← Product Code [IYN](/productcode/IYN) · K122710

# TRANSCRANIAL DOPPLER WITH ROBOTIC PROBE HEADBAND (K122710)

_Shenzhen Delicate Electronics Co., Ltd. · IYN · Feb 15, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K122710

## Device Facts

- **Applicant:** Shenzhen Delicate Electronics Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Feb 15, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system: - For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values; - 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye. The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during: - Diagnostic exams; 1) - 2) Surgical interventions. The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

## Device Story

Transcranial Doppler (TCD) system with robotic probe headband; inputs ultrasound Doppler signals from cerebral/peripheral arteries; transforms signals to measure blood flow velocities and detect micro-embolic signals; robotic headband automatically adjusts probe angle based on signal intensity for continuous monitoring; output displayed to clinician for evaluation of neurological vascular diseases (e.g., vasospasm, intracranial stenosis) and intraoperative monitoring of flow changes/embolic events; used in clinical/surgical settings; aids clinical decision-making by identifying potential stroke risks/obstructions; benefits patient through non-invasive monitoring of cerebral blood supply.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and acoustic output measurement (NEMA UD 2-2004).

## Technological Characteristics

Diagnostic ultrasound fluid flow analysis system; includes robotic probe headband for automated signal tracking; supports multiple transducer models; energy source: electrical; connectivity: not specified; sterilization: not specified; software: embedded firmware for signal processing and robotic control.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- EMS9UA Transcranial Doppler with Robotic Probe Headband ([K092164](/device/K092164.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Additional Information I for K122710 - Attachment II 510(k) Summary

# Attachment II 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: __

1. Date of Submission: October 29, 2012

2. Sponsor Identification

> Shenzhen Delicate Electronics Co., Ltd. 6C, Block 8, Tian-an Ind. Area, Nanshan District Shenzhen, Guangdong, 518054, China

Establishment Registration Number: 3006441164

Contact Person: Min Li Position: General Manager Tel: +86-755-26413482 Fax: +86-755-26425970 Email: minnli@delica-sz.com

Submission Correspondent 3.

> Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023 Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{1}------------------------------------------------

- 4. Type and Reason of Submission
Special 510(k) Submission for the design change to the existed device, which is EMS9UA Transcranial Doppler with Robotic Probe Headband as cleared in K092164.

- Proposed Device Identification 5.
Proposed Device Name: Transcranial Doppler with Robotic Probe Headband Proposed Device Model: EMS9UA, EMS 9PB

Classification: Class II Product Codes: IYN, ITX and OQQ Regulation Number: 21 CFR 892.1570 Review Panel: Radiology

#### Intended Use Statement:

Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:

- For the measurement of cerebral artery blood velocities to determine the presence of hemo 1) dynamically significant deviations from normal values;
- 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.

The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during:

- Diagnostic exams; 1)
- 2) Surgical interventions.

The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

- Predicate Device Identification 6.
510(k) Number: K092164

Product Name: EMS9UA Transcranial Doppler with Robotic Probe Headband Manufacturer: Shenzhen Delicate Electronics Co., Ltd.

- Device Description 7.
The proposed device, Transcranial Doppler (TCD) with Robotic Probe Headband.is intended for obtaining the information of blood flow velocities throughout the body by using non-invasive technique. This method of measurement is particularly useful for examining the major arteries supplying blood to the brain.

{2}------------------------------------------------

. .

Robotic Probe is developed by Delicate for monitoring brain blood application. It was previously cleared in K092164 at May 10, 2010. The system can adjust the angle of probe automatically according to signal intensity when monitoring.

TCD is used to evaluation of numerous neurological vascular diseases such as vasospasm and intracranial stenosis. It is also used for intraoperative monitoring to help detect sudden changes in flow and potential embolic events. Emboli are small particles of foreign matter (air, particulate, thrombin, etc.) within the bloodstream that can potentially cause obstructions in various arteries in the body and the brain. Such obstructions can often lead to stroke.

