GRAM POSTITVE CLINDAMYCIN

K122547 · bioMerieux, Inc. · LON · Oct 15, 2012 · Microbiology

Device Facts

Record IDK122547
Device NameGRAM POSTITVE CLINDAMYCIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateOct 15, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 AST-GP Clindamycin is designed for antimicrobial susceptibility testing of Staphylococcus aureus and Staphylococcus epidermidis. VITEK® 2 AST-GP Clindamycin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Clindamycin has been shown to be active against most strains of the microorganism listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Staphylococcus aureus (methicillin-susceptible strains) and Staphylococcus epidermidis (methicillin-susceptible strains). The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Device Story

VITEK 2 AST-GP Clindamycin is a miniaturized card for antimicrobial susceptibility testing (AST) of Staphylococcus aureus and Staphylococcus epidermidis. Input: bacterial isolate diluted in 0.45-0.5% saline, used to rehydrate antimicrobial medium in card. Operation: VITEK 2 system automatically fills, seals, and incubates cards; VITEK 2 Compact uses manual filling/sealing. System monitors growth in card wells over time. Output: MIC value and interpretive category result. Used in clinical laboratories by technicians/microbiologists. Healthcare providers use results to guide antibiotic therapy decisions for patients with bacterial infections. Benefits: standardized, automated susceptibility testing compared to manual broth microdilution.

Clinical Evidence

Performance established via clinical study of 492 clinical isolates and 139 challenge isolates (S. aureus and S. epidermidis). Compared against CLSI broth microdilution reference method. Combined results (AutoDilution): 96.0% Essential Agreement (EA) and 99.4% Category Agreement (CA). Reproducibility >95% within +/- one dilution. Bench testing confirmed QC performance within expected ranges for ATCC 29212 and ATCC 29213.

Technological Characteristics

Miniaturized card format for broth microdilution. Automated incubation and optical growth monitoring. Compatible with VITEK 2 and VITEK 2 Compact instrument platforms. Quantitative MIC determination.

Indications for Use

Indicated for antimicrobial susceptibility testing of isolated colonies of Staphylococcus aureus and Staphylococcus epidermidis (methicillin-susceptible strains) using VITEK 2 and VITEK 2 Compact Systems. For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KI22547 Image /page/0/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a stylized globe above it. The globe is divided into two halves, one with horizontal lines and the other solid black. A curved line passes through the globe and the text. OCT 1 5 2012 # 510(k) SUMMARY ## VITEK® 2 AST- Gram Positive Clindamycin #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |-------------------|-----------------------------------------| | Address: | 100 Rodolphe Street<br>Durham, NC 27712 | Contact Person: Phone Number: Fax Number: Date of Preparation: tham, NC 27712 Elizabeth (Betty) Landon Staff Regulatory Affairs Specialist VITEK® 2 AST - Gram Positive Clindamycin VITEK® 2 Gram Positive Linezolid (k032766) 919-620-2329 919-620-2548 Auqust 20, 2012 21 CFR 866.1645 Product Code LON Antimicrobial Susceptibility Test VITEK® 2 AST - GP Clindamycin #### B. Device Name: · Formal/Trade Name: Classification Name: Common Name: C. Predicate Device: ## D. 510(k) Summary: VITEK 2 AST - Gram Positive Clindamycin is designed for an antimicrobial susceptibility testing of Staphylococcus aureus and Staphylococcus epidermidis. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcq/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The bacterial isolate to be tested is diluted to a standardized concentration with 0.45 - 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. VITEK 2 AST - Gram Positive Clindamycin for Staphylococcus aureus and Staphylococcus epidemidis demonstrated substantially equivalent performance with the CLSI broth microdilution reference method, as defined in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued August 28, 2009. The Premarket Notification (510(k)) presents data in support of VITEK 2 AST - Gram Positive Clindamycin for Staphylococcus aureus and Staphylococcus epidermidis. An external evaluation was > bioMérieux, Inc. 100 Rodolphe Street, Durham, Allyge Zag & Tridge Phone: 919 620 2000 http://www.biomerieux-usa.com {1}------------------------------------------------ conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK AST - Gram Positive Clindamycin for Staphylococcus aureus and Staphylococcus epidermidis by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK 2 and VITEK 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), k050002. VITEK 2 AST - Gram Positive Clindamycin for Staphylococcus aureus and Staphylococcus epidermidis demonstrated acceptable performance of 99.4% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 October 15, 2012 CT 15 2012 bioMérieux, Inc. c/o Elizabeth Landon Staff Regulatory Affairs Specialist 595 Anglum Road Hazelwood, MO 63042 Re: K122547 Trade/Device Name: VITEK®2 AST- Gram Positive Clindamycin (≤ 0.12-≥4 µg/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Antimicrobial Susceptibility Test Regulatory Class: Class II Product Code: LON Dated: August 20, 2012 Received: August 21, 2012 Dear Landon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Elizabeth Landon forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to. http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ## Uwe Scherf for Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: VITEK® 2 AST - Gram Positive Clindamycin ( < 0.12 - ≥ 4 µg/mL ) Indications For Use: VITEK® 2 AST-GP Clindamycin is designed for antimicrobial susceptibility testing of Staphylococcus aureus and Staphylococcus epidermidis. VITEK 2 AST-GP Clindamycin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Clindamycin has been shown to be active against most strains of the microorganism listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Staphylococcus aureus (methicillin-susceptible strains) and Staphylococcus epidermidis (methicillin-susceptible strains) The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) ![signature](signature.png) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1513(k) k 122547 Page 7 of 352
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...