle spasm due to occlusal interference; for the prevention of chronic tension and temporal mandibular joint (TMJ) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscle. - Temporary treatment of Temporal Mandibular Disorder (TMD) along with the relief of associated headaches and pains.
Device Story
SmartGuard is a thermoplastic dental night guard composed of Elvax; molded at home or in-office by submerging in hot water to conform to anterior maxillary teeth (first bicuspid to first bicuspid). Device worn on upper 6-8 anterior teeth; prevents posterior teeth occlusion; relaxes jaw muscles; reduces clenching/grinding symptoms. Lower anterior teeth bite against substrate; allows free sliding to reduce posterior cuspal interferences. Worn at night (max 12 hours). Available in two designs: Original (21mm palatal extension) and Elite (30mm palatal extension). Benefits include reduced tooth damage, noise reduction, and relief from TMD/bruxism-related pain.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and material identity to legally marketed predicate devices.
Technological Characteristics
Material: Elvax thermoplastic. Form factor: Anterior maxillary dental guard. Dimensions: Original (21mm palatal extension), Elite (30mm palatal extension). Principle: Mechanical occlusal separation to prevent posterior tooth contact and relax jaw muscles. Customization: Heat-moldable by user.
Indications for Use
Indicated for patients requiring protection against bruxism, night-time teeth grinding, and jaw clenching. Used for protection of restorations, relief of bruxism-related headaches/pains, short-term pain relief from muscle spasms due to occlusal interference, prevention of chronic tension/TMJ syndrome, and temporary treatment of TMD.
Predicate Devices
Submission Summary (Full Text)
{1}------------------------------------------------
MAR. 1 4 2013
# .HUMPHREY&ASSOCIATESPA
CHRISTINE M. HUMPHREY, ESO CHUMPHREY OCHPALAW.CO ITTED IN FLORIDA
501 BRICKELL KEY DRIV SUITE 104 MIAMI. FLORIDA 33131 P: 305.372.2421 : 305.372.7983
W.CHPALAW.COM
This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 C.F.R. 807.92.
### General Information
A. Submitted By:
Dr. David L. Spainhower, DDS SmartGuard, Inc. 2360 Corporate Circle, Suite 400 Henderson, NV 89074
Contact Person:
Date Prepared:
B. Device Trade Name:
Common Name:
Classification Name:
C. Predicate Devices: (SleepRight -
Dr. David L. Spainhower
August 24, 2012
SmartGuard® Night Guard, SmartGuard® Original, SmartGuard® Night Guard Original and SmartGuard Elite"
Dental Guard/Night Guard/Mouth Guard
Unclassified
SleepRight Products (K071404 and K022809)
Select; SleepRight - Low Profile; SleepRight -Advance; SleepRight - Low Profile Rx, Advance Rx)
NTI Tension Suppression System (K010876 & K981546)
Freeway Comfort Bite Guard (K110468)
GrindGuardn (K082723)
Brux-TMD QuickSplint (K111066)
Doctor's Night Guard (K073220)
.
{2}------------------------------------------------
August 28, 2012 Page 2 of 3
#### D. Device Description:
SmartGuard® The SmartGuard® is a dental night guard composed of Elvax and once this thermoplastic material is submerged in hot water (see molding instructions) it softens and conforms to the individual's anterior maxillary teeth from first bicuspid. SmartGuard® is designed to be worn on the anterior, upper 6 to 8 teeth, preventing the posterior teeth from coming into occlusion, which relaxes the jaw muscles and helps to reduce jaw clenching and grinding and reduces the associated symptoms. The lower bottom anterior teeth bite against the bottom of the substrate with the ability to slide around freely as the device reduces posterior teeth cuspal interferences. The design features of SmartGuard® allow patients to easily mold this device at home or in a prescribing office, to fit their teeth properly. The SmartGuard® is designed to be worn at night and not longer than 12 consecutive hours in a 24 hour period.
SmartGuard® Original and SmartGuard Elite® differ slightly in design, with the primary difference being in the width of the palatal ex-tension, (Original 21mm and Elite 30 mm) as well as the distance between the outer rim and mesial/occlusal surface of the palatal extension (Original 10 mm and Elite 9 mm). The SmartGuard Elite® was designed to fit the majority of individuals mouths. However, if the patient has a small mouth or has a sensitive gag reflex, the SmartGuard® Original might be a better fit.
