RAPIDFRET ORAL FLUID ASSAY FOR PCP, PCP CALIBRATOR SET, PCP CONTROL SET AND RAPIDEASE ORAL FLUID COLLECTOR
Applicant
Biophor Diagnostics, Inc.
Product Code
LCM · TX
Decision Date
Apr 25, 2013
Decision
SESE
Submission Type
Traditional
Device Class
Class U
Intended Use
The RapidFRET Oral Fluid Assay for PCP is a homogeneous time-resolved fluorescence assay that is intended for use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Phencyclidine at 10 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For in vitro diagnostic use only. The RapidFRET Oral Fluid PCP Calibrator Set and the RapidFRET Oral Fluid PCP Control Set are intended for use only with the RapidFRET Oral Fluid Assay for PCP and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For in vitro diagnostic use only.
Device Story
The RapidFRET Oral Fluid Assay for PCP is a homogeneous time-resolved fluorescence immunoassay used for qualitative drug screening. It utilizes the RapidEASE Oral Fluid Collector to obtain neat oral fluid samples. The assay operates on the RapidFRET Integrated Workstation, which automates all pipetting and reading steps. The principle of operation is a competitive immunoassay based on Fluorescence Resonance Energy Transfer (FRET); drug in the sample competes with donor-fluorophore-labeled drug for binding sites on an acceptor-fluorophore-labeled antibody. When binding occurs, energy transfer produces a signal inversely proportional to the drug concentration. The workstation software processes the fluorescence signal, comparing it against a 10 ng/mL cutoff calibrator to provide a qualitative positive or negative result. The device is intended for use in central laboratories; results are preliminary and require confirmation by GC/MS or LC/MS/MS. The system aids healthcare providers in identifying potential PCP use, necessitating professional judgment for clinical decision-making.
Clinical Evidence
No clinical data. Performance was established via bench testing, including precision studies (n=88 per concentration level), analytical specificity (testing 40,000 ng/mL for cross-reactants), and a method comparison study (n=246 samples) against GC/MS reference methods. Method comparison showed 100% agreement among positives and 99% agreement among negatives.
Technological Characteristics
Homogeneous time-resolved fluorescence immunoassay. Reagents: liquid, ready-to-use PCP-specific antibody and drug conjugate. Platform: RapidFRET Integrated Workstation. Sample collection: RapidEASE Oral Fluid Collector (direct expectoration, no diluent, glass tube). Cutoff: 10 ng/mL. Detection principle: FRET signal measurement.
Indications for Use
Indicated for qualitative screening of Phencyclidine (PCP) in neat oral fluid at a 10 ng/mL cutoff. Intended for use in central laboratories by professional personnel. Not for point-of-care settings.
Predicate Devices
Thermo Scientific CEDIA® Phencyclidine (PCP) OFT Assay (k101746)
Related Devices
K142129 — RAPIDFRET ORAL FLUID ASSAY FOR COCAINE, RAPIDFRET ORAL FLUID COCAINE CALIBRATOR SET, RAPIDFRET ORAL FLUID COCAINE CONTROL SET · Biophor Diagnostics, Inc. · Jul 24, 2015
{0}------------------------------------------------
Biophor Diagnostics, Inc. Traditional Premarket Notification 510(k) Submission: K122703 _ RapidFRET Oral Fluid Assay for PCP
#### 510(k) Summary for the RapidFRET Oral Fluid Assay for PCP
# APR 2 5 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is:
#### 807,92(a)(1); Contact Information
- Name: Biophor Diagnostics, Inc. 1201 Douglas Avenue Address: Redwood City, CA 94063
Nathaniel G. Butlin, Ph.D. Contact: Phone: 650-367-4954 650-364-4985 Fax:
Date prepared: April 23, 2013
#### 807,92(a)(2); Device Name, Common Name and Classification
RapidFRET Oral Fluid Assay for PCP (Enzyme Immunoassay for Phencyclidine) RapidFRET Oral Fluid PCP Calibrator Set (Clinical Toxicology Calibrator) RapidFRET Oral Fluid PCP Control Set (Drug Mixture Control Materials) RapidEASE Oral Fluid Collector RapidFRET Integrated Workstation
| Product | Code | Class | Regulation<br>Section | Panel |
|-----------------------------------------|------|-------|-----------------------|-----------------|
| RapidFRET Oral Fluid Assay for PCP | LCM | II | 862.3100 | 91 - Toxicology |
| RapidFRET Oral Fluid PCP Calibrator Set | DKB | II | 862.3200 | 91 - Toxicology |
| RapidFRET Oral Fluid PCP Control Set | DIF | I | 862.3280 | 91 - Toxicology |
### 807.92(a)(3): Identification of Legally Marketed Predicate Devices
Thermo Scientific CEDIA® Phencyclidine (PCP) OFT Assay (k101746).
