K122981 · Ray Co., Ltd. · OAS · Mar 12, 2013 · Radiology
Device Facts
Record ID
K122981
Device Name
RAYSCAN A-EXPERT 3D
Applicant
Ray Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Mar 12, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
RAYSCAN α - Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dentomaxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Image is obtained using the standard narrow beam technique. CAUTION: Federal (US) law restricts the sale of this device to, or on the order of, licensed professionals.
Device Story
RAYSCAN α-Expert 3D is an extraoral dental X-ray system for 3D CBCT, panoramic, and cephalometric imaging. Device uses a rotating C-arm with a high-voltage generator, X-ray tube, and flat-panel detector to capture projections of the dentomaxillofacial region. Data is reconstructed into 3D volumes or 2D images for viewing on external workstations. Operated by dentists, radiologists, or qualified healthcare professionals in clinical settings. System includes SMARTDent software for processing and archiving. Output assists clinicians in diagnosis and treatment planning for implantology, TMJ studies, and general dental radiography. Benefits include high-resolution visualization of hard tissues for improved clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated via electrical, mechanical, and environmental safety testing per IEC 60601-1, 60601-1-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, 60601-2-44, and EMC testing per IEC 60601-1-2. Non-clinical considerations addressed per FDA guidance for solid-state X-ray imaging devices. All test results were satisfactory.
Technological Characteristics
Extraoral X-ray system; high-frequency generator; flat-panel detector (100 μm pixel size for CT/Pano, 150 μm for Ceph); 0.5mm focal spot; 60-90kVp; 4-17mA; 360° rotation; DICOM 3.0 compatible; Giga-Ethernet connectivity; wall or floor mount; Class I with Type B applied parts per IEC 60601-1.
Indications for Use
Indicated for dental radiographic examination of teeth, jaw, and oral structures in patients requiring panoramic, implantology, TMJ, or cephalometric imaging. Intended for use by licensed professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
MAR 1 2 2013
K122981
page 1 of 5
510(k) Submission- RAYSCAN a-Expert 3D
# 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
Date:
| APPLICANT | RAY Co.,Ltd |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ADDRESS | #362-43 (218 Maeyoung St.) 3rd & 4th Floor,<br>Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Manufacturer | RAY Co.,Ltd<br>#362-43 (218 Maeyoung St.) 3rd & 4th Floor,<br>Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea<br>TEL : +82-31-605-1000<br>FAX : +82-2-6280-5534 |
| Contact Person | Yun Jung. HA<br>e-mail : yunjung.ha@raymedical.co.kr |
## Device Name
Trade/Proprietary Name : RAYSCAN α-Expert 3D
Common Name: Dental panoramic/tomography and cephalometric x-ray system
### Classification
x-ray, tomography, computed, dental (21 CFR 892.1750)
Class : Il
Product code : OAS
Panel : Radiology
#### Predicate device
Rotagraph EVO 3D(K111152)
{1}------------------------------------------------
K122981
Page 2 of 5
510(k) Submission- RAYSCAN a-Expert 3D
## Description
RAYSCAN a-Expert 3D is a 3D computed tomography for scanning hard tissues such as bones and teeth. By rotating the c-arm which includes the high voltage generator all-in-one x-ray tube and a detector on each end, CBCT image of whole dentomaxillofacial is attained by recombining data from the same level that are scanned from different angles.
Panoramic image scanning function for attaining images of the entire teeth and cephalometric scanning option for attaining the cephalic images are included.
The system includes processing, and archiving "SMARTDent "software(Optional)
#### Indication for use
RAYSCAN a-Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dento-maxillo-facial 3D images.
The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations.
2D Image is obtained using the standard narrow beam technique.
"CAUTION: Federal (US) law restricts the sale of this device to, or on the order of, licensed professionals."