- 8. Non-Clinical Test Conclusion
Bench tests were conducted to verify that all the risks identified associated with the design change were acceptable. The tests include:

- a) IEC 60601-1:2005 Ed3.0, Medical Electrical Equipment - Part 1: General Requirements for Safety:
- b) IEC 60601-1-2:2007 Ed3.0, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests;
- NEMA UD 2-2004 (R2009), Acoustic Output MeasurementStandard for Diagnostic c) Ultrasound Equipment Revision 3.
- 9. Substantially Equivalent Conclusion

The proposed device, Transcranial Doppler with Robotic Probe Headband, EMS 9UA and EMS9PB,are determined to be Substantially Equivalent (SE) to the predicate device (existed device), EMS9UA Transcranial Doppler with Robotic Probe Headband as cleared in K092164, in respect of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

# February 15, 2013

Shenzhen Delicate Electronics Co., Ltd. c/o Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 237-023. Shanghai. 200237 CHINA

### Re: K122710

Trade/Device Name: Transcranial Doppler with Robotic Probe Headband; Models: EMS9UA and EMS9PB Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN. ITX. and OOO Dated: January 22, 2013 Received: January 25, 2013

# Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Transcranial Doppler with Robotic Probe Headband; Models: EMS9UA and EMS9PB, as described in your premarket notification:

### Transducer Model Number

02.0001.0214.01 02.0001.0213.01 02.0001.0805.01 AP99-0815-PW1.60

AP99-0607-PW2.0 02.0001.0408.01 02.0128.1616.01 02.0001.1613.02

{4}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely yours.

FDA

for

Janine Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

# Exhibit #2 Indications for Use

510(k) Number: K122710

Device Name: Transcranial Doppler with Robotic Probe Headband Model: EMS9UA

Indications for Use:

Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:

- 1) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values;
- To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for 2) observation include, but are not limited to the middle, anterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.

The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during:

- 1) Diagnostic exams;
- 2) Surgical interventions.

The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

ZPRESCRIPTION USE . (Part 21 CFR 801 Subpart D) OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Image /page/5/Picture/18 description: The image shows a black and white graphic that appears to be a stylized signature or logo. The design incorporates a mix of cursive and geometric shapes, with some areas filled in with solid black and others outlined. The overall impression is abstract and somewhat illegible, suggesting a unique and artistic representation of a name or brand.

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

Page E2-1/12

{6}------------------------------------------------

SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband

# Transducer: 02.0001.0214.01

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application           |                                 | Mode of Operation |   |   |     |     |                  |                   |                    |
|--------------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General                        | Specific                        | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only)                 | (Track 1 and Track 3)           |                   |   |   |     |     |                  |                   |                    |
| Ophthalimic                    | Ophthalmic                      |                   |   |   | P   |     |                  |                   |                    |
|                                | Fetal                           |                   |   |   |     |     |                  |                   |                    |
|                                | Abdominal                       |                   |   |   |     |     |                  |                   |                    |
|                                | Intraoperative (specify)        |                   |   |   |     |     |                  |                   |                    |
|                                | Intraoperative (Neuro)          |                   |   |   |     |     |                  |                   |                    |
|                                | Laparoscopic                    |                   |   |   |     |     |                  |                   |                    |
|                                | Pediatric                       |                   |   |   |     |     |                  |                   |                    |
|                                | Small Organ (specify)           |                   |   |   |     |     |                  |                   |                    |
| Fetal<br>Imaging<br>&<br>Other | Neonatal Cephalic               |                   |   |   |     |     |                  |                   |                    |
|                                | Adult Cephalic                  |                   |   |   | P   |     |                  |                   |                    |
|                                | Trans-rectal                    |                   |   |   |     |     |                  |                   |                    |
|                                | Trans-vaginal                   |                   |   |   |     |     |                  |                   |                    |
|                                | Trans-urethral                  |                   |   |   |     |     |                  |                   |                    |
|                                | Trans-esoph. (non-Card.)        |                   |   |   |     |     |                  |                   |                    |
|                                | Musculo-skeletal (Conventional) |                   |   |   |     |     |                  |                   |                    |
|                                | Musculo-skeletal (Superficial)  |                   |   |   |     |     |                  |                   |                    |
|                                | Intravascular                   |                   |   |   |     |     |                  |                   |                    |
|                                | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
|                                | Cardiac Adult                   |                   |   |   |     |     |                  |                   |                    |
|                                | Cardiac Pediatric               |                   |   |   |     |     |                  |                   |                    |
|                                | Intravascular (Cardiac)         |                   |   |   |     |     |                  |                   |                    |
| Cardiac                        | Trans-esoph. (Cardiac)          |                   |   |   |     |     |                  |                   |                    |
|                                | Intra-Cardiac                   |                   |   |   |     |     |                  |                   |                    |
|                                | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
| Peripheral                     | Peripheral vessel               |                   |   |   | P   |     |                  |                   |                    |
| Vessel                         | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

