K122875 · Oricare, Inc. · FSY · Dec 18, 2012 · General, Plastic Surgery
Device Facts
Record ID
K122875
Device Name
ORICARE L2700
Applicant
Oricare, Inc.
Product Code
FSY · General, Plastic Surgery
Decision Date
Dec 18, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Oricare L2700 Series Surgical Lamps provide illumination at varying illumination levels to the patient surgical area and provide video-visual procedural support for medical staff during surgical procedures.
Device Story
Ceiling-mounted surgical lighting system; utilizes LEDs for illumination; powered by 115VAC. Features adjustable intensity (10 levels); optical lens design provides extended light depth without manual focus. Configurable with single or dual light heads for major/ancillary illumination; provides low-level background lighting for endoscopy. Optional integrated SD or HD video camera systems capture surgical procedures for training, live transmission, or remote consultation; outputs video to TFT monitors. Used in operating rooms by surgeons and medical staff. Benefits include shadowless illumination, improved visualization of surgical field, and procedural documentation/consultation capabilities.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60601-2-41:2009, ISO 14971:2009, and IEC 60601-1-6 confirmed device safety and essential performance.
Technological Characteristics
Ceiling-mounted LED surgical luminaire; 115VAC power; 10-level intensity adjustment; optical lens focusing; optional integrated SD/HD video system. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-41, ISO 14971. Software classified as Minor Level of Concern.
Indications for Use
Indicated for providing surgical field illumination and video-visual procedural support for medical staff during surgical and non-surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
Maquet PowerLED Surgical Lighting System (K070442)
Steris Harmony LED-1 Surgical Lighting System (K072072)
Submission Summary (Full Text)
{1}------------------------------------------------
| ORICARE | DEC 18 2012 | 1900 AM Drive<br>Quakertown, PA. 18951<br>Phone: 215-538-2470<br>Fax: +1-215-538-2578<br>Website: www.oricaremed.com |
|--------------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| 510(k) Summary | | |
| Date Prepared: | September 11, 2012 | K122875 |
| Applicant: | Oricare, Inc.<br>1900 AM Drive<br>Quakertown, Pa 18951<br>Registration Number: 3009129579 | |
| Contact Person: | Fred Cowdery<br>QA/RA Manager<br>Ph 267-347-4544<br>Fax 215-538-2578 | |
| Device Trade/Proprietary Name: | L2700 Series OR Surgical Luminaires | |
| Device Name: Common / Usual: | Light, Surgical, Ceiling Mounted | |
| Classification Name: | Surgical Lamp (21 CFR 878.4580, Product Code: FSY) | |
| Predicate Devices: | Maquet PowerLED Surgical Lighting System (K070442)<br>Steris Harmony LED-1 Surgical Lighting System (K072072) | |
# Device Description / Technological Characteristics
ਦੇ ਪੈ
OL 2700 series Surgical Lamps are ceiling mounted Surgical Lamps, suitable for providing supplemental illumination, with reduced shadowing effects during surgical and non-surgical procedures.
The lighting system utilizes LED's for illumination and is powered from a standard 115VAC voltage source. The head design is comprised of multiple LED's (Light Emitting Diodes). It provides a cool, color corrected light which is adjustable with 10 levels of intensity. The light path design provides an extended light depth so that the beam requires no manual focus. Light generated by the LEDs is focused by optical lens to achieve an ideal irradiance pattern.
The OL 2700 series LED surgical light can be configured with one or two light heads, the primary light head is useful for major illumination while the secondary light head is for ancillary illumination. The lights can also provide low level background lighting for endoscopic surgery.
Oricare L2700 series Operating Room Lamps can be easily used in conjunction with the optional SD or HD video systems, to document surgical procedures for training purposes, make use of live transmissions or consult with external surgeons during operations. The camera system is available in different versions and can be used with integrated OR systems. It delivers images in SD or HD quality straight to the TFT monitor.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "ORICARE" in bold, black letters. The font is sans-serif and slightly slanted to the right. The letters are tightly spaced, creating a solid block of text.
