K122625 · Getinge Sourcing, LLC · FLE · May 22, 2013 · General Hospital
Device Facts
Record ID
K122625
Device Name
GETINGE 800HC- SERIES STEAM STERILIZER
Applicant
Getinge Sourcing, LLC
Product Code
FLE · General Hospital
Decision Date
May 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The Getinge 800HC-E Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
Device Story
The Getinge 800HC-E Series Steam Sterilizer (Model 833HC-E) is a large-capacity steam sterilizer for healthcare facilities. It utilizes gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. The device processes heat and moisture-stable items, including surgical instruments and linens. It features two chamber sizes (42-inch and 76-inch lengths). The control system sequences and monitors cycle phases, providing audible and visual notifications for parameter deviations. Operators select from up to 17 cycles, including Prevac, Gravity, and Liquid cycles. The device ensures sterility assurance through controlled thermal conditions. It benefits patients by providing sterilized surgical equipment, reducing infection risk. The system is designed for use in clinical environments by trained healthcare personnel.
Clinical Evidence
No clinical data required. Bench testing performed per ANSI/AAMI ST8:2008. Validation included empty chamber testing, fabric process challenge packs, and wrapped instrument process challenge devices. Results demonstrated SAL of at least 10^-6, Fo values of at least 12, and complete biological indicator (BI) kill. Liquid load cycles validated using 1L flasks with water loss <50mL. Bowie-Dick cycle validated per standard.
Technological Characteristics
Steam sterilizer utilizing gravity/downward displacement and pressure/vacuum pulsing. Chamber sizes: 42 in (39.6 cu ft) and 76 in (72.0 cu ft). Materials are identical to predicate. Connectivity: control system monitors cycle phases. Sterilization method: pressurized steam. Software: cycle operation control.
Indications for Use
Indicated for use by healthcare facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items, including surgical instruments and linens, using pressurized steam. Liquid cycles are not for sterilization of liquids intended for direct patient contact.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
Getinge Model 700HC-E Series Steam Sterilizer (K120441)
Submission Summary (Full Text)
{0}------------------------------------------------
K122625
# 510(k) SUMMARY
# MAY 2 2 2013
# Getinge 800HC-E Series Steam Sterilizer
| Submitted by: | Getinge Sourcing LLC<br>1777 E Henrietta Road<br>Rochester, NY 14623-3133 |
|------------------------|----------------------------------------------------------------------------------------------------|
| Contact Person: | Barb Smith, RAC<br>Sr. Manager, Regulatory Affairs<br>Phone: (585) 214-6049<br>Fax: (585) 272-5299 |
| Date prepared: | August 15, 2012 |
| Proprietary Name: | Model 800HC-E Series Steam Sterilizer |
| Common Name: | Steam Sterilizer |
| Device Classification: | Steam Sterilizer (80 FLE)<br>Class II, as listed per 21 CFR 880.6880 |
| Predicate Device: | Getinge Model 700HC-E Series Steam Sterilizer [K120441] |
## · Description of Device:
The Getinge 800HC-E Series Steam Sterilizer is designed for sterilization of heat and moisture stable materials used in healthcare facilities. The only model designation provided in the 800HC-E Series Steam Sterilizer is 833HC-E. The model 833HC-E is available in 2 chamber sizes; 42 inches long (39.6 cu ft) and 76 inches long (72.0 cu ft).
The Getinge 800HC-E Series Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Up to 17 cycles can be easily accessed and custom cycle names can be designated by the user (duplicate cycles are provided to allow for user designated names). All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters.