{7}------------------------------------------------

SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband

# Transducer: AP99-0607-PW2.0

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                   |   | Mode of Operation |   |     |     |                  |                   |                    |
|---------------------------|-----------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic               | Ophthalmic                        |   |                   |   |     |     |                  |                   |                    |
|                           | Fetal                             |   |                   |   |     |     |                  |                   |                    |
|                           | Abdominal                         |   |                   |   |     |     |                  |                   |                    |
|                           | Intraoperative (specify)          |   |                   |   |     |     |                  |                   |                    |
|                           | Intraoperative (Neuro)            |   |                   |   |     |     |                  |                   |                    |
|                           | Laparoscopic                      |   |                   |   |     |     |                  |                   |                    |
|                           | Pediatric                         |   |                   |   |     |     |                  |                   |                    |
|                           | Small Organ (specify)             |   |                   |   |     |     |                  |                   |                    |
|                           | Neonatal Cephalic                 |   |                   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other  | Adult Cephalic                    |   |                   |   | N   |     |                  |                   |                    |
|                           | Trans-rectal                      |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-vaginal                     |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-urethral                    |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)          |   |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Conventional)   |   |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Superficial)    |   |                   |   |     |     |                  |                   |                    |
|                           | Intravascular                     |   |                   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |
|                           | Cardiac Adult                     |   |                   |   |     |     |                  |                   |                    |
|                           | Cardiac Pediatric                 |   |                   |   |     |     |                  |                   |                    |
| Cardiac                   | Intravascular (Cardiac)           |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (Cardiac)            |   |                   |   |     |     |                  |                   |                    |
|                           | Intra-Cardiac                     |   |                   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |
| Peripheral<br>Vessel      | Peripheral vessel                 |   |                   |   | N   |     |                  |                   |                    |
|                           | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |

N= new indication; P=previously cleared by FDA; E= added under Appendix E

{8}------------------------------------------------

# K122710

# Diagnostic Ultrasound Indications for Use Form

SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband

### Transducer: 02.0001.0213.01

.

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                   |   | Mode of Operation |   |     |     |                  |                   |                    |
|---------------------------|-----------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic               | Ophthalmic                        |   |                   |   |     |     |                  |                   |                    |
|                           | Fetal                             |   |                   |   |     |     |                  |                   |                    |
|                           | Abdominal                         |   |                   |   |     |     |                  |                   |                    |
|                           | Intraoperative (specify)          |   |                   |   |     |     |                  |                   |                    |
|                           | Intraoperative (Neuro)            |   |                   |   |     |     |                  |                   |                    |
|                           | Laparoscopic                      |   |                   |   |     |     |                  |                   |                    |
|                           | Pediatric                         |   |                   |   |     |     |                  |                   |                    |
|                           | Small Organ (specify)             |   |                   |   |     |     |                  |                   |                    |
|                           | Neonatal Cephalic                 |   |                   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other  | Adult Cephalic                    |   |                   |   | P   |     |                  |                   |                    |
|                           | Trans-rectal                      |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-vaginal                     |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-urethral                    |   |                   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)          |   |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Conventional)   |   |                   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Superficial)    |   |                   |   |     |     |                  |                   |                    |
|                           | Intravascular                     |   |                   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |
|                           | Cardiac Adult                     |   |                   |   |     |     |                  |                   |                    |
|                           | Cardiac Pediatric                 |   |                   |   |     |     |                  |                   |                    |
|                           | Intravascular (Cardiac)           |   |                   |   |     |     |                  |                   |                    |
| Cardiac                   | Trans-esoph. (Cardiac)            |   |                   |   |     |     |                  |                   |                    |
|                           | Intra-Cardiac                     |   |                   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |
| Peripheral                | Peripheral vessel                 |   |                   |   | P   |     |                  |                   |                    |
| Vessel                    | Other (specify)                   |   |                   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Comments:

.

