ARC ENDOCUFF

K122565 · Boddingtons Plastics, Ltd. · FED · Sep 4, 2012 · Gastroenterology, Urology

Device Facts

Record IDK122565
Device NameARC ENDOCUFF
ApplicantBoddingtons Plastics, Ltd.
Product CodeFED · Gastroenterology, Urology
Decision DateSep 4, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

To be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following: - Keeping the suitable depth of endoscope's view field - Helping the endoscope with being inserted into the gastrointestinal tract

Device Story

Arc EndoCuff is a single-use, sterile, cylindrical polymer attachment for the distal end of an endoscope. It features flexible 'hair' structures that fold during forward movement (intubation) and open when the device is withdrawn to manipulate colonic folds, maintain a central position in the lumen, and prevent slippage. Used by gastroenterologists during endoscopic procedures to stabilize the endoscope and optimize the field of view. The device provides mechanical assistance to the physician, potentially improving visualization and procedural efficiency. It is a passive mechanical accessory; no electronic or software components are involved.

Clinical Evidence

Bench testing only. Performance data confirmed the device maintains field of view and aids insertion similarly to predicate devices. Biocompatibility testing (USP Class VI) demonstrated no safety concerns. No clinical data provided.

Technological Characteristics

Cylindrical polymer device; flexible 'hair' features for lumen stabilization; radiation sterilized; single-use. Materials conform to USP Class VI. Mechanical attachment mechanism; no energy source or software.

