The intended use of the Progeny Vantage Extra Oral X-Ray System is to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evaluation.
Device Story
The Progeny Vantage Panoramic X-Ray System is an extraoral radiographic imaging system used in dental clinics. It captures panoramic and cephalometric views of teeth, jaws, and oral structures. The system comprises an X-ray tubehead with integrated collimation, a digital image receptor (CCD), a rotating C-arm, an elevating column, a patient positioning table, and a computer display workstation. Operated by dental professionals, the device transforms X-ray inputs into digital radiographs displayed on a monitor. These images assist clinicians in diagnosing oral diseases and planning orthodontic treatments. The system utilizes standard radiographic techniques similar to predicate devices, with key components often sourced from identical suppliers. The output allows for image manipulation, archiving, and communication via the workstation, facilitating clinical decision-making and patient care.
Clinical Evidence
No clinical trials on human subjects. Bench testing performed per IEC 61223-3-4 for cephalometric equipment. Comparative study conducted using a skull phantom; two licensed dentists evaluated images from the Vantage system against those from the Instrumentarium OP200D and Planmeca ProMax. Dentists concluded Vantage images were of acceptable quality and at least as effective as the predicate devices for cephalometric diagnostic purposes.
Technological Characteristics
Extraoral digital X-ray system. Components: X-ray tubehead, CCD digital image receptor, rotating C-arm, elevating column, patient positioning table. Technical specs: 60-84 kVp, 4-14 mA, 0.5mm focal spot, 96 um pixel size. Connectivity: Computer display workstation for image manipulation/archiving. Testing: IEC 61223-3-4. No specific software algorithm class or advanced AI/ML features described.
Indications for Use
Indicated for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. With cephalometric option, indicated for cephalometric radiographic examinations for orthodontic treatment planning and evaluation.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
K122643
・
# JAN 0 7 2013
1 ﺮ
:
.
## 510(k) Summary
| Date: | November 8, 2012 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Midmark Corporation<br>675 Heathrow Drive<br>Lincolnshire, IL 60069 USA |
| Contact Person: | Mark Greenwood<br>Director, Engineering and Product Development<br>(847) 415-9800 |
| Device Name: | Vantage Panoramic X-Ray System |
| Classification Name: | System, x-ray, extraoral source, digital<br>(21 CFR 872.1800, Product Code MUH) |
| Predicate Devices: | Instrumentarium Orthopantomograph OP200D (K043612)<br>Planmeca ProMax (K011619) |
| Description of Device: | The Progeny Vantage Panoramic X-Ray System is an extraoral<br>radiographic imaging system for producing digital radiographs in<br>panoramic and cephalometric views of the teeth, jaw and oral structure. |
| | The Progeny Vantage Panoramic Extraoral Radiographic Imaging System<br>consists of the following main components: |
| | X-ray tubehead with integrated collimation |
| | Digital Image Receptor |
| | Rotating C-Arm for tubehead and image receptor mounting |
| | Overhead Arm |
| | Elevating Column |
| | Patient Positioning Table |
| | Electronic Control Unit<br>Computer Display Workstation |
| | 8 ft. coil cord with exposure switch |
| | Optional cephalometric extension arm |
| Intended Use: | The intended use of the Progeny Vantage Extra Oral X-Ray System is to<br>provide dental radiographic examination and diagnosis of diseases of<br>the teeth, jaw and oral structures. When the system is equipped with<br>the cephalometric option, the system will also provide cephalometric |
.
the state of the states of the states
:
.
·
{1}------------------------------------------------
radiographic examinations for use in orthodontic treatment planning and evaluation.