{9}------------------------------------------------

### K122710

## Diagnostic Ultrasound Indications for Use Form

SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband

### Transducer: 02.0001.0408.01

イベント

:

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                                 | Mode of Operation |   |   |     |     |                  |                   |                    |
|-----------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General                     | Specific                        | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only)              | (Track 1 and Track 3)           |                   |   |   |     |     |                  |                   |                    |
| Ophthalimic                 | Ophthalmic                      |                   |   |   |     |     |                  |                   |                    |
|                             | Fetal                           |                   |   |   |     |     |                  |                   |                    |
|                             | Abdominal                       |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (specify)        |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (Neuro)          |                   |   |   |     |     |                  |                   |                    |
|                             | Laparoscopic                    |                   |   |   |     |     |                  |                   |                    |
|                             | Pediatric                       |                   |   |   |     |     |                  |                   |                    |
|                             | Small Organ (specify)           |                   |   |   |     |     |                  |                   |                    |
|                             | Neonatal Cephalic               |                   |   |   |     |     |                  |                   |                    |
| Fetal<br>Imaging &<br>Other | Adult Cephalic                  |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-rectal                    |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-vaginal                   |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-urethral                  |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-esoph. (non-Card.)        |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Conventional) |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Superficial)  |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular                   |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
|                             | Cardiac Adult                   |                   |   |   |     |     |                  |                   |                    |
|                             | Cardiac Pediatric               |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular (Cardiac)         |                   |   |   |     |     |                  |                   |                    |
| Cardiac                     | Trans-esoph. (Cardiac)          |                   |   |   |     |     |                  |                   |                    |
|                             | Intra-Cardiac                   |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
| Peripheral                  | Peripheral vessel               |                   |   |   | P   |     |                  |                   |                    |
| Vessel                      | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

{10}------------------------------------------------

SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband

Transducer: 02.0001.0805.01

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                                 | Mode of Operation |   |   |     |     |                  |                   |                    |
|-----------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General                     | Specific                        | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only)              | (Track 1 and Track 3)           |                   |   |   |     |     |                  |                   |                    |
| Ophthalimic                 | Ophthalmic                      |                   |   |   |     |     |                  |                   |                    |
|                             | Fetal                           |                   |   |   |     |     |                  |                   |                    |
|                             | Abdominal                       |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (specify)        |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (Neuro)          |                   |   |   |     |     |                  |                   |                    |
|                             | Laparoscopic                    |                   |   |   |     |     |                  |                   |                    |
|                             | Pediatric                       |                   |   |   |     |     |                  |                   |                    |
|                             | Small Organ (specify)           |                   |   |   |     |     |                  |                   |                    |
|                             | Neonatal Cephalic               |                   |   |   |     |     |                  |                   |                    |
| Fetal<br>Imaging &<br>Other | Adult Cephalic .                |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-rectal                    |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-vaginal                   |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-urethral                  |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-esoph. (non-Card.)        |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Conventional) |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Superficial)  |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular                   |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
|                             | Cardiac Adult                   |                   |   |   |     |     |                  |                   |                    |
| Cardiac                     | Cardiac Pediatric               |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular (Cardiac)         |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-esoph. (Cardiac)          |                   |   |   |     |     |                  |                   |                    |
|                             | Intra-Cardiac                   |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
| Peripheral                  | Peripheral vessel               |                   |   |   | P   |     |                  |                   |                    |
| Vessel                      | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

{11}------------------------------------------------

#### Indications for Use

510(k) Number: K122710

Device Name: Transcranial Doppler with Robotic Probe Headband Model: EMS9PB

Indications for Use:

Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:

- 3) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values;
- To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for 4) observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.