#### 807,92(a){4): Device Description
The RapidFRET Oral Fluid Assay for PCP is an In Vitro Diagnostic competitive immunoassay used to detect PCP in human oral fluid. This is a ready-to-use homogenous system that involves energy transfer between an acceptor fluorophore labeled to an antibody and a donor fluorophore labeled to drug. The assay is based on competition between drug in the sample and drug labeled with the donor fluorophore for a fixed number of binding sites on the antibody reagent. When acceptor and donor fluorophores are brought into close proximity through a binding event, energy transfer occurs. The fluorescence resonance energy transfer (FRET) signal is measured at the wavelength of the acceptor fluorophore and is inversely
{1}------------------------------------------------
Biophor Diagnostics, Inc.
# Traditional Premarket Notification 510(k) Submission: K122703-S002
RapidFRET Oral Fluid Assay for PCP proportional to the amount of drug in the sample. A Cutoff Calibrator is used to translate the sample measurement into a positive or negative result. Controls are used to establish and monitor precision and accuracy. The assay is performed on the RapidFRET Integrated Workstation.
## 807,92(a){5): Intended Use
The RapidFRET Oral Fluid Assay for PCP is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Phencyclidine at 10 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only.
The RapidFRET Oral Fluid PCP Calibrator Set and RapidFRET Oral Fluid PCP Control Set are intended for use only with theRapidFRET Oral Fluid Assay for PCP and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.
| | Thermo PCP | RapidFRET PCP |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Qualitative determination of<br>phencyclidine in human oral<br>fluid. | Same |
| Methodology | Homogeneous competitive<br>immunoassay. | Same |
| Kit Components | 1 PCP specific antibody reagent<br>(marketed in combination as a<br>lyophilized reagent and<br>reconstitution buffer).<br>1 PCP drug conjugate reagent<br>(marketed in combination as a<br>lyophilized reagent and<br>reconstitution buffer). | 1 PCP specific antibody reagent in<br>liquid, ready to use format.<br>1 PCP drug conjugate reagent in liquid,<br>ready to use format. |
| Controls | 4 different Control and<br>Calibrator concentrations are<br>available for qualitative screens. | Same |
