TYSONBIO AC500, AC800 BLOOD GLUCOSE MONITORING SYSTEM, TYSONBIO AC500 AC800 PRO BLOOD GLUCOSE MONITORING SYSTEM

K123087 · Tyson Bioresearch, Inc. · NBW · Jun 18, 2013 · Clinical Chemistry

Device Facts

Record IDK123087
Device NameTYSONBIO AC500, AC800 BLOOD GLUCOSE MONITORING SYSTEM, TYSONBIO AC500 AC800 PRO BLOOD GLUCOSE MONITORING SYSTEM
ApplicantTyson Bioresearch, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateJun 18, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The TysonBio AC500 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio AC500 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio AC500 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC500 Test Strips are for use with the TysonBio AC500 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC500 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

Device Story

System comprises glucose meter, test strips, and control solutions. Operates via electrochemical biosensor technology using glucose oxidase; measures glucose in 0.5uL capillary blood samples within 5 seconds. Meter performs auto-coding upon strip insertion. AC800/AC800 Pro models include voice guidance (not for vision-impaired). Used by patients at home or professionals in clinical settings to monitor diabetes control. Results displayed on meter; used by patients/providers to adjust diabetes management. Benefits include rapid, quantitative glucose monitoring to aid glycemic control.

Clinical Evidence

Performance evaluated in laboratory and clinical settings by healthcare professionals and lay users. Studies confirmed system accuracy and performance meet intended use requirements. No specific numerical clinical metrics (e.g., MARD, bias) provided in summary.

Technological Characteristics

Electrochemical biosensor using glucose oxidase. Hematocrit range 20-60%. Sample volume 0.5uL. Measurement range 20-600 mg/dL. Auto-coding via strip insertion. Powered by two AAA batteries. Connectivity via USB protocol. Memory capacity 500 results. Features include hypoglycemic/hyperglycemic alarms and meal flagging.

