← Product Code [GCJ](/productcode/GCJ) · K123102

# SOVEREIGN MINI (K123102)

_Aesculap, Inc. · GCJ · Jan 3, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123102

## Device Facts

- **Applicant:** Aesculap, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 3, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Aesculap's Sovereign® mini system is indicated for use in adult and pediatric diagnostic and therapeutic general endoscopy and laparoscopy surgery.

## Device Story

Sovereign® mini system consists of modular laparoscopic instruments (forceps, scissors) and reusable trocars/pins; used in adult and pediatric general surgery. Instruments available in 200mm and 290mm lengths with interchangeable ratcheting/non-ratcheting handles. Trocars available in 3.5mm and 5mm diameters with varying lengths; 5mm trocars accommodate 4mm and 5mm endoscopes. Devices feature monopolar capabilities. Operated by surgeons in clinical settings to perform diagnostic and therapeutic procedures. System provides mechanical access and tissue manipulation capabilities during minimally invasive surgery.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Modular laparoscopic instruments and trocars; materials: stainless steel and PEEK; monopolar electrical capability; dimensions: 200/290mm instruments, 3.5/5mm diameter trocars; manual operation; no software/electronics.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Aesculap Needlescopic Instrument System ([K982623](/device/K982623.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123102 Page 1 of 2

#### B. Special 510(k) Summary (as required by 21 CFR 807.92)

|                      | Sovereign® mini System<br>September 28, 2012                                                                       | JAN 0 3 2013 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|--------------|
| COMPANY:             | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |              |
| CONTACT:             | Denise R. Adams<br>610-984-9076<br>610-791-6882 (fax)<br>denise.adams@aesculap.com                                 |              |
| COMMON NAME:         | Laparoscope, General and Plastic Surgery                                                                           |              |
| CLASSIFICATION NAME: | Endoscope and Accessories                                                                                          |              |
| REGULATION NUMBER:   | 21 CFR 876.1500                                                                                                    |              |
| PRODUCT CODE:        | GCJ                                                                                                                |              |
|                      |                                                                                                                    |              |

## SUBSTANTIAL EQUIVALENCE

Aesculap®, Inc. believes that the modified components of the Sovereign® mini system are substantially equivalent to the components of the Aesculap Needlescopic Instrument System (K982623).

## DEVICE DESCRIPTION

Aesculap's Sovereign® mini system can be used in adult and pediatric laparoscopic general surgery. The system consists of modular forceps and scissors in lengths of 200 and 290 millimeters with interchangeable ratcheting and non-ratcheting handles. The system also includes reusable trocars and trocar pins. The trocars and trocar pins are available in 3.5 millimeter diameter and 60 and 110 millimeter lengths, and 5 millimeter and 60,110 and 150 millimeter lengths with or without a stopcock. The 3.5mm trocars are for use with the 3.5mm instruments, and the 5mm trocars accommodate 4 and 5mm endoscopes. The instruments and trocars are composed of stainless steel and PEEK materials.

### INDICATIONS FOR USE

Aesculap's Sovereign® mini system is indicated for use in adult and pediatric diagnostic and therapeutic general endoscopy and laparoscopy surgery.

{1}------------------------------------------------

K123102

Page 2 of 2

ರ

# TECHNOLOGICAL CHARACTERISTICS

The modified instruments have the same technological characteristics as the predicate devices. They are similar in design and size, manufactured from the same materials, and have monopolar capabilities. .

# PERFORMANCE DATA

No applicable performance standards have been promulgated under Section 514 of the Food, Drug, and Cosmetic Act for these devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged around the top half of the circle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Aesculap, Incorporated % Ms. Denise Adams Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

January 3, 2012

· Re: K123102

Trade/Device Name: Aesculap Sovereign® mini System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: December 01, 2012 Received: December 04, 2012

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Denise Adams

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Peter D. Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### INDICATIONS FOR USE STATEMENT . A.

510(k) Number:_K123102

Device Name: Aesculap Sovereign® mini System

Indications for Use:

Aesculap's Sovercign® mini system is indicated for use in adult and pediatric diagnostic and therapcutic general endoscopy and laparoscopy surgery.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X = and/or Over-the-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dwight Yen 2013.01.02 15:52:57 -05'00'

(Division Sign-Off) Division of Surgical Devices 510(k) Number ___K123102

---

**Source:** [https://fda.innolitics.com/device/K123102](https://fda.innolitics.com/device/K123102)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).