LUCENT

K122967 · Spinal Elements, Inc. · MAX · Feb 21, 2013 · Orthopedic

Device Facts

Record IDK122967
Device NameLUCENT
ApplicantSpinal Elements, Inc.
Product CodeMAX · Orthopedic
Decision DateFeb 21, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Lucent® intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device.

Device Story

Lucent® is an intervertebral body fusion device (interbody cage) used in lumbar spinal surgery. Box-shaped with internal voids for autograft placement; superior/inferior surfaces feature teeth/sharp engagement members to prevent migration. Manufactured from titanium (Ti-6Al-4V) or PEEK, with optional plasma-sprayed titanium coating. Implanted by surgeons during spinal fusion procedures; requires supplemental fixation (e.g., pedicle screws/rods). Provides structural support to intervertebral space to facilitate fusion; benefits patients by stabilizing spinal segments and reducing pain associated with DDD.

Clinical Evidence

Bench testing only. Static torsion testing performed per ASTM F 2077 to verify performance of new device sizes.

Technological Characteristics

Box-shaped interbody cage. Materials: Titanium (Ti-6Al-4V) per ASTM F 136/ISO 5832-3, PEEK per ASTM F 2026, or PEEK with plasma-sprayed titanium coating per ASTM F 1580. Features surface teeth for fixation. Mechanical stabilization principle.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with up to Grade 1 spondylolisthesis or retrolisthesis and those with prior non-fusion spinal surgery. Requires 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 2 # 510(k) Summary Lucent® # 510(k) Number K122967 # FEB 2 1 2013 | Manufacturer Identification<br>Submitted by: | Spinal Elements, Inc.<br>2744 Loker Ave. W., Suite 100<br>Carlsbad, CA 92010<br>760-607-0121 | |--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Information: | Benjamin A. Kimball<br>Regulatory Affairs Manager<br>Spinal Elements, Inc.<br>2744 Loker Ave. W., Suite 100<br>Carlsbad, CA 92010<br>760-607-1816<br>Bkimball@spinalelements.com | | Date Prepared: | November 9, 2012 | | Proprietary Name<br>Common Name<br>Device Classification<br>Proposed Regulatory Class<br>Device Product Code | Lucent®<br>Intervertebral Body Fusion Device.<br>21 CFR 888.3080 (Intervertebral Body Fusion Device)<br>Class II<br>MAX | ### Purpose of this Special 510(k) This Special 510(k) seeks clearance for a modification (larger sizes - line addition) to Lucent® intervertebral body fusion devices previously cleared under K071724, K073348, and K110632. # Device Description This product is an intervertebral body fusion device for use in lumbar spinal surgery. It may also be referred to as an interbody device or interbody cage. The device is generally box-shaped with various holes throughout its design to allow for the placement of autograft. The exterior of the device has "teeth" or other generally sharp engagement members on the superior and inferior surfaces to help prevent the device from migrating once it is surgically positioned. The devices may be manufactured from titanium (Ti-6Al-4V) conforming to ASTM F 136 or ISO 5832-3, polyetheretherketone materials conforming to ASTM F 2026 or polyetheretherketone with a plasma sprayed coating of commercially pure titanium (per ASTM F 1580) on their superior and inferior surfaces. {1}------------------------------------------------ # Intended Use of the Device Lucent® intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device. #### Performance Data The following testing was performed: Static torsion, ASTM F 2077 #### Substantial Equivalence The subject Lucent intervertebral body fusion device was shown to be substantially equivalent to previously cleared devices in indications for use, design, function, and materials used. > Lucent 510(k) K071724, K073348 and K110632 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 February 21, 2013 Spinal Elements, Incorporated % Mr. Benjamin A. Kimball Regulatory Affairs Manager 2744 Loker Avenue West. Suite 100 Carlsbad, California 92010 Re: K122967 Trade/Device Name: Lucent® Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: September 24, 2012 Received: September 25, 2012 Dear Mr. Kimball: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications . for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act . or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - Mr. Benjamin A. Kimball device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Erin Jakeith Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use K122967 Page 1 of 1 510(k) K: Device Name: Lucent® #### Indications for Use: Lucent intervertebral body fusion devices are intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). This device is intended to be used with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months non-operative treatment prior to being treated with this device. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINÙE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Stephanie Bechtold -S Page 1 of 1 (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K122967
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...