Indications for Use

Indicated for the quantitative measurement of glucose in capillary whole blood as an aid in monitoring the effectiveness of diabetes control programs.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY ## JUN 1 8 2013 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. - 1. Date prepared May 8, 2013 # 2. Sponsor information Tyson Bioresearch, Inc. 5 F., # 22, Ke E. Road III., science based industrial park Chun-Nan, Miao-Li county, Taiwan 350 Correspondent: WEN-HAI TSAI Phone: 886-37-585988 EXT 720 Facsimile: 886-37-585996 #### 3. Device name and classification Device Name: TysonBio AC500 Blood Glucose Monitoring System TysonBio AC800 Blood Glucose Monitoring System TysonBio AC500 Pro Blood Glucose Monitoring System TysonBio AC800 Pro Blood Glucose Monitoring System TysonBio AC500/AC800 Control Solution TysonBio AC500 Test Strip TysonBio AC800 Test Strip TysonBio AC500 Pro Test Strip TysonBio AC800 Pro Test Strip Common Names: Blood Glucose Monitoring System #### Classification: Classification Regulation: 21 CFR 862.1345 and 21CFR 862.1660 Classification: Class II (Glucose Test System) Class I (Quality control material ) Product Code: NBW- System, Test, Blood Glucose, Over-the-counter CGA - Glucose Oxidase, Glucose - JJX Single (specified) analyte controls (assayed and unassayed) - Panel Clinical Chemistry and Toxicology {1}------------------------------------------------ ### 4. Device description The TysonBio AC500/AC800 Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio AC500/AC800 Test Strips and TysonBio AC500/AC800 Control Solution with the TysonBio AC500/AC800 Blood Glucose Monitoring System. The TysonBio AC500 Pro/AC800 Pro Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio AC500 Pro/AC800 Pro Test Strips and TysonBio AC500/AC800 Control Solution with the TysonBio AC500 Pro/AC800 Pro Blood Glucose Monitoring System. The TysonBio AC500/AC800/AC500 Pro/AC800 Pro Blood Glucose Monitoring System consists of: - a. TysonBio AC500/AC800/AC500 Pro/AC800 Pro Blood Glucose Meter - b. TysonBio AC500/AC800/AC500 Pro/AC800 Pro Test Strips - c. TysonBio AC500/AC800 Control Solution #### 5. Intended use ### TysonBio AC500 Blood Glucose Monitoring System : The TysonBio AC500 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio AC500 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio AC500 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC500 Test Strips are for use with the TysonBio AC500 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC500 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. {2}------------------------------------------------ ### TysonBio AC800 Blood Glucose Monitoring System : The TysonBio AC800 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC800 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio AC800 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio AC800 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC800 contains a voice function but this function is not intended for users with impaired vision. The TysonBio AC800 Test Strips are for use with the TysonBio AC800 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC800 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. #### TysonBio AC500 Pro Blood Glucose Monitoring System : The TysonBio AC500 Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, or forearm. The TysonBio AC500 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with single-use lancing devices. The TysonBio AC500 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC500 Pro Blood Glucose Test Strips are for use with the TysonBio AC500 Pro Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC500 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. #### TysonBio AC800 Pro Blood Glucose Monitoring System : The TysonBio AC800 Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, or forearm. The TysonBio AC800 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the {3}------------------------------------------------ effectiveness of diabetes control program. This system should only be used with single-use lancing devices. The TysonBio AC800 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC800 Pro contains a voice function but this function is not intended for users with impaired vision. The TysonBio AC800 Pro Blood Glucose Test Strips are for use with the TysonBio AC800 Pro Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC800 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. ## 6. Test principle The test principle is based on electrochemical biosensor technology using glucose oxidase. There has been no change to the fundamental scientific technology. ## 7. Predicate device TysonBio AC100 Blood Glucose Monitoring System (K101543) | | Proposed Device | | Predicate Device | |------------------|-----------------------------|----------------------------------------|-----------------------------| | Item | TysonBio AC500 | TysonBio AC800 | TysonBio AC100 | | Meter appearance | Image: TysonBio AC500 meter | Image: TysonBio AC800 meter | Image: TysonBio AC100 meter | | Backlight | NA | Selectable backlight<br>on/off by user | NA | | Voice guidance | NA | Selectable on/off<br>voice guidance | IDENTICAL | | Hematocrit range | 20~60% | 20~60% | 35-55% | | Meal flag | Apple | Apple | Hamburger | #### Difference: {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows three digital displays, likely from a medical device such as a glucose meter. The first two displays show similar readings with the time as 88-88 and a value of 88.8 mg/dL. The third display shows a similar time, but the value is 888 mg/dL. . ## Similarities: . | | Proposed Device | | Predicate Device | |-----------------------------------------|----------------------------------------------------------------------|----------------|------------------| | Item | TysonBio AC500 | TysonBio AC800 | TysonBio AC100 | | Test principle | Electrochemical<br>biosensor with<br>glucose oxidase. | IDENTICAL | IDENTICAL | | Blood sample | Fingertip, palm and<br>forearm | IDENTICAL | IDENTICAL | | Sample volume | 0.5uL | IDENTICAL | IDENTICAL | | Test time | 5 seconds | IDENTICAL | IDENTICAL | | Detecting range | 20 ~ 600 mg/dL | IDENTICAL | IDENTICAL | | Unit of measurement | mg/dL or