| Calibrators | 2 Calibration levels are available. | Same |
# 807-92(a){6): Technological Similarities and Differences to the Predicate
{2}------------------------------------------------
### Biophor Diagnostics, Inc.
# Traditional Premarket Notification 510(k) Submission: K122703-S002 RapidFRET Oral Fluid Assay for PCP
| | Performance<br>Characteristics | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Precision, accuracy, cross<br>reacting/interfering studies are<br>similar to the RapidFRET Oral<br>Fluid Assay for PCP | Precision, accuracy, cross<br>reacting/interfering studies<br>demonstrate equivalence to the<br>predicate devices |
| Safety and<br>Effectiveness | Demonstrated in bench testing<br>and described in PI. | Demonstrated in bench testing and<br>described in PI, equivalent to<br>predicate. |
| Neat Oral Fluid<br>Cutoff Level | 3 ng/mL neat oral fluid. | 10 ng/mL neat oral fluid. |
| Platform | MGC240 analyzer | RapidFRET Integrated<br>Workstation |
| Sample Collection | Oral fluid is collected with the<br>Oral-Eze Saliva Collection<br>System. This device uses an<br>absorbent swab and diluent.<br>Sample is stored in plastic tube<br>with snap cap. | Neat oral fluid is collected with the<br>RapidEASE Oral Fluid Collector via<br>direct expectoration. No diluent is<br>used and sample is stored in glass<br>sample tube with inert screw cap. |
| Principle and<br>Procedure | The assay is based on the sample<br>analytes competing with analyte<br>conjugates to one inactive<br>fragment of β- galactosidase for<br>antibody binding sites.<br>The amount of drug in the<br>specimen is inversely<br>proportional to the assay signal<br>as measured by absorbance. | Drugs in the oral fluid sample compete<br>with the PCP conjugate donor<br>fluorophore for a fixed number of<br>binding sites on the individual drug<br>antibody acceptor reagents. When<br>acceptor and donor fluorophores are<br>brought into close proximity, through<br>the binding event, fluorescent energy<br>transfer is measured.<br>The amount of drug in the specimen<br>sample is inversely proportional to the<br>assay signal as measured by time<br>resolved fluorescence. |
| Controls and<br>Calibrator Levels | Calibrators are available at 0<br>ng/mL, 1 ng/mL, 3 ng/mL and 20<br>ng/mL | Calibrators are available at 0 ng/mL<br>and 10 ng/mL. Controls are available<br>at 5 ng/ml and 15 ng/ml |
### 807.92(b)(1): Brief Description of Study Data:
A series of studies were performed that evaluated the device performance characteristics including precision and analytical sensitivity, correlation with GC/MS, cross reactivity, and analytical specificity that are summarized below.
### Precision
Three lots of the RapidFRET Oral Fluid Assay for PCP were analyzed, four times daily, for a minimum of 20 days. Negative oral fluid pools were spiked with PCP at 0%, 25%, 50%, 75%, 100%, 125%, 150%, 175% and 200% of the cutoff level corresponding to approximately 0, 2.5, 5.0, 7.5, 10, 12.5, 15.0, 17.5 and 20 ng/mL. The aggregate data is summarized in the
{3}------------------------------------------------
Biophor Diagnostics, Inc. Traditional Premarket Notification 510(k) Submission: K122703-S002 RapidFRET Oral Fluid Assay for PCP
table below:
| РСР | 0% | 25% | 50% | 75% | 100% | 125% | 150% | 175% | 200% |
|-----|-----|-----|------|-----|------|------|------|------|------|
| POS | | | | | | 278 | 263 | 294 | 278 |
| NEG | 279 | 278 | 279 | 279 | 277 | | 0 | | 0 |
| 2 | 279 | 278 | 279. | 279 | 279 | 278 | 263 | 294 | 278 |
#### Correlation with GC/MS
Neat oral fluid was collected with the RapidEASE Oral Fluid Collection Device from volunteers potentially positive and negative for PCP. The samples (n=246) were randomized and blinded to the instrument operator and assayed using RapidFRET PCP reagents. Following screening, positive and negative samples were sent to a reference laboratory for confirmatory testing. The summarized data are shown below.
| n = 246 | GC/MS POS | GC/MS NEG |
|---------------|-----------|-----------|
| RapidFRET POS | 119 | 1† |
| RapidFRET NEG | 0 | 126 |
| % Agreement | 100% | 99% |
Sample contained 9 ng/mL PCP by GC/MS.
The data indicate that the RapidFRET Oral Fluid Assay for PCP was accurate >99% of the time in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector.