| Parameter | RAYSCAN α-Expert 3D<br>RAY Co.,Ltd | Rotograph EVO 3D<br>K111152 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common<br>Name<br>Indications<br>for use | Dental panoramic/tomography and<br>cephalometric x-ray system<br>RAYSCAN α-Expert 3D, panoramic x-ray<br>imaging system with cephalostat, is an<br>extraoral source x-ray system, which is<br>intended for dental radiographic<br>examination of the teeth, jaw, and oral<br>structures, specifically for panoramic<br>examinations and implantology and for<br>TMJ studies and cephalometry, and it has<br>the capability, using the CBVT technique,<br>to generate dento-maxillo-facial 3D<br>images.<br>The device uses cone shaped x-ray beam<br>projected on to a flat panel detector, and<br>the examined volume image is<br>reconstructed to be viewed in 3D viewing<br>stations.2D Images are obtained using<br>the standard narrow beam technique.<br>"CAUTION: Federal (US) law<br>restricts the sale of this device to,<br>or on the order of, licensed<br>professionals." | Dental panoramic/tomography and<br>cephalometric x-ray system<br>Rotograph EVO 3D, panoramic x-ray<br>imaging system with cephalostat, is an<br>extraoral source x-ray system, which is<br>intended for dental radiographic<br>examination of the teeth, jaw, and oral<br>structures, specifically for panoramic<br>examinations and implantology and for<br>TMJ studies and cephalometry, and it<br>has the capability, using the CBVT<br>technique, to generate<br>dentomaxillofacial 3D images.<br>The device uses cone shaped x-ray<br>beam projected on to a flat panel<br>detector, and the examined volume<br>image is reconstructed to be viewed in<br>3D viewing stations. 2D Images are<br>obtained using the standard narrow<br>beam technique.<br>The device is to be operated and used<br>by dentists, radiologists and other |
| | | legally qualified health care<br>professionals. |
| | | |
## Statement of Substantial Equivalence
{2}------------------------------------------------
K122981
page 3 of 5
510(k) Submission- RAYSCAN α--Expert 3D
| Mode of<br>Operation | | Continuous operation with<br>intermittent, stated permissible<br>loading | Continuous operation with<br>intermittent load | |
|------------------------------|-------|--------------------------------------------------------------------------|------------------------------------------------|----------------------|
| | | CBCT Cone beam Computed | CBCT Cone beam Computed | |
| 3D technology | | Tomography | Tomography | |
| | | CBCT Computed tomography | CBCT Computed tomography | |
| Performance<br>Specification | | Panoramic | Panoramic | |
| | | Cephalometric(optional) | Cephalostic(optional) | |
| | | | Full Dentition | |
| | | Standard | | |
| | | Implant | Left Jaws | |
| | CT | Surgical guide | | |
| | | Endo<br>Sinus | Right Jaws | |
| | | TMJ | | |
| | | | Sinus<br>Adult Panoramic | |
| | | Normal | Child Panoramic | |
| | PANO | TMJ | Open-closed mouth lateral TMJ | |
| Exam | | Sinus | P-A Sinus (rotational) | |
| mode | | PA,AP | | |
| | | Lateral | Lateral Ceph | |
| | Ceph | SMV | A-P and P-A Ceph | |
| | | Carpus | Carpus (hand) | |
| | | Reverse Town's<br>Waters | | |
| | | | Half Panoramic adult | |
| | | | Half Panoramic child | |
| | Other | | Orthogonal Projection | |
| | | | Low Dose Panoramic | |
| | | Computed Tomography(CT) : | Frontal Dentition<br>Computed Tomography(CT) : | |
| Detector | | Flat panel X-ray sensor | Flat panel X-ray sensor | |
| Type | | Pano : Flat panel X-ray sensor | Pano : Flat panel X-ray sensor | |
| | | Ceph : Flat panel X-ray sensor | Ceph : CCD Sensor | |
| | | Ceph Apparatus | Ceph Apparatus | |
| | | Vertical Carriage | Vertical Carriage | |
| | | Rotator | Rotator | |
| | | X-RAY Generator | X-ray Generator | |
| | | X-ray tube | X-ray tube | |
| | | High Frequency Generator | High Frequency Generator | |
| Main | | Column | Column | |
| Components | | Touch monitor (panal) | Keyboard | |
| | | Detector(CT/Panoramic, | Detector(CT/Panoramic, | |
| | | Cephalometric) | Cephalometric) | |
| | | Chinrest<br>Head rest | Chinrest<br>Head rest | |
| | | Automatic Collimator | Collimator | |
| | | Exposure switch | Exposure switch | |
| | | Emergency stop switch | Emergency stop switch | |
| | | | CT exams | CT exams(Volumetric) |
| Automatic | | Panoramic exams | Panoramic exams | |
| Collimator | | | Cephalometric exams | Cephalometric exams |
.
·
{3}------------------------------------------------
K122981
Page 4 of 5
510(k) Submission- RAYSCAN a-Expert 3D
.