The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal.

Transcranial Doppler is intended for use during:

- 3) Diagnostic exams:
- 4) Surgical interventions.

The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

XPRESCRIPTION USE (Part 21 CFR 801 Subpart D) _OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{12}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband

# Transducer: 02.0128.1616.0】「

K122710

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application     |                                 | Mode of Operation |   |   |     |     |                  |                   |                    |
|--------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General                  | Specific                        | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only)           | (Track 1 and Track 3)           |                   |   |   |     |     |                  |                   |                    |
| Ophthalimic              | Ophthalmic                      |                   |   |   | N   |     |                  |                   |                    |
|                          | Fetal                           |                   |   |   |     |     |                  |                   |                    |
|                          | Abdominal                       |                   |   |   |     |     |                  |                   |                    |
|                          | Intraoperative (specify)        |                   |   |   |     |     |                  |                   |                    |
|                          | Intraoperative (Neuro)          |                   |   |   |     |     |                  |                   |                    |
|                          | Laparoscopic                    |                   |   |   |     |     |                  |                   |                    |
|                          | Pediatric                       |                   |   |   |     |     |                  |                   |                    |
|                          | Small Organ (specify)           |                   |   |   |     |     |                  |                   |                    |
|                          | Neonatal Cephalic               |                   |   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other | Adult Cephalic                  |                   |   |   | N   |     |                  |                   |                    |
|                          | Trans-rectal                    |                   |   |   |     |     |                  |                   |                    |
|                          | Trans-vaginal                   |                   |   |   |     |     |                  |                   |                    |
|                          | Trans-urethral                  |                   |   |   |     |     |                  |                   |                    |
|                          | Trans-esoph. (non-Card.)        |                   |   |   |     |     |                  |                   |                    |
|                          | Musculo-skeletal (Conventional) |                   |   |   |     |     |                  |                   |                    |
|                          | Musculo-skeletal (Superficial)  |                   |   |   |     |     |                  |                   |                    |
|                          | Intravascular                   |                   |   |   |     |     |                  |                   |                    |
|                          | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
|                          | Cardiac Adult                   |                   |   |   |     |     |                  |                   |                    |
|                          | Cardiac Pediatric               |                   |   |   |     |     |                  |                   |                    |
| Cardiac                  | Intravascular (Cardiac)         |                   |   |   |     |     |                  |                   |                    |
|                          | Trans-esoph. (Cardiac)          |                   |   |   |     |     |                  |                   |                    |
|                          | Intra-Cardiac                   |                   |   |   |     |     |                  |                   |                    |
|                          | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |
| Peripheral               | Peripheral vessel               |                   |   |   | N   |     |                  |                   |                    |
| Vessel                   | Other (specify)                 |                   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

