← Product Code [HRS](/productcode/HRS) · K123000

# INTEGRA TOTAL FOOT SYSTEMS (K123000)

_Integra LifeSciences Corporation · HRS · Dec 21, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K123000

## Device Facts

- **Applicant:** Integra LifeSciences Corporation
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Dec 21, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

INTEGRA Total Foot System is indicated for skeletally mature patients for the following: - Stabilization and fixation of fresh fractures - Intra and extra articular fractures, joint depression, and multi-fragmentary fractures - Revision procedures, joint fusion, and reconstruction of small bones of the feet.

## Device Story

System consists of variety of bone plates and screws for fixation of fractures, osteotomies, and fusions in adult foot. Plates cover forefoot, midfoot, and rearfoot applications; designs include MPJ fusion, Opening Wedge, Calcaneus, Dwyer, Interpositioning, Lapidus, Reconstruction, 2-10 Hole Tarsalis, fibula, and fibula tubular plates. Screws include locking and non-locking designs. Used by surgeons in clinical settings for orthopedic reconstruction and fracture management. Provides mechanical stabilization of bone segments to facilitate healing.

## Clinical Evidence

Bench testing only. Verification activities demonstrated that new or redesigned components are substantially equivalent to predicate devices.

## Technological Characteristics

Materials: ASTM F136 Ti-6Al-4V ELI alloy. Components: Bone plates (various configurations) and locking/non-locking screws. Mechanical fixation principle.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Ascension Total Plate System ([K100502](/device/K100502.md))
- Osetomed Foot Plating System ([K091614](/device/K091614.md))
- TC Plating System ([K094037](/device/K094037.md))
- VariAx Locked Plating System ([K102282](/device/K102282.md))
- Synthes (USA) Modular Mini Fragment LCP System ([K063049](/device/K063049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123000(pg 1/2)

Integra LifeSciences Corporation Traditional 510(k) Premarket Notification INTEGRA® Total Foot System

# 510(K) SUMMARY

# INTEGRA® Total Foot System

### Submitter's name and address:

Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536 USA

## Contact person and telephone number:

Facsimile: 609.275.9445

Stephen Beier Senior Specialist, Regulatory Affairs Telephone: 609.936.5436

# Date Summary was prepared:

September 26, 2012

#### Name of the device:

Proprietary Name: INTEGRA Total Foot System Common Name: Plate, Fixation, Bone Product Code: HRS Subsequent Product Code: HWC (Screw, Fixation, Bone) Classification Panel: Orthopedic

#### Substantial Equivalence:

INTEGRA Total Foot System is substantially equivalent in function and intended use to the predicate device detailed in the following table.

| 510(k) Number | Product Code | Trade Name                                           | Manufacturer                                                                    |
|---------------|--------------|------------------------------------------------------|---------------------------------------------------------------------------------|
| K100502       | HRS; HWC     | Ascension Total Plate<br>System                      | Integra LifeSciences<br>Corporation, formally<br>Ascension Orthopedics,<br>Inc. |
| K091614       | HRS; HWC     | Osetomed Foot Plating<br>System                      | Osteomed, L.P.                                                                  |
| K094037       | KTT          | TC Plating System                                    | Orthopro, LLC                                                                   |
| K102282       | HRS          | VariAx Locked Plating<br>System                      | Howmedica Osteonics<br>Corp                                                     |
| K063049       | HRS          | Synthes (USA) Modular<br>Mini Fragment LCP<br>System | Synthes (USA)                                                                   |

#### Device Description:

The INTEGRA "Total Foot System consists of a variety of bone plates and screws intended to be used in fixating fractures, osteotomies, and fusions in the adult foot. The various plate designs are divided into forefoot, midfoot, and rearfoot applications. Plate designs include MPJ fusion, Opening Wedge, Calcaneus, Dwyer, Interpositioning, Lapidus, Reconstruction, 2-10 Hole Tarsalis,

DEC 2 1 2012

{1}------------------------------------------------

# Integra LifeSciences Corporation Traditional 510(k) Premarket Notification INTEGRA® Total Foot System

fibula, and fibula tubular plates. Screws include both locking and non-locking designs. All plates and screws are manufactured from ASTM F136 Ti-6Al-4V ELI alloy.

## Intended Use:

INTEGRA Total Foot System is indicated for skeletally mature patients for the following:

- Stabilization and fixation of fresh fractures
- Intra and extra articular fractures, joint depression, and multi-fragmentary fractures
- Revision procedures, joint fusion, and reconstruction of small bones of the feet.

## Substantial Equivalence Comparison:

Components of the INTEGRA Total Foot System are similar in design and materials to the predicate devices: (K100502, K091614, K094037, K102282, and K063049).

#### Testing and Test Results:

The INTEGRA® Total Foot System and Ascension `Total Foot System (K100502) are both systems comprised of titanium plates and screws with identical indications for use. Additional predicate devices have been included to justify the inclusion of new components to the system.

All new or redesigned components to this system were examined, and results of the verification activities demonstrated that the tests were acceptable and the components substantially equivalent to the predicate devices identified.

#### Conclusion:

The INTEGRA Total Foot System is substantially equivalent to the commercially marketed devices, Ascension Total Foot System (K100502). Additional predicate devices to this system include Osteomed Foot Plating System (K091614), OrthoPro TC Plating System (K094037), Stryker VariAx Locked Plating System (K102282), and the Synthes Modular Mini Fragment LCP System (K063049).

The designs and design modifications expressed in this 510(k) Premarket Notification do not change the intended use or fundamental scientific technology of the predicate devices, nor do they raise any new issues of safety or effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Integra LifeSciences Corporation % Mr. Stephen H. Beier Senior Specialist, Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536

December 21, 2012

Re: K123000

Trade/Device Name: INTEGRA® Total Foot System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 26, 2012

Received: September 27, 2012

Dear Mr. Beier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

# Page 2 - Mr. Stephen H. Beier

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Peter D. Rumm -S

Mark N. Melkerson Director . Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

# 510(k) Number (if known):

INTEGRA® Total Foot System Device Name:

Indications For Use: INTEGRA® Total Foot System is indicated for skeletally mature patients for the following:

- Stabilization and fixation of fresh fractures
- Intra and extra articular fractures, joint depression, and multi-fragmentary fractures
- Revision procedures, joint fusion, and reconstruction of small bones of the feet.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Krishna Asundi

Division of Orthopedic Devices

Page 1 of 1

viii

---

**Source:** [https://fda.innolitics.com/device/K123000](https://fda.innolitics.com/device/K123000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