#### OTC:
- Protection against bruxism or night time teeth grinding
- . Intended to reduce damage to the teeth and reduce noise associated with bruxing and/or grinding.
#### Rx:
- . Protection of teeth grinding, bruxism, and jaw clenching.
- . Protection of restoration from injury due to bruxism or clenching.
- . Relief of bruxism related headaches and pains.
- . Short-term pain relief from muscle spasm due to occlusal interference; for the prevention of chronic tension and temporal mandibular joint (TMI) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscles.
- Temporary treatment of Temporal Mandibular Disorder (TMD) along with the relief of . associated headaches and pains.
{3}------------------------------------------------
#### E. Comparison of Technical Characteristics of Predicate Devices
See Attachment Comparison of predicate devices
## OTC:
The SmartGuard® devices are substantially equivalent to the SleepRight Select, SleepRight Low Profile, SleepRight Advance (cleared under K071404) and Freeway Comfort Bite Guard (K110468) with product code OBR. SmartGuard® is substantially equivalent to the above indicated devices cleared under the referenced 510K numbers. SmartGuard® is identical in design and material to the OTC devices cleared under K071404 and K110468. It has the same intended uses, indications, and similar technological characteristics as its predicate devices. The minor technological differences between SmartGuard® and the predicate devices raise no new questions of safety or effectiveness. Thus, SmartGuard® is substantially equivalent to the above referenced products and product code OBR noted above.
## Rx:
The SmartGuard® devices are also substantially equivalent to SleepRight Low Profile Rx, SleepRight Advance Rx, Freeway Comfort Bite Guard, GrindGuardn and Brux-TMD QuickSplint with product code MQC. SmartGuard® has the same intended uses, indications and similar technological characteristics as its predicate devices. The minor technological differences between SmartGuard® and the predicate devices raise no new questions of safety or effectiveness. Thus, SmartGuard® is substantially equivalent to the above referenced products with the Product Code MOC.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing segments, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## March 14, 2013
SmartGuard, Incorporated C/O Ms. Shelly Garg Attorney Humphrey & Associates, P.A. 601 Brickell Key Drive, Suite 104 MIAMI FL 33131
## Re: K123161
Trade/Device Name: SmartGuard® Night Guard, SmartGuard® Original, SmartGuard® Night Guard Original and SmartGuard Elite™
Regulation Number: Unclassified Regulation Name: Prescription Mouthguard Regulatory Class: Unclassified Product Code: MQC, OBR Dated: January 10, 2013 Received: February 12, 2013
### Dear Ms. Garg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the text "Kwame O. Ulmer for" in a simple, sans-serif font. The name "Kwame" is in bold, followed by "O. Ulmer" which is also in bold but with a decorative outline. The word "for" is smaller and in a regular font, positioned to the right of the name.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## HREY&ASSOCIATES ...
Christine M. Humphre CHUMPHREY@CHPALAW.CO ADMITTED IN FLORIDA
501 BRICKELL KEY DRIVE SUITE 104 MAMI. FLORIDA 33131 P: 305.372.2421 F: 305.372.7983
WWW.CHPALAW.COM
October 1, 2012
#### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: SmartGuard® Night Guard® Original, SmartGuard® Night Guard Original and SmartGuard Elite™
Indications For Use:
OTC:
- Protection against bruxism or night time teeth grinding ♥
- Reduce damage to the teeth and reduce the noise associated with bruxing and/or grinding .
Rx:
- Protection of teeth grinding, bruxism, and jaw clenching. o
- Protection of restorations from injury due to bruxism or clenching. o
- Relief of bruxism related headaches and pains. .
- Short-term pain relief from muscle spasm due to occlusal interference; for the prevention of . chronic tension and temporal mandibular joint (TMJ) syndrome that is caused by chronic jaw clenching of the mandibular and maxillary teeth by the temporalis muscle.
- Temporary treatment of Temporal Mandibular Disorder (TMD) along with the relief of associated o headaches and pains.
| Prescription Use | X |
|------------------|---|
|------------------|---|
AND/OR
| Over-The-Counter Use | X |
|----------------------|---|
|----------------------|---|
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Mary S. Runner -5 Concurrence of CDRH, Office of Device Evaluation (ODE) 2013.03.14 1.1 :57:49 -04'00'
Division Sign-Off) )ivision of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number:
k123161
(Part 21 CFR 801 Subpart D)
Page 1 of 1
4 . 1