{13}------------------------------------------------

### K122710

### Diagnostic Ultrasound Indications for Use Form

# SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband

# Transducer: AP99-0815-PW1.60

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                                 |   | Mode of Operation |   |     |     |                  |                   |                    |  |
|-----------------------------|---------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|--|
| General                     | Specific                        | A | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |  |
| (track 1 Only)              | (Track 1 and Track 3)           |   |                   |   |     |     |                  |                   |                    |  |
| Ophthalimic                 | Ophthalmic                      |   |                   |   |     |     |                  |                   |                    |  |
|                             | Fetal                           |   |                   |   |     |     |                  |                   |                    |  |
| Fetal<br>Imaging &<br>Other | Abdominal                       |   |                   |   |     |     |                  |                   |                    |  |
|                             | Intraoperative (specify)        |   |                   |   |     |     |                  |                   |                    |  |
|                             | Intraoperative (Neuro)          |   |                   |   |     |     |                  |                   |                    |  |
|                             | Laparoscopic                    |   |                   |   |     |     |                  |                   |                    |  |
|                             | Pediatric                       |   |                   |   |     |     |                  |                   |                    |  |
|                             | Small Organ (specify)           |   |                   |   |     |     |                  |                   |                    |  |
|                             | Neonatal Cephalic               |   |                   |   |     |     |                  |                   |                    |  |
|                             | Adult Cephalic                  |   |                   |   | N   |     |                  |                   |                    |  |
|                             | Trans-rectal                    |   |                   |   |     |     |                  |                   |                    |  |
|                             | Trans-vaginal                   |   |                   |   |     |     |                  |                   |                    |  |
|                             | Trans-urethral                  |   |                   |   |     |     |                  |                   |                    |  |
|                             | Trans-esoph. (non-Card.)        |   |                   |   |     |     |                  |                   |                    |  |
|                             | Musculo-skeletal (Conventional) |   |                   |   |     |     |                  |                   |                    |  |
|                             | Musculo-skeletal (Superficial)  |   |                   |   |     |     |                  |                   |                    |  |
|                             | Intravascular                   |   |                   |   |     |     |                  |                   |                    |  |
|                             | Other (specify)                 |   |                   |   |     |     |                  |                   |                    |  |
| Cardiac                     | Cardiac Adult                   |   |                   |   |     |     |                  |                   |                    |  |
|                             | Cardiac Pediatric               |   |                   |   |     |     |                  |                   |                    |  |
|                             | Intravascular (Cardiac)         |   |                   |   |     |     |                  |                   |                    |  |
|                             | Trans-esoph. (Cardiac)          |   |                   |   |     |     |                  |                   |                    |  |
|                             | Intra-Cardiac                   |   |                   |   |     |     |                  |                   |                    |  |
|                             | Other (specify)                 |   |                   |   |     |     |                  |                   |                    |  |
| Peripheral<br>Vessel        | Peripheral vessel               |   |                   |   | N   |     |                  |                   |                    |  |
|                             | Other (specify)                 |   |                   |   |     |     |                  |                   |                    |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

. .

{14}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

# Transducer: 02.0001.1613.02

K122710

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                   | Mode of Operation |   |   |     |     |                  |                   |                    |
|---------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic               | Ophthalmic                        |                   |   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other  | Fetal                             |                   |   |   |     |     |                  |                   |                    |
|                           | Abdominal                         |                   |   |   |     |     |                  |                   |                    |
|                           | Intraoperative (specify)          |                   |   |   |     |     |                  |                   |                    |
|                           | Intraoperative (Neuro)            |                   |   |   |     |     |                  |                   |                    |
|                           | Laparoscopic                      |                   |   |   |     |     |                  |                   |                    |
|                           | Pediatric                         |                   |   |   |     |     |                  |                   |                    |
|                           | Small Organ (specify)             |                   |   |   |     |     |                  |                   |                    |
|                           | Neonatal Cephalic                 |                   |   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other  | Adult Cephalic                    |                   |   |   | N   |     |                  |                   |                    |
|                           | Trans-rectal                      |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-vaginal                     |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-urethral                    |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)          |                   |   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Conventional)   |                   |   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Superficial)    |                   |   |   |     |     |                  |                   |                    |
|                           | Intravascular                     |                   |   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
|                           | Cardiac Adult                     |                   |   |   |     |     |                  |                   |                    |
|                           | Cardiac Pediatric                 |                   |   |   |     |     |                  |                   |                    |
|                           | Intravascular (Cardiac)           |                   |   |   |     |     |                  |                   |                    |
| Cardiac                   | Trans-esoph. (Cardiac)            |                   |   |   |     |     |                  |                   |                    |
|                           | Intra-Cardiac                     |                   |   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
|                           | Peripheral                        | Peripheral vessel |   |   |     | N   |                  |                   |                    |
|                           | Vessel                            | Other (specify)   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