Indications for Use

Indicated for use in patients undergoing endoscopic therapy to assist in maintaining endoscope field of view and aiding insertion into the gastrointestinal tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ SEP 4 2012 # 510(k) Summary ## Summary of Safety and Effectiveness ## 510(k) number_ كا2256) 510 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. ### Summary | Submitter's Identification: | Boddingtons Plastics Ltd | | |-----------------------------|----------------------------------------|--| | | | | | | Unit 6 Wheelbarrow Park Estate | | | | Pattenden Lane | | | | Tonbridge | | | | Kent | | | | TN12 9QJ | | | | United Kingdom | | | Contact: | John Bovis | | | Telephone: | +44 1622 833700 | | | Fax: | +44 1622 833701 | | | Email: | John.bovis@boddingtons.co.uk | | | Date of Summary: | 11th May 2012 | | | Device Name: | EndoCuff | | | Proprietary Name: | Arc EndoCuff AEC 110; AEC 120; AEC 140 | | | Classification: | Endoscope and/or accessories | | | Product Code: | FED | | | Product Class: | II | | | Code of Federal Regulation: | 21 CFR 876.1500 | | | Panel | Gastroenterology and Urology | | | Substantial Equivalence: | | | | | |----------------------------|------------------------------------------|-------------------|---------------|--| | Manufacturer | Trade Name | Model | 510(k) Number | | | Olympus Optical Co.<br>Ltd | Olympus Distal<br>Attachment | MH & D-201 Series | K984358 | | | Fujinon Corporation | Fujinon EC-450815<br>(BS-2 Balloon only) | BS-2 Balloon | K090116 | | | Device Cross-Reference | | | | | |------------------------|------------------------|--------------------|---------------------------|--| | Arc EndoCuff | MH-Series<br>(K984358) | D-201<br>(K984358) | BS-2 Balloon<br>(K090116) | | | Model No. AEC 110 | MH-463 | D-201-13404 | BS-2 Balloon | | | Model No. AEC 120 | MH-466 | D-201-14304 | BS-2 Balloon | | | Model No. AEC 140 | MH-466 | D-201-15004 | BS-2 Balloon | | lssue 2 {2}------------------------------------------------ ### Description ### Description of Device The Arc EndoCuff has a short tube like shape with flexible hairs and is attached to the distal end of the endoscope to facilitate endoscopic therapy. The Arc EndoCuff is designed to fit specific endoscopes (as designated on the packaging), and is supplied sterile following radiation sterilization and is single use only. #### Indications for Use: To be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following: - Keeping the suitable depth of endoscope's view field - Helping the endoscope with being inserted into the gastrointestinal tract � Summary of the Technological Characteristics of the Device compared to the Predicate Device #### Design The EndoCuff are a cylindrical shaped all polymer products. It utilises the combined properties of two polymer materials to self-retain to the designated endoscope(s). The product features a number of flexible "hair" features that fold within the structure of the product during intubation and forward movement when in use, and open out when drawn backward to control field of view. #### Materials EndoCuff raw materials have been purpose chosen as conforming to USP class VI and tested for biocompatibility (See Section 15). #### Size: The product is similar in bodily size and cylindrical shape to the products seen in both the Olympus Reusable (MH Series K984358) and Single-use (D-201 Series K984358) Distal Attachment model ranges. The flexible hair features work at a similar diameter to the size of the 200 predicate Fujinon BS-2 balloon (when inflated). {3}------------------------------------------------ ### Summary of the Determination of Substantial Equivalence & Performance Data The EndoCuff is used to maintain and maximise the endoscopes field of view during endoscopic therapy by manipulating colonic folds, maintaining a central position in the lumen and avoiding sudden slippage. The nominated predicate devices maximise the field of view in the same way, in particular: - . The Olympus Distal Attachment (MH Series and D-201 series K984358), is used to manipulate colonic folds; and - . The Fujinon BS-2 balloon (K090116) maintains central position in the lumen and avoids slippage by exerting circumferential pressure. The bench testing demonstrates that the Endocuff design combines some of the benefits of both being a slim cylindrical device (like the Olympus product) and being able to stabilise and view the mucosa effectively on a greater diameter (as does the Fujinon balloon) but within the one product. The design of the device is similar in length and is cylindrical in shape. Like both types of predicate devices, flexible polymers are the means for attachment/retention to the distal end of an endoscope. The device contacts the patient in the same way and for the same period of time however they are made from different materials and are sterilised (where appropriate) using different methods. The performance data showed that the sterilisation method selected (Arc EndoCuff is irradiated and the D-201 Series is Ethylene Oxide sterilised K984358) has not introduced any additional risks and the patient contacting materials have been tested for biocompatibility with reports demonstrating no negatives in safety or effectiveness The bench testing has demonstrated that the device does not introduce any additional risks when undertaking endoscopic therapy and meeting the intended use. #### Conclusion After a review of the bench testing, it is concluded that the Arc EndoCuff is as safe and effective as the predicate devices for facilitating endoscopic therapy. There are a few minor technological differences, in both materials used and the design. However the bio-compatibility testing shows that no additional risks are introduced from the materials used, and the performance testing demonstrates that there are no design weaknesses that could affect safety or effectiveness. There are no new indications for use and therefore by following the FDA 510(k) "Substantial Equivalence" decision making process, http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourD evice/PremarketSubmissions/PremarketNotification510k/ucm134783.htm it is substantially equivalent to the predicate device. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus. The logo is black and white. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 4 2012 Mr. Satish Champaneri Technical Reviewer BSI Healthcare Kitemark Court, Davy Avenue, Knowlhill MILTON KEYNES MK5 8PP UNITED KINGDOM Re: K122565 Trade/Device Name: Arc EndoCuff AEC110; AEC120; AEC140 Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: August 22, 2012 Received: August 22, 2012 Dear Mr. Champaneri: We have reviewed your Section 510(k) premarket notification of intent to market the device w oneve a rowe and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 abs been and to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act that do not require approval of a premarket approval application (PMA). Tou may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may), ate of novisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adviseration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r source a view a determination that your device complies with other requirements of the Act that I Dr Hab Intacted and regulations administered by other Federal agencies. You must or ury reachard the Act's requirements, including, but not limited to: registration and listing Configry with an the Pre-Brequirence) at 801); medical device reporting (reporting of medical {5}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rico-related daverse ovenas) (21 CFR Part 820); and if applicable, the electronic for in the quality by belies (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire operatio ad 110 coutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for go to mep., www.radiological Health's (CDRH's) Office of Compliance. Also, please the Conton Do 1000 and Radiovanding by reference to premarket notification" (21CFR Part note the regalation onlines and ing the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the rou may ocain outsi generers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerély yours, Sincerely yours, Benjamin K. Kirk Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Arc EndoCuff ## 510(k) number_K 122565 ## Indications for Use 510(k) Number (if known): Device Name: Arc EndoCuff AEC110; AEC120; AEC140 Indications for Use: To be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following: - Keeping the suitable depth of endoscope's view field . - Helping the endoscope with being inserted into the gastrointestinal tract . No Over the Counter Use And/Or Prescription Use Yes (21 CFR Part 801 Subpart D) (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) (Division Sign-off) currence of CDRH, Office of Device Evaluation (ODE) Division of Reproductive, Gastro-Renal, and Urological Devices | 510(k) Number | K122565 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...