{1}------------------------------------------------
Getinge Sourcing LLC FDA 510(k) Summary Device: 800HC-E Series Steam Sterilizer
## List of available cycles:
# Getinge 800HC-E Series Steam Sterilizer Model 833HC-E Cycles and Load Chart
| Cycle Type | No. of<br>Cycles | Factory Settings | | | Load Configuration<br>(Note 1) | Maximum Items per<br>Chamber Length | |
|-----------------------------|------------------|----------------------|------------------|------------------------------------------|------------------------------------------------------------------------------|-------------------------------------|---------------------|
| | | Exp.<br>Temp. | Exp.<br>Time | Drying<br>Time | | 42 in.<br>(1067 mm) | 76 in.<br>(1930 mm) |
| vac<br>PREVAC 1 | 6 | 135.0°C<br>(275.0°F) | 3 min | 16 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric Packs | 36 | 72 |
| vac<br>PREVAC 4 | 1 | 132.2°C<br>(270.0°F) | 4 min | 16 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| vac<br>B & D TEST | 1 | 133.9°C<br>(273.0°F) | 3 min,<br>30 sec | 0 min | S.M.A.R.T. Pack or<br>equivalent (1 max.) in<br>an EMPTY chamber | 1 Test<br>Pack | 1 Test<br>Pack |
| grv<br>GRAVITY 1 | 3 | 121.1°C<br>(250.0°F) | 30 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| grv<br>GRAVITY 2 | 3 | 135.0°C<br>(275.0°F) | 10 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| grv<br>GRAVITY 3 | 1 | 132.2°C<br>(270.0°F) | 15 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| liq<br>LIQUIDS 1 | 1 | 121.1°C<br>(250.0°F) | 45 min | 5.17<br>kPa/min<br>(0.75<br>psi/min) | Each container<br>1000 mL (34 fl oz) or<br>smaller<br>(Notes 4, 5) | 160 | 320 |
| | | | | (Note 3) | | | |
| lk<br>LEAK TEST<br>(Note 2) | 1 | 131.1°C<br>(268.0°F) | 3 min. | 15 min dry, 5 min equalize , 15 min test | Empty Chamber<br>(other than loading accessories) | - | - |
| | | | | | | | |
{2}------------------------------------------------
# 00HC-E Series Steam Sterilizer
## NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.
#### TABLE NOTES:
- 1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAM! Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8).
For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterilization and sterility assurance in health care facilities.
2. Vacuum leak test parameters are not adjustable.
#### 3. Cooldown rate
- 4. Your facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).
5. Use vented or open containers only.
## Intended Use:
The Getinge 800HC-E Series Steam Sterilizer is intended for use by health care facilities and to be used to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
## Comparisons to Predicate Device:
Similarities between the Getinge 800HC-E Series Steam Sterilizer and the identified predicate are:
- Intended use is the same: Intended for use by health care facilities to sterilize � wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
- Operating Principle is the same: Saturated steam is the sterilizing agent. o
- Materials of construction are the same. There is no direct patient contact . associated with this device.
{3}------------------------------------------------
Getinge Sourcing LLC FDA 510(k) Summary Device: 800HC-E Series Steam Sterilizer
- Cycle Types: The cycle types offered are the same; Prevac (135°C, 132°C), . Gravity (121°C, 132°C, 135°C), and Liquids 121°C (not for sterilization of liquids used directly for patient contact).
- Performance Testing: Factory recommended cycles were tested per industry . standards and guidelines and effectiveness of sterilizer function was demonstrated by complete kill of biological indicators. Getinge Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8:2008 Hospital Steam Sterilizers.
The differences between the Getinge 800HC-E Series Steam Sterilizer and the predicate device (Getinge 700HC-E Series Steam Sterilizer) are:
- The Getinge 800HC-E Series Steam Sterilizer has larger vessel sizes. o Because of the larger vessel size the chamber closure (door operation) is different. The larger chamber size also allows for larger loads to be processed. Factory recommended maximum load sizes are tested for effectiveness and clearly identified in product labeling.
## Summary of Performance Testing:
Factory recommended cycles were tested per industry standards and guidelines and effectiveness of sterilizer function was demonstrated by complete kill of biological indicators. Getinge Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8:2008 Hospital Steam Sterilizers.
The results of Getinge 800HC-E Series Steam Sterilizer validation testing demonstrate that the sterilizer performs as intended. Summary of testing:
- Empty chamber testing performed for all cycles as described in ANSI/AAMI . ST8:2008 Hospital Steam Sterilizers section 5.4.2.5. The results demonstrated that the sterilizer is capable of providing steady-state thermal conditions within the chamber that are consistent with the predicated sterility assurance level (SAL) in the load.
- All PREVAC and GRAVITY cycles were validated using fabric process o challenge packs as described in ANSI/AAMI ST8:2008 section 5.5.2. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12, complete Bl kill and moisture retention of less than 3% increase in pre-sterilization test pack weight including no visible wet spots.
- All PREVAC and GRAVITY cycles were validated using wrapped instrument 0 process challenge devices as described in ANSI/AAMI ST8:2008 section 5.5.4. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12, complete BI kill and moisture retention of less than 20% increase in pre-sterilization weight of the towel including no visible wet spots on the outer wrapper.
{4}------------------------------------------------
Getinge Sourcing LLC FDA 510(k) Summary Device: 800HC-E Series Steam Sterilizer
- Liquid loads cycles were validated using 3 one liter flasks as described in � ANSI/AAMI ST8:2008 section 5.5.3. The results from this testing demonstrated a sterility assurance level of at least 10° through achievement of time at temperature sufficient to produce an Fo value of at least 12, complete Bl kill and water loss not exceeding 50ml.