{15}------------------------------------------------

SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband

Transducer: 02.0001.0408.01

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                   | Mode of Operation |   |   |     |     |                  |                   |                    |
|---------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic               | Ophthalmic                        |                   |   |   |     |     |                  |                   |                    |
| Fetal Imaging &<br>Other  | Fetal                             |                   |   |   |     |     |                  |                   |                    |
|                           | Abdominal                         |                   |   |   |     |     |                  |                   |                    |
|                           | Intraoperative (specify)          |                   |   |   |     |     |                  |                   |                    |
|                           | Intraoperative (Neuro)            |                   |   |   |     |     |                  |                   |                    |
|                           | Laparoscopic                      |                   |   |   |     |     |                  |                   |                    |
|                           | Pediatric                         |                   |   |   |     |     |                  |                   |                    |
|                           | Small Organ (specify)             |                   |   |   |     |     |                  |                   |                    |
|                           | Neonatal Cephalic                 |                   |   |   |     |     |                  |                   |                    |
|                           | Adult Cephalic                    |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-rectal                      |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-vaginal                     |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-urethral                    |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (non-Card.)          |                   |   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Conventional)   |                   |   |   |     |     |                  |                   |                    |
|                           | Musculo-skeletal (Superficial)    |                   |   |   |     |     |                  |                   |                    |
|                           | Intravascular                     |                   |   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
| Cardiac                   | Cardiac Adult                     |                   |   |   |     |     |                  |                   |                    |
|                           | Cardiac Pediatric                 |                   |   |   |     |     |                  |                   |                    |
|                           | Intravascular (Cardiac)           |                   |   |   |     |     |                  |                   |                    |
|                           | Trans-esoph. (Cardiac)            |                   |   |   |     |     |                  |                   |                    |
|                           | Intra-Cardiac                     |                   |   |   |     |     |                  |                   |                    |
|                           | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
| Peripheral<br>Vessel      | Peripheral vessel                 |                   |   |   |     | N   |                  |                   |                    |
|                           | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Comments;

{16}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband

### Transducer: 02.0001.0805.01

Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                                   | Mode of Operation |   |   |     |     |                  |                   |                    |
|-----------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only)   | Specific<br>(Track 1 and Track 3) | A                 | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic                 | Ophthalmic                        |                   |   |   |     |     |                  |                   |                    |
|                             | Fetal                             |                   |   |   |     |     |                  |                   |                    |
|                             | Abdominal                         |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (specify)          |                   |   |   |     |     |                  |                   |                    |
|                             | Intraoperative (Neuro)            |                   |   |   |     |     |                  |                   |                    |
|                             | Laparoscopic                      |                   |   |   |     |     |                  |                   |                    |
|                             | Pediatric                         |                   |   |   |     |     |                  |                   |                    |
|                             | Small Organ (specify)             |                   |   |   |     |     |                  |                   |                    |
| Fetal<br>Imaging &<br>Other | Neonatal Cephalic                 |                   |   |   |     |     |                  |                   |                    |
|                             | Adult Cephalic                    |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-rectal                      |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-vaginal                     |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-urethral                    |                   |   |   |     |     |                  |                   |                    |
|                             | Trans-esoph. (non-Card.)          |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Conventional)   |                   |   |   |     |     |                  |                   |                    |
|                             | Musculo-skeletal (Superficial)    |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular                     |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
|                             | Cardiac Adult                     |                   |   |   |     |     |                  |                   |                    |
|                             | Cardiac Pediatric                 |                   |   |   |     |     |                  |                   |                    |
|                             | Intravascular (Cardiac)           |                   |   |   |     |     |                  |                   |                    |
| Cardiac                     | Trans-esoph. (Cardiac) .          |                   |   |   |     |     |                  |                   |                    |
|                             | Intra-Cardiac                     |                   |   |   |     |     |                  |                   |                    |
|                             | Other (specify)                   |                   |   |   |     |     |                  |                   |                    |
|                             | Peripheral<br>Vessel              | Peripheral vessel |   |   |     |     | N                |                   |                    |
|                             |                                   | Other (specify)   |   |   |     |     |                  |                   |                    |

N= new indication; P= previously cleared by FDA: E= added under Appendix E

Comments:

FDA

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

---

**Source:** [https://fda.innolitics.com/device/K122710](https://fda.innolitics.com/device/K122710)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
