← Product Code [DPS](/productcode/DPS) · K122632

# GUAVA II (K122632)

_Cardiocomm Solutions, Inc. · DPS · Feb 11, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K122632

## Device Facts

- **Applicant:** Cardiocomm Solutions, Inc.
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Feb 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

Guava II is intended to be used by cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multilead ECG devices. Guava II is intended to be electronically interfaced, and perform analysis with data transferred from other computer based ECG systems. Guava II includes an algorithm that will analyze the 12 Lead ECG and produce measurements of the ECG recording as well as textual interpretation. The product also includes an automatic analysis and interpretation software library that provides ECG signal processing and analysis on a beat by beat basis. QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis for up to twelve (12) leads of historic ambulatory ECG data. The Guava II algorithm analysis can detect and classify arrhythmias and may be used to triage data, however, Guava II is not intended to offer independent diagnosis or provide medical alarms. Clinical judgment and experience are used to check and interpret the data as part of a diagnosis. Guava II is not for use in lifesupporting or sustaining systems or EGG monitoring devices. Guava II is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate.

## Device Story

Guava II is medical device software with an API for capturing, storing, retrieving, viewing, and analyzing 1-, 2-, 3-channel, and 12-lead ECGs and biosignals. Used in clinics, hospitals, and physician offices by cardiologists, technicians, and nurses. Inputs include ECG data from external computer-based systems; outputs include ECG measurements, textual interpretations, and arrhythmia classifications. The device utilizes the Monebo Automated ECG Analysis and Interpretation Software Library and the QRS Louvain Algorithm to perform beat-by-beat signal processing, QRS feature extraction, interval/heart rate measurement, and rhythm analysis. It functions as a triage tool; clinical judgment is required for diagnosis. It does not provide medical alarms or independent diagnosis. Benefits include streamlined ECG data management and automated analysis of historic ambulatory data.

## Clinical Evidence

Bench testing only. Verification and validation activities included scenario validations, algorithm confirmation testing, and device functional testing. The device complies with AAMI EC11 and AAMI EC38 standards. No clinical data was required for this submission.

## Technological Characteristics

Medical device software with formal API. Features include ECG signal processing, QRS feature extraction, and arrhythmia classification. Complies with AAMI EC11 and AAMI EC38 standards. Utilizes the Monebo Automated ECG Analysis and Interpretation Software Library and QRS Louvain Algorithm (identical to predicate). Supports storage of raw input data and multi-language UI. Operates on current software platforms.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- GlobalCardio ([K111320](/device/K111320.md))
- Monebo Automated ECG Analysis and Interpretation Software Library ([K062282](/device/K062282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122632

Pg. 1 of 3

# 510(k) SUMMARY (as required by 807.92(c))

FEB 1 1 2014

| Regulatory Correspondent: | AJW Technology Consultants, Inc.<br>445 Apollo Beach Blvd.<br>Apollo Beach, FL 33572<br>Jon Ward<br>wardjp@ajwtech.com<br>Phone: 813-645-2855<br>Fax: 813-677-4787 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of 510(k):      | CardioComm Solutions, Inc.<br>919 Fort St. Suite 203<br>Victoria, BC V8V 3K3<br>Mona Palfreyman<br>mona@cardiocomm.com<br>Phone: 250-744-1122<br>Fax: 250-744-1866 |
| Date of Summary:          | January 31, 2014                                                                                                                                                   |
| Trade/Proprietary Name:   | Guava II                                                                                                                                                           |
| Classification Name:      | Class II                                                                                                                                                           |
| Product Code:             | DPS and DSH                                                                                                                                                        |
| Regulation:               | 870.2340                                                                                                                                                           |

#### Intended Use:

Guava II is intended to be used by cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multilead ECG devices. Guava II is intended to be electronically interfaced, and perform analysis with data transferred from other computer based ECG systems. Guava II includes an algorithm that will analyze the 12 Lead ECG and produce measurements of the ECG recording as well as textual interpretation. The product also includes an automatic analysis and interpretation software library that provides ECG signal processing and analysis on a beat by beat basis. QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis for up to twelve (12) leads of historic ambulatory ECG data. The Guava II algorithm analysis can detect and classify arrhythmias and may be used to triage data, however, Guava II is not intended to offer independent diagnosis or provide medical alarms. Clinical judgment and experience are

{1}------------------------------------------------

Pg. 2 of 3

used to check and interpret the data as part of a diagnosis. Guava II is not for use in lifesupporting or sustaining systems or EGG monitoring devices. Guava II is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate.