mmol/L | IDENTICAL | IDENTICAL | | Coding | Auto coding by<br>inserting the test strip | IDENTICAL | IDENTICAL | | Operating temperature | 10 to 40 °C (50-104<br>°F) | IDENTICAL | IDENTICAL | | Operating humidity | 10 to 90% | IDENTICAL | IDENTICAL | | Average result | 7,14, 30 days pre<br>meal and post meal<br>test average | IDENTICAL | IDENTICAL | | Hypoglycemic and<br>hyperglycemic alarm | 2 user setting alarms | IDENTICAL | IDENTICAL | | Reminder alarm | 4 user setting alarms | IDENTICAL | IDENTICAL | | PC link Protocol | USB protocol | IDENTICAL | IDENTICAL | | Control solution | 3 Levels available | IDENTICAL | IDENTICAL | | Strip storage<br>temperature | 4 to 30 °C (40-86°F) | IDENTICAL | IDENTICAL | | Memory capacity | 500 test result<br>memories | IDENTICAL | IDENTICAL | | Supply power | Two AAA batteries | IDENTICAL | IDENTICAL | | Button | Three operating<br>button (M, up and<br>down)<br>One ejection button | IDENTICAL | IDENTICAL | {5}------------------------------------------------ The hematocrit range of proposed devices is 20%~60% by sensing voltage and algorithm change from predicted device. ## 8. Performance characteristic summary The performance of TysonBio AC500/AC800/AC500 Pro/AC800 Pro Blood Glucose Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the performance of this system meets its intended use. ### 9. Conclusion The TysonBio AC500/AC800/AC500 Pro/AC800 Pro Blood Glucose Monitoring System is substantially equivalent to the predicate device. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three tail feathers, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. The text is in all caps and is relatively small compared to the eagle symbol. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 18, 2013 Tyson Bioresearch. Inc. C/O Wen-Hai Tsai, Ph.D. 5F., #22, Ke E.Road III., Science-Based Industrial Park CHUN-NAN, MIAO-LI COUNTY, TAIWAN, R.O.C. 350 Re: K123087 Trade/Device Name: TysonBio AC500 Blood Glucose Monitoring System TysonBio AC800 Blood Glucose Monitoring System Tyson Bio AC500 Pro Blood Glucose Monitoring System TysonBio AC800 Pro Blood Glucose monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: May 18, 2013 Received: May 23, 2013 #### Dear Dr. Tsai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general.controls.provisions.of.the.Act.include.requirements.for.annual.registration..listing.off devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {7}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): K123087 Device Name: TysonBio AC500 Blood Glucose Monitoring System Indications for Use: The TysonBio AC500 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio AC500 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio AC500 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC500 Test Strips are for use with the TysonBio AC500 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC500 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. AND/OR Over-The-Counter Use X Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) ' (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OIR) # Katherine /Serrano Division Sign-Off Office of In Vitro Diagnostics and Radiologic Health 510(k) K123087 {9}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): K123087 Device Name: TysonBio AC800 Blood Glucose Monitoring System Indications for Use: The TysonBio AC800 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC800 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio AC800 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio AC800 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC800 contains a voice function but this function is not intended for users with impaired vision. The TysonBio AC800 Test Strips are for use with the TysonBio AC800 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC800 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. | Prescription Use | AND/OR Over-The-Counter Use X | |-----------------------------|-------------------------------| | (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OlR) # Katherine //Serrano Division Sign-Off Office of In Vitro Diagnostics and Radiologic Health Page 2 of 4 {10}------------------------------------------------ ## 510(k) K123087 # Indications for Use Form 510(k) Number (if known): K123087 Device Name: TysonBio AC500 Pro Blood Glucose Monitoring System Indications for Use: 1 The TysonBio AC500 Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, or forearm. The TysonBio AC500 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. The TysonBio AC500 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC500 Pro Blood Glucose Test Strips are for use with the TysonBio AC500 Pro Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm, The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC500 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. Prescription Use X (Part 21 CFR 801 Subpart D) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ AND/OR Over-The-Counter Use X_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OIR) # Katherine /Serrano Division Sign-Off Office of In Vitro Diagnostics and Radiologic Health Page 3 of 4 {11}------------------------------------------------ 510(k) K123087 # Indications for Use Form 510(k) Number (if known): K123087 Device Name: TysonBio AC800 Pro Blood Glucose Monitoring System Indications for Use: The TysonBio AC800 Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, or forearm. The TysonBio AC800 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. The TysonBio AC800 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio AC800 Pro contains a voice function but this function is not intended for users with impaired vision. The TysonBio AC800 Pro Blood Glucose Test Strips are for use with the TysonBio AC800 Pro Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio AC500/AC800 Control Solutions is for use on the TysonBio AC800 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results. AND/OR Over-The-Counter Use X Prescription Use X (21 CFR 801 Subpart C) . (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OIR) # Katherine Serrano Division Sign-Off Office of In Vitro Diagnostics and Radiologic Health 510(k) K123087
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...