Substantial Equivalence Comparison:
| Technical<br>Characteristic | Progeny<br>Vantage | Instrumentarium<br>OP200D | Planmeca<br>ProMax |
|-----------------------------|--------------------|---------------------------|--------------------|
| kVp | 60-84 kVp | 57-85 kVp | 50-84 kVp |
| mA | 4-14 mA | 2-16 mA | .5-16 mA |
| Sensor type | CCD | CCD | CCD |
| Image Pixel size | 96 um | 96 um | 66, 99, or 132 um |
| Panoramic Scan Time | 2.5-16s | 2.7-14s | 2.7-16s |
| X-Ray tube focal spot | 0.5mm | 0.5mm | 0.5mm |
| Ceph option | Yes | Yes | Yes |
Intended Use for Each Device:
Progeny Vantage:
The intended use of the Progeny Vantage Extra Oral X-Ray System is to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evaluation.
Instrumentarium OP200D:
Planmeca ProMax:
Orthopantomograph OP200 (film unit) and OP200D (digital unit) devices are intended to be used for producing x-ray radiographs of dentition, TM-joints, and other oral structures. The units are capable of taking panoramic, TM-joint, maxillary sinus radiographs from patients. When the units are equipped with cephalometric option Orthoceph OC200 (film unit) and OC200D (digital unit) units can be used for cephalometric radiography and examinations related thereto. OP200 or OC200 units can also be equipped with Ortho Trans (OT) option, which is capable of taking both cross and longitudinal slices of region of interest. Ortho Trans uses linear tomography imaging principle.
The ProMax, Panoramic X-Ray Imaging System with Cephalostat, is an extraoral source X-Ray System, which is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. The device can be equipped with accessories to fulfill different diagnostic needs. In digital configuration the images are
{2}------------------------------------------------
displayed on a monitor, and image manipulation, archiving, and communication are performed via a computer.
Technical Analysis: A review of the technical characteristics indicates no significant differences between the Vantage unit and the predicate devices. The radiographic technique factors which would determine patient radiation dosage are almost identical. In fact, many of the key components (x-ray tubes, detectors, etc) are sourced from the same suppliers.
> The intended uses of the Vantage unit and the predicate devices also have no significant differences. All three devices' intended use focuses on providing radiograph examination of the teeth, jaw, and oral structures. The Progeny Vantage and Instrumentarium IFU statements specifically describe cephalometric while the Planmeca refers to more general "different diagnostic needs." The Instrumentarium unit IFU describes a linear tomography feature which is NOT present in Vantage unit or IFU and therefore should not be considered.
Non-Clinical Tests: The Progeny Vantage Panoramic unit was tested per IEC 61223-3-4 which is an FDA recognized standard for acceptance testing of dental xray equipment, specifically Section 7 which is for Cephalometric equipment. The Progeny Vantage with ceph option passed this testing which is a confirmation that the product is safe and effective.
Clinical Tests: Two licensed dentists reviewed images taken of a skull phantom with the Progeny Vantage with Ceph. Both dentists deemed that the image quality was acceptable and that they would be effective as effective as cephalometric images. In addition, the two dentists reviewed images taken of the skull by the Instrumentarium OP200D of the skull phantom and Planmeca sample cephalometric images and concluded that the Progeny Vantage images were at least as effective as the Instrumentarium and Planmeca images.
Conclusion: The Vantage Panoramic X-Ray system is substantially equivalent to other legally marketed devices in the United States. Specifically, the Vantage Panoramic X-Ray system is substantially equivalent in technical characteristics, intended use, and effectiveness to the OP200D marketed by Instrumentarium and the ProMax marketed by Planmeca.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 7, 2013
Progeny Dental, A Midmark Company Mr. Mark Greenwood Director, Engineering and Product Development 675 Heathrow Drive LINCOLNSHIRE IL 60069
Re: K122643
Trade/Device Name: Progeny Vantage Panoramic X-Ray System Regulation Number: 21 CFR §872.1800 Regulation Name: System, x-ray, extraoral source, digital Regulatory Class: II Product Code: MUH Dated: December 19, 2012 Received: December 26, 2012
Dear Mr. Greenwood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2-Mr. Greenwood
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):_____K122643
Device Name:___ Progeny Vantage Panoramic X-Ray System
Indications For Use:
The indications for use of the Progeny Vantage Panoramic X-Ray System is to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometic radiographic examinations for use in orthodontic treatment planning and evaluation.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K/22643
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.