| | | Bite stick |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | Bite block<br>Chinrest(Patient support of the<br>stand type) | Chinrest<br>(Patient support of the stand<br>type) |
| | Chinrest(TMJ, Sinus and etc.) | Head rest |
| | Head rest | X-ray push button with extensible<br>cable |
| Accessories | X-ray push button with<br>extensible cable | Disposable byte protective<br>sleeves |
| | Disposable byte protective<br>sleeves | 10ear centring pins for ceph |
| | Remote controller | 10 disposable head strips for 3D<br>exams |
| Display Type | TFT LCD type(Normally black)<br>*1280x800 pixel | Alphanumeric"OLED"Display of<br>2 lines of 20 characters |
| | Touch monitor (panel). | Alpanumeric "OLED" Display. |
| | Exposure button. | x-ray push button with extensible<br>cable. |
| | Operating PC screen. | Operating PC screen. |
| | Every operation is guided by<br>messages shown on the display. | Every operation is guided by<br>messages shown on the display. |
| | Selectable language : English | Selectable languages : Italian,<br>English, French, Spanish, German,<br>Turkish, Portuguese, Dutch. |
| | Operation S/W-<br>RayScan(Workstation) | Operation S/W(Workstation) |
| Rated power | 110-240 V~, 50 / 60 Hz, 2.5kVA | 110-120/220-240V~,50/60Hz,1.5 kVA |
| Class | Class I with type B applied parts<br>according to IEC 60601-1 | Class I with type B applied parts<br>according to IEC 60601-1 |
| Focal size | 0.5mm | 0.5mm |
| Field of<br>View(CT) | 90x90mm | 85x85mm |
| X-ray Voltage | 60~90kVp | 60~86kVp |
| X-ray Current | 4~17mA | 6~12 mA |
| Total Filtration | 2.6 mm Al equivalent | 2.5 mm Al equivalent |
| Detector Pixel<br>size | Computed Tomography(CT) : 100 μm<br>Pano : 100 μm<br>Ceph : 150 μm | Computed Tomography(CT) : 127 μm<br>Pano : 127 μm<br>Ceph : 48 μm |
| | | CT<br>Open/close mouth TMJ : 1.25 |
| Magnification | CT : 1.39<br>Pano : 1.31<br>Ceph : 1.13 | Sinus : 1.27<br>Pano : 1.28<br>Ceph : 1.10 in the mid-sagittal plane in<br>LL projection |
| Scan time | CT : 14sec<br>Pano : 14sec<br>Ceph : 0.3sec~3.0sec | CT : Max 20sec<br>Pano : Max 13.8sec<br>Ceph : 15sec |
| Grey level | CT : 14bit<br>Pano : 14bit<br>Ceph : 14bit | CT : 14bit<br>Pano : 14bit<br>Ceph : 12bit |
| Format<br>compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
| Image acquisition | Giga-Ethernet Network | Giga-Ethernet Network |
| Rotation angle | 360° | 200° |
| | | |
| PC technical<br>specifications | - OS : Windows 7, 32Bit<br>- CPU : Intel Dual core and over<br>- RAM : 4GB and over<br>- HDD : 250GB and over<br>- Network : Gigabit Ethernet<br>- Resolution : 1366x768 and over | - OS : Windows 7, 32Bit<br>- CPU: Intel® Core™ i5-750<br>- RAM: 4GB<br>- HDD: 1TB<br>- Network : Gigabit Ethernet<br>- Resolution : 1440x900 |
| Total Height | Max 2,296mm | Max 2,450mm |
| Weight | Computed Tomography(CT)+<br>Panoramic<br>: 148kg(326lb)<br>Computed Tomography(CT)+<br>Panoramic+Cephalostic(optional)<br>: 165kg(363lb) | Computed Tomography(CT)+<br>Panoramic<br>: 161kg(354lb)<br>Computed Tomography(CT)+<br>Panoramic<br>+Cephalostic(optional)<br>: 186kg(409lb) |
| Type of<br>installation | Wall or floor mount | Wall or floor mount |
| Patient position | Standing / Wheelchair | Standing / Wheelchair |
| Applicable<br>Standards | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-3<br>IEC 60601-2-7<br>IEC 60601-2-28<br>IEC 60601-2-32<br>IEC 60601-2-44<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-3<br>IEC 60601-2-7<br>IEC 60601-2-28<br>IEC 60601-2-32<br>IEC 60601-2-44<br>IEC 60601-1-2 |
| Certificate<br>Product | CE0120(MDD93/42/EEC) | CE0051(MDD93/42/EEC), FDA |
{4}------------------------------------------------
K122981)
Page 5 of 5
510(k) Submission- RAYSCAN a-Expert 3D
RAYSCAN a-Expert 3D has the same indication for use as the predicate devices. It shares the same technological characteristics as the predicate devices. Minor technological differences do not raise any new questions regarding safety or effectiveness of the device.
### Safety and Effectiveness Information
Electrical, mechanical, environmental safety and performance testing according to standards IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32 and IEC 60601-2-44 was performed, and EMC testing was conducted in accordance with the standard IEC 60601-1-2.
Non-clinical & Clinical considerations according to FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed.
All test results were satisfactory.
#### Conclusions
Based on a comparison of intended use, indications, constructions, construction materials, principal of Operation, features and technical data, the RAYSCAN α-Expert 3D system are safe and effective to perform its intended use as well as substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 12, 2013
Ray, Co., Ltd. % Mr. Andrew Paeng Consultant 8920 Wilshire Blvd., Suite 603 BEVERLY HILLS CA 90211
Re: K122981
Trade/Device Name: RAYSCAN a-Expert 3D Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: January 14, 2013 Received: February 06, 2013
Dear Mr. Paeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Mr. Paeng
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
\$\left(m\right)\$
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: RAYSCAN a-Expert 3D
Indications For Use:
RAYSCAN α - Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dentomaxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations.
2D Image is obtained using the standard narrow beam technique.
AND/OR Over-The-Counter Use Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
sinh.7)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510 (k) :
Page 1 of ﻢ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.