- Bowie Dick cycle was validated using the Bowie-Dick test pack as described in o ANSI/AAMI ST8:2008 section 5.6.1.1.
- The software validation for the cycle operation was performed according to o FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (5/11/2005)".
#### Clinical Data:
No clinical data is required for this device classification submission.
#### Conclusion:
The 800HC-E Series Steam Sterilizer has the same intended use and technological characteristics as the predicate device. The 800HC-E Series Steam Sterilizer meets the applicable requirements of AAMI ST8:2008 performance standards. Based on the information provided in this premarket notification, it can be concluded that the subject device is substantially equivalent to the predicate device and is safe and effective when used as intended.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized lines forming its body and wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 22, 2013
Ms. Barb Smith Getinge Sourcing LLC Senior Manager Regulatory Affairs 1777 East Henrietta Road ROCHESTER, New York 14623-3133
Re: K122625
Trade/Device Name: Getinge 800HC-E Series Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: May 13, 2013 Received: May 14, 2013
#### Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/6/Picture/7 description: The image shows a signature and contact information. The text "Sincerely yours" is at the top of the image. Below that is the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The acronym "DAGRID" is also present.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K i 2 2 G 2 5
Device Name: Getinge 800HC-E Series Steam Sterilizer
## Indications for Use: .
The Getinge 800HC-E Series Steam Sterilizer is intended for use by health care facilites to sterlizes I he Gelinge Boortes Beatle Stean Cremocous heat and moisture stable items such as surgical it writished and linens by means of pressurized steam. The 800HC-E Series Steam Sterliizer is thistriments and michs by means of prossanted by chamber and 833HC-E 42 inch chamber and 833HC-E 76 inch chamber.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
013
# Sreekanth Gut 2013.05.2012-2
Division Sign-Off) ·vision of Anesthesiology, General Hospital · nection Control, Dental Devices
K122625 ్10(k) Number:
{8}------------------------------------------------
# List of available cycles:
# Getinge 800HC-E Series Steam Sterilizer Model 833HC-E Cycles and Load Chart
KiZZ 625
IFU P -- 50 2 - F 3
| Cycle Type | No. of<br>Cycles | Factory Settings | | | Load Configuration<br>(Note 1) | Maximum Items per<br>Chamber Length | |
|-----------------------------|------------------|----------------------|------------------|---------------------------------------------------------|------------------------------------------------------------------------------|-------------------------------------|---------------------|
| | | Exp.<br>Temp. | Exp.<br>Time | Drying<br>Time | | 42 in.<br>(1067 mm) | 76 in.<br>(1930 mm) |
| vac<br>PREVAC 1 | 6 | 135.0°C<br>(275.0°F) | 3 min | 16 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric Packs | 36 | 72 |
| vac<br>PREVAC 4 | 1 | 132.2°C<br>(270.0°F) | 4 min | 16 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| vac<br>B & D TEST | 1 | 133.9°C<br>(273.0°F) | 3 min,<br>30 sec | 0 min | S.M.A.R.T. Pack or<br>equivalent (1 max.) in<br>an EMPTY chamber | 1 Test<br>Pack | 1 Test<br>Pack |
| grv<br>GRAVITY 1 | 3 | 121.1°C<br>(250.0°F) | 30 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| grv<br>GRAVITY 2 | 3 | 135.0°C<br>(275.0°F) | 10 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| grv<br>GRAVITY 3 | 1 | 132.2°C<br>(270.0°F) | 15 min | 30 min | Double-wrapped<br>instrument trays, up<br>to 11.3 kg (25.0 lb)<br>(per tray) | 12 | 24 |
| | | | | | Fabric packs | 36 | 72 |
| liq<br>LIQUIDS 1 | 1 | 121.1°C<br>(250.0°F) | 45 min | 5.17<br>kPa/min<br>(0.75<br>psi/min)<br>(Note 3) | Each container<br>1000 mL (34 fl oz) or<br>smaller<br>(Notes 4, 5) | 160 | 320 |
| lk<br>LEAK TEST<br>(Note 2) | 1 | 131.1°C<br>(268.0°F) | 3 min. | 15 min<br>dry, 5<br>min<br>equalize<br>, 15 min<br>test | Empty Chamber<br>(other than loading<br>accessories) | - | - |
{9}------------------------------------------------
K122625 IFU C 3
NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.
#### TABLE NOTES:
- 1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAM Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8).
For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterilization and sterility assurance in health care facilities.
2. Vacuum leak test parameters are not adjustable.
- 3. Cooldown rate
4. Your facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).
- 5. Use vented or open containers only.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.