## Device Description:

The Guava II is a medical device software with a formal Application Programming Interface (API). The API can be invoked from Host Applications to provide services for capturing, storing, retrieving, viewing, editing, and analyzing ECGs (1-, 2-, 3-Channel and 12-Lead) and other Biosignals. Guava II may also be licensed to 3rd party organizations interested in embedding the capabilities within their own products. When the Guava II is used in other medical products, manufacturers will identify the indications for use depending on the application of their device.

The Guava II is not a life-supporting or life-sustaining system. It is intended that competent human intervention be involved before any impact on health occurs. Clinical judgment and experience are used to check and interpret the data.

| Predicate Device(s): | GlobalCardio (K111320)<br>Monebo Automated ECG Analysis and<br>Interpretation Software Library (K062282) |
|----------------------|----------------------------------------------------------------------------------------------------------|
|----------------------|----------------------------------------------------------------------------------------------------------|

Guava II has the same classification information, the same performance safety, the same design principle, similar intended use and technological characteristics, similar product design and specifications. similar effectiveness. and similar function as compared to the predicate devices.

Guava II represents, specifically, the subset of the functionality embedded within the predicate GlobalCardio (K111320) - Guava, the predecessor of Guava II, which is the subject of this 510(k) application.

GlobalCardio makes full use of the Guava functionality to capture and manage ECG's the same functionality found in Guava II. Beyond that, GlobalCardio contains additional functionality outside the scope of ECG processing such as Patient Management, System Administration, Device Inventory Control, Cardiac Rehabilitation Data Management, etc. These are capabilities of the predicate device that are not contained in Guava II.

Guava II extends its predecessor, Guava, through the inclusion of features such as;

- . The integration of the Monebo Automated ECG Analysis and Interpretation Software Library (K062282) for use in analysis and triage.
- Technological enhancements such as a "globalization" capability to simplify . the translation of the software user interface into other languages.
- . The ability to run on current software platforms.
- . Incremental features such as the storage of the raw input data (e.g. audio waveform) into the ECG representation.
- And support for additional annotation types. .

{2}------------------------------------------------

The Guava II also contains a version of the QRS Louvain Algorithm; which is also embedded within the GlobalCardio predicate. The Guava II Louvain Data Processor performs automated ECG analysis and interpretation in the same manner as GlobalCardio. The two products use the identical machine code library (DLL) to perform the analysis, and testing demonstrates the ORS Algorithm embedded in the Guaya II Louvain Data Processor is identical in function to the QRS Algorithm embedded in GlobalCardio.

All other features of Guava II are present in at least one of the predicate devices. Therefore, the proposed Guava II device is substantially equivalent to GlobalCardio (K111320) and Monebo Automated ECG Analysis and Interpretation Software Library (K062282).

# Performance Testing:

The device complies with AAMI EC11 and AAMI EC38 standards. Verification and validation activities related to the device modification were performed on the applicant device, and the predetermined acceptance criteria were met in all cases. The activities included scenario validations, algorithm confirmation testing, and device functional testing.

### Conclusion:

The proposed device is substantial equivalent to the GlobalCardio (K111320) Monebo Automated ECG Analysis and Interpretation Software Library (K062282) devices. The proposed device has a similar intended use, technological, and design characteristics as the predicate device. Any minor differences do not introduce new issues of safety or effectiveness.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing or being embraced, represented by three curved lines.

February 11, 2014

Cardiocomm Solutions, Inc. Jonathan Ward 962 Allegro Lane Apollo Beach, FL 33572 US

Re: K122632

> Trade/Device Name: Guava II Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DSH Dated: October 25, 2013 Received: August 29, 2012

### Dear Jonathan Ward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Owen-Paris-S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

. . . . ......

510(k) Number (if known): K122632

Device Name: Guava II

Guava II is intended to be used by cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multilead ECG devices. Guava II is intended to be electronically interfaced, and perform analysis with data transferred from other computer based ECG systems. Guava II includes an algorithm that will analyze the 12 Lead ECG and produce measurements of the ECG recording as well as textual interpretation. The product also includes an automatic analysis and interpretation software library that provides ECG signal processing and analysis on a beat by beat basis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis for up to twelve (12) leads of historic ambulatory ECG data. The Guava II algorithm analysis can detect and classify arrhythmias and may be used to triage data, however, Guava II is not intended to offer independent diagnosis or provide medical alarms. Clinical judgment and experience are used to check and interpret the data as part of a diagnosis. Guava II is not for use in lifesupporting or sustaining systems or EGG monitoring devices. Guava II is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate.

Prescription Use X Prescription Ose A
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| Digitally signed by                                    |
| Owen P. Faris -S                                       |

Digitally signed By
Owen P. Faris -S
Date: 2014.02.11
16:23:45'-05'00'

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K122632](https://fda.innolitics.com/device/K122632)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
