Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system: - For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values; - 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye. The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during: - Diagnostic exams; 1) - 2) Surgical interventions. The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
Device Story
Transcranial Doppler (TCD) system with robotic probe headband; inputs ultrasound Doppler signals from cerebral/peripheral arteries; transforms signals to measure blood flow velocities and detect micro-embolic signals; robotic headband automatically adjusts probe angle based on signal intensity for continuous monitoring; output displayed to clinician for evaluation of neurological vascular diseases (e.g., vasospasm, intracranial stenosis) and intraoperative monitoring of flow changes/embolic events; used in clinical/surgical settings; aids clinical decision-making by identifying potential stroke risks/obstructions; benefits patient through non-invasive monitoring of cerebral blood supply.
Clinical Evidence
Bench testing only. No clinical data provided. Verification testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and acoustic output measurement (NEMA UD 2-2004).
Technological Characteristics
Diagnostic ultrasound fluid flow analysis system; includes robotic probe headband for automated signal tracking; supports multiple transducer models; energy source: electrical; connectivity: not specified; sterilization: not specified; software: embedded firmware for signal processing and robotic control.
Indications for Use
Indicated for diagnostic ultrasound fluid flow analysis in patients requiring measurement of cerebral artery blood velocities and assessment of arterial cerebral blood flow for micro embolic signals. Vessels include middle, anterior, posterior cerebral arteries, vertebral mid basilar arteries, ophthalmic artery, and intracranial internal carotid artery. Used during diagnostic exams and surgical interventions. Not for fetal applications or use inside the sterile field.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
EMS9UA Transcranial Doppler with Robotic Probe Headband (K092164)
Submission Summary (Full Text)
{0}------------------------------------------------
Additional Information I for K122710 - Attachment II 510(k) Summary
# Attachment II 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: __
1. Date of Submission: October 29, 2012
2. Sponsor Identification
> Shenzhen Delicate Electronics Co., Ltd. 6C, Block 8, Tian-an Ind. Area, Nanshan District Shenzhen, Guangdong, 518054, China
Establishment Registration Number: 3006441164
Contact Person: Min Li Position: General Manager Tel: +86-755-26413482 Fax: +86-755-26425970 Email: minnli@delica-sz.com
Submission Correspondent 3.
> Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023 Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
- 4. Type and Reason of Submission
Special 510(k) Submission for the design change to the existed device, which is EMS9UA Transcranial Doppler with Robotic Probe Headband as cleared in K092164.
- Proposed Device Identification 5.
Proposed Device Name: Transcranial Doppler with Robotic Probe Headband Proposed Device Model: EMS9UA, EMS 9PB
Classification: Class II Product Codes: IYN, ITX and OQQ Regulation Number: 21 CFR 892.1570 Review Panel: Radiology
#### Intended Use Statement:
Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
- For the measurement of cerebral artery blood velocities to determine the presence of hemo 1) dynamically significant deviations from normal values;
- 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during:
- Diagnostic exams; 1)
- 2) Surgical interventions.
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
- Predicate Device Identification 6.
510(k) Number: K092164
Product Name: EMS9UA Transcranial Doppler with Robotic Probe Headband Manufacturer: Shenzhen Delicate Electronics Co., Ltd.
- Device Description 7.
The proposed device, Transcranial Doppler (TCD) with Robotic Probe Headband.is intended for obtaining the information of blood flow velocities throughout the body by using non-invasive technique. This method of measurement is particularly useful for examining the major arteries supplying blood to the brain.
{2}------------------------------------------------
. .
Robotic Probe is developed by Delicate for monitoring brain blood application. It was previously cleared in K092164 at May 10, 2010. The system can adjust the angle of probe automatically according to signal intensity when monitoring.
TCD is used to evaluation of numerous neurological vascular diseases such as vasospasm and intracranial stenosis. It is also used for intraoperative monitoring to help detect sudden changes in flow and potential embolic events. Emboli are small particles of foreign matter (air, particulate, thrombin, etc.) within the bloodstream that can potentially cause obstructions in various arteries in the body and the brain. Such obstructions can often lead to stroke.
- 8. Non-Clinical Test Conclusion
Bench tests were conducted to verify that all the risks identified associated with the design change were acceptable. The tests include:
- a) IEC 60601-1:2005 Ed3.0, Medical Electrical Equipment - Part 1: General Requirements for Safety:
- b) IEC 60601-1-2:2007 Ed3.0, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests;
- NEMA UD 2-2004 (R2009), Acoustic Output MeasurementStandard for Diagnostic c) Ultrasound Equipment Revision 3.
- 9. Substantially Equivalent Conclusion
The proposed device, Transcranial Doppler with Robotic Probe Headband, EMS 9UA and EMS9PB,are determined to be Substantially Equivalent (SE) to the predicate device (existed device), EMS9UA Transcranial Doppler with Robotic Probe Headband as cleared in K092164, in respect of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# February 15, 2013
Shenzhen Delicate Electronics Co., Ltd. c/o Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 237-023. Shanghai. 200237 CHINA
### Re: K122710
Trade/Device Name: Transcranial Doppler with Robotic Probe Headband; Models: EMS9UA and EMS9PB Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN. ITX. and OOO Dated: January 22, 2013 Received: January 25, 2013
# Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Transcranial Doppler with Robotic Probe Headband; Models: EMS9UA and EMS9PB, as described in your premarket notification:
### Transducer Model Number
02.0001.0214.01 02.0001.0213.01 02.0001.0805.01 AP99-0815-PW1.60
AP99-0607-PW2.0 02.0001.0408.01 02.0128.1616.01 02.0001.1613.02
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours.
FDA
for
Janine Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
# Exhibit #2 Indications for Use
510(k) Number: K122710
Device Name: Transcranial Doppler with Robotic Probe Headband Model: EMS9UA
Indications for Use:
Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
- 1) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values;
- To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for 2) observation include, but are not limited to the middle, anterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during:
- 1) Diagnostic exams;
- 2) Surgical interventions.
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
ZPRESCRIPTION USE . (Part 21 CFR 801 Subpart D) OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Image /page/5/Picture/18 description: The image shows a black and white graphic that appears to be a stylized signature or logo. The design incorporates a mix of cursive and geometric shapes, with some areas filled in with solid black and others outlined. The overall impression is abstract and somewhat illegible, suggesting a unique and artistic representation of a name or brand.
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
Page E2-1/12
{6}------------------------------------------------
SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband
# Transducer: 02.0001.0214.01
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|--------------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General | Specific | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only) | (Track 1 and Track 3) | | | | | | | | |
| Ophthalimic | Ophthalmic | | | | P | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| Fetal<br>Imaging<br>&<br>Other | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | P | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | P | | | | |
| Vessel | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
{7}------------------------------------------------
SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband
# Transducer: AP99-0607-PW2.0
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | |
|---------------------------|-----------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal Imaging &<br>Other | Adult Cephalic | | | | N | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | N | | | | |
| | Other (specify) | | | | | | | | |
N= new indication; P=previously cleared by FDA; E= added under Appendix E
{8}------------------------------------------------
# K122710
# Diagnostic Ultrasound Indications for Use Form
SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband
### Transducer: 02.0001.0213.01
.
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | |
|---------------------------|-----------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal Imaging &<br>Other | Adult Cephalic | | | | P | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | P | | | | |
| Vessel | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
.
{9}------------------------------------------------
### K122710
## Diagnostic Ultrasound Indications for Use Form
SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband
### Transducer: 02.0001.0408.01
イベント
:
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General | Specific | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only) | (Track 1 and Track 3) | | | | | | | | |
| Ophthalimic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | P | | | | |
| Vessel | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
{10}------------------------------------------------
SYSTEM: EMS9UA Transcranial Doppler with Robotic Probe Headband
Transducer: 02.0001.0805.01
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General | Specific | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only) | (Track 1 and Track 3) | | | | | | | | |
| Ophthalimic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Adult Cephalic . | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | P | | | | |
| Vessel | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
{11}------------------------------------------------
#### Indications for Use
510(k) Number: K122710
Device Name: Transcranial Doppler with Robotic Probe Headband Model: EMS9PB
Indications for Use:
Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
- 3) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values;
- To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for 4) observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal.
Transcranial Doppler is intended for use during:
- 3) Diagnostic exams:
- 4) Surgical interventions.
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
XPRESCRIPTION USE (Part 21 CFR 801 Subpart D) _OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{12}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband
# Transducer: 02.0128.1616.0】「
K122710
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|--------------------------|---------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General | Specific | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| (track 1 Only) | (Track 1 and Track 3) | | | | | | | | |
| Ophthalimic | Ophthalmic | | | | N | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal Imaging &<br>Other | Adult Cephalic | | | | N | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | N | | | | |
| Vessel | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
{13}------------------------------------------------
### K122710
### Diagnostic Ultrasound Indications for Use Form
# SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband
# Transducer: AP99-0815-PW1.60
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | | |
|-----------------------------|---------------------------------|---|-------------------|---|-----|-----|------------------|-------------------|--------------------|--|
| General | Specific | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) | |
| (track 1 Only) | (Track 1 and Track 3) | | | | | | | | | |
| Ophthalimic | Ophthalmic | | | | | | | | | |
| | Fetal | | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Abdominal | | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (specify) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | N | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-Cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | N | | | | | |
| | Other (specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
. .
{14}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
# Transducer: 02.0001.1613.02
K122710
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal Imaging &<br>Other | Adult Cephalic | | | | N | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Peripheral | Peripheral vessel | | | | N | | | |
| | Vessel | Other (specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
{15}------------------------------------------------
SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband
Transducer: 02.0001.0408.01
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | N | | | |
| | Other (specify) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments;
{16}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
SYSTEM: EMS9PB Transcranial Doppler with Robotic Probe Headband
### Transducer: 02.0001.0805.01
Indented Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|-----------------------------|-----------------------------------|-------------------|---|---|-----|-----|------------------|-------------------|--------------------|
| General<br>(track 1 Only) | Specific<br>(Track 1 and Track 3) | A | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalimic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intraoperative (specify) | | | | | | | | |
| | Intraoperative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) . | | | | | | | | |
| | Intra-Cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| | Peripheral<br>Vessel | Peripheral vessel | | | | | N | | |
| | | Other (specify) | | | | | | | |
N= new indication; P= previously cleared by FDA: E= added under Appendix E
Comments:
FDA
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.