. (
Table 1 below defines the available configurations of the L2700 Series Surgical Lamps.
| Table 1 | |
|------------------------|---------------------------------------------|
| Oricare Part<br>Number | Model Number / Description |
| LA5003 | L2770 Operating Room Lamp |
| LA5004 | L2750 Operating Room Lamp |
| LA5005 | L2770L2750 Operating Room Lamp |
| LA5006 | L2770L2770 Operating Room Lamp |
| LA5007 | L2750L2750 Operating Room Lamp |
| LA5012 | L27702750 OR Lamp with SD Video and Display |
| LA5013 | L27702750 OR Lamp with HD Video and Display |
Table 1
## Intended Use:
The Oricare L2700 Series Surgical Lamps provide illumination at varying illumination levels to the patient surgical area and provide video-visual procedural support for medical staff during surgical procedures.
# Performance Testing:
Comprehensive performance testing has been conducted on the L2700 Series Surgical Luminaires in accordance with various recognized industry standards by a recognized third party organization. Product compliance with the following standards included: IEC 60601-1:2005 30 edition, IEC 60601-1-2:2007, IEC 60601-2-41:2009, ISO 14971:2009, IEC 60601-1-6.
Software was evaluated in accordance with relevant FDA Software Guidance Document, per the requirements assigned for software classified as a "Minor Level of Concern".
The combined testing and analysis of results provides assurance that the device meets the requirements for safety and essential performance and is effective for its intended use.
## Non-clinical Comparisons to Predicate Device:
The differences between the proposed and predicate devices are limited to differences in design, material, and operational. These differences do not raise any new issues of safety and efficacy.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "ORICARE" in a bold, sans-serif font. The letters are all capitalized and appear to be black. The font has a slightly distressed or textured appearance, giving it a somewhat rough or worn look.
1900 AM Drive Quakertown, PA. 18951 Phone: 215-538-2470 Fax: +1-215-538-2578 Website: www.oricaremed.com
## Substantial Equivalence:
The Oricare L2700 Surgical Luminaires are believed to be substantially equivalent to currently marketed Surgical Lamps with regards to intended use, safety and effectiveness, technology, and performance.
Similar to the predicate devices, the L2700 Surgical Luminaires provide a shadowless light source using LEDs as the main light source. The L2700 Surgical Luminaires provide a variable intensity light source designed to provide visible illumination of the surgical field and the patient and to provide video-visual procedural support for the hospital staff during surgical procedures.
# Conclusions:
Based upon the information provided herein this 510(k) Premarket Notification, we conclude that the L2700 Series Surgical Luminaires are substantially equivalent to the identified predicate devices and are safe and effective when used as intended.
RLC
Fred Cowdery Manager, Regulatory Affairs and Quality Assurance Oricare, Inc. 1900 AM Drive Quakertown, PA. 18951
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 18, 2012
Oricare, Inc. % Mr. Fred Cowdery QA/RA Manager 1900 AM Drive Quakertown, Pennsylvania 18951
Re: K122875
Trade/Device Name: L2700 Series OR Surgical Luminaires Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: September 12, 2012 Received: September 19, 2012
Dear Mr. Cowdery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Fred Cowdery
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# ORICARE
1900 AM Drive Quakertown, PA. 18951 Phone: 267-347-4544 Fax: +1-215-538-2578 Website: www.oricaremed.com
# Indications for Use Statement
510 (k) Number (if known)
Device Name: Oricare L2700 Series Surgical Lamp
Indications for Use:
The Oricare L2700 Series Surgical Lamps provide illumination at varying illumination levels to the patient surgical area and provide video-visual procedural support for medical staff during surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden 2012.12.18 08:48:17 -05'00' 11. 11.
(Division Sign-Off)
Division of Surgical Devices
510(k) Number __ K122875
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.