#### Cross Reactivity and Analytical Specificity
A compound library of 175 different structurally related and unrelated compounds including metabolites, OTC and prescription medications and drugs of abuse was used to evaluate the device cross reactivity and specificity. Compounds were spiked at 30,000 ng/mL into neat oral fluid pool aliquots with 0 ng/mL, 5 ng/mL and 15 ng/mL PCP, processed with the RapidEASE Collector, and tested with the RapidFRET PCP assay. Those compounds that gave an unexpected result were further titrated to determine the concentration at which the cross-reacting compound yielded a result approximately equivalent to the cutoff. Only 4-HydroxyPCP and PCM were found to cross react below 10,000 ng/mL with cutoff equivalent concentrations determined to be 620 and 310 ng/mL, respectively, in the absence of PCP.
A second study evaluated common substances such as foods and dental products as well as pH variations. HSA, ethanol, baking soda, whole blood, hemoglobin, hydrogen peroxide, sodium chloride, cholesterol, denture adhesive, ascorbic acid, bilirubin, lgA, lgG and IgM were spiked into neat oral fluid pool aliquots that contained either 5 ng/mL or 15 ng/mL PCP. Neat oral fluid pool was titrated to pH values of 5, 6, 7, 8 and 9, spiked with PCP at 5 ng/mL or 15 ng/mL and assayed with the RapidFRET PCP Assay. The effects of antiseptic mouthwash, cough syrup, cranberry juice, orange juice, tooth paste, chewing tobacco, cigarettes, chewing gum, hard candy, teeth whitening strips, cola, water, antacid, coffee and tea were evaluated by asking volunteers to use a specific item and provide an oral fluid
{4}------------------------------------------------
Biophor Diagnostics, Inc.
Traditional Premarket Notification 510(k) Submission: K122703-S002 RapidFRET Oral Fluid Assay for PCP
sample. These samples were then spiked with PCP at 5 ng/mL or 15 ng/mL, processed with a RapidEASE Collector and assayed with the RapidFRET PCP assay. All compounds at the listed concentrations gave a NEG result when spiked with 5 ng/mL PCP and a POS result when spike with 15 ng/mL PCP.
### 807,92(b){3): Conclusions
The RapidFRET Oral Fluid Assay for PCP including the RapidFRET Oral Fluid Negative and Cutoff Calibrators, the RapidFRET Oral Fluid Negative and Positive Controls and the RapidEASE Oral Fluid Collector were determined to be safe and effective for their intended use.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a staff with a serpent entwined around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Biophor Diagnostics, Inc. C/O Nathaniel G. Butlin 1201 Douglas Avenue REDWOOD CITY CA 94063
April 25, 2013
Re: K122703
Trade/Device Name: RapidFRET Oral Fluid Assay for PCP RapidFRET Oral Fluid PCP Calibrator Set RapidFRET Oral Fluid PCP Control Set RapidEASE Oral Fluid Collector RapidFRET Integrated Workstation Regulation Number: 21 CFR 862.3100
Regulatory Class: Unclassified Product Code: LCM, DKB, DIF Dated: April 09, 2013 Received: April 17, 2013
Dear Dr. Butlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, interne of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract lightlity warranties. We remind you, however, that device labeling must be truthful and not misleding.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device an be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA max publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other mauron wour not nection of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1000.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Katherine Serrano
For
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): k122703
Device Names:
RapidFRET Oral Fluid Assay for PCP RapidFRET Oral Fluid PCP Calibrator Set RapidFRET Oral Fluid PCP Control Set RapidEASE Oral Fluid Collector RapidFRET Integrated Workstation
### Indications for Use:
The RapidFRET Oral Fluid Assay for PCP is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Phencyclidine at 10 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only.
The RapidFRET Oral Fluid PCP Calibrator Set and RapidFRET Oral Fluid PCP Control Set are intended for use only with the RapidFRET Oral Fluid Assay for PCP and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.
Prescription Use X (21 CFR Part 801 Subpart D)
Over the Counter Use (21 CFR Part 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
And/Or
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Denise Johnson=lyles 2013.04.24 14:06:21 =04'00'
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
